key: cord-261253-btwx2vxo authors: Yip, Timothy T. C.; Cho, William C. S.; Cheng, Wai Wai; Chan, Johnny W. M.; Ma, Victor W. S.; Yip, Tai-Tung; Lau Yip, Christine N. B.; Ngan, Roger K. C.; Law, Stephen C. K. title: Application of ProteinChip Array Profiling in Serum Biomarker Discovery for Patients Suffering From Severe Acute Respiratory Syndrome date: 2007 journal: Microarrays DOI: 10.1007/978-1-59745-304-2_20 sha: doc_id: 261253 cord_uid: btwx2vxo A new strain of coronavirus has caused an outbreak of severe acute respiratory syndrome (SARS) from 2002 to 2003 resulting in 774 deaths worldwide. By protein chip array profiling technology, a number of serum biomarkers that might be useful in monitoring the clinical course of SARS patients were identified. This book chapter describes how the protein chip array profiling was carried out for this study. Briefly, SARS patients’ serum samples were first fractionated in Q Ceramic HyperD ion exchange sorbent beads by buffers at different pH. Serum protein fractions thus obtained were then bound onto a copper (II) immobilized metal affinity capture (IMAC30 Cu [II]) ProteinChip® Array or a weak cation-exchange (CM10) ProteinChip Array. After washing and addition of sinapinic acid, the chips were read in a Protein Biological System (PBS) IIc mass spectrometer. Ions were generated by laser shots and flied in a time of flight mode to the ion detector according to their mass over charge (m/z) ratio. The serum profiling spectra in SARS patients were acquired, baseline subtracted and analyzed in parallel with those from the control subjects by Ciphergen ProteinChip Software 3.0.2 with their peak intensities compared by a nonparametric two sample Mann-Whitney-U test. More than twelve peaks were differentially expressed in SARS patients with one at m/z of 11,695 (later identified to be serum amyloid A protein), which had increase in peak intensity correlating with the extent of SARS-coronavirus induced pneumonia as defined by a serial chest X-ray opacity score. The remaining biomarkers could also be useful in the study of other clinical parameters in SARS patients. A new strain of coronavirus (CoV) caused a pandemic outbreak of severe acute respiratory syndrome (SARS) in China, Hong Kong, Singapore, Toronto, and Taiwan from 2002 to 2003, resulting in 8098 individuals being infected and 774 deaths (1) . As compared to other upper viral respiratory infections (2), SARS-CoV can rapidly induce pneumonia. The accompanying adult respiratory distress syndrome can quickly progress resulting in rapid death of the patients (3,4) . The most commonly used laboratory diagnostic tests for SARS-CoV-infection are reverse transcription (RT)-PCR and anti-SARS-CoV antibody serological tests (5-7), but their roles in monitoring the extent of pneumonia are rather limited. Although, serial chest radiography is helpful for monitoring the progress of disease (8, 9) , it is also limited by its subjectivity and variability in the interpretation of the imaging results and occasionally suboptimal quality of portable films taken in isolation ward to avoid the spread of the disease. In a recent article, the discovery of 12 upor downregulated serum biomarkers were reported in SARS patients by a novel protein chip array profiling approach (10- 14) with one biomarker appears to be useful in monitoring the extent of pneumonia (15) . In this chapter, how this protein chip array profiling technique is carried out is described. The technique involves four steps with the first step being serum fractionation in ceramic ion exchange sorbent beads. This step separates serum proteins into several fractions aiming at reducing their complexity before protein chip profiling is performed. The second step involves the binding of serum protein biomarkers onto protein chip array surfaces. The third step is desorption of bound serum biomarkers by laser shots in the mass spectrometer generating a time of flight spectrum for each sample. The fourth step concerns with data acquisition, data processing and statistical analyses of the spectra in SARS patients in comparison with the control groups. These procedures will be described in full details in the following chapter and summarized in the flow chart diagram as in Fig. 1 1. 44 Serum samples from 24 SARS patients for differential mapping of SARS associated biomarkers. Fig. 1 . Sample fractionation, chip binding, and reading in protein chip array profiling technique. Serum samples from SARS patients and various control groups were fractionated in Q Ceramic Hyper D F anion exchange sorbent beads. The eluted fractions were in turn spotted onto two types of protein chips, namely, copper IMAC30 Cu(II) ProteinChip array and weak cation-exchange (CM10) ProteinChip array. Each sample bound protein chip was in turn inserted into a Protein Biology System II SELDI-TOF-mass spectrometer and the protein biomarkers were ionized and desorbed by laser beam. The ions flied to the ion detector generating a biomarker peak intensity 7. 50% Sinapinic acid (Ciphergen Biosystems Inc.) in 500 mL/L acetonitrile and 5 mL/L trifluoroacetic acid (Sigma-Adrich). 8. Protein Biological System (PBS) IIc mass spectrometer reader (Ciphergen Biosystems). 9. All-in-1 peptide molecular mass standard (Ciphergen Biosystems Inc.). 10. Ciphergen ProteinChip Software 3.0.2 (Ciphergen Biosystems Inc.). The extent of pneumonia in the SARS patients with longitudinal follow-up is assessed by a radiologist according to a serial chest radiographic score, modified from a score system initially proposed to assess computer tomography of the chest (16) and summarized as in Fig. 2. 1. After X-ray chest radiography is taken, divide the frontal chest X-ray radiograph into six lung zones, namely, left upper zone, left middle zone, left lower zone, right upper zone, right middle zone, and right lower zone. 2. The upper zone (left or right) represents area above carina (including the apex). 3. The middle zone (left or right) represents area from carina to the level of inferior pulmonary veins. 4. The lower zone (left or right) represents area from the lower margin of middle zone to the lung base. 5. Score the opacity in each lung zone by a "coarse semiquantitative method" with a five points'scale of grades 0-4. 6. Grade 0 represents no opacity involved area. 7. Grade 1 represents 5-24% opacity involved areas. 8. Grade 2 represents 25-49% opacity involved areas. 9. Grade 3 represents 50-74% opacity involved areas. 10 . Grade 4 represents 75-100% opacity involved areas. 11. Add the grading from each of the six lung zones to provide a 0-24 point summation scale for assessing the extent of pneumonia. 1. 8 mL Blood is collected from each patient or control subject. 2. Incubate the blood for approx 2 h at 4°C to clot the blood cells. 3. Separate the serum portion from the clotted cells by centrifugation at 230g for 15 min. 4. Aliquot the sera for various routine laboratory diagnostic tests. 5. Freeze the remaining sera at −70°C for protein chip array profiling analysis. Owing to the protein complexity in the serum sample, an initial fractionation in Q Ceramic Hyper D F ion exchange sorbent beads is performed to generate serum fractions for chip binding experiments ( Fig. 1) (10,11) . Assessing the extent of pneumonia by chest X-ray opacity score. The opacity in each zone (apex+upper, middle, and lower zones) from the left and right regions of the lung was scored by a "coarse semiquantitative method" with a 5 points'scale of grades 0-4 representing involved areas of 0, 5-24, 25-49, 50-74, and 75-100%, respectively. A total score was then calculated by adding up all the grades in the six zones (Apex and upper zones were considered as one zone) to provide a 0-24 points'scale with the higher number representing more severe pneumonic disease. 1. Thaw serum from -70°C freezer immediately before serum fractionation. 2. Aliquot 20 μL of serum to each well in a 96-well sample plate. 3. Add 30 μL of D1 denaturing buffer to the well containing the serum to denature the serum proteins. 4. Shake the sample plate on a Micromix 5-01 shaker vigorously (in an amplitude of 7 and form of 20 Hg) for 20 min at 4°C to mix the serum with the D1 buffer well. 1. Add 180 μL of 50% suspension of ion exchange beads to each well of a 96-well Silent Screen filter plate and drain off the W1 bead-washing buffer by vacuum. 2. Wash the ion exchange beads in each well three more times by adding 200 μL of Q1 bead equilibration buffer to the beads and draining off the buffer by vacuum. 3. After the last vacuum drain, transfer 50 μL of the denatured serum sample from the well in the sample plate to the corresponding well in the filter plate containing the equilibrated ion exchange beads. 4. Add 50 μL of Q1 buffer to the well in the sample plate where serum was originally placed and rinse the residual serum by pipetting up and down five times. 5. Transfer the 50 μL of the serum rinse to the well in the filter plate containing the mixture of denatured serum sample and ion exchange beads. 6. Shake the filter plate vigorously on the Micromix shaker (in an amplitude of 7 and form of 20 Hg) for 30 min at 4°C to mix the serum and beads well. 1. After shaking, place a 96-well F1 collection plate under the filter plate and centrifuge at 1600g for 1 min to collect the flow-through fraction in the wells of the F1 collection plate. 2. Add 100 μL of E1 elution buffer (at pH 9.0) to each well in the filter plate containing the serum bound beads and shake vigorously again on orbital shaker (speed and time as before) for 10 min at RT. 3. Collect the eluate in the same collection plate F1 by centrifugation at 1600g for 1 min again. 4. This represents fraction 1, which contains both the flow-through fraction at pH 9.0 and the pH 9.0 eluate. 5. Add 100 μL of E2 elution buffer (at pH 7.0) to each well in the filter plate and again shake at the same speed and time. 6. Collect the eluate in another new F2 collection plate by centrifugation as before. 7. Add 100 μL of E2 elution buffer again to each filter well and elute a second time by shaking and centrifugation as before onto F2 collection plate again. 8. This represents fraction 2, which contains the pH 7.0 eluate. 9. Elute the serum protein bound beads similarly in turn with E3, E4, and E5 elution buffers as above resulting in fractions 3, 4, and 5 containing pH 5.0, pH 4.0, and pH 3.0 eluates, respectively, in three new F3, F4, and F5 collection plates. 10. Final elution is achieved by adding E6 organic solvent elution buffer and centrifugation at 2000g for 5 min giving rise to fraction 6 containing the organic solvent eluate in collection plate F6. 11. Freeze fractions 1-6 in collection plates E1-6 at -70°C until the chip binding protocol proceeds. The ProteinChip Array Profiling technique is also called surface enhanced laser desorption and ionization time of flight mass spectrometry (SELDI-TOF-MS, Ciphergen Biosystems Inc., Fremont, CA). The chip array used is a 10 mm wide × 80 mm long metal chip having eight 2-mm spots with a specific chromatographic surface for binding of biomarkers of interest (see Fig. 1 ). For IMAC30 Cu(II) ProteinChip Array, a pretreatment procedure is required for loading copper ions onto the chip for binding protein biomarkers with affinity to copper ions. This metal loading step is, however, not required for CM10 ProteinChip Array, which is impregnated with carboxylate ions as weak cation exchanger for biomarker binding. CM10 chips only requires buffer washing instead. Both types of chips are tested to be ideal for binding serum proteins/peptides for the study. 9. Remove the neutralizing buffer after shaking. 10. Rinse the well again with 100 μL of Milli-Q water by shaking for 1 min at room temperature. 11. Add 150 μL of 100 mM sodium phosphate buffer containing 0.5 mol/L NaCl (IMAC30 chip binding buffer) onto each well and shake for 5 min at RT to wash the chip spot. 12. Remove the buffer after shaking. 13 . Wash with the same IMAC30 chip-binding buffer one more time. 14. Remove the buffer after washing. 15 . Proceed immediately to Subheading 3.6.1. for sample binding without letting the chip surface to dry. to each chip spot. 11. Air-dry the chips for about 10 min. 12. Add 1 μL of 50% sinapinic acid and air-dry again. 13 . The chip is ready to be read in the PBS IIc mass spectrometer reader. 14. Remember to process at least one reference control serum concurrently with the patients'samples on each chip for quality control of chip-to-chip variability. 1. Add 90 μL of 100 mM sodium acetate buffer at pH 4.0 (CM10 chip binding buffer) into each well in the bioprocessor set up containing the CM10 chips. 2. Add 10 μL of each ion exchange bead eluate (sample fractions 1-6) to the well for sample binding. 3. Mix well by shaking for 30 min at RT. 4. Remove the sample fractions. 5. Wash the chip spot in the well by adding 150 μL of CM10 chip binding buffer again into each well and shake for 5 min at RT. 6. Remove buffer. 7. Repeat washing (steps 5 and 6) once more. 8. Rinse with water two times by adding 200 μL of Milli-Q water to each well and discard immediately. 9. Remove the chips from bioprocessor and air-dry the chips for 5 min. 10 . Add 1 μL of 50% sinapinic acid to each chip spot. 11. Air-dry the chip for around 10 min. 12. Add 1 μL of 50% sinapinic acid and air-dry again. 13. The chip is ready to be read in the PBS IIc mass spectrometer reader. 14. Process at least one reference control serum concurrently with the patients'samples on each chip for quality control of chip-to-chip variability as before. The PBS IIc SELDI-TOF mass spectrometer reader is a laser desorption ionization time of flight mass spectrometer equipped with a pulsed ultraviolet nitrogen laser source. When the laser activates the bound serum biomarkers on the chip surface, the biomarkers become desorbed and ionized. Ionized molecules fly along the mass spectrometer to the ion detector in a time of flight manner according to their mass over charge ratio (m/z). When the ion signal is detected in the ion detector, signal processing is accomplished by high-speed analog-todigital converter linking to a computer. Detected protein biomarkers are displayed in spectral, map or gel view formats by the Ciphergen ProteinChip software 3.0.2. The following steps show how the chips are read in the mass spectrometer. e. Focus lag time at 800 ns. f. Mass deflector to Auto. g. Data acquisition method to SELDI quantitation. h. 338 laser shots per sample. 12. Click "Start Running" button from the menu bar. 13. By the time 338 laser shots are fired, the collected data is automatically saved. 14. It is important to first carry out external calibration of the equipment using the All-In-1 Peptide molecular mass standard (according to the instruction sheet from the manufacturer) to ensure mass accuracy of the spectra before running the samples. 15. Carry out baseline subtraction and calibration for the spectra. 16 . Generate labeled peak groups (clusters) across multiple spectra by the Biomarker Wizard mode. 17. Compare peak groups detected in the SARS patients with those of the controls by nonparametric two sample Mann-Whitney U-test in the Biomarker Wizard mode. 18. Biomarker Wizard operates in two passes, with the first pass uses low sensitivity settings to detect obvious and well-defined peaks and the second pass uses higher sensitivity settings to search for smaller peaks. 19. This will identify both strong and weak peaks that are significantly increased or decreased in the patients vs the controls. 4. Separating the controls into individual groups, a highly significant increase of biomarker at 11,695 is also observed in SARS patients vs each control group (Fig. 4) . patients, we further monitor the biomarker-11,695 level in a SARS patient longitudinally (Fig. 5) and correlate the biomarker level with the extent of pneumonia by chest X-ray opacity score (Fig. 2) . 6. Figure 5 illustrates, in this patient, an elevation of the biomarker-11,695 level preceding the development of pneumonia at the onset of disease (as indicated by an increasing opacity score) followed by a rapid drop of the biomarker level down to the background on recovery from the illness after stringent antiviral and steroid treatments (see Note 1 for details of clinical presentation of the patients). 7. On the other hand, Fig. 6 shows low biomarker-11,695 level during the whole monitoring period in another SARS patient who has low X-ray scores and whose clinical course is relatively mild (see Note 2). 8. By means of peptide mapping and tandem MS/MS mass spectrometry, this biomarker is identified to be Serum Amyloid A protein (SAA), which is an acute phase reactant frequently and rapidly induced in abundance at pneumonia (see ref. 15 for the protein ID results and ref. 17-20 for the techniques adopted). 9. This serial study demonstrates the potential of the biomarkers discovered by protein chip array profiling in monitoring the disease manifestation in SARS patients. Clinical and protein chip monitoring of the first SARS patient (see Fig. 5 ): one day after SARS-CoV infection was diagnosed, the radiographic score of this patient was increased from 6 to a peak value of >16 and then dropped to 12, 9, and finally to 4 demonstrating a progressive recovery. The biomarker-11,695 level by protein chip array profiling study showed an elevation that peaked earlier than the radiographic score but it then gradually subsided along with the score to a nadir when the patient was discharged. 2. Clinical and protein chip monitoring of the second SARS patient (see Fig. 6 ): the second patient had all the typical clinical symptoms of SARS-CoV infection with left upper lobe consolidation in chest radiographs. Despite negative SARS-CoV RT-PCR, titer elevation in paired anti-SARS CoV serum antibody tests performed in two laboratories confirmed the viral infection. He was managed as SARS on the next day. The clinical course was uneventful and uncomplicated. His pneumonia had never been extensive and hence his radiographic score showed just a low level of lung involvement from May 8th to 15th and then totally subsided from May 16th onwards. Biomarker-11,695 level was essentially low throughout the monitoring period. Copper (II) immobilized metal affinity capture Weak cation-exchange (CM10) ProteinChip Array (Ciphergen Biosystems, Inc.). 3. 96-Well ProteinChip bioprocessor mM Copper sulfate mM Sodium phosphate containing 0.5 mol/L NaCl (IMAC30 chip binding buffer) mM Sodium acetate buffer pH 4.0 (serves as both IMAC30 chip neutralizing buffer and CM10 chip binding buffer) Sample Exchange Dialog" button from the manu bar of the Ciphergen ProteinChip software 3.0.2 in the PBS IIc SELDI-TOF mass spectrometer reader Open Lid Place the sample treated chips into the slot of the chip chamber Close Lid" button The chip is automatically inserted into the chip chamber Sample Exchange Dialog" box, input the serial number, chip type, and chip format Chip Protocol" button from the menu bar Highest mass of detection to 200,000 Da. d. Optimal mass range of detection from m/z of 2000 to 20,000 (signals from m/z of 0-2000 are not analyzed as artifacts can be produced by energy-absorbing molecules or Data Analyses 1. As an example, Fig. 3 illustrates a comparison of multiple biomarkers in serum fraction 1 (pH 9.0 eluate) of SARS patients vs all the other controls in a gel view format in the m/z range of C) with arrows are biomarkers at m/z of 11,695, 9159, and 7784, respectively with biomarker at 11,695 being significantly increased in the SARS patients vs the controls and biomarkers 9159 and 7784 being significantly decreased Nine biomarker were found within m/z range of 4900-15,000 with highly significant increase in their normalized peak intensities in the SARS patients vs those of the controls and three biomarkers with highly significant decrease Monitoring of the biomarker-11,695 level and chest X-ray opacity score (an indicator of the extent o pneumonia) was shown in the lower left panel. The lower right panel illustrates the biomarker-11,695 in a gel view measured at different time points. CoV IgG <25 and 100, Serum SARS-Coronavirus IgG antibody titers of <1/25 and 1/100; CXR, Lung consolidation as shown by chest X-ray imaging Admitted to Intensive care unit; i.v. Ig, Intravenous injection of immune globulin Daily treatment by 500 mg × 2 of methylprednisolone; NP Swab CoV -ve, RT-PCR negative for SARS Coronavirus in nasopharyngeal swab Rectal Swab CoV +ve, RT-PCR positive for SARS Coronavirus in rectal swab Daily treatment by 400 mg Ribavirin RT-PCR positive for SARS Coronavirus in the stool; Throat Swab CoV +ve, RT PCR positive for SARS Coronavirus in throat swab. References 1. World Health Organization Viral upper respiratory infections Short term outcome and risk factors for adverse clinical outcomes in adults with severe acute respiratory syndrome (SARS) Epidemiological determinants of spread of casual agent of severe acute respiratory syndrome in Hong Kong Evaluation of reverse transcriptase-PCR assays for rapid diagnosis of severe acute respiratory syndrome associated with a novel coronavirus Primary investigation on the changing mode of plasma specific IgG antibody in SARS patients and their physicians and nurses Rapid diagnosis of a coronavirus associated with severe acute respiratory syndrome (SARS) Severe acute respiratory syndrome: radiographic review of 40 probable cases in Toronto Severe acute respiratory syndrome: radiographic and CT findings Protein biochips for differential profiling SELDI proteinchip MS: a platform for biomarker discovery and cancer diagnosis Current achievements using ProteinChip Array technology The SELDI-TOF MS approach to proteomics: protein profiling and biomarker identification Protein quantification by the SELDI-TOF-MS-based ProteinChip ® system protein chip array profiling analysis in patients with severe acute respiratory syndrome identified serum amyloid a protein as a biomarker potentially useful in monitoring the extent of pneumonia Visual quantitation and observer variation of signs of small airways disease at inspiratory and expiratory CT Peptide mapping by limited proteolysis in sodium dodecyl sulfate and analysis by gel electrophoresis Protein chemical characterization of Mucor pusillus aspartic proteinase amino acid sequence homology with the other aspartic proteinases disulfide bond arrangement and site of carbohydrate attachment Sequencing of a branched peptide using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry Probabilitybased protein identification by searching sequence databases using mass spectrometry data The authors thank our Kowloon Central Cluster/Queen Elizabeth Hospital Chief Executive, Dr. C. T. Hung and the SARS Management Committee for their support in this project. We also express our appreciation to Dr. Dominic N. C.