key: cord-0684982-p3lb44xm authors: Kassis, Nicholas; Tanaka-Esposito, Christine; Chung, Roy; Kalra, Ankur; Shao, Mingyuan; Kumar, Ashish; Alzubi, Jafar; Chung, Mina K.; Khot, Umesh N. title: Validating and implementing cardiac telemetry for continuous QTc monitoring: A novel approach to increase healthcare personnel safety during the COVID-19 pandemic date: 2021-04-27 journal: J Electrocardiol DOI: 10.1016/j.jelectrocard.2021.04.014 sha: cdf373332db63aa41e7c9c8bb34d9f527f0e76dd doc_id: 684982 cord_uid: p3lb44xm BACKGROUND: Minimizing direct patient contact among healthcare personnel is crucial for mitigating infectious risk during the coronavirus disease 2019 (COVID-19) pandemic. The use of remote cardiac telemetry as an alternative to 12‑lead electrocardiography (ECG) for continuous QTc monitoring may facilitate this strategy, but its application has not yet been validated or implemented. METHODS: In the validation component of this two-part prospective cohort study, a total of 65 hospitalized patients with simultaneous ECG and telemetry were identified. QTc obtained via remote telemetry as measured by 3 independent, blinded operators were compared with ECG as assessed by 2 board-certified electrophysiologists as the gold-standard. Pearson correlation coefficients were calculated to measure the strength of linear correlation between the two methods. In a separate cohort comprised of 68 COVID-19 patients treated with combined hydroxychloroquine and azithromycin, telemetry-based QTc values were compared at serial time points after medication administration using Friedman rank-sum test of repeated measures. RESULTS: Telemetry-based QTc measurements highly correlated with QTc values derived from ECG, with correlation coefficients of 0.74, 0.79, 0.85 (individual operators), and 0.84 (mean of all operators). Among the COVID-19 cohort, treatment led to a median QTc increase of 15 milliseconds between baseline and following the 9th dose (p = 0.002), with 8 (12%) patients exhibiting an increase in QTc ≥ 60 milliseconds and 4 (6%) developing QTc ≥ 500 milliseconds. CONCLUSIONS: Cardiac telemetry is a validated clinical tool for QTc monitoring that may serve an expanding role during the COVID-19 pandemic strengthened by its remote and continuous monitoring capability and ubiquitous presence throughout hospitals. The coronavirus disease 2019 (COVID- 19) pandemic has transformed individual lives and societal framework on a global scale, and in no other sector is this more evident than in healthcare. As healthcare personnel have faced a significantly higher risk of infection, particularly in the early stages of the outbreak 1-2 , mitigating such susceptibility is vital for the overall health of the healthcare delivery system and to avoid nosocomial transmission. The use of personal protective equipment (PPE) by those with the greatest exposure risk, implementation of infection control strategies, and rigorous compliance with such guidelines have reduced rates of infection in healthcare personnel. Yet despite such measures, the relatively long incubation phase coupled with the variable virulence [3] [4] [5] , speak to the insidious nature of SARS-CoV-2. Delivering necessary medical care to all hospitalized patients while minimizing individual risk among healthcare personnel has been a major challenge during this time. Unprecedented multilevel disruption of supply chain networks has led to severe shortages of vital supplies such as PPE 6 , in turn hampering efforts and transiently limiting the delivery of services only to those deemed medically essential. Physical distancing is a powerful protective mechanism to reduce the transmission of SARS-CoV-2 7 , and therefore, novel strategies to minimize contact between healthcare personnel and patients while still providing essential care universally must be implemented in order to reopen the healthcare system. Opportunity for such an intervention presented early in the outbreak when small observational studies described the use of hydroxychloroquine and azithromycin to treat infected patients. [8] [9] [10] Both drugs are known to increase the risk of drug-induced Torsades de pointes and sudden cardiac death in at-risk patients [11] [12] [13] , thereby necessitating monitoring of the QT interval. 14 While the use of these therapies has fallen out of favor, the significance of continuous J o u r n a l P r e -p r o o f Journal Pre-proof QTc monitoring in COVID-19 patients remains paramount as prior work reveals that QTc strongly predicts in-hospital mortality in this cohort 15 , and that irrespective of QTc-prolonging medications, infected patients have longer baseline QTc intervals than the general population. 16 The available strategies for in-hospital QTc monitoring are suboptimal for the current context. Although 12-lead electrocardiography (ECG) has traditionally been used for this purpose, it requires direct contact between the caregiver and the patient, and obtaining multiple readings in high-risk patients is inefficient and labor intensive. While the first smartphoneenabled personal ECG, AliveCor's KardiaMobile 6L, was issued allowance by the Food and Drug Administration for QTc monitoring in response to COVID-19, this technology relies on user involvement and routine use in hospitalized patients would necessitate considerable upfront expense. 17 In addition, though the use of a wearable remote monitoring system, the BodyGuardian, has previously identified patients at risk for drug-induced long QT syndrome, this was only validated in the out-of-hospital setting. 18 Cardiac telemetry offers several advantages, including 1) data acquisition that demands no action on the part of the patient or the caregiver, 2) remote capability for data analysis, 3) continuous monitoring for immediate detection of cardiac arrhythmia, and 4) ubiquitous presence in all hospitals worldwide. In our two-part study, we sought to first validate a "no-touch" method using cardiac telemetry for serial tracking of QTc intervals, and second, implement telemetry in a clinical QTc intervals in hospitalized patients. The protocol was approved by the Cleveland Clinic Institutional Review Board. Phillips real-time cardiac telemetry was used for QT interval measurement and compared with traditional standard 12-lead ECG. The Philips real-time cardiac telemetry system digitized the signal at a sample rate of 500 samples/second and a high-pass frequency limit of 0.05 Hz. Utilizing 5 surface electrodes, the system provides 7 ECG waveform leads: limb (I, II, III), augmented (aVL, aVF, aVR), and precordial lead (V1). Enabled by the AirStrip ONE integrative platform, patient waveforms were accessed virtually, obviating the need for direct patient contact. Automated QTc intervals normalized for heart rate were not available via telemetry at the time of our study, and therefore, QTc values were calculated from measured QT intervals obtained using freeze capture of rhythm ECG strips. Using only lead II, QT intervals were manually measured from the onset of the Q wave to the offset of the T wave, defined by the threshold method as the point at which the T wave merges with the isoelectric baseline. [19] [20] [21] QTc intervals were then calculated using the following approaches: 1) Bazzett equation, correcting the QT with the RR interval of the preceding beat, if the heart rate was 60-100 bpm, 2) Framingham equation if the heart rate was <60 or >100 bpm, 3) average of multiple QT measurements in the presence of atrial fibrillation, and 4) correction of the QTc, by subtracting the difference in QRS lengthening, in the presence of a wide QRS (>120 ms). 22 Sample size estimation was made prior to conducting the study to achieve a power of 0.9 with a one-sided alpha of 0.025. Accordingly, a total of 65 patients hospitalized at Cleveland Clinic Fairview Hospital between May 4-8, 2020, with a need for 12-lead ECG for any indication and available concurrent cardiac telemetry, were identified at random. Using only lead J o u r n a l P r e -p r o o f Journal Pre-proof II, three operators (TS, TF, CL) independently and in a blinded fashion, sampled three random QT interval measurements for each patient from a remotely accessed digital telemetry rhythm strip coinciding in time )±60 seconds) with the time-date-stamped 12-lead ECG. Telemetry waveforms were displayed at 25 mm/sec sweep speed and 10 mm/mV voltage scale. QTc values were subsequently calculated from each of the three measured QT intervals and reported as a mean value per individual operator and collectively. QTc values were similarly derived from 12lead ECG. Manual measurements of QT intervals were independently performed by two boardcertified electrophysiologists (RC, CT), using lead II displayed at 25 mm/sec sweep speed and 10 mm/mV voltage scale. When the offset of the T wave could not be defined with certainty in lead II, an alternative ECG lead (V5 or V6) was used for QT interval measurement. 23 QTc values were calculated, reported as a mean, and used as a gold-standard for comparison. To assess the accuracy in the telemetry-based QTc acquisition method, we compared QTc values obtained simultaneously via remote telemetry and 12-lead ECG. To evaluate the precision of the telemetry-based QTc approach, mean QTc values by each of the operators were compared for inter-rater reliability. Utilizing telemetry-based remote monitoring, serial QTc intervals were compiled for analysis in a cohort of 68 symptomatic patients with COVID-19 treated with combined hydroxychloroquine and azithromycin. This population was identified by reviewing the Figure 1 ). For all patients, QTc values were recorded at baseline and 2 hours after each drug dose administration. This timing was selected based upon previously reported peak plasma concentrations of hydroxychloroquine and azithromycin at 2-4 hours. [24] [25] QTc monitoring continued throughout hospitalization while being monitored via telemetry. To assess the ability of telemetry to detect the applied effect of hydroxychloroquine and azithromycin on QTc intervals, mean QTc values were compared at serial time points of drug administration. Continuous variables are expressed as mean ± standard deviation. Normal Q-Q plots and normality tests showed that QTc derived from both telemetry and 12-lead ECG were normally distributed. A comparison of telemetry-based QTc measurements was made with the goldstandard QTc. Pearson correlation coefficient was calculated to measure the strength of linear correlation. Agreement was quantified utilizing Lin's concordance correlation coefficient. Concordance correlation coefficients and associated 95% confidence intervals (CI) were Mean QTc measurements obtained by telemetry and 12-lead ECG are reported in Table 1 . Telemetry-based QTc measurements highly correlated with gold-standard QTc values derived incorporation of the U wave into the QT interval; both data sets were removed from the analysis. Equivalence tests for paired samples comparing mean differences in QTc values between the gold-standard and all telemetry operators, with predefined equivalence margins of -25 to 25 milliseconds, are shown in Table 2 . The null hypothesis of nonequivalence was rejected (p<0.001) in all cases, indicating that mean QTc differences did not exceed the preset margin. In the cohort of patients with COVID-19 treated with combination hydroxychloroquine and azithromycin and who underwent telemetry-based remote monitoring (n=68), serial QTc intervals were compiled for analysis. The median QTc at baseline and following each dose of hydroxychloroquine and azithromycin or hydroxychloroquine alone are presented in Figure 2 , with a statistically significant increase in QTc of 15 milliseconds [interquartile range (-2, 41)] between baseline and following the 9th dose (p=0.002) by Friedman rank-sum tests of repeated measures. Figure 3 reports both the parallel coordinates plot for telemetry-measured QTc at baseline and following each dose of hydroxychloroquine and azithromycin, as well as box plots indicating the distribution of differences between repeated measures of QTc with serial doses. Significant differences in QTc were found after the 9 th drug dose administration compared with the QTc at baseline (p=0.02) and following the 1 st (p=0.004), 2 nd (p=0.004), 4 th (p=0.006), 5 th (p=0.01), 6 th (p=0.02), and 7 th doses (p=0.02), separately. Additionally, eight (12%) patients showed an increase in QTc ≥ 60 milliseconds, and four (6%) developed QTc ≥ 500 milliseconds. Using a novel "no-touch" telemetry-based approach to QTc monitoring, our primary findings are twofold: 1) telemetry-based QTc measurements are highly accurate relative to goldstandard 12-lead ECG, and precise among three trained independent operators competent in measuring cardiac intervals; 2) initially validated in a general hospitalized population, this method can be successfully implemented in the clinical setting as demonstrated by its utilization in COVID-19 patients who experienced a significant increase in QTc after treatment with hydroxychloroquine and azithromycin. Our results lend credence to telemetry as a valid clinical tool for QTc monitoring in place of standard 12-lead ECG, thereby minimizing direct patient contact and mitigating infectious risk among healthcare personnel and patients alike. Notwithstanding perfect agreement, the mean QTc difference of 11.6 ± 19.3 milliseconds between telemetry and ECG in our study is inconsequential, coinciding with less than half of a millimeter measured on standard ECG paper speed (25 millimeter/second). Moreover, this discrepancy is in line with prior studies comparing QTc monitoring using wireless smartphone ECG acquisition technology (AliveCor KardiaMobile) and 12-lead ECG. [26] [27] Collectively, our findings support telemetry as a method for QTc monitoring in place of standard 12-lead ECG. In the separate prospective cohort analysis applying the validated method for temporal QTc monitoring in COVID-19 patients, our results are consistent with prior studies that demonstrate a significant increase in the QTc interval with treatment 28-32 , further supporting the clinical application of telemetry-based monitoring, particularly in cohorts posing exposure risk and who warrant minimal direct contact. COVID-19 has taken a dramatic toll on the health of societies worldwide. At the onset of the outbreak, the initial response by health systems was to reduce the intensity of care delivery. Although transient and necessary, an unintended outcome was subsequent deferment in health care needs. The Kaiser Family Foundation poll reported that 48% of Americans delayed medical care due to the pandemic, of whom 11% noted worsening in their medical condition. 33 In another survey by the American College of Emergency Physicians, 80% feared contracting COVID-19 from visiting the emergency room, while 29% avoided seeking medical care specifically because of these concerns. 34 Centers around the world observed a significant reduction in hospitalizations for acute myocardial infarction and ischemic stroke, and consequently reported adverse yet potentially avoidable consequences. [35] [36] [37] [38] [39] Thus, eliminating the provision of healthcare services is not a sustainable solution during the COVID-19 pandemic. In its current state, the primary goal of the health system is to restore the public's confidence and sense of security in seeking medical care during the COVID-19 pandemic; prioritizing healthcare personnel safety is at the forefront of rebuilding this trust. The Centers for Disease Control and Prevention issued updated recommendations for infection prevention and control for healthcare personnel, with emphasis on universal screening, PPE use, and physical distancing. 40 While avoiding exposure entirely is the most effective means of mitigating risk, this is not practical, and potentially harmful in the clinical setting. Instead, it is incumbent upon the healthcare community to seek and maximize opportunities to limit non-essential physical contact between healthcare personnel and hospitalized patients. We have successfully demonstrated a novel "no-touch" cardiac telemetry-based approach for QTc monitoring in COVID-19-infected patients treated with drugs with arrhythmogenic potential. Importantly, we understood the sharp decline in the utility of these agents to treat the infection, and instead aimed to test the ability to clinically implement telemetry as QTc monitoring becomes increasingly critical [15] [16] and for valid options in future unforeseen circumstances. While the primary application of cardiac telemetry in current practice is arrhythmia detection in hospitalized patients, we believe it holds greater potential and expanding clinical use. Our focus on telemetry as a non-contact method for QTc monitoring in patients with COVID-19 treated with proarrhythmic drugs stands as an important application. Similarly, telemetry may further replace serial 12-lead ECGs used in antiarrhythmic drug loading protocols, allowing safe uninterrupted delivery of necessary elective services throughout the duration of the pandemic. Furthermore, telemetry holds great promise as a tool for predicting ventricular arrhythmias. Heightened QTc variability has been described in certain pathophysiologic states and shown to predict increased risk for potentially lethal arrhythmias and cardiovascular death. 41- J o u r n a l P r e -p r o o f Journal Pre-proof 49 The continuous recording nature of telemetry lends itself to advanced diagnostics that can provide granular insight beyond mere static measurement of QTc, and may offer automated serial analysis of simultaneous ECG leads as previously demonstrated by digital Holter recordings. 50 With increasing metadata that necessitates a standardized exchange between a growing number of ECG devices, the use of digital and continuous recordings is paramount for accurate and precise ECG measurements such as QTc. The advent of the internationallyrecognized Standard Communication Protocol for Computer-Assistant ECG, for instance, has facilitated open, secure storage and exchange of full or select ECG sequences between operating systems, in turn optimizing the interpretation and serial analysis of large data, and ultimately allowing for more reliable and efficient diagnostics. 51 The main limitation of the method validated and employed in our study is the laborintensive nature of manual QT interval measurements and QTc calculations. However, the handoperated nature of our method with attention to use the most appropriate QT correction formula in each instance, resulted in highly accurate QTc values that were made readily available and easily applicable by all caregivers irrespective of experience or knowledge in the nuances of QTc determination. The future of automated QTc via telemetry holds great promise if algorithms consider the application of the most appropriate correction formula in a given circumstance. J o u r n a l P r e -p r o o f Epidemiology of and risk factors for coronavirus infection in health care workers: a living rapid review Risk of COVID-19 among frontline healthcare workers and the general community: a prospective cohort study Early transmission dynamics in Wuhan, China, of novel coronavirus-infected pneumonia SARS-CoV-2 viral load in upper respiratory specimens of infected patients Pathogenicity and transmissibility of 2019-nCoV-A quick overview and comparison with other emerging viruses Navigating healthcare supply shortages during the COVID-19 pandemic: a cardiologist's perspective on behalf of the COVID-19 Systematic Urgent Review Group Effort (SURGE) study authors. Physical distancing, face masks, and eye protection to prevent person-to-person transmission of SARS-CoV-2 and COVID-19: a systematic review and meta-analysis Hydroxychloroquine and azithromycin as a treatment of COVID-19: results of an openlabel non-randomized clinical trial A pilot study of hydroxychloroquine in treatment of patients with common coronavirus disease-19 (COVID-19) Efficacy of hydroxychloroquine in patients with COVID-19: results of a randomized clinical Chronic hydroxychloroquine use associated with QT prolongation and refractory ventricular arrhythmia Life threatening severe QTc prolongation in patient with systemic lupus erythematosus dues to hydroxychloroquine. Case Rep Cardiol Azithromycin and the risk of cardiovascular death Urgent guidance for navigating and circumventing the QTc-prolonging and torsadogenic potential of possible pharmacotherapies for Coronavirus Disease-19 (COVID-19) San Matteo COVID Cardiac Injury Task Force. QTc interval and mortality in a population of SARS-2-COV infected patients Interleukin-6 inhibition of hERG underlies risk for acquired long QT in cardiac and systemic inflammation Wireless smartphone ECG enables large-scale screening in diverse populations A wearable remote monitoring system for the identification of subjects with a prolonged QT interval or at risk for drug-induced long QT syndrome Intra-and interreader variability in QT interval measurement by tangent and threshold methods in a central electrocardiogram laboratory New formula for evaluation of the QT interval in patients with left bundle branch block The measurement of the QT interval Zithromax (azithromycin for injection Wireless smartphone ECG enables large-scale screening in diverse populations Comparison of QT interval readings in normal sinus rhythm between a smartphone heart monitor and a 12-lead ECG for healthy volunteers and in patients receiving sotalol or dofetilide Assessment of QT intervals in a case series of patients with coronavirus disease 2019 (COVID-19) infection treated with hydroxychloroquine alone or in combination with azithromycin in an intensive care unit The QT interval in patients with COVID-19 treated with hydroxychloroquine and azithromycin Arrhythmic profile and 24 hour QT interval variability in COVID-19 patients treated with hydroxychloroquine and azithromycin Public Poll: Emergency Care Concerns Amidst COVID-19 Impact of the COVID-19 epidemic on stroke care and potential solutions Collateral effect of Covid-19 on stroke evaluation in the United States Reduction of hospitalization for myocardial infarction in Italy in the COVID-19 era The COVID-19 pandemic and the incidence of acute myocardial infarction Out-of-hospital cardiac arrest during the COVID-19 outbreak in Italy Interim Infection Prevention and Control Recommendations for Health Care Personnel During the Coronavirus Disease 2019 (COVID-19) Pandemic, updated Beat-tobeat QT interval variability: novel evidence for repolarization lability in ischemic and nonischemic dilated cardiomyopathy Age and QT variability index during free breathing, controlled breathing and tilt in patients with chronic heart failure and healthy control subjects Beat-to-beat QT interval variability associated with acute myocardial ischemia Temporal repolarization lability in hypertrophic cardiomyopathy caused by beta-myosin heavychain gene mutations QT-interval variability and autonomic control in hypertensive subjects with left ventricular hypertrophy Temporal repolarization lability differences among genotyped patients with the long QT syndrome Increased QT variability in patients with panic disorder and depression QT interval variability and spontaneous ventricular tachycardia or fibrillation in the Multicenter Automatic Defibrillator Implantation Trial (MADIT) II patients QT variability strongly predicts sudden cardiac death in asymptomatic subjects with mild or moderate left ventricular systolic dysfunction: a prospective study Electrocardiographic identification of drug-induced QT prolongation: assessment by different recording and measurement methods 0: an enhanced Standard Communication Protocol for Computer-assisted Electrocardiography J o u r n a l P r e -p r o o f