key: cord-0709783-oqgwol61 authors: Tai, Wei-Chun; Chang, Yu-Chi; Chou, Dean; Fu, Lung-Ming title: Lab-on-Paper Devices for Diagnosis of Human Diseases Using Urine Samples—A Review date: 2021-08-03 journal: Biosensors (Basel) DOI: 10.3390/bios11080260 sha: 0e2f01ea67e55f41a2fd977549e3677d4ee1d106 doc_id: 709783 cord_uid: oqgwol61 In recent years, microfluidic lab-on-paper devices have emerged as a rapid and low-cost alternative to traditional laboratory tests. Additionally, they were widely considered as a promising solution for point-of-care testing (POCT) at home or regions that lack medical infrastructure and resources. This review describes important advances in microfluidic lab-on-paper diagnostics for human health monitoring and disease diagnosis over the past five years. The review commenced by explaining the choice of paper, fabrication methods, and detection techniques to realize microfluidic lab-on-paper devices. Then, the sample pretreatment procedure used to improve the detection performance of lab-on-paper devices was introduced. Furthermore, an in-depth review of lab-on-paper devices for disease measurement based on an analysis of urine samples was presented. The review concludes with the potential challenges that the future development of commercial microfluidic lab-on-paper platforms for human disease detection would face. Urine is a by-product of kidney metabolism and is rich in many nitrogen-containing substances, including urea, uric acid and creatinine, which are excreted from the body as water-soluble chemicals during urination. The urine volume of normal healthy adults ranges from 0.6 to 2.6 L per day, where approximately 91-96% of this urine is composed of water. However, urine also contains various inorganic salts and organic compounds, such as proteins, hormones, and metabolites. The chemical composition of fresh urine mainly consists of nitrogen, ammonium, ammonia nitrogen, nitrate, nitrite, phosphorus, potassium, sulfate, sodium, magnesium, chloride, and calcium. Moreover, the urine of healthy individuals is clear or light yellow in color. However, in the presence of certain diseases or disorders, such as hematuria, diabetes, or kidney stones, distinct changes in the color, composition or smell of urine may occur. Therefore, urine serves as an important bio-rich resource for health monitoring [1, 2] . Unlike blood, urine is a non-invasive sample that can be easily collected without pain, or the need for special equipment. As a result, it has significant potential for point-of-care testing (POCT) or home health monitoring and diagnosis. However, current urinalysis diagnosis techniques still need the use of sophisticated laboratory apparatus and skilled personnel, which precludes their use in the home or in undeveloped areas of the world with poor medical infrastructure and lack of resources. Consequently, lab-on-paper diagnostic platforms have aroused great interest in recent years. Compared to traditional In many cases, it is impossible to process and analyze testing samples immediately on collection due to insufficient detection facilities, limited professional technicians, unfavorable on-site conditions, and so on. Consequently, the samples must be stored in some way before being transported to the testing laboratory for and testing and analysis. Traditional sample collection and storage methods usually require complicated procedures and/or large and expensive equipment. Thus, the use of paper chips, suitably modified through surface treatments such as nitrification and/or the addition of appropriate surfactants has attracted growing interest as an alternative sample collection and storage technology in recent years [39, 40] . For example, fast technical analysis (FTA) cards have now been successfully commercialized for the collection and storage of blood samples, urine samples, plasmids, cells, and viruses [41] [42] [43] [44] . In addition, many lab-on-paper platforms have also been developed for the collection and storage of blood, urine, saliva, sweat and tear samples [45] [46] [47] [48] [49] . For instance, Shay et al. developed a wearable lab-on-paper platform with a capillary-evaporative transport function for the collection of sweat samples for long-term measurement and analysis [49] . As shown in Figure 1a , typical paper-based sample collection and storage devices comprise three parts, namely a region for storing the sample fluid, a closed paper channel for transporting the sample fluid, and a paper pad with a large surface area for driving the sample fluid continuously through the device under the effects of evaporation. For many lab-on-paper applications, it is necessary to pre-treat the raw sample prior to analysis using separation, purification and extraction methods, in order to minimize the effects of interference errors and possible diagnostic distortion. For example, whole blood samples comprise a mixture of red blood cells (RBCs), white blood cells (WBCs), platelets, and plasma, and if the plasma is not properly separated from the whole blood, the determination of the detection signal during the detection process may be severely impaired. Accordingly, many lab-on-paper platforms for the separation of whole blood samples have been proposed in recent years. Generally speaking, these devices perform the separation process using filter paper [50, 51] , filtration membranes [52] [53] [54] [55] , chitosan polymer structure separation [56] , or combined dielectrophoretic (DEP) and capillary force separation [57] . For example, Laurenciano et al. [51] fabricated a sliding hybrid PMMA/paper microchip with an integrated separation membrane to filter urine samples or separate whole blood (Figure 1b) . The feasibility of the proposed device was demonstrated by filtering human urine samples to measure the concentration of total protein (TP) and separating whole blood samples to measure the concentrations of albumin (ALB) and creatine (CRE), respectively. In the detection process, the extraction step always requires separation of the target analyte (e.g., cells, virus, ions, proteins, DNA, drugs, RNA, and so on) from the collected sample (e.g., biological, environmental, or food) for subsequent measurement and analysis. Many paper devices have been proposed for sample extraction [58] [59] [60] [61] [62] [63] [64] [65] [66] [67] [68] , including commercial FTA and FTA elute cards [41, 43] . However, while these cards are easy to use, their integration with lab-on-paper platforms still presents a significant challenge. Accordingly, many studies have attempted to implement the extraction function on the paper platform itself extraction device of lab-on-paper can be used for DNA, RNA, dug, heavy metal ions, and pesticide residues, and so on [61] [62] [63] [64] [65] [66] [67] [68] . For instance, Batule et al. [63] presented a lab-on-paper platform based on nucleic acid testing for the rapid extraction and detection of virus RNA from blood serum samples. The device consisted of an assembly of seven pads mounted on a backing card, namely a washing pad, three transfer pads, a sample pad, a binding pad and a wicking pad (see Figure 1c ). The binding pad was modified by DNA probes designed to capture SS virus RNA, and the RNA extraction process involved a lysis of the virus particles, followed by hybridization and elution. It was shown that the proposed platform successfully extracted the RNA from various viruses, including Zika virus, dengue fever and chikungunya fever, within 5 min. To enhance the detection limit of paper-based platforms, it is frequently necessary to perform some form of sample preconcentration or amplification procedure prior to the analysis and detection process. Sample pre-concentration methods typically employ ion concentration polarization (ICP) [69, 70] , isotachophoresis (ITP) [71, 72] , electrokinetic stacking (EKS) [73, 74] , or multiplex electroanalytical techniques [75] . Meanwhile, sample amplification methods generally utilize loop-mediated isothermal amplification (LAMP) [76] [77] [78] [79] [80] , catalytic hairpin assembly (CHA) amplification [81, 82] , or hybrid chain reaction (HCR) [82, 83] . Cai et al. [84] presented a lab-on-paper device for the preconcentration of urine microalbuminuria (MAU) proteins by the coalesced effects of a pH gradient and an electric field. The proposed device achieved a 100-fold amplification of ALB in artificial urine medium within 70 s (see Figure 1d ). Moreover, the detection results obtained for the MAU content of urine samples collected from real diabetic patients were found to be in excellent agreement with those obtained using a traditional immunoturbidimetry method. and pesticide residues, and so on [61] [62] [63] [64] [65] [66] [67] [68] . For instance, Batule et al. [63] presented a labon-paper platform based on nucleic acid testing for the rapid extraction and detection of virus RNA from blood serum samples. The device consisted of an assembly of seven pads mounted on a backing card, namely a washing pad, three transfer pads, a sample pad, a binding pad and a wicking pad (see Figure 1c ). The binding pad was modified by DNA probes designed to capture SS virus RNA, and the RNA extraction process involved a lysis of the virus particles, followed by hybridization and elution. It was shown that the proposed platform successfully extracted the RNA from various viruses, including Zika virus, dengue fever and chikungunya fever, within 5 min. To enhance the detection limit of paper-based platforms, it is frequently necessary to perform some form of sample preconcentration or amplification procedure prior to the analysis and detection process. Sample pre-concentration methods typically employ ion concentration polarization (ICP) [69, 70] , isotachophoresis (ITP) [71, 72] , electrokinetic stacking (EKS) [73, 74] , or multiplex electroanalytical techniques [75] . Meanwhile, sample amplification methods generally utilize loop-mediated isothermal amplification (LAMP) [76] [77] [78] [79] [80] , catalytic hairpin assembly (CHA) amplification [81, 82] , or hybrid chain reaction (HCR) [82, 83] . Cai et al. [84] presented a lab-on-paper device for the preconcentration of urine microalbuminuria (MAU) proteins by the coalesced effects of a pH gradient and an electric field. The proposed device achieved a 100-fold amplification of ALB in artificial urine medium within 70 s (see Figure 1d ). Moreover, the detection results obtained for the MAU content of urine samples collected from real diabetic patients were found to be in excellent agreement with those obtained using a traditional immunoturbidimetry method. [49] . Copyright 2020 AIP Publishing. (b) Schematic diagram of PMMA/paper microchip for sample separation. Reprinted with permission from ref. [51] . Copyright 2021 Elsevier. (c) Schematic illustration showing working principle of lateral-flow-based extraction of viral RNA on paper device. Reprinted with permission from ref. [63] . Copyright 2020 Elsevier. (d) Schematic illustration showing working principle and neutralization reaction of double E/pH gradient concentrator. Reprinted with permission from ref. [84] . Copyright 2020 Elsevier. Reprinted with permission from ref. [49] . Copyright 2020 AIP Publishing. (b) Schematic diagram of PMMA/paper microchip for sample separation. Reprinted with permission from ref. [51] . Copyright 2021 Elsevier. (c) Schematic illustration showing working principle of lateral-flow-based extraction of viral RNA on paper device. Reprinted with permission from ref. [63] . Copyright 2020 Elsevier. (d) Schematic illustration showing working principle and neutralization reaction of double E/pH gradient concentrator. Reprinted with permission from ref. [84] . Copyright 2020 Elsevier. Urine samples are easy to obtain and have particular biological richness (including urea, amino acids, chlorides, potassium, phosphates, sodium, sulfates and other trace chemical elements). A comparison of detection using human samples is shown in Table 1 . Furthermore, the average person produces a large quantity of urine every day (400 to 2000 mL). Consequently, urine represents an ideal non-invasive sample for biomarker detection purposes [85] . For example, abnormal quantities of urinary oxalate and calcium can provide an indication of kidney stones [86] , while excessive quantities of urinary glucose and protein may be a sign of diabetes [87] . Similarly, high levels of urinary serotonin can be an early indicator of melancholia [88] . The literature thus contains many proposals for lab-on-paper platforms based on urine samples for the detection of diabetes, kidney disease, liver and heart function disorders, pregnancy, drugs, heavy metals, and so on. Diabetes is a chronic metabolic abnormality characterized by elevated blood sugar levels over a long period of time. When the blood sugar level exceeds 180 mg/dL, glucose appears in the urine, causing a so-called glucose in urine (GLU) disorder [89, 90] . GLU is easily detected through simple urine tests and is regarded as a reliable indicator of early diabetes and kidney disease [91] . Li et al. [92] developed a paper-based electrochemical device (PED) for the detection of glucose in artificial urine, in which the carbon working electrode was patterned using a pressure-assisted ballpoint pen and the silver reference electrode was deposited directly on the paper. The PED was used as a test strip in combination with a commercial blood glucometer and was found to achieve a linear glucose detection range of 2.0 to 20 mM with a LOD of 2.0 mM. Chen et al. [93] developed a handheld paper-based bipolar electrode-electrochemiluminescence (P-BPE-ECL) chip for the detection of glucose in phosphate-buffered saline (PBS) solution and artificial urine samples. As shown in Figure 2a , the BPE incorporated an electronic conductor between the anode and the cathode and produced oxidation and reduction reactions under the effects of an external DC voltage when immersed in electrolyte solution. In the proposed detection method, the glucose level was determined by the P-BPE-ECL assay. The experimental results showed that the P-BPE-ECL system achieved LODs of 0.017 mM and 0.030 mM in PBS and artificial urine, respectively. In other recent studies [87, [94] [95] [96] lab-on-paper platforms were integrated with EC or ECL sensors to perform glucose detection in urine with LODs ranging from 0.35 mM to 3 × 10 −5 mM. Many lab-on-paper platforms with integrated colorimetric detection [97] [98] [99] [100] [101] have been proposed for determining the glucose concentration in human urine samples. de Oliveira et al. [102] developed 2D and 3D lab-on-paper platforms for the integrated colorimetric detection of glucose, total protein and nitrite in serum and artificial urine. The two lab-on-paper platforms were manufactured using a paper cutter printer and contained separate detection areas for the colorimetric detection of the three analytes of interest. The results obtained using artificial urine samples, showed that the 2D lab-on-paper platform achieved LODs of 0.54 mM, 5.19 µM and 2.34 µM for glucose, protein and nitrite, respectively. The corresponding LODs of 3D lab-on-paper platform used to detect glucose, protein and nitrite in artificial serum were 0.44 mM, 1.26 µM and 4.35 µM. Neris et al. [103] presented a 3D lab-on-paper device and multilayer microfluidic line/paper-based analysis device (3D-µTPAD) for the colorimetric detection of glucose in artificial urine samples. The 3D lab-on-paper device comprised three layers, i.e., wax, hot-pressed wax printing paper and single-sided tape, while the 3D-µTPAD consisted of four layers, namely a holepunched layer, a chromatography layer, a heat-pressed wax-printed paper layer, and a hole-punched layer containing trifurcated thread (see Figure 2b ). Glucose oxidase solution was flowed through each layer; resulting in the formation of a yellow-brown reaction complex following contact with glucose. The 3D lab-on-paper device and 3D-µTPAD devices achieved LODs of 1.0 mM and 0.5 mM, respectively, when applied to artificial urine samples. Table 2 briefly reviews several other lab-on-paper platforms presented in recent studies for the detection of glucose in urine samples. tively. The corresponding LODs of 3D lab-on-paper platform used to detect glucose tein and nitrite in artificial serum were 0.44 mM, 1.26 μM and 4.35 μM. Neris et al presented a 3D lab-on-paper device and multilayer microfluidic line/paper-based an device (3D-μTPAD) for the colorimetric detection of glucose in artificial urine sam The 3D lab-on-paper device comprised three layers, i.e., wax, hot-pressed wax pr paper and single-sided tape, while the 3D-μTPAD consisted of four layers, namely a punched layer, a chromatography layer, a heat-pressed wax-printed paper layer, hole-punched layer containing trifurcated thread (see Figure 2b ). Glucose oxidase so was flowed through each layer; resulting in the formation of a yellow-brown re complex following contact with glucose. The 3D lab-on-paper device and 3D-μTPA vices achieved LODs of 1.0 mM and 0.5 mM, respectively, when applied to artificial samples. Table 2 briefly reviews several other lab-on-paper platforms presented in studies for the detection of glucose in urine samples. Chronic kidney disease (CKD) usually has no obvious symptoms during the development stage, and hence often passes undetected until finally diagnosed. As a the effective treatment period is curtailed, leading to serious lesions and a significa terioration in quality of life [104] . However, recent research has shown that lab-onplatforms provide a feasible means of detecting early-stage kidney disease by evalu the concentrations of creatinine (Cre) and albumin (Alb) [105] [106] [107] [108] [109] [110] [111] , micro-albumi protein [112] [113] [114] [115] [116] , and uric acid [117] [118] [119] [120] [121] [122] [123] in urine samples. Chronic kidney disease (CKD) usually has no obvious symptoms during the early development stage, and hence often passes undetected until finally diagnosed. As a result, the effective treatment period is curtailed, leading to serious lesions and a significant deterioration in quality of life [104] . However, recent research has shown that lab-on-paper platforms provide a feasible means of detecting early-stage kidney disease by evaluating the concentrations of creatinine (Cre) and albumin (Alb) [105] [106] [107] [108] [109] [110] [111] , micro-albumin and protein [112] [113] [114] [115] [116] , and uric acid [117] [118] [119] [120] [121] [122] [123] in urine samples. Healthy kidneys play a vital role in maintaining human health by removing waste and excess fluid from the blood and expelling them with extra water in the form of urine. An excess quantity of creatinine (Cre) in the urine is a reliable indicator of early-stage kidney disease [105] . Thus, in recent years, various lab-on-paper devices have been proposed for Cre detection in urine samples [106, 107] . For instance, Sununta et al. [106] demonstrated a colorimetric lab-on-paper device based on the Jaffé reaction of Cre with picric acid under alkaline conditions. The intensity of the resulting orange color Cre-alkaline picrate complex was analyzed using ImageJ software to detect the Cre concentration. The device achieved a wide linear detection range of 0.2 to 1 mM with limits of detection and quantitation of 0.08 and 0.26 mM, respectively. Mathaweesansurn et al. [108] presented a contactimprinted lab-on-paper device for the multi-point determination of Cre in human urine. As shown in Figure 3a , a single-step embossing method was first performed using a rubber stamp to fabricate eight individual hydrophobic barrier structure patterns on the lab-on-paper platform for multi-point standard analysis. As in the device proposed by Sununta et al. [106] , Cre detection was performed colorimetrically by inducing a Jaffé reaction between the Cre content in the urine sample and the picric acid embedded in the reaction zones of the lab-on-paper device. The device was shown to have a linear Cre detection range of 50 to 1000 mg/L and a LOQ of 16.9 mg/L. The urine Alb to Cre ratio can also be used as a screening strategy for kidney disease [109] . Hiraoka et al. [110] developed a Drawing lab-on-paper device for urinary Alb-to-Cre ratio analysis to support the early detection of renal insufficiency. In the proposed device, the Alb-to-Cre ratio was determined simply by drawing a straight line between the top of the two color-changed assay channel sections (Alb and Cre) and then observing the position of the intercept point on an Alb-to-Cre scale placed between them. The results obtained for 12 urine samples collected from a hospital showed that the detection results for the Alb-to-Cre ratio were within 15% of the theoretical values. Healthy kidneys play a vital role in maintaining human health by removing waste and excess fluid from the blood and expelling them with extra water in the form of urine. An excess quantity of creatinine (Cre) in the urine is a reliable indicator of early-stage kidney disease [105] . Thus, in recent years, various lab-on-paper devices have been proposed for Cre detection in urine samples [106, 107] . For instance, Sununta et al. [106] demonstrated a colorimetric lab-on-paper device based on the Jaffé reaction of Cre with picric acid under alkaline conditions. The intensity of the resulting orange color Cre-alkaline picrate complex was analyzed using ImageJ software to detect the Cre concentration. The device achieved a wide linear detection range of 0.2 to 1 mM with limits of detection and quantitation of 0.08 and 0.26 mM, respectively. Mathaweesansurn et al. [108] presented a contact-imprinted lab-on-paper device for the multi-point determination of Cre in human urine. As shown in Figure 3a , a single-step embossing method was first performed using a rubber stamp to fabricate eight individual hydrophobic barrier structure patterns on the lab-on-paper platform for multi-point standard analysis. As in the device proposed by Sununta et al. [106] , Cre detection was performed colorimetrically by inducing a Jaffé reaction between the Cre content in the urine sample and the picric acid embedded in the reaction zones of the lab-on-paper device. The device was shown to have a linear Cre detection range of 50 to 1000 mg/L and a LOQ of 16.9 mg/L. The urine Alb to Cre ratio can also be used as a screening strategy for kidney disease [109] . Hiraoka et al. [110] developed a Drawing lab-on-paper device for urinary Alb-to-Cre ratio analysis to support the early detection of renal insufficiency. In the proposed device, the Alb-to-Cre ratio was determined simply by drawing a straight line between the top of the two colorchanged assay channel sections (Alb and Cre) and then observing the position of the intercept point on an Alb-to-Cre scale placed between them. The results obtained for 12 urine samples collected from a hospital showed that the detection results for the Alb-to-Cre ratio were within 15% of the theoretical values. (a) Lab-on-paper device fabricated by contact stamping for multi-point standard addition analysis. Reprinted with permission from ref. [108] . Copyright 2020 Elsevier. (b) Lab-on-paper device integrated with field-amplified stacking technology and desalination during stacking process. Reprinted with permission from ref. [112] . Copyright 2019 Elsevier. (c) Schematic illustration showing working principle of lab-on-paper EC sensor device for the detection of Cre and uric acid. Reprinted with permission from ref. [122] . Copyright 2019 Elsevier. Reprinted with permission from ref. [108] . Copyright 2020 Elsevier. (b) Lab-on-paper device integrated with field-amplified stacking technology and desalination during stacking process. Reprinted with permission from ref. [112] . Copyright 2019 Elsevier. (c) Schematic illustration showing working principle of lab-on-paper EC sensor device for the detection of Cre and uric acid. Reprinted with permission from ref. [122] . Copyright 2019 Elsevier. Micro-albumin and urinary protein measurement also provide a feasible means of performing the early-stage detection of kidney disease [112] [113] [114] . Cai et al. [112] presented a field-amplification lab-on-paper device for the sensitive colorimetric detection of microalbuminuria (MAU) in human urine. As shown in Figure 3b , the geometric shape of a 2D paper-based microfluidic channel was used to introduce an electric field gradient and achieve a field amplification effect by progressively reducing the sample conductivity. In the detection process, albumin (colorless) was selectively stained with bromophenol blue (BPB), and after passing through the stacking band of the lab-on-paper device, the MAU intensity was analyzed using ImageJ software and the corresponding micro-albumin concentration was determined. The experimental results showed that the device achieved a linear MAU concentration range of 10 to 100 mg/L with a LOD of 6.5 mg/L. Gao et al. [115] proposed a lab-on-paper platform incorporating a cation exchange membrane for the ion concentration polarization (ICP) stacking and colorimetric detection of total protein in urine. It was shown that the ICP stacking effect enhanced the sensitivity of the proposed device by about 60 times. Moreover, the linear detection ranges of human serum albumin (HSA) extended from 50 to 350 mg/L, while the recovery rate was 93 to 108% compared to the total protein detected in clinical urine samples. The same group [84] later demonstrated a second lab-on-paper platform for HSA concentration determination, in which the detection sensitivity was enhanced by a factor of around 100 times by purifying and enriching the protein content under a combined electric field and pH gradient. The device showed a linear HSA detection range of 10 to 300 mg/L in artificial urine samples and a LOD of 4.9 mg/L. Uric acid is a heterocyclic consisted of oxygen, nitrogen, carbon and hydrogen, and is a natural product of the metabolic breakdown of purine nucleotides in the human body. A high concentration of uric acid in urine is not only a forewarning of gout [116, 117] , but also a strong indicator of kidney disease [118] as well as diabetes, hyperuricemia and urolithiasis [119] . Thus, many lab-on-paper platforms for the detection of uric acid in urine samples have been proposed in recent years [117] [118] [119] [120] [121] [122] [123] . Ali et al. [119] demonstrated a colorimetric lab-on-paper device for the quantitative detection of uric acid using citrateterminated Pt nanoparticles (PtNPs) as sensing probes. The PtNPs served as a catalyst in producing an oxidation effect, which was subsequently reduced in the presence of uric acid, resulting in a detectable color change from blue to yellow. Compared to the detection results obtained using a standard colorimetric technique without the citrate-capped PtNPs, the proposed device showed a wider linear detection range of 0 to 8 mM and a LOD almost one order higher. Villa and Poppi [120] developed a quantitative method for detecting uric acid in urine using a surface-enhanced Raman spectroscopy (SERS) technique in combination with a lab-on-paper device coated with AuNPs. The device had a linear detection range of 0 to 3.5 mM (mmol/L) and a LOD of 0.11 mM. Several studies have attempted to increase the sensitivity of the uric acid concentration detection process by integrating lab-on-paper platforms with EC sensors [121, 122] . For example, Cincotto et al. [122] demonstrated a simple disposable lab-on-paper device consisting of two EC sensors for the simultaneous determination of uric acid and Cre, respectively, in human urine samples. In the proposed platform, graphene quantum dots (QDs) were used to enhance the oxidation of the uric acid and Cre on the working electrode and the concentrations of the two analytes were then determined via EC analysis with square-wave voltammetry (see Figure 3c) . The results showed that the platform was capable of detecting uric acid and creatinine concentrations in the range of 0.010 to 3.0 µM (µmol/L) with LODs of 8.4 nM (nmol/L) and 3.7 nM, respectively. Table 2 briefly reviews several other lab-on-paper platforms presented in recent studies for the detection of kidney disease using urine samples. Healthy urine contains no bacteria. However, bacteria in the anus can easily retrograde to the bladder through the urethra. For normal healthy individuals, the bladder can remove these bacteria on its own. However, in some cases, the bacteria may remain in the urinary system and cause infection [124, 125] . In fact, most urinary tract infections (UTI) are caused by Escherichia coli (E. coli) [126] [127] [128] , which normally appears in the rectum. Consequently, several lab-on-paper platforms aimed at the detection of E. coli in urine have been proposed in recent years [125] [126] [127] [128] [129] . For example, Gumustas et al. [125] proposed a paper-based test strip for the sensitive detection and quantification of E. coli in urine samples by means of SERS-based and colorimetric lateral flow immunoassay (LFIA) techniques. In developing the strip, the detection sensitivity was improved by using AuNPs modified with Raman labels to enhance the SERS response. The results showed that the colorimetric and SERS methods achieved LOD values of 78 and 45 CFU/mL, respectively. Ilhan et al. [126] presented a lab-on-paper platform with SERS detection-coupled immunomagnetic enrichment for the detection of E. coli in urine samples. As shown in Figure 4a . Fe 3 O 4 @Au nanoparticles combined with E. coli antibodies were used for selective separation and enrichment, and the casein-modified Fe 3 O 4 /Au-PEI nanoparticles were then cleaved; enabling the SERS detection process to achieve a more sensitive quantitative analysis. The platform achieved a linear E. coli detection range of 101 to 107 CFU/mL and a LOD of 0.52 CFU/mL. Calabretta et al. [128] developed a bioluminescent sensing lab-on-paper platform based on a luciferase/D-luciferin reaction for detecting bacterial adenosine-5 -triphosphate (ATP) for UTI diagnostics. The proposed platform used an innovative freeze-drying procedure and stable nano-lantern structured ATP-sensing paper biosensor for the effective detection of E. coli cells in urine samples within 10 min. In experimental tests, the device showed a LOD of 10 −14 mol ATP (corresponding to 3.8 × 10 4 CFU/mL). Lab-on-paper platforms can also be used to detect many different types of cancer cells in urine, including Hela cells, bladder cancer cells, prostate cancer cells and microRNA-21 [130] [131] [132] [133] [134] . Zhang et al. [131] presented a fluorescence detection method based on polyacrylamide dendrimer-activated chromatography-based paper for the detection of telomerase activity in HeLa cell lysates. The paper platform was manufactured using a manual drawing technique and exploited the hybridization of Cy5-modified single-stranded DNA probes with telomerase extension products for the detection of telomerase activity. The results show that the platform can detect telomerase activity in HeLa cell lysates of only 10 cells. Ma et al. [132] demonstrated a lab-on-paper based on Rox-DNA functionalized CdZnTeS quantum dots (QDs) with fluorescent probes to detect bladder cancer cells in urine samples. As shown in Figure 4b , the Rox-DNA functionalized QDs were first quenched by H 2 O 2 and then catalyzed as G-quadruplex reactants; thereby minimizing the interference of the H 2 O 2 on the green fluorescence emissions. The LOD and response time were shown to be 10 cells and less than 60 min, respectively. Abarghoei et al. [133] presented a colorimetric lab-on-paper sensor based on the peroxidase-like activity of cysteine-capped gold nanoclusters (CysAuNCs) for citrate detection in patients with prostate cancer. The Cys-AuNCs had an inherent peroxidase mimicking activity, and hence served both to promote catalytic oxidation and generate blue dye. As a result, they provided the means to measure the citrate ions directly without modification. The proposed sensor showed a good linear response over the citrate concentration range of 1.0 µM to 10 mM and achieved a LOD of 0.4 µM. Table 3 briefly reviews several other lab-on-paper platforms presented in recent studies for the detection of bacteria and cancer cells in urine samples. In experimental tests, the device showed a LOD of 10 mol ATP (corresponding 10 4 CFU/mL). Lab-on-paper platforms can also be used to detect many different types o cells in urine, including Hela cells, bladder cancer cells, prostate cancer cells croRNA-21 [130] [131] [132] [133] [134] . Zhang et al. [131] presented a fluorescence detection metho on polyacrylamide dendrimer-activated chromatography-based paper for the det telomerase activity in HeLa cell lysates. The paper platform was manufactured manual drawing technique and exploited the hybridization of Cy5-modified stranded DNA probes with telomerase extension products for the detection of tel activity. The results show that the platform can detect telomerase activity in H lysates of only 10 cells. Ma et al. [132] demonstrated a lab-on-paper based on R functionalized CdZnTeS quantum dots (QDs) with fluorescent probes to detect cancer cells in urine samples. As shown in Figure 4b , the Rox-DNA functionaliz were first quenched by H2O2 and then catalyzed as G-quadruplex reactants; there imizing the interference of the H2O2 on the green fluorescence emissions. The L response time were shown to be 10 cells and less than 60 min, respectively. Abar al. [133] presented a colorimetric lab-on-paper sensor based on the peroxidase-like of cysteine-capped gold nanoclusters (CysAuNCs) for citrate detection in patie prostate cancer. The Cys-AuNCs had an inherent peroxidase mimicking activ hence served both to promote catalytic oxidation and generate blue dye. As a res provided the means to measure the citrate ions directly without modification. T posed sensor showed a good linear response over the citrate concentration ran μM to 10 mM and achieved a LOD of 0.4 μM. Table 3 briefly reviews several othe paper platforms presented in recent studies for the detection of bacteria and can in urine samples. Reprinted with permission from ref. [132] . Copyright 2019 Elsevier. In general, the term "drug" refers to any substance that causes physiological or psychological changes in an organism when used. Moreover, "prohibited drugs" are psychotropic or narcotic drugs that are used for recreational "non-medical purposes" and may cause dependency and eventual addiction. Many lab-on-paper platforms have been proposed for the detection of drugs such as caffeine, paracetamol, adenosine, cocaine, ketamine, cannabidiol (CBD), methamphetamine (MA), 3,4-methylenedioxy -N-methamphetamine (MDMA) and tricyclic antidepressants [135] [136] [137] [138] [139] [140] [141] [142] . Petroni et al. [135] demonstrated a lab-on-paper platform with a CE sensor coupled to an external graphene-Cu nanoparticlemodified solid electrode through a meniscus configuration for the detection of paracetamol and caffeine in urine samples. The working electrode had the form of a glassy carbon electrode consisting of graphene oxide-copper nanoparticles (GO-CuNPs) synthesized via the chemical reduction of GO and Cu(II) salt. The device was shown to be capable of performing the quantitative analysis of paracetamol and caffeine in real urine samples with LODs of 24.6 nM and 36.1 nM, respectively. de Oliveira et al. [137] presented an electric paper-based spray (PS) platform combined with ionization mass spectrometry (MS) for the supramolecular microextraction and sensitive determination of four antidepressants (amitriptyline, doxepin, imipramine and nortriptyline) in urine (see Figure 5a ). The supramolecular microextraction process increased the pre-concentration factor of the tricyclic antidepressants in the urine sample and improved the sensitivity of the detection process as a result. The proposed platform achieved a LOD and LOQ of 5.2 to 8.6 µg/L and 7.4 to 28.7 µg/L, respectively. The literature contains various other PS-MS platforms for the rapid and sensitive analysis of psychiatric or anesthetic drugs in urine [138] [139] [140] . Narang et al. [141] developed an ultrasensitive technique for ketamine detection in three samples (whiskey, urine and fruit juice) using a zeolite nanocrystal-modified lab-on-paper platform coupled with GO nanosheets and an EC sensor. The lab-on-paper platform had a linear range of 0.001 to 5 nM/mL for the detection of ketamine in the three different samples and achieved a LOD of 0.001 nM/mL. Moreover, when applied to real-world samples such as alcohol and non-alcoholic beverages, the correlation between the detection results and the ground truth results was found to be as high as 99%. The ion content and composition in human urine provide many useful insights into the health of the individual. For example, excessive sodium content is often associated with disorders such as blood pressure and cardiovascular disease, while a high potassium content may indicate low blood pressure and arrhythmia [107] . Similarly, a high chlorine concentration is frequently associated with kidney function disorders [143] . Heavy metal ions, such as Cu 2+ , Pb + and Au 3+ , are easily deposited in the body through foods and drinks and cause chronic poisoning when present in excessive quantities. Accordingly, the problem of detecting electrolytes and heavy metal ions in human urine samples using lab-on-paper platforms has attracted significant attention in recent years [144] [145] [146] [147] [148] [149] [150] . Rahbar et al. [143] proposed a trapezoidal distance-based lab-on-paper platform with reverse Mohr precipitation titration for the determination of microscale quantities of chloride in urine samples (see Figure 5b ). In the proposed device, Ag 2 CrO 4 crystals were initially formed along the microfluidic detection channel and when the Cl − was then introduced at the microchannel inlet, the Cl − progressively replaced the CrO 4 2− in the complex and formed white AgCl crystals in its place. The length of the AgCl crystal band along the microchannel was then used to measure the concentration of chloride within the sample. It was shown that for a microchannel with the same area and length, the sensitivity achieved using a trapezoidal ascent channel (i.e., a diffusion type channel) was better than that achieved using a rectangular and descending trapezoid channel (i.e., a tapered type channel). When using a channel of the former type, the device achieved a linear concentration range of 0.05 to 25 mM and a LOD of 0.05 mM. Cai et al. [146] presented a color-evolution-based lab-on-paper platform with a tricolorratiometric fluorescent sensor for the visual detection of endogenous Cu 2+ ions in human urine. As shown in Figure 5c , the three-color probe sensor comprised blue-emission carbon dots (bCDs), red-emissive QDs (rQDs) and green-emissive QDs (gQDs). The fluorescence of the gQDs and rQDs was quenched in the presence of Cu 2+ . However, the fluorescence of the bCDs was unaffected by the Cu 2+ ions, and hence served as an internal standard for the light stability. It was shown that the color of the probe gradually changed from light pink to dark blue as the Cu 2+ concentration increased. The device presented a linear detection range for Cu 2+ ions in urine samples of 3 to 430 nM and a LOD of just 1.3 nM. Faham et al. [149] demonstrated a lab-on-paper colorimetric platform with an optical probe based on AuNPs modified with 2,2 -thiodiacetic acid (TDA) for the determination of Cr(III) and Cr(VI) ions in urine and diluted human plasma samples. The Cr(III) ions induced the aggregation of the modified AuNPs and prompted a color change of the TDA-AuNPs probe from red to blue, which was subsequently detected using a colorimetry technique implemented on a smartphone. Meanwhile, the Cr(VI) concentration was inferred by reducing the concentration of the Cr(III) ions using ascorbic acid and then quantifying the total Cr(III) concentration and subtracting this value from the total chromium concentration. The lab-on-paper platform had a linear detection range of 1.0 nM to 1.0 mM for Cr(III) and a LOD of 0.64 nM. Table 3 briefly reviews several other lab-on-paper platforms presented in the recent literature for drug detection and ion analysis in urine samples. green-emissive QDs (gQDs). The fluorescence of the gQDs and rQDs was quenched in the presence of Cu 2+ . However, the fluorescence of the bCDs was unaffected by the Cu 2+ ions, and hence served as an internal standard for the light stability. It was shown that the color of the probe gradually changed from light pink to dark blue as the Cu 2+ concentration increased. The device presented a linear detection range for Cu 2+ ions in urine samples of 3 to 430 nM and a LOD of just 1.3 nM. Faham et al. [149] demonstrated a lab-on-paper colorimetric platform with an optical probe based on AuNPs modified with 2,2′-thiodiacetic acid (TDA) for the determination of Cr(III) and Cr(VI) ions in urine and diluted human plasma samples. The Cr(III) ions induced the aggregation of the modified AuNPs and prompted a color change of the TDA-AuNPs probe from red to blue, which was subsequently detected using a colorimetry technique implemented on a smartphone. Meanwhile, the Cr(VI) concentration was inferred by reducing the concentration of the Cr(III) ions using ascorbic acid and then quantifying the total Cr(III) concentration and subtracting this value from the total chromium concentration. The lab-on-paper platform had a linear detection range of 1.0 nM to 1.0 mM for Cr(III) and a LOD of 0.64 nM. Table 3 briefly reviews several other lab-on-paper platforms presented in the recent literature for drug detection and ion analysis in urine samples. Figure 5 . (a) Paper spray ionization mass spectrometry detection device for sensitive determination of tricyclic antidepressants in urine. Reprinted with permission from ref. [137] . Copyright 2020 Elsevier. (b) Lab-on-paper device based on argentometric distance-based for the detection of chloride ion concentration. Reprinted with permission from ref. [143] . Copyright 2019 Elsevier. (c) Lab-on-paper device with fluorescent sensor for visual detection of Cu 2+ using three-color probe. Reprinted with permission from ref. [146] . Copyright 2018 Elsevier. In addition to the lab-on-paper platforms described above, paper-based devices have been proposed for the detection of many other analytes in urine samples, including serotonin (symptoms: regulating physiological processes) [151, 152] , folic acid (also known as vitamin B9; symptoms: physical weakness, irritability, anemia and leucopenia) [153, 154] , Paper spray (c) Figure 5 . (a) Paper spray ionization mass spectrometry detection device for sensitive determination of tricyclic antidepressants in urine. Reprinted with permission from ref. [137] . Copyright 2020 Elsevier. (b) Lab-on-paper device based on argentometric distance-based for the detection of chloride ion concentration. Reprinted with permission from ref. [143] . Copyright 2019 Elsevier. (c) Lab-on-paper device with fluorescent sensor for visual detection of Cu 2+ using three-color probe. Reprinted with permission from ref. [146] . Copyright 2018 Elsevier. In addition to the lab-on-paper platforms described above, paper-based devices have been proposed for the detection of many other analytes in urine samples, including serotonin (symptoms: regulating physiological processes) [151, 152] , folic acid (also known as vitamin B9; symptoms: physical weakness, irritability, anemia and leucopenia) [153, 154] , urobilinogen and billirubin (symptoms: liver disease, jaundice, hemolysis) [155, 156] , 8hydroxy-2-deoxyguanosine (8-OHdG; symptoms: aging, diabetes, inflammation) [157, 158] , and phenylketonuria (PKU; symptoms: metabolic diseases, delayed growth and development). Amatatongchai et al. [151] demonstrated a 3D folded lab-on-paper platform for the highly sensitive detection of serotonin in urine samples using a three-electrode EC sensor consisting of a working electrode (WE), a reference electrode (RE), and a counter electrode (CE). The WE was modified with Fe 3 O 4 @Au@SiO 2 and coated with molecularly imprinted polymers (MIP) in order to improve its selectivity for serotonin. The results showed that the device had a linear serotonin detection range of 0.01 to 1000 µM under the anodic peak current and a LOD of 0.002 µM. Nghia et al. [154] developed a lab-on-paper platform with colorimetric probes for the detection of folic acid (FA) in serum and urine samples. As shown in Figure 6a [154] . Copyright 2020 Elsevier. (b) Lab-on-paper device with in-situ formation of AuNPs for colorimetric determination of bilirubin. Reprinted with permission from ref. [156] . Copyright 2020 Springer. (c) A stacked four-layer cellulose paper lab-on-paper device for detection of BSA, glucose, nitrite and pH value in urine sample. Reprinted with permission from ref. [101] . Copyright 2020 Elsevier. [154] . Copyright 2020 Elsevier. (b) Lab-on-paper device with in-situ formation of AuNPs for colorimetric determination of bilirubin. Reprinted with permission from ref. [156] . Copyright 2020 Springer. (c) A stacked four-layer cellulose paper lab-on-paper device for detection of BSA, glucose, nitrite and pH value in urine sample. Reprinted with permission from ref. [101] . Copyright 2020 Elsevier. Edachana et al. [156] demonstrated a lab-on-paper device based on colorimetric method with AuNPs for the detection of bilirubin in urine samples. When chloroauric acid was added to the detection zone, the resulting yellow Au(III) complex reduced to purple AuNPs in the presence of bilirubin, and the color change was used to analyze the bilirubin concentration (see Figure 6b ). The intensity of the purple AuNPs increased linearly with an increasing bilirubin concentration in the range of 5.0 to 1000 mg/L. Moreover, the detection limit was 1.0 mg/L. Meng et al. [157] employed a 3-film paper-based solidphase microextraction (p-SPME) device for the analysis of 8-OHdG in urine samples using a capillary electrophoresis laser-induced fluorescence detection (CE-LIF) technique. The p-SPME device leveraged the large surface area of the cellulose molecules and their preferential bonding with hydrogen to enhance the concentration of the 8-OHdG sample and minimize the interference of the CE-LIF separation effect. The device had a linear detection range of 10 to 1000 nM and a LOD of 5 nM. Messina et al. [159] demonstrated a lab-on-paper platform to quantify the concentration of phenylalanine in patients with PKU from urine samples. In preparing the device, phenylalanine ammonia lyase (PAL) enzymes and reagents were immobilized on the surface matrix of the lab-on-paper platform and when PKU was subsequently added, it reacted with the PAL and reagents under temperature control to produce a detectable change in the color intensity. The device exhibited a linear response over the range of 20 µM to 3000 µM and achieved a LOD of 20 µM. In addition, the detection of pH in urine is also very important, and it is also one of the main indicators for determining urine detection [160] . Galanis et al. [101] developed a stacked four-layer cellulose paper and made a 3D multi-layer lab-on-paper device (see Figure 6c ). The lab-on-paper device can detect BSA, glucose, nitrite and the pH value of test samples mixed in artificial urine. The pH value test uses bromothymol blue as the indicator, dilutes the red chemical dye with deionized water, and the detectable range is 6.0~7.6. Table 3 briefly reviews several other lab-on-paper platforms presented in the literature for the detection of various other analytes in urine samples. Microfluidic lab-on-paper platforms are a rapidly developing and promising solution for the realization of next-generation preventive health care and detection analysis tools. Microfluidic lab-on-paper devices have many practical advantages over traditional laboratory systems, including a low cost, a simple manufacturing process, a straightforward operating procedure, good portability, high reliability, and a diagnostic performance close to that of benchtop methods. Furthermore, their use is compatible with many common biomedical samples, such as blood and urine. As a result, they have aroused great attention for POCT in the home or areas of the world with underdeveloped medical resources and infrastructures. Urine has many advantages as a biomedical sample, including natural abundance, noninvasiveness and biological richness. Consequently, many microfluidic lab-on-paper platforms for urine analysis and disease detection have been proposed in recent years [100, 108, 164, 165] . Many devices have also been successfully commercialized around the world, including the HiPee S2 smart urine analyzer (Tianjin Guo-Shih Tech. Co., Tianjin, China) and YH-1200 portable urine analyzer (Yao-Hua Co., Hebei, China). The MSLUA13 automatic urine analyzer (Medsinglong Medical Equipment Co., GuangZhou, China) has the ability to perform 14 different urine tests, including creatine, protein, microalbumin, specific gravity, calcium, nitrite, urobilinogen, pH, occult blood, glucose, bilirubin, ketone bodies, white blood cells, and vitamin C. Similarly, the Multistix 10 SG reagent strip device (Siemens Medical Solutions Inc., Malvern, PA, USA) can perform 10 urine tests, such as leukocytes, nitrite, urobilinogen, protein, pH, occult blood, specific gravity, ketone, bilirubin, and glucose. The test multi-drogues device produced by NarcoCheck Co. (Montluçon, France) provides the ability to test for 12 common drugs in urine, namely cannabis, cocaine, morphine, heroin, amphetamine, ecstasy, fentanyl, synthetic cannabinoids, ketamine, lysergic acid diacetamide, methylcystine and methamphetamine. Many simple urine test strips have also been commercialized for the detection of leukocytes and nitrites (Scanwell Health Co., San Diego, CA, USA), nicotine (Easy Healthcare Co., Willowbrook, IL, USA), pregnancy (Wondfo Biotech Co., GuangZhou, China) and ovulation (Femometer Co., Hong Kong, China). However, most of these devices provide only qualitative outcomes. Therefore, further research on the development of quantitative methods for the rapid analysis of urine samples using microfluidic lab-on-paper devices is still required. The problem of improving the analysis accuracy of microfluidic lab-on-paper devices is also an ongoing concern. Studies have shown that the detection performance of colorimetric methods can be improved by inducing the aggregation of nanomaterials in order to enhance the detection resolution [119, 149, 156] . Similarly, in the EC and SERS methods, the detection performance can be improved by patterning nanomaterials with good electrical conductivity on the electrode surface [120, 125, 135, 166, 167] . In CL and fluorescence methods, the intensity of the detection signal can be amplified through the catalytic reaction of nanomaterials and the induced fluorescence quenching [132, [168] [169] [170] , respectively. Many studies have confirmed that the detection performance of microfluidic lab-on-paper platforms can be improved through the use of sample pre-concentration techniques (such as ICP, ITP, EKS, and so on) [72, 73, 75, 115, 171] , or amplification devices (e.g., LAMP, PCR, CHR, HCR, and so forth.) [78, 81, 83, 172, 173] . For example, the authors in [84] showed that the detection performance of a lab-on-paper platform for urine microalbuminuria diagnosis in diabetic patients could be significantly improved by incorporating EKS with a 100-fold amplification capability in the sample pre-treatment stage. Although lab-on-chip devices that use human urine samples for human disease diagnosis have achieved rapid development in recent years, there are still several key challenges that need to be resolved. For instance, due to the irregularity of the fiber structure of paperbased materials, achieving precise fluid control and quantitative sample transportation is extremely difficult, and leads to potentially serious reproducibility and stability issues. Accordingly, the development of porous materials with a uniform pore size distribution is essential in improving the reproducibility and detection sensitivity of future lab-on-paper devices. Recent studies have reported that glass fiber paper and nitrocellulose membranes have a uniform pore size distribution [5, 10, [16] [17] [18] , and therefore offer an improved detection sensitivity and reproducibility. However, the manufacturing process for such materials is complicated and is more expensive than that for regular paper. In addition, although paper-based devices have many advantages over traditional benchtop systems for clinical diagnosis purposes, many of the sample pretreatment processes required to enhance the detection performance of these devices still require the use of large-scale laboratory equipment. As a result, further research is required to develop on-chip sample pretreatment methods in order to fully realize the commercial potential of paper-based microfluidic devices for POCT applications. Finally, there are still several remaining target analytes for detection by using urine samples and lab-on-paper platforms that have not been considered, including ketone body, Leu. esterase, epithelial cells, occult blood, and casts. In addition, lap-on-paper has insufficient research on virus infection or COVID in urine sample, mainly due to the detection limit. If lap-on-paper can be combined with amplification system [77] or immunosensor [174] , it is research that can be developed. Hence, future studies are also needed to develop effective lab-on-paper platforms to detect such analytes. In summary, this review describes all the key developments and concepts achieved in the microfluidic lab-on-paper platform field over the past five years for diagnosing and analyzing common human diseases with urine samples. It is expected that this review will help improve the availability and quality of medical services in developing areas of the world and accelerate the development of microfluidic lab-on-paper platforms in the field of human disease detection. The data presented in this study are available upon request from the corresponding author. The authors declare no conflict of interest. Non-invasive disease diagnosis using surface-enhanced Raman spectroscopy of urine and saliva Detection and quantification of proteins in human urine Highly sensitive paper-based electrochemical sensor for reagent free detection of bisphenol A Paper-based point-of-care immunoassays: Recent advances and emerging trends DNA-Gold Nanozyme-Modified Paper Device for Enhanced Colorimetric Detection of Mercury Ions Recent Advances in Microfluidic Devices for Contamination Detection and Quality Inspection of Milk Three-Dimensional Paper-Based Microfluidic Analysis Device for Simultaneous Detection of Multiple Biomarkers with a Smartphone Recent advances in microfluidic paper-based assay devices for diagnosis of human diseases using saliva, tears and sweat samples Smartphone and Paper-Based Fluorescence Reader: A Do It Yourself Approach Simple and inexpensive electrochemical paper-based analytical device for sensitive detection of Pseudomonas aeruginosa A paper-based analytical device coupled with electrochemical detection for the determination of dexamethasone and prednisolone in adulterated traditional medicines Paper-based human neutrophil elastase detection device for clinical wound monitoring Pushing the limits of spatial assay resolution for paper-based microfluidics using low-cost and high-throughput pen plotter approach Recent Advances in Cellulose-Based Biosensors for Medical Diagnosis Usability as a guiding principle for the design of paper-based, point-of-care devices-A review Paper-based Vertical Flow Immunoassay (VFI) for detection of bio-threat pathogens Blister-actuated LIFT printing for multiparametric functionalization of paper-like biosensors Paper-based colorimetric spot test utilizing smartphone sensing for detection of biomarkers Screen-printed microfluidic paper-based analytical device (µPAD) as a barcode sensor for magnesium detection using rubber latex waste as a novel hydrophobic reagent Ion-Exchange Based Immobilization of Chromogenic Reagents on Microfluidic Paper Analytical Devices Screening of highly-specific aptamers and their applications in paper-based microfluidic chips for rapid diagnosis of multiple bacteria A Review of the Construction of Nano-Hybrids for Electrochemical Biosensing of Glucose Programmable paper-based microfluidic devices for biomarker detections Characterization and Inkjet Printing of an RNA Aptamer for Paper-Based Biosensing of Ciprofloxacin Instrument-free fabrication of microfluidic paper-based analytical devices through 3D pen drawing Electrochemical paper-based devices: Sensing approaches and progress toward practical applications Paper-Based Biosensors: Frontiers in Point-of-Care Detection of COVID-19 Disease A Simple Distance Paper-Based Analytical Device for the Screening of Lead in Food Matrices Detection methods and applications of microfluidic paper-based analytical devices Green synthesis of fluorescent carbon dots for determination of glucose in biofluids using a paper platform Paper-based speedy separation of amplified DNA (PASS-DNA): Potential for molecular point-of-care testing Microfluidic paper-based analytical devices for environmental analysis of soil, air, ecology and river water A PET/paper chip platform for high resolution sulphur dioxide detection in foods Microfluidic colorimetric system for nitrite detection in foods 3D-PAD: Paper-Based Analytical Devices with Integrated Three-Dimensional Features Rapid electrochemical-biosensor microchip platform for determination of microalbuminuria in CKD patients Research and application progress of paper-based microfluidic sample preconcentration. Chin A review on advances in methods for modification of paper supports for use in point-of-care testing Advances in paper-based sample pretreatment for point-of-care testing Emerging trends in paper spray mass spectrometry: Microsampling, storage, direct analysis, and applications Fully integrated and slidable paper-embedded plastic microdevice for point-of-care testing of multiple foodborne pathogens Optimisation of DNA recovery and analysis of urine samples stored on FTA ® card A simple filter paper-based method for transporting and storing Enterocytozoon hepatopenaei DNA from infected Litopenaeus vannamei tissues The peroxidase-mimicking function of acetate and its application in singleenzyme-based glucose test paper Paper-based microfluidic system for tear electrolyte analysis UPLC-MS/MS analysis of keratan sulfate from urine samples collected on filter paper for monitoring & follow-up of Morquio A patients Paper Stacks for Uniform Rehydration of Dried Reagents in Paper Microfluidic Devices Exploring the applicability of nano-selenium for capture of mercury vapor: Paper based sorbent and a chemical modifier in graphite furnace atomic absorption spectrometry Principles of long-term fluids handling in paper-based wearables with capillaryevaporative transport SNAPflex: A paper-andplastic device for instrument-free RNA and DNA extraction from whole blood Microfluidic colorimetric detection platform with sliding hybrid PMMA/paper microchip for human urine and blood sample analysis Paper-based antibody detection devices using bioluminescent BRET-switching sensor proteins Precision medicine in Alzheimer's disease: An origami paper-based electrochemical device for cholinesterase inhibitors Emerging applications of paper-based analytical devices for drug analysis: A review Recent Advances of Fluid Manipulation Technologies in Microfluidic Paper-Based Analytical Devices (µPADs) toward Multi-Step Assays Integrated paper-based microfluidic devices for point-of-care testing An integrated platform for fibrinogen quantification on a microfluidic paper-based analytical device Countercurrent liquid-liquid extraction on paper Shaping up field-deployable nucleic acid testing using microfluidic paper-based analytical devices Paper-based device with on-chip reagent storage for rapid extraction of DNA from biological samples An innovative paper-based device for DNA extraction from processed meat products Lab-on-paper for all-in-one molecular diagnostics (LAMDA) of zika, dengue, and chikungunya virus from human serum Paper-based nucleic acid testing system for simple and early diagnosis of mosquito-borne RNA viruses from human serum Chromium speciation using paper-based analytical devices by direct determination and with electromembrane microextraction Flexible fabrication of a paper-fluidic SERS sensor coated with a monolayer of core-shell nanospheres for reliable quantitative SERS measurements Silane-based modified papers and their extractive phase roles in a microfluidic platform Detection and extraction of heavy metal ions using paper-based analytical devices fabricated via atom stamp printing A facile microfluidic paper-based analytical device for acetylcholinesterase inhibition assay utilizing organic solvent extraction in rapid detection of pesticide residues in food Simultaneous electrokinetic stacking and separation of anionic and cationic species on a paper fluidic channel Rapid pre-concentration of Escherichia coli in a microfluidic paper-based device using ion concentration polarization Rapid bacteria detection at low concentrations using sequential immunomagnetic separation and paper-based isotachophoresis Paper-based ITP technology: An application to specific cancer-derived exosome detection and analysis Origami-paper-based device for microvesicle/exosome preconcentration and isolation Electrokinetic stacking of particle zones in confined channels enabling their UV absorbance detection on microchips Preconcentration and sensitive determination of the anti-inflammatory drug diclofenac on a paper-based electroanalytical platform Roll-to-roll fabrication of integrated PDMS-paper microfluidics for nucleic acid amplification Development of a self-priming PDMS/paper hybrid microfluidic chip using mixed-dye-loaded loop-mediated isothermal amplification assay for multiplex foodborne pathogens detection An integrated one-step assay combining thermal lysis and loop-mediated isothermal DNA amplification (LAMP) in 30 min from E. coli and M. smegmatis cells on a paper substrate Fully integrated nucleic acid pretreatment, amplification, and detection on a paper chip for identifying EGFR mutations in lung cancer cells Origami-based "Book" shaped three-dimensional electrochemical paper microdevice for sample-to-answer detection of pathogens Ultrasensitive biosensing pathogenic bacteria by combining aptamer-induced catalysed hairpin assembly circle amplification with microchip electrophoresis Distance-based quantification of miRNA-21 by the coffee-ring effect using paper devices Donor/acceptor-induced ratiometric photoelectrochemical paper analytical device with a hollow double-hydrophilic-walls channel for microRNA quantification Online sample clean-up and enrichment of proteins from salty media with dynamic double gradients on a paper fluidic channel Paper-based assays for urine analysis A simple paper-based colorimetric device for rapid and sensitive urinary oxalate determinations Electrochemical paper-based microfluidic device for high throughput multiplexed analysis A decoupler-free simple paper microchip capillary electrophoresis device for simultaneous detection of dopamine, epinephrine and serotonin An equipment-free, paper-based electrochemical sensor for visual monitoring of glucose levels in urine Microfluidic paper-based device for colorimetric determination of glucose based on a metal-organic framework acting as peroxidase mimetic Paper-based disk-type self-powered glucose biosensor based on screen-printed biofuel cell array Direct writing electrodes using a ball pen for paper-based point-of-care testing Paper-based bipolar electrode-electrochemiluminescence (BPE-ECL) device with battery energy supply and smartphone read-out: A handheld ECL system for biochemical analysis at the point-of-care level A novel screen-printed microfluidic paper-based electrochemical device for detection of glucose and uric acid in urine New Disposable Electrochemical Paper-based Microfluidic Device with Multiplexed Electrodes for Biomarkers Determination in Urine Sample Large area, one step synthesis of NiSe2 films on cellulose paper for glucose monitoring in bio-mimicking samples for clinical diagnostics Selective colorimetric urine glucose detection by paper sensor functionalized with polyaniline nanoparticles and cell membrane Thread/paper-and paper-based microfluidic devices for glucose assays employing artificial neural networks Inkjet printed microfluidic paper-based analytical device (µPAD) for glucose colorimetric detection in artificial urine New single-layered paper-based microfluidic devices for the analysis of nitrite and glucose built via deposition of adhesive tape Local photo-polymer deposition-assisted fabrication of multilayer paper-based devices A simple method to produce 2D and 3D microfluidic paper-based analytical devices for clinical analysis 3D Multilayered paper-and thread/paper-based microfluidic devices for bioassays Bioplasmonic paper-based assay for perilipin-2 non-invasively detects renal cancer Paper-based Platform for Urinary Creatinine Detection Microfluidic paper-based analytical devices for determination of creatinine in urine samples Feasibility of measuring sodium, potassium and creatinine from urine sample dried on filter paper Simple and fast fabrication of microfluidic paper-based analytical device by contact stamping for multiple-point standard addition assay: Application to direct analysis of urinary creatinine A novel paper-based colorimetry device for the determination of the albumin to creatinine ratio Paper-based Device for Naked Eye Urinary Albumin/Creatinine Ratio Evaluation Collaboration of cyclometalated platinum complexes and metallic nanoclusters for rapid discrimination and detection of biogenic amines through a fluorometric paperbased sensor array Novel field amplification for sensitive colorimetric detection of microalbuminuria on a paper-based analytical device Monitoring cellulose oxidation for protein immobilization in paper-based low-cost biosensors Paper-based immunosensor utilizing dielectrophoretic trapping of microprobes for quantitative and label free detection using electrochemical impedance spectroscopy Detection of urine protein by a paper-based analytical device enhanced with ion concentration polarization effect Determination of salivary uric acid by using poly(3,4-ethylenedioxythipohene) and graphene oxide in a disposable paper-based analytical device Simultaneous determination of renal function biomarkers in urine using a validated paper-based microfluidic analytical device Developing paper based diagnostic technique to detect uric acid in urine Paper-based selective and quantitative detection of uric acid using citrate-capped Pt nanoparticles (PtNPs) as a colorimetric sensing probe through a simple and remote-based device A portable SERS method for the determination of uric acid using a paper-based substrate and multivariate curve resolution Fabrication of a three-dimensional electrochemical paper-based device (3D-ePAD) for individual and simultaneous detection of ascorbic acid, dopamine and uric acid A new disposable microfluidic electrochemical paper-based device for the simultaneous determination of clinical biomarkers Chitosan-Fe 3 O 4 nanoparticle enzymatic electrodes on paper as an efficient assay for glucose and uric acid detection in biological fluids Paper-based ELISA to rapidly detect Escherichia coli Paper based lateral flow immunoassay for the enumeration of Escherichia coli in urine Tamer, U. The coupling of immunomagnetic enrichment of bacteria with paper-based platform Multifunctional paper-based analytical device for in situ cultivation and screening of Escherichia coli infections Nano-lantern on paper for smartphone-based ATP detection Laser-patterned paper-based sensors for rapid point-of-care detection and antibiotic-resistance testing of bacterial infections Surface-enhanced Raman spectroscopy and MCR-ALS for the selective sensing of urinary adenosine on filter paper Polyamidoamine starburst dendrimer-activated chromatography paper-based assay for sensitive detection of telomerase activity Highly sensitive ratiometric fluorescent paper sensor for the urine assay of cancer A colorimetric paper sensor for citrate as biomarker for early stage detection of prostate cancer based on peroxidase-like activity of cysteine-capped gold nanoclusters Paper based colorimetric detection of miRNA-21 using Ag/Pt nanoclusters Paper-based Electrochemical Devices Coupled to External Graphene-Cu Nanoparticles Modified Solid Electrode through Meniscus Configuration and their Use in Biological Analysis Microfluidic distance readout sweet hydrogel integrated paper-based analytical device (µDiSH-PAD) for visual quantitative point-of-care testing Supramolecular microextraction combined with paper spray ionization mass spectrometry for sensitive determination of tricyclic antidepressants in urine The use of conductive polymers as a substrate for paper spray ionization mass spectrometry Rapid and on-site analysis of amphetamine-type illicit drugs in whole blood and raw urine by slug-flow microextraction coupled with paper spray mass spectrometry Polymeric ionic liquid immobilized onto paper as sorptive phase in microextraction Point of care with micro fluidic paper based device integrated with nano zeolite-graphene oxide nanoflakes for electrochemical sensing of ketamine Naked-eye quantitative assay on paper device for date rape drug sensing via smart phone Instrument-free argentometric determination of chloride via trapezoidal distance-based microfluidic paper devices Target-responsive DNA hydrogel mediated "stop-flow" microfluidic paper-based analytic device for rapid, portable and visual detection of multiple targets Paper-based scanometric assay for lead ion detection using DNAzyme Profuse color-evolution-based fluorescent test paper sensor for rapid and visual monitoring of endogenous Cu 2+ in human urine Determination of gold ions in human urine using genetically engineered microorganisms on a paper device Green design of a paper test card for urinary iodine analysis A paper-based optical probe for chromium by using gold nanoparticles modified with 2, 2 -thiodiacetic acid and smartphone camera readout A Fluorescent detection method for copper ions based on a direct redox route and desk study of wax-printed paper-based probes Highly sensitive and selective electrochemical paper-based device using a graphite screen-printed electrode modified with molecularly imprinted polymers coated Fe3O4@ Au@ SiO2 for serotonin determination A novel disposable self-adhesive inked paper device for electrochemical sensing of dopamine and serotonin neurotransmitters and biosensing of glucose Fluorescent paper-based sensor based on carbon dots for detection of folic acid Paper-based colorimetric probe for highly sensitive detection of folic acid based on open-ring form amplification of rhodamine B derivative Continuous-ink, multiplexed pen-plotter approach for low-cost, high-throughput fabrication of paper-based microfluidics Paper-based device for the colorimetric assay of bilirubin based on in-situ formation of gold nanoparticles Paper-based solid-phase microextraction for analysis of 8-hydroxy-2 -deoxyguanosine in urine sample by CE-LIF Diabetic retinopathy screening using a gold nanoparticle-based paper strip assay for the at-home detection of the urinary biomarker 8-hydroxy-2 -deoxyguanosine A facile method for urinary phenylalanine measurement on paper-based lab-on-chip for PKU therapy monitoring Polyaniline Functionalized Impedimetric Paper Sensor for Urine pH Measurement Molecularly imprinted polymer grafted paper-based method for the detection of 17β-estradiol Vertical flow paper-based plasmonic device for cysteine detection Novel paper-based colorimetric immunoassay (PCI) for sensitive and specific detection of salbutamol residues in flesh of swine and urine using Ag3PO4/Ag nanocomposite as label Understanding partial saturation in paper microfluidics enables alternative device architectures Programmable microfluidic flow for automatic multistep digital assay in a single-sheet 3-dimensional paper-based microfluidic device A disposable paper-based hydrophobic substrate for highly sensitive surface-enhanced Raman scattering detection Hybrid paper and 3D-printed microfluidic device for electrochemical detection of Ag nanoparticle labels A simple paper-based aptasensor for ultrasensitive detection of lead (II) ion Distance-dependent visual fluorescence immunoassay on CdTe quantum dot-impregnated paper through silver ion-exchange reaction A low-cost paper-based aptasensor for simultaneous trace-level monitoring of mercury (II) and silver (I) ions Colorimetric speciation of Cr on paper-based analytical devices based on field amplified stacking A microfluidic paper-based laser-induced fluorescence sensor based on duplex-specific nuclease amplification for selective and sensitive detection of miRNAs in cancer cells Multiplexed chemiluminescence determination of three acute myocardial infarction biomarkers based on microfluidic paper-based immunodevice dual amplified by multifunctionalized gold nanoparticles Early diagnosis of Zika infection using a ZnO nanostructures-based rapid electrochemical biosensor