key: cord-332557-qm3qfvry authors: Lau, Susanna K.P.; Che, Xiao-Yan; Woo, Patrick C.Y.; Wong, Beatrice H.L.; Cheng, Vincent C.C.; Woo, Gibson K.S.; Hung, Ivan F.N.; Poon, Rosana W.S.; Chan, Kwok-Hung; Peiris, J.S. Malik; Yuen, Kwok-Yung title: SARS Coronavirus Detection Methods date: 2005-07-17 journal: Emerg Infect Dis DOI: 10.3201/eid1107.041045 sha: doc_id: 332557 cord_uid: qm3qfvry Using clinical samples from patients with severe acute respiratory syndrome, we showed that the sensitivities of a quantitative reverse transcription–polymerase chain reaction (80% for fecal samples and 25% for urine samples) were higher than those of the polyclonal (50% and 5%) and monoclonal (35% and 8%) antibody-based nucleocapsid antigen capture enzyme-linked immunosorbent assays. Fecal specimens (n = 40, from 40 patients 1-27 days after symptom onset) and urine specimens (n = 133, from 101 patients 2-57 days after symptom onset) were collected from SARS patients hospitalized in Hong Kong from March to May 2003. SARS was confirmed by the presence of serum immunoglobulin (Ig) G against SARS-CoV by an immunofluorescence assay (4) . Specimens were tested with polyclonal and monoclonal antibody-based capture ELISAs for SARS-CoV nucleocapsid protein and realtime qRT-PCR. Control urine (n = 100) and fecal (n = 100) specimens were obtained from hospitalized patients without SARS. SARS-CoV nucleocapsid protein was detected by polyclonal antibody-based ELISA according to published protocols (7, 11) . SARS-CoV nucleocapsid protein was detected by monoclonal antibody-based ELISA using a modified protocol for serum samples (10) . Briefly, fecal and urine specimens were inactivated with 2% and 0.5% phenol, respectively, for 15 min before centrifugation and dilution in phosphate-buffered saline with 2% skim milk. One hundred microliters of 1:10 diluted fecal specimens or 1:2 diluted urine specimens was added to wells previously coated with antinucleocapsid monoclonal antibodies. Plates were incubated, washed, treated with antinucleocapsid rabbit monoclonal antibodies, and analyzed as described previously (10, 11) . RNA extraction and realtime qPCR assay specific for the 1b region of SARS-CoV were conducted as described previously (3, 9) . We compared the detection rates of 2 ELISAs and realtime qRT-PCR using the McNemar test and studied the correlation between the optical density values at 450 nm (OD 450 ) of the 2 ELISAs and log 10 viral concentrations, as determined by real-time qRT-PCR, by linear regression (SPSS version 11.0, SPSS Inc., Chicago, IL, USA). A p value <0.05 was regarded as significant. A comparison of the 2 ELISAs is shown in the Figure and Table 1 . The cutoffs of the polyclonal antibody-based ELISA have been determined previously, with specificities of 96% and 99% for fecal and urine specimens, respectively (11) . The baselines of the monoclonal antibody-based ELISA were determined by using 100 control fecal and urine specimens, with mean OD 450 values of 0.089 and 0.05 and standard deviation (SD) values of 0.074 and 0.03, respectively. The specificities of the monoclonal antibody-based ELISA were 93% for fecal specimens and 98% for urine specimens, as determined using cutoffs defined as the mean + 2 SD. Of 40 fecal samples obtained from SARS patients, 20 (50%) obtained on days 9 to 23 after onset of symptoms were positive by the polyclonal antibody-based ELISA, and 14 (35%) obtained on days 2 to 21 were positive by the monoclonal antibody-based ELISA. Of 133 urine samples, 6 (5%) obtained on days 16 to 32 after onset of symptoms were positive by the polyclonal antibody-based ELISA, and 11 (8%) obtained on days 6 to 45 were positive by the monoclonal antibody-based ELISA. Results of the polyclonal antibody-based ELISA were comparable with our previous data on different specimens (11) . The OD 450 values of both fecal (Pearson correlation 0.610, p<0.0005) and urine specimens (Pearson correlation 0.475, p<0.0005) detected by the 2 ELISAs were significantly correlated. The method of choice for early diagnosis of SARS-CoV infection should be the qRT-PCR. The sensitivity of qRT-PCR is superior to that of both ELISAs. Moreover, qRT-PCR can detect SARS-CoV earlier in fecal specimens (Tables 1 and 2 ). Among the 40 fecal samples from SARS patients, 32 (80%) were positive by qRT-PCR, which was significantly higher than that of the polyclonal (50%) and monoclonal (35%) antibody-based ELISAs (McNemar test, p<0.005 and p<0.001, respectively). Of the 133 urine samples from SARS patients, 33 (25%) were positive by qRT-PCR, which was also significantly higher than that of the polyclonal (5%) and monoclonal (8%) antibody-based ELISAs (McNemar test, p<0.001 for both comparisons). When qRT-PCR was used as a standard, the sensitivities of the polyclonal and monoclonal antibody-based ELISAs were 53.1% (17/32) and 43.8% (14/32) in fecal specimens, and 12.1% (4/33) and 15.2% (5/33) in urine specimens, respectively. The qRT-PCR can detect SARS-CoV in fecal specimens obtained on days 1 to 27 after onset of symptoms and in urine specimens obtained on days 9 to 45. Moreover, 6 (75%) of the 8 fecal specimens obtained on days 1 to 10 were positive by qRT-PCR. All 3 tests had the highest detection rates in fecal specimens collected on days 16 to 20, which suggested that this was the period of peak viral shedding in stool. The detection rates in urine specimens were much lower than those in fecal specimens in all 3 assays. SARS-CoV can be detected during the late phase of illness. Since SARS-CoV cannot be readily isolated from SARS patients after week 3 of illness (14) , the detection of SARS-CoV beyond this time may be due to prolonged shedding of nonviable viruses in these patients or the presence of neutralizing immunoglobulins in clinical specimens, which has prevented viral replication in cell cultures. SARS-CoV RNA concentration and ELISA results were correlated. Higher detection rates by both ELISAs were found in specimens with higher viral concentrations ( Table 2 ). There was also a significant correlation between viral load and ELISA OD 450 values in fecal specimens tested with the monoclonal antibody-based ELISA (Pearson correlation 0.424, p = 0.003), and in urine specimens tested with both the polyclonal and monoclonal antibody-based ELISAs (Pearson correlation 0.386 and 0.331, respectively, p<0.0005 in both analysis). Although the correlation between viral load and ELISA OD 450 values in fecal specimens tested with the polyclonal antibody-based ELISA was not significant, there was a trend for such a correlation (Pearson correlation 0.229, p = 0.078). In this study, fecal and urine samples were used because they are easier and safer to obtain and more readily available. In our previous reports, nucleocapsid protein was detected by the polyclonal antibody-based ELISA in 83% of nasopharyngeal aspirates collected on days 11 to 15 after symptom onset and by the monoclonal antibody-based ELISA in 85% of serum obtained during the first 10 days (10, 11) . These findings suggest that ELISA may be more useful when used with nasopharyngeal aspirate and serum specimens. However, these specimens were not included in the current study because only small amounts were available. Similar studies should be conducted if such samples are available. A novel coronavirus associated with severe acute respiratory syndrome Clinical progression and viral load in a community outbreak of coronavirus-associated SARS pneumonia -a prospective study The severe acute respiratory syndrome Coronavirus as a possible cause of severe acute respiratory syndrome Detection of specific antibodies to SARS coronavirus nucleocapsid protein for serodiagnosis of SARS coronavirus pneumonia Longitudinal profile of immunoglobulin G (IgG), IgM, and IgA antibodies against SARS coronavirus nucleocapsid protein in patients with pneumonia due to the SARS coronavirus Relative rates of non-pneumonic SARS coronavirus infection and SARS coronavirus pneumonia Isolation and characterization of viruses related to the SARS coronavirus from animals in southern China Rapid diagnosis of a coronavirus associated with severe acute respiratory syndrome (SARS) Sensitive and specific monoclonal antibody-based capture enzyme immunoassay for detection of nucleocapsid antigen in sera from patients with severe acute respiratory syndrome Detection of SARS coronavirus nucleocapsid protein in SARS patients by enzyme-linked immunosorbent assay Early diagnosis of human immunodeficiency virus infection in children less than 6 months of age: comparison of polymerase chain reaction, culture, and plasma antigen capture techniques Sensitivity of HCV core antigen and HCV RNA detection in the early infection phase Detection of SARS coronavirus in patients with suspected SARS