key: cord-0738172-jtqy0big authors: Hui, David S.; Chow, Benny K.; Lo, Thomas; Ng, Susanna S.; Ko, Fanny W.; Gin, Tony; Chan, Matthew T.V. title: Exhaled Air Dispersion During Noninvasive Ventilation via Helmets and a Total Facemask date: 2015-12-22 journal: Chest DOI: 10.1378/chest.14-1934 sha: 83b125f80f01c2febdc9233e325f95f4310f2ad6 doc_id: 738172 cord_uid: jtqy0big BACKGROUND: Noninvasive ventilation (NIV) via helmet or total facemask is an option for managing patients with respiratory infections in respiratory failure. However, the risk of nosocomial infection is unknown. METHODS: We examined exhaled air dispersion during NIV using a human patient simulator reclined at 45° in a negative pressure room with 12 air changes/h by two different helmets via a ventilator and a total facemask via a bilevel positive airway pressure device. Exhaled air was marked by intrapulmonary smoke particles, illuminated by laser light sheet, and captured by a video camera for data analysis. Significant exposure was defined as where there was ≥ 20% of normalized smoke concentration. RESULTS: During NIV via a helmet with the simulator programmed in mild lung injury, exhaled air leaked through the neck-helmet interface with a radial distance of 150 to 230 mm when inspiratory positive airway pressure was increased from 12 to 20 cm H(2)O, respectively, while keeping the expiratory pressure at 10 cm H(2)O. During NIV via a helmet with air cushion around the neck, there was negligible air leakage. During NIV via a total facemask for mild lung injury, air leaked through the exhalation port to 618 and 812 mm when inspiratory pressure was increased from 10 to 18 cm H(2)O, respectively, with the expiratory pressure at 5 cm H(2)O. CONCLUSIONS: A helmet with a good seal around the neck is needed to prevent nosocomial infection during NIV for patients with respiratory infections. We examined exhaled air dispersion during NIV using a human patient simulator reclined at 45° in a negative pressure room with 12 air changes/h by two diff erent helmets via a ventilator and a total facemask via a bilevel positive airway pressure device. Exhaled air was marked by intrapulmonary smoke particles, illuminated by laser light sheet, and captured by a video camera for data analysis. Signifi cant exposure was defi ned as where there was Ն 20% of normalized smoke concentration. During NIV via a helmet with the simulator programmed in mild lung injury, exhaled air leaked through the neck-helmet interface with a radial distance of 150 to 230 mm when inspiratory positive airway pressure was increased from 12 to 20 cm H 2 O, respectively, while keeping the expiratory pressure at 10 cm H 2 O. During NIV via a helmet with air cushion around the neck, there was negligible air leakage. During NIV via a total facemask for mild lung injury, air leaked through the exhalation port to 618 and 812 mm when inspiratory pressure was increased from 10 to 18 cm H 2 O, respectively, with the expiratory pressure at 5 cm H 2 O. CONCLUSIONS: A helmet with a good seal around the neck is needed to prevent nosocomial infection during NIV for patients with respiratory infections. 6, 7 Patients may progress rapidly to ARDS and multiorgan failure, requiring intensive care support. [1] [2] [3] [4] [5] [6] [7] [8] Noninvasive ventilation (NIV) may play a supportive role in patients with early ARDS or acute lung injury due to SARI as a bridge to invasive mechanical ventilation. [8] [9] [10] However, a systematic review has shown that mask ventilation, tracheal intubation, tracheotomy, and NIV may increase the risk of nosocomial transmission of respiratory infections to health-care workers (HCWs). 11 Following the outbreak of SARS and emergence of the A(H1N1) infection, it has been recommended that when NIV is required for patients with acute hypoxemic respiratory failure due to SARI, infection control measures such as the use of helmets or full facemasks, double circuit tubes, and addition of viral-bacterial fi lters be considered. 10, 12 However, whether these infection control measures are eff ective in minimizing exhaled air leakage has not been objectively evaluated. Th is study aimed to examine the dispersion of exhaled air during application of NIV via helmets and total facemask. Knowledge about the extent of exhaled air leakage from diff erent masks will facilitate the development of preventive measures to reduce the risk of nosocomial transmission during application of NIV to high-risk patients hospitalized with SARI. We examined the extent of exhaled air dispersion during application of NIV on a high-fi delity human patient simulator (HPS) (HPS 6.1; CAE Healthcare, Inc) via two diff erent helmets (PN530L; Sea-Long Medical Systems Inc, and StarMed CaStar R; Intersurgical Ltd) using a SERVO-i ventilator (MAQUET) with double-limb circuit and fi lters. In addition, we studied the deliberate leakage from the exhalation port of a total facemask (Koninklijke Philips N.V.) during NIV via a bilevel positive airway pressure device and a single-limb circuit fi rmly attached to the HPS ( Fig 1 ) . Th e experiments were conducted in a negative pressure room, with 12 air changes/h (ACH) ( Fig 2 ) . Th e experimental design and method of data analysis have been described in detail in our previous studies on exhaled air dispersion related to the application of NIV, 12,13 oxygen masks, 14,15 jet nebulizer, 16 and mask ventilation. 17 Th e HPS represented a 70-kg adult man sitting on a 45°-inclined hospital bed. It was programmed to breathe spontaneously to mimic diff erent degrees of lung injury ( Table 1 ) . [12] [13] [14] [15] [16] [17] [18] [19] Th e HPS contains a realistic airway and a lung model that undergoes gas exchange by removing oxygen and adding CO 2 to the system simultaneously. Th e lung compliance and airway resistance respond in a realistic manner to relevant respiratory challenges. In addition, the HPS produces an airfl ow pattern that is close to the in vivo situation and has been applied in previous studies to simulate human respiration. 20-23 During examination of air leakage via the helmets, the inspiratory positive airway pressure (IPAP) was increased from 12 to 20 cm H 2 O, while keeping the expiratory positive airway pressure (EPAP) at 10 cm H 2 O. During examination of exhaled air dispersion from the total facemask, NIV was applied using a bilevel positive airway pressure device (ResMed VPAP III ST). Th e IPAP was initially set at 10 cm H 2 O and gradually increased to 14 and then 18 cm H 2 O, whereas the EPAP was maintained at 5 cm H 2 O throughout the study. 12, 13 Th e EPAP was higher for the helmets than the total face mask, as higher pressure was required to prevent CO 2 rebreathing in the helmets to overcome the larger anatomic dead space. Visualization of airflow around the helmets and the total facemask was facilitated by marking air with smoke particles produced by an M-6000 smoke generator (N19; Dick Smith Electronics Pty Limited) as described in our previous studies. [12] [13] [14] [15] [16] [17] Th e oil-based smoke particles, measuring , 1 m m in diameter, are known to follow the airfl ow pattern precisely with negligible slip. 24 Th e smoke was introduced continuously to the right main bronchus of the HPS. It mixed with alveolar gas and then exhaled through the airway. Sections through the leakage jet plume were then revealed by a thin laser light-sheet (Green, 532 nm wavelength, continuous-wave mode) created by a diode-pumped solid state laser (OEM UGH-800 mW, Lambdapro Technologies Ltd), with custom cylindrical optics for two-dimensional laser light-sheet generation. [12] [13] [14] [15] [16] [17] Th e light-sheet was initially positioned in the median sagittal plane of the HPS and was subsequently shift ed to the paramedian sagittal planes. Th is allowed us to investigate the regions directly above and lateral to the mask and the patient. [12] [13] [14] [15] [16] [17] All leakage jet plume images revealed by the laser light-sheet were captured by the high-defi nition video camera (Sony high-defi nition digital video camcorder, HDR-SR8E ClearVid complementary metal oxide semiconductor Sensor, Carl Zeiss Vario-Sonnar T* Lens), with optical resolution of 1,440 3 1,080 pixels per video frame. Normalized smoke concentration in the plume was estimated from the light intensity scattered by smoke particles. [12] [13] [14] [15] [16] [17] We estimated normalized smoke concentration in the mask leakage air from the light scattered by the particles. Th e analysis was based on scattered light intensity being proportional to particle concentration under the special conditions of constant intensity laser light-sheet illumination and monodisperse, small (submicron) particles. 24 In short, the thin laser light-sheet of near constant intensity illuminated smoke particle markers in the mask airfl ow leakage. Smoke particles scattered laser light perpendicular to the light-sheet, and this was collected and integrated by the video camera element and lens ( Fig 2 ) . [12] [13] [14] [15] [16] [17] Th e motion video of breathing cycles for each NIV setting was captured and individual frames extracted as gray scale bitmaps for intensity analysis. Frames were extracted at times initiated from the beginning of each inspiration, to generate an ensemble average for the corresponding instance of the respiratory cycle. [12] [13] [14] [15] [16] [17] Th e time at which the normalized concentration contours spread over the widest region from the NIV mask was chosen for the ensemble average to estimate the greatest dispersion distance. Th is was found to be approximately at the midrespiratory cycle. 12, 13 All gray scale frames were read into a program specifically developed for this study (MathCad 8.0), 25 along with background intensity images taken with the laser switched off. The background intensity image was subtracted from each frame, pixel by pixel, to remove any stray background light, and the pixel intensity values were averaged over all frames to determine the ensemble averaged intensity. Th e resulting image was the total intensity of light scattered perpendicular to the light-sheet by the smoke particles and was directly proportional to smoke concentration under the conditions mentioned previously. Th e image was normalized against the highest intensity found within the leakage jet plume to generate normalized particle concentration contours. [12] [13] [14] [15] [16] [17] As the smoke particles marked air that originated from the HPS airways, before leaking from the mask, the concentration contours eff ectively represent the probability of encountering air around the patient that has come from within the mask and/or the patient's respiratory system. Th e normalized concentration contours are made up of data collected from at least 20 breaths. A normalized concentration of 100% or a contour value of 1 indicates a region that consists entirely of air exhaled by the patient, where there is a very high chance of exposure to the exhaled air, such as at the mask exhaust vents. A normalized concentration of 0% or a contour value of 0 indicates no measurable air leakage in the region and a small chance of exposure to the exhaled air. Signifi cant exposure was arbitrarily defi ned as where there was Ն 20% of normalized smoke concentration. [12] [13] [14] [15] [16] [17] Th e study received nonionizing radiation safety approval by the Chinese University of Hong Kong (N/DSCH/RFCID 2012). Results are presented with reference to the median sagittal plane . During application of NIV via the Sea-Long oxygen head tent, exhaled air was observed to leak through the neck-tent interface ( Fig 1A ) . Using a normalized smoke concentration of 20% as a cutoff , the radial dispersion distance was 170 mm in normal lung condition with IPAP of 12 cm H 2 O, and the dispersion distance was 150 mm when the HPS was programmed in either mild or severe lung injury. When IPAP was increased gradually from 12 to 20 cm H 2 O, the dispersion distance increased to 270 mm, 230 mm, and 180 mm in normal condition, mild lung injury, and severe lung injury, respectively ( Fig 3 , e-Fig 1 ) . Th e StarMed CaStar R helmet had a tight air cushion around the neck-helmet interface. Th erefore, negligible air leak was noted during application of NIV when IPAP was increased from 12 to 20 cm H 2 O ( Fig 1B ) . Figure 4 shows the dispersion distance of exhaled smoke during application of NIV using the Respironics total facemask in diff erent lung conditions. It was observed that exhaled air jet leaked through the mask exhalation port ( Fig 1C ) Figure 1 legend for expansion of other abbreviation. lung injury worsened, there was a decrease in the dispersion distance ( Fig 4 , e-Fig 2 ) . Because of the lack of any reliable and safe marker that can be introduced into human lungs for study, we have examined the maximum distribution of exhaled air, marked by fi ne smoke particles, from the HPS during application of NIV using two diff erent helmets and a total facemask. We have shown that leakage of exhaled air was negligible when NIV was applied to the HPS via a helmet with double limb circuit, fi lters and a good seal at the neck-helmet interface, whereas leakage at the neck interface could reach a maximum radial distance of 270 mm through another helmet without a tight seal in the interface. In addition, exhaled air jet through the exhalation port could reach a distance of 916 mm when NIV was applied to the HPS via the total face mask using single circuit. In addition to reporting the exhaled air dispersion during manual ventilation with and without addition of a viral-bacterial fi lter on the HPS model, 17 we have previously examined the maximum exhaled air distances from the HPS receiving NIV via several orofacial masks using a single limb circuit. 12, 13 Th e ResMed mirage mask could leak through its exhalation port up to 500 mm when the HPS was programmed in mild lung injury, 12 whereas leakage from the Respironics ComfortFull 2 mask would increase from 650 to 850 mm through the exhalation diff user at a direction perpendicular to the head of the HPS along the sagittal plane, when IPAP was increased from 10 to 18 cm H 2 O, respectively. 13 In contrast, even when a low IPAP of 10 cm H 2 O was applied to the HPS via the Respironics Image 3 mask connected to the whisper swivel exhalation port, the exhaled air leaked far more diff usely than via the ComfortFull 2 mask, dispersing a low normalized concentration to 950 mm along the median sagittal plane of the HPS, whereas higher IPAP resulted in wider spread of a higher normalized concentration of smoke around the HPS in the isolation room with negative pressure. 13 NIV is eff ective in the treatment of patients with respiratory failure due to COPD, acute cardiogenic pulmonary edema, and pneumonia in immunocompromised patients, but the evidence supporting its use in hypoxic respiratory failure due to SARI is limited. 26 NIV was applied successfully to some patients with SARS in a hospital in Hong Kong, with reduced needs for tracheal intubation and decreased mortality without causing any clinical and serological evidence of nosocomial transmission among the HCWs. 27 Th e hospital involved managed to install exhaust ventilation fans in the windows of the treatment rooms to create a negative pressure environment with . 8 ACH, and the HCWs involved complied well with airborne precautions. 27 However, other reports suggested that NIV might have led to nosocomial transmission of SARS involving HCWs, 28, 29 whereas NIV was shown to be an independent risk factor of super-spreading events in the hospital setting in a large case-controlled study. 30 Indeed, a systematic review has shown that NIV is one of the factors that may lead to increased risk of nosocomial infection. 11 In immunocompromised patients with lung infi ltrates and acute hypoxic respiratory failure, NIV delivered via helmet appeared as eff ective as NIV via face mask in avoiding tracheal intubation and improving gaseous exchange, whereas fewer NIV discontinuations and fewer complications were observed in the helmet group. 31 NIV has been applied to some patients with severe H7N9 infection 5 and Middle East respiratory syndrome, 32 but the majority of patients required invasive mechanical ventilation. Several groups had applied NIV to patients hospitalized with infl uenza A(H1N1) and acute hypoxemic respiratory failure with variable success. [33] [34] [35] Infl uenza A viruses may spread between humans through contact, large respiratory droplets, and small particle droplet nuclei (aerosols). 36, 37 Th ere was evidence of possible aerosol transmission in a nosocomial outbreak of seasonal infl uenza temporally related to the use of NIV in an index patient with acute exacerbation of COPD due to infl uenza A(H3N2) in our acute hospital medical ward with an imbalanced indoor airfl ow. 38 Patients with mild ARDS due to SARI may be considered for a trial of NIV if there is suffi cient local experience. 10, 39 However, NIV is generally regarded as one of the aerosol-generating procedures in which there is possibly increased risk of respiratory pathogen transmission. 11, 40 Th us, it is advisable to apply NIV carefully with airborne precautions in an adequately ventilated room (with 6-12 ACH) when caring for patients with SARI of infectious nature. 10, 12, 13, 40, 41 Based on our study fi ndings, NIV via the helmet with double limb circuit and a good seal at the neck-helmet interface would be a safe option for managing infectious patients with hypoxemic respiratory failure due to SARI. Our study was limited by the use of smoke particles as markers for exhaled air. However, evaporation of water content in some droplets during NIV may produce droplet nuclei suspended in air, whereas the larger droplets will fall to the ground in a trajectory pathway. 42 As the smoke particles in this study mark the continuous air phase, our data contours are referring to exhaled air. Our results would therefore represent the "upper bound" estimates for the dispersion of droplets, which would be expected to follow a shorter trajectory than the air jet because of gravitational eff ects, but not fully refl ect the risk of droplet transmission. [12] [13] [14] [15] [16] [17] In summary, we have shown that leakage of exhaled air was negligible when NIV was applied to the HPS via a helmet with double limb circuit, fi lters, and a good seal at the neck interface, whereas leakage at the neck interface could reach a maximum radial distance of 270 mm through another helmet without a tight neck seal. In addition, leakage of exhaled air jet through the exhalation port could reach a distance of 916 mm when NIV was applied via the total face mask and a single circuit. HCWs should take adequate precautions when providing NIV support to patients with SARI complicated by respiratory failure. A major outbreak of severe acute respiratory syndrome in Hong Kong Treatment of severe acute respiratory syndrome Clinical management of pandemic 2009 infl uenza A(H1N1) infection Writing Committee of the Second World Health Organization Consultation on Clinical Aspects of Human Infection with Avian Infl uenza A (H5N1) Virus . Update on avian infl uenza A (H5N1) virus infection in humans Clinical fi ndings in 111 cases of infl uenza A (H7N9) virus infection State of knowledge and data gaps of Middle East respiratory syndrome coronavirus (MERS-CoV) in humans Severe acute respiratory syndrome vs. the Middle East respiratory syndrome Th e critically ill avian infl uenza A (H5N1) patient Infl uenza A/H5N1 infection: other treatment options and issues On the role of non-invasive ventilation (NIV) to treat patients during the H1N1 influenza pandemic Aerosol generating procedures and risk of transmission of acute respiratory infections to healthcare workers: a systematic review Noninvasive positive-pressure ventilation: An experimental model to assess air and particle dispersion Exhaled air dispersion distances dur ing noninvasive ventilation via diff erent Respironics face masks Airfl ows around oxygen masks: a potential source of infection? Exhaled air dispersion during oxygen delivery via a simple oxygen mask Exhaled air and aerosolized droplet dispersion during application of a jet nebulizer Mask ventilation and dispersion of exhaled air Breathing pattern and workload during automatic tube compensation, pressure support and T-piece trials in weaning patients Pulmonary pathophysiology of pneumococcal pneumonia Bellows-less lung system for the human patient simulator Performance of an oxygen delivery device for weaning potentially infectious critically ill patients A model for educational simulation of infant cardiovascular physiology An audible indication of exhalation increases delivered tidal volume during bag valve mask ventilation of a patient simulator Fluid Dynamics of Multiphase Systems MathSoft Inc. Mathcad 8.0 for Windows Noninvasive positive pressure ventilation in the acute care setting: where are we? Eff ectiveness of noninvasive positive pressure ventilation in the treatment of acute respiratory failure in severe acute respiratory syndrome Transmission of severe acute respiratory syndrome during intubation and mechanical ventilation A retrospective study of 78 patients with severe acute respiratory syndrome Why did outbreaks of severe acute respiratory syndrome occur in some hospital wards but not in others? Noninvasive ventilation by helmet or face mask in immunocompromised patients: a case-control study Clinical course and outcomes of critically ill patients with Middle East respiratory syndrome coronavirus infection Early non-invasive ventilation treatment for severe infl uenza pneumonia Respiratory support during the influenza A (H1N1) pandemic flu in Sweden Eff ectiveness and predictors of success of noninvasive ventilation during H1N1 pandemics: a multicenter study Infl uenza virus in human exhaled breath: an observational study Aerosol transmission is an important mode of infl uenza A virus spread Possible role of aerosol transmission in a hospital outbreak of infl uenza Interim guidance document on clinical management of severe acute respiratory infections when novel coronavirus is suspected: what to do and what not to do. World Health Organization website Infection control and MERS-CoV in health-care workers How far droplets can move in indoor environments-revisiting the Wells evaporation-falling curve