key: cord-0738344-76buf5cc authors: Gajic, Ina; Kabic, Jovana; Kekic, Dusan; Jovicevic, Milos; Milenkovic, Marina; Mitic Culafic, Dragana; Trudic, Anika; Ranin, Lazar; Opavski, Natasa title: Antimicrobial Susceptibility Testing: A Comprehensive Review of Currently Used Methods date: 2022-03-23 journal: Antibiotics (Basel) DOI: 10.3390/antibiotics11040427 sha: b540303ffb3d23cc6d7524de0333e2b962e8835f doc_id: 738344 cord_uid: 76buf5cc Antimicrobial resistance (AMR) has emerged as a major threat to public health globally. Accurate and rapid detection of resistance to antimicrobial drugs, and subsequent appropriate antimicrobial treatment, combined with antimicrobial stewardship, are essential for controlling the emergence and spread of AMR. This article reviews common antimicrobial susceptibility testing (AST) methods and relevant issues concerning the advantages and disadvantages of each method. Although accurate, classic technologies used in clinical microbiology to profile antimicrobial susceptibility are time-consuming and relatively expensive. As a result, physicians often prescribe empirical antimicrobial therapies and broad-spectrum antibiotics. Although recently developed AST systems have shown advantages over traditional methods in terms of testing speed and the potential for providing a deeper insight into resistance mechanisms, extensive validation is required to translate these methodologies to clinical practice. With a continuous increase in antimicrobial resistance, additional efforts are needed to develop innovative, rapid, accurate, and portable diagnostic tools for AST. The wide implementation of novel devices would enable the identification of the optimal treatment approaches and the surveillance of antibiotic resistance in health, agriculture, and the environment, allowing monitoring and better tackling the emergence of AMR. Antimicrobial resistance (AMR) remains the world's most urgent public health concern [1] . According to the World Health Organization (WHO), Geneva, Switzerland, antibiotic resistance is rising to dangerously high levels in all parts of the world, leading to increased morbidity and mortality [2] . Hence, the six leading mortality-causing pathogens-Escherichia coli, Staphylococcus aureus, Klebsiella pneumoniae, Streptococcus pneumoniae, Acinetobacter baumannii, and Pseudomonas aeruginosa-were responsible for 929,000 deaths attributable to AMR and 3.57 million deaths associated with AMR in 2019 [3] . This number could rise to 10 million by 2050 according to estimates by the WHO [4] . Furthermore, the SARS-CoV-2 pandemic has exacerbated the existing global crisis of AMR, mostly due to the mis-and over-use of antibiotics, treatments that induce immunosuppression, and prolonged hospitalisation [5] . Besides, during the COVID-19 pandemic, limited ability to work with AMR partnerships, decreases in funding, and reduced availability of nursing, medical, and public health staff affected AMR surveillance, prevention, and control [6] . In addition, increased use of disinfectants, including hand sanitisers and surface cleaners, is anticipated to cause increased rates of antimicrobial The choice of the best therapeutic option for the treatment of bacterial infections relies on the results of AST, a part of the routine work of all clinical microbiological laboratories. These reports provide insight into local patterns of antimicrobial susceptibility, helping physicians to choose the most effective antibiotic therapy [13] . For instance, if the AMR rate of a pathogen is above 20%, that drug should not be administered as a single empiric therapy for infection treatment [14] . Evaluation of the effectiveness of prevention and infection control measures relies as well on the results of AST, e.g., monitoring of resistant pathogens such as MRSA (methicillin-resistant Staphylococcus aureus), VRE (vancomycin-resistant enterococci), extended-spectrum beta-lactamase (ESBL)-and carbapenemase-producing Enterobacterales, carbapenem-resistant Acinetobacter baumannii (CRAB), carbapenem-resistant Pseudomonas aeruginosa (CRPA), colistin-resistant bacteria, etc. [15] . Finally, surveillance of antimicrobial resistance is based on routine clinical antimicrobial susceptibility data from microbiological laboratories. Numerous AMR surveillance systems exist, of which the WHOs Global Antimicrobial Resistance and Use Surveillance System (GLASS), European Antimicrobial Resistance Surveillance Network (EARS-Net), and Antibiotic Resistance Laboratory Network (AR Lab Network) of the Centers for Disease Control and Prevention are the most recognizable networks of national surveillance systems providing information on the actual burden of resistance at the international level. Policymakers and health administrators revise the recommendations for empirical treatment for community or hospital-acquired infections according to the local, national, and international AMR data. In addition, prevention and infection control measures are implemented based on the same data as a part of AMS programs [16, 17] . Likewise, continuous monitoring provides early warnings of emerging threats and identifies long-term resistance trends. Although new technologies have been introduced to obtain data on bacterial susceptibility to antimicrobial agents, conventional technologies are still in widespread use. Besides disc diffusion susceptibility tests, the most widely used methods include broth macro-and microdilution and agar dilution, representing the reference methods [28] . By using broth and agar dilution methods, the minimum inhibitory concentrations (MICs) of antimicrobial agents (i.e., the lowest concentration at which the agent inhibits the growth of microorganisms) can be determined [29, 30] . The MIC value serves as the basis for assessing the susceptibility category of the pathogen to a given antibiotic, of organisms that give ambiguous results, and especially when no clinical breakpoints for disk diffusion are available. Contrary to a qualitative method, the MIC value allows assessing the degree of susceptibility or resistance to the antibiotic [31] . Besides the determination of MICs, the advantage of broth dilution methods is the possibility of obtaining the minimum bactericidal concentration (MBC), which is the lowest concentration of an antimicrobial substance that kills 99.9% of bacteria [32] . Although new technologies have been introduced to obtain data on bacterial susceptibility to antimicrobial agents, conventional technologies are still in widespread use. Besides disc diffusion susceptibility tests, the most widely used methods include broth macro-and microdilution and agar dilution, representing the reference methods [28] . By using broth and agar dilution methods, the minimum inhibitory concentrations (MICs) of antimicrobial agents (i.e., the lowest concentration at which the agent inhibits the growth of microorganisms) can be determined [29, 30] . The MIC value serves as the basis for assessing the susceptibility category of the pathogen to a given antibiotic, of organisms that give ambiguous results, and especially when no clinical breakpoints for disk diffusion are available. Contrary to a qualitative method, the MIC value allows assessing the degree of susceptibility or resistance to the antibiotic [31] . Besides the determination of MICs, the advantage of broth dilution methods is the possibility of obtaining the minimum bactericidal concentration (MBC), which is the lowest concentration of an antimicrobial substance that kills 99.9% of bacteria [32] . The macrodilution method, also known as the in-tube dilution test, uses serial two-fold dilution of antimicrobial substances in corresponding media. A known concentration of suspended bacteria is added to the tubes prepared, as described in [32] . After 24 h of incubation at 37 • C, bacterial growth is measured by turbidity of media, allowing visual determination of MIC values. Another macrodilution method is the time-kill methodology. This test allows monitoring of the effect of different concentrations of antimicrobial substances by examining the rate at which antimicrobials lead to bacterial death-i.e., the bactericidal activity of antimicrobial agents is determined depending on the concentration and time. Bacterial viability is determined by counting colonies on agar plates at regular time points for 24 h [33] . The rate of bacterial growth is monitored via changes in log CFU/mL during the first 24 h time-kill test. Based on the results, experimental curves which represent the absence of growth or the killing effect can be constructed and give us insight into the interaction between the bacteria and the antimicrobial agent. The data can be further analysed using different mathematical models [30, 34, 35] . The BMD method is standardised, accurate, and inexpensive. Since it is performed in 96-well microtiter plates, it allows the testing of several antimicrobial substances in a row and eight series of two-fold dilutions of antimicrobial agents in one plate. After the dilutions are made, each well is inoculated with standardised bacterial inoculum and incubated for at least 16-24 h. Although this procedure is used as a reference method, it has been improved by the addition of a resazurin colour redox indicator. Resazurin is a blue colour that turns into pink, fluorescent resorufin in the presence of metabolically active bacterial cells. The reduction of resazurin to fluorescent resorufin can be measured fluorimetrically [27, 32, [36] [37] [38] . Nowadays, there are several commercially available easy to perform BMD systems such as MBD Sensititre System (Thermo Fisher Scientific, Waltham, MA, USA) and ComASP Colistin (Liofilhem, Roseto degli Abruzzi, Italy), formerly Sen-siTest Colistin. The MBD Sensititre System can be performed manually or automatically. ComASP Colistin is a compact panel containing the antibiotic colistin in seven two-fold serial dilutions and allows for four samples to be tested simultaneously with the BMD method [39] . The agar dilution method involves adding different concentrations of antimicrobial substances to the non-selective medium before solidification [40] . Afterwards, the standardised bacterial inoculum is spotted on the agar surface. Following overnight incubation, plates are evaluated visually, determining whether growth has occurred at the inoculated sites. The lowest concentration of antibiotics that prevent bacterial growth is considered to be the MIC. This method allows simultaneous testing of different bacterial strains [41] . The gradient strip test is a combination of disk-diffusion and dilution method of AST, having advantageous properties of both methods. It allows the MIC to be determined while keeping it simple and easy to use. The method is based on the diffusion of an antibiotic through agar with a continuous gradient. A concordance of the susceptibility categories and MIC values obtained by gradient test and BMD method, a "gold standard" recommended by the European Committee on Antimicrobial Susceptibility Testing (EUCAST) and CLSI, were observed [28, 42] . For example, the new ceftazidime-avibactam and ceftolozanetazobactam gradient tests (Etests, bioMérieux, Marcy-l'Étoile, France) have shown a high categorical agreement between gradient test and BMD, of 96% and 94%, respectively [43, 44] . On the other hand, for some antibiotics, such as colistin and tigecycline [45] , controversial results have been obtained. Some agar-related factors, i.e., the content of divalent cations, can affect the diffusion of colistin, resulting in false susceptibility. Consequently, BMD remains the only appropriate method for MIC determination for certain antibiotics [46] . Currently, a few commercial gradient strip tests, such as Etest (bioMérieux, France), MIC Test Strip (Liofilchem, Roseto degli Abruzzi, Italy), M.I.C.Evaluator (Thermo Scientific, Waltham, MA, USA), and Ezy MIC Strip (HiMedia Laboratories, Mumbai, India), are available [47] . They can be used for susceptibility testing of microorganisms to antibiotics and antifungals [48] [49] [50] . A gradient strip test is performed according to the manufacturer's instructions: a short plastic or paper strip impregnated with antibiotic is placed on inoculated agar ( Figure 2 ). On the standardised 100 mm Petri dish, two strips may be placed, while on the larger Antibiotics 2022, 11, 427 6 of 26 150 mm Petri dish, up to six antibiotics may be tested simultaneously. The MIC of a tested agent is determined by the intersection of a zone of inhibition with the strip and measured using labelled concentrations on the strip. If the intersection is between two values on a scale, a higher value is reported as MIC. In addition, if beta-haemolysis is present on the plate, careful examination of the strip is required since the reporting of the intersection of haemolysis leads to false higher MICs values. Automated systems for reading the results of gradient tests are also available (ADAGIO Automated System, Bio-Rad Laboratories, Hercules, CA, USA). Antibiotics 2022, 11, x FOR PEER REVIEW 6 of 27 available [47] . They can be used for susceptibility testing of microorganisms to antibiotics and antifungals [48] [49] [50] . A gradient strip test is performed according to the manufacturer's instructions: a short plastic or paper strip impregnated with antibiotic is placed on inoculated agar (Figure 2 ). On the standardised 100 mm Petri dish, two strips may be placed, while on the larger 150 mm Petri dish, up to six antibiotics may be tested simultaneously. The MIC of a tested agent is determined by the intersection of a zone of inhibition with the strip and measured using labelled concentrations on the strip. If the intersection is between two values on a scale, a higher value is reported as MIC. In addition, if beta-haemolysis is present on the plate, careful examination of the strip is required since the reporting of the intersection of haemolysis leads to false higher MICs values. Automated systems for reading the results of gradient tests are also available (ADAGIO Automated System, Bio-Rad Laboratories, Hercules, CA, USA). A variation of gradient tests exists for the detection of various AMR phenotypes. Currently, Etests for phenotypic detection of ESBL production in enterobacteria are available, such as strips with cefotaxime+clavulanic acid, ceftazidime with clavulanic acid, and cefepime with clavulanic acid [51] . The gradient tests for ESBL detection are two-sided strips that contain antibiotic on one end, while on the other is the same antibiotic with clavulanic acid. Reduction in MIC equal to or greater than eight times by the combination of antibiotic and clavulanate refers to ESBL production [52] . Similar to the double-disk synergy test, the phantom zone below the clavulanic end also indicates a positive result. Identification of metallo-beta-lactamase (MBL)-producing bacteria can be carried out using a gradient test. These tests contain carbapenem antibiotic on one side of the strip and the same carbapenem with EDTA on the other side. Imipenem with EDTA for detection of MBL in Acinetobacter spp. and Pseudomonas spp. is available, although sensitivity and specificity may vary [53, 54] . For detection of AmpC beta-lactamase-producing enterobacteria, Etest impregnated with cefotetan on one end and cefotetan-cloxacillin on the other end can be used [55] . Gradient tests with a predefined gradient of vancomycin and teicoplanin on each side of the strip can be used for the detection of glycopeptide resistance in Staphylococcus aureus [56] . Since these tests are easy to perform, they could be used as "screening" tests for the detection of emerging resistance patterns among clinically relevant bacteria. A variation of gradient tests exists for the detection of various AMR phenotypes. Currently, Etests for phenotypic detection of ESBL production in enterobacteria are available, such as strips with cefotaxime+clavulanic acid, ceftazidime with clavulanic acid, and cefepime with clavulanic acid [51] . The gradient tests for ESBL detection are two-sided strips that contain antibiotic on one end, while on the other is the same antibiotic with clavulanic acid. Reduction in MIC equal to or greater than eight times by the combination of antibiotic and clavulanate refers to ESBL production [52] . Similar to the double-disk synergy test, the phantom zone below the clavulanic end also indicates a positive result. Identification of metallo-beta-lactamase (MBL)-producing bacteria can be carried out using a gradient test. These tests contain carbapenem antibiotic on one side of the strip and the same carbapenem with EDTA on the other side. Imipenem with EDTA for detection of MBL in Acinetobacter spp. and Pseudomonas spp. is available, although sensitivity and specificity may vary [53, 54] . For detection of AmpC beta-lactamase-producing enterobacteria, Etest impregnated with cefotetan on one end and cefotetan-cloxacillin on the other end can be used [55] . Gradient tests with a predefined gradient of vancomycin and teicoplanin on each side of the strip can be used for the detection of glycopeptide resistance in Staphylococcus aureus [56] . Since these tests are easy to perform, they could be used as "screening" tests for the detection of emerging resistance patterns among clinically relevant bacteria. Plenty of advantages of gradient tests are known: simple performance, flexibility in the testing of any combination of antibiotics, and the fact that they do not require expertise and special technologies. Moreover, their use is suited when only a couple of antibiotics are needed to be tested. The price of each strip is relatively high, compared with the price of disks; therefore, gradient tests are usually used to test only a few antibiotics per strain. The incubation length of 16-24 h for gradient tests may represent a disadvantage, as more rapid automated systems are available with the reliable determination of MIC. Since its development in 1940, the disk diffusion (DD) test has remained the most widely used routine AST in clinical microbiological laboratories [57] . It has been standardised to test the susceptibility of the most common and clinically relevant bacteria that cause human diseases [42, 58] . The standardisation is a continuous process, and DD for many microorganisms/antimicrobials is an ongoing process [59] . The method is based on placing different antibiotic-impregnated disks on previously inoculated agar with bacterial suspension. The antibiotic diffuses radially outward through the agar medium, producing an antibiotic concentration gradient. After the inhibition zones are established within 24 h of incubation at 35 ± 1 • C, the zone diameters of each tested antibiotic are measured by the naked eye or using an automated system [60] . Obtained results should be interpreted and categorised according to the recommended clinical breakpoint of the standard in use [42, 61] . Disk diffusion is the most widely used AST method in microbiology laboratories because of its low cost and ease of performance and applicability of numerous bacterial species and antibiotics [32] . The choice of antibiotic disks is flexible and enables the clinical laboratory to make different combinations according to the bacterial species and the type of sample the isolate was obtained from [62] . Simple interpretation allows the detection of atypical phenotypes and visibility of contamination. However, the main disadvantages are the inability to determine the MIC and delays in getting the results. Reduction in turnaround time and timely treatment are of great importance for critically ill patients. In addition, the biological properties of lag and log phase of bacterial growth and their expression on antibiotic influence should be considered [63] . Nevertheless, methods to reduce incubation time for DD were suggested decades ago [64] [65] [66] . A revival of that idea led to the development of automated systems (WASPLab, Copan, Murrieta, CA, USA and BD Kiestra, Becton, Dickinson and Company, Franklin Lakes, NJ, USA) for an acceleration of AST by DD [67, 68] . The automatisation of the AST by DD leads to a shortening of the required time to obtain results and produce the final report [68] . EUCAST has defined a methodology of disk diffusion rapid AST (RAST), which is performed directly from positive blood culture bottles, with breakpoints for short incubations of 4, 6, and 8 h [69] [70] [71] . RAST can be implemented in routine laboratories without major investments. The method has been validated for a limited number of bacterial species and antibiotics so far. Furthermore, the combined use of a MALDI-TOF MS for the identification of bacteria and RAST directly from positive blood bottles enables reporting AST results within less than 24 h, which significantly reduced the turnaround time compared with the 24-48 h needed for culturing and classical AST methods, such as DD [72] . The predictive value of direct DD testing from positive blood cultures has been reported to have an important influence on AMS [73] . Quality control testing of media and antibiotics is of great importance for ensuring that disk diffusion is providing accurate and reliable results [59, 74] . In some cases, the DD method is more reliable than MIC determination. For instance, in the case of detecting penicillinase-producing S. aureus strains, the inhibition zone diameter combined with the EUCAST-based zone edge test is the most sensitive and specific phenotypic method [42, 75] . The DD method can be used for screening of susceptibility to a larger number of antibiotics or a whole class of antibiotics; detection of certain important resistance phenotypes, such as ESBL, carbapenemases, inducible resistance to macrolides ( Figure 2 ); or the presence of a heteroresistant population of bacterial species in a sample that cannot be detected by other phenotypic AST methods. The above-mentioned suggests that the DD method will remain a widely used AST method in the future. Since the introduction of the first chromogenic media for the detection of antibioticresistant bacteria 20 years ago, a variety of different media for the detection of clinically important resistant pathogens-such as MRSA, VRE, and ESBL-and carbapenemasesproducing or colistin-resistant Gram-negative bacteria-have been developed [76] [77] [78] [79] [80] . The main purpose of the development of chromogenic media was to enable more rapid detection and identification of resistant microorganisms. The target organisms are characterised by specific enzyme systems that metabolise the substrates to release the chromogen. The chromogen can then be visually detected by direct observation of a distinct colour change in the medium. Thus, these selective and differential media enable target pathogens to grow as coloured colonies. Compared with the use of conventional culture media, the use of chromogenic agar often reduces the costs and labour time [81] . Their primary use is for screening of patients colonised with various pathogens, and therefore they are valuable in infection prevention [82] and control of hospital-acquired infections. The sensitivity and specificity of chromogenic media depend on the manufacturer and the type of microorganism detected; thus, additional identification confirmation of the resistant bacteria is sometimes needed. Because of their wide applicability, new chromogenic media are being developed [83, 84] . Colourimetric tests represent phenotypic methods developed for the detection of AMR. They are based upon the bacterial enzymes hydrolysing test component, which is detected by the changes in pH values and the colour of chromogenic substances. Briefly, bacterial suspension or bacterial lysate suspension is added to a detection solution containing antibiotic and pH indicator dyes, such as phenol red, and incubated for a short period of time, no longer than a couple of hours. The pH of the detection solution changes due to the growth of antibiotic-resistant bacteria, or bacterial enzymes activity, and subsequently, the colour of the solution changes, which can then be visually observed. These tests were shown to be fast, easy to perform and interpret, and highly sensitive and specific. A good example of such colourimetric tests is Carba NP (bioMérieux, Marcy-l'Étoile, France), which detects carbapenemase-producing bacteria. The test gives reliable results in 30 min to 2 h, making it the quick and easy way to control carbapenemase producers [85] . Clinical microbiology laboratories are under increasing pressure to provide fast and reliable microbial identification (ID) and AST [86] . Automated and semi-automated devices for bacterial ID and AST are worthy of the task and have significantly improved laboratory efficiency. Nowadays, automation has been successfully implemented in most clinical microbiological laboratories to reduce turnaround times, increase efficiency, and improve cost-effectiveness [86, 87] . Various test systems-such as the VITEK 2 (bioMérieux, France), MicroScan Walkaway (Dade-Behring MicroScan, Deerfield, IL, USA), and Phoenix system (BD Diagnostic Systems, Baltimore, MD, USA)-have been widely used over the last decades. These instruments, using optical systems for measuring subtle changes, determine bacterial growth and antimicrobial susceptibility [88] and can produce results in a shorter time (6-12 h) than conventional manual assessment [36] . • VITEK 2 Systems-The first generation of VITEK system with a turnaround time of 13 h was developed for enumeration and identification of bacteria and yeasts in 1973. The VITEK 2 System, the next-generation of an instrument, is a BMD-based AST system that uses 64-well plastic cards containing 17-20 antimicrobial agents. If the bacterial isolate is not previously identified, one card is used for bacterial identification (ID card) and the other for antimicrobial susceptibility testing (AST card • Phoenix System-The Phoenix System is widely accepted and used in clinical microbiology laboratories for identification testing (ID) and antimicrobial susceptibility testing (AST). The principle of determining the susceptibility is based on the use of an oxidation-reduction indicator (resazurin dye or Alamar blue) and the detection of bacterial growth in the presence of various concentrations of the antimicrobial agent. In the Phoenix instrument, a maximum of 100 tests can be performed by using Phoenix ID/AST combination panels (51 for ID and 85 for AST). The instrument performs automatic reading at 20 min intervals during incubation for up to 18 h and provides accurate and rapid susceptibility results with easy workflow for the laboratory worker. In 2014, the new panel for susceptibility of Gram-negative bacteria was introduced for the Phoenix system to be used in combination with the BD Bruker MALDI-TOF [91] . Each of the above-mentioned systems has inherent advantages and limitations, and the results vary widely by antimicrobial drugs, software versions, and cards used. Hence, some of the systems are not reliable for correct categorisation of susceptibility profiles for certain drugs, leading to wrong classifications of susceptibility categories [92] . It seems that low inoculum size has a major influence on the outcome of these systems, with false susceptibilities being reported. Additionally, software updates and synchronisation of breakpoints according to the current standards are mandatory. Thus, it is incumbent upon the instrument manufacturer to keep pace with the breakpoint updates and make relevant improvements, such as extending the detection limit and verifying the performance of the AST system with the revised breakpoints internally, to avoid the problem of uncategorised results [93] . Panels usually contain only several concentrations of each antimicrobial agent, and the resulting MIC is not always given as an exact value. In contrast, classical BMD contains a wide range of doubling dilution antimicrobial concentrations for the determination of the MIC, thus obtaining the more precise value. In addition, according to the previously published reports, many of the resistance phenotypes are not easily detected using the automated susceptibility testing methods so prevalent in today's clinical laboratories [94] . Nonetheless, interestingly, the ability of automated systems to detect inducible resistance to clindamycin in 524 isolates of Staphylococcus spp. revealed sensitivity and specificity of 100% and 99.6%, respectively, for Phoenix, and 91.1% and 99.8%, respectively, for Vitek 2 [95] . The multicentre evaluation showed that categorical agreement between the Phoenix system and a BMD reference method for 2013 streptococcal isolates including Streptococcus pneumoniae, viridans group streptococci, and beta-haemolytic Streptococcus groups A, B, C, and G ranged from 92% to 100%, with one exception for viridans streptococci and penicillin, which was 87% [96] . However, according to the results of the evaluation of ASTs obtained using Vitek 2, Phoenix, and MicroScan, caution should be taken for AST of Stenotrophomonas maltophilia, as a high rate of errors may be observed [97] . Molecular AST directly detects specific resistance genes, as well as mutations in and expression of these genes. These molecular methods have been developed and tested as an alternative for or complementary to conventional AST and are generally faster than classic culture-based assays, with the test results available within one to a few hours [98] (Figure 3) . Most of the molecular AST methods fall into one of the three categories: amplification-based, hybridization-based, or sequence-based. In amplification-based methods, the target gene sequence is amplified to allow detection; in hybridization-based techniques, hybridized nucleic acid probes target gene sequences allowing detection; and in sequence-based approaches, genome sequences are analysed to detect resistance-conferring mutations or resistance genes. ity and specificity of 100% and 99.6%, respectively, for Phoenix, and 91.1% and 99.8%, respectively, for Vitek 2 [95] . The multicentre evaluation showed that categorical agreement between the Phoenix system and a BMD reference method for 2013 streptococcal isolates including Streptococcus pneumoniae, viridans group streptococci, and beta-haemolytic Streptococcus groups A, B, C, and G ranged from 92% to 100%, with one exception for viridans streptococci and penicillin, which was 87% [96] . However, according to the results of the evaluation of ASTs obtained using Vitek 2, Phoenix, and MicroScan, caution should be taken for AST of Stenotrophomonas maltophilia, as a high rate of errors may be observed [97] . Molecular AST directly detects specific resistance genes, as well as mutations in and expression of these genes. These molecular methods have been developed and tested as an alternative for or complementary to conventional AST and are generally faster than classic culture-based assays, with the test results available within one to a few hours [98] ( Figure 3) . Most of the molecular AST methods fall into one of the three categories: amplification-based, hybridization-based, or sequence-based. In amplification-based methods, the target gene sequence is amplified to allow detection; in hybridization-based techniques, hybridized nucleic acid probes target gene sequences allowing detection; and in sequence-based approaches, genome sequences are analysed to detect resistance-conferring mutations or resistance genes. The most widely used nucleic acid amplification-based method for the detection of specific resistance genes is polymerase chain reaction (PCR). Both real-time and conventional PCR rely on the amplification of nucleic acid sequences that encode resistance to an antibiotic. New PCR-based methods are being developed for the detection of genetic determinants of resistance to a variety of antibiotics for various bacterial species, as our knowledge about the genetic basis of antibiotic resistance increases [99] . Multiplex assays for simultaneous testing of multiple genetic determinants in various bacterial species have also been developed, i.e., multiplex assays for identifying numerous cephalosporinaseand carbapenemase-encoding genes, such as bla KPC , bla NDM , bla IMP , bla VIM , bla AmpC , bla TEM , bla SHV , and bla OXA , or mecA gene-encoding methicillin resistance in MRSA [100, 101] . Op-Gen, Inc. (Rockville, MD, USA) has recently released the multiplex-based Acuitas ® AMR Gene Panel that detects 28 genetic AMR markers, covering select drugs in nine classes of antibiotics, from 26 different pathogens [102] . The advantage of this test in comparison with other commercially available molecular tests is that it also detects non-beta-lactam resistance genes and those for what would be considered "last-resort antibiotics", such as colistin. Real-time PCR (quantitative PCR, qPCR) is one of the most ubiquitous methods found throughout clinical microbiology. Although costlier, qPCR offers several advantages over conventional PCR, including the measurement of data in real-time, greater sensitivity, reduced risk of carryover contamination, and greater amenity to multiplexing. Further advantages are that many systems are partially or even completely automated, such as GeneXpert ® Instrument Systems (Cepheid Corp., Sunnyvale, CA, USA) and BD MAX System platform (Becton Dickinson, Franklin Lakes, NJ, USA), which are easily operated and can be used for the detection of carbapenemases, ESBLs, MRSA, VRE, etc. [103] [104] [105] . The downside is that they are limited to using test assays only from specific manufacturers, with GeneXpert ® Instrument Systems requiring GeneXpert assays (Cepheid Corp., Sunnyvale, CA, USA) and BD MAX System using Check-Points ® qPCR assays (Wageningen, The Netherlands). Tests based on qPCR can also be used for phenotypic differentiation of resistant and susceptible strains due to its ability to measure genome copy numbers during bacterial growth in the presence of antibiotics [106] . The major disadvantage is that the system cannot provide information about the mechanism of resistance and that it requires the previous culture, meaning that the primary clinical samples cannot be used. DNA-microarrays are used to identify the presence of specific nucleic acid sequences using complementary short oligonucleotides immobilised on a solid surface. Since these oligonucleotides can be assembled onto solid surfaces in close proximity, this method could detect numerous sequences in a single assay, which would allow simultaneous, in parallel detection of different pathogens and detection of vast numbers of different resistance genes, as well as detecting numerous distinct mechanisms of resistance or variants of a single mechanism present in bacterial isolates, as opposed to PCR-based approaches [107] . The Verigene system (Luminex Corporation, Austin, TX, USA) has developed Blood Culture Multiplex Microarray-Based Molecular assays for rapid diagnostics of 12 Gram-positive and 9 Gram-negative bacteria, along with their associated resistance genes (i.e., mecA, vanA, vanB, bla KPC , bla NDM , bla IMP , and bla VIM ) [108] . In addition to qPCR-based assays, Check-Points ® has also developed the CHECK-MDR CT103 DNA microarray for the detection of the clinically most prevalent ESBLs and carbapenemases, as well as mobile colistin resistance (mcr) genes in Gram-negative bacteria [109] . Both of these microarray tests have shown high sensitivity (94-100%) and specificity (94-98%), and CHECK-MDR CT103 DNA microarray also showed the ability to discriminate between carbapenemase and ESBL variants of GES-type beta-lactamase [108, 109] . The advantages of currently available molecular-based methods are that they are direct, rapid, highly sensitive, and specific, thus potentially allowing the earlier administration of targeted therapy [98] . Furthermore, for some methods, direct clinical samples can be used. However, it should be noted that the presence of a resistance marker does not always have to correlate with phenotypic resistance. Additionally, the extent and intensity of gene expression are important parameters, as some genes need different expression levels to produce resistance. A potential solution to this issue would be the use of reverse transcription qPCR, which relies on the measurement of gene transcripts (RNA levels) instead of the presence of a gene [110, 111] . Another drawback is that these methods can only detect resistances that are searched for, and not novel or uncharacterised mechanisms of resistance, which could lead to false-negative results and inappropriate classification of resistant isolates as susceptible. A final consideration is that these methods are not capable of defining MIC values. As such, these methods have to be validated against phenotypic data to be useful, and extensive resistance marker databases and innovative bioinformatics methodologies are mandatory requirements. Nevertheless, molecular-based AST methods are a safe, efficient, and reliable screening tool in clinical settings. As experience with these tests grows, and as data are gathered on their efficacy and clinical impact, they will likely be more widely adopted. As DNA sequencing technology and bioinformatics pipelines for genome assembly and analysis advance, the possibility of using these techniques for the detection of antibiotic resistance opens. Applying whole-genome sequencing (WGS) would essentially enable the detection of all genes involved in AMR, which would help make comprehensive databases of all species-specific resistance factors (i.e., CARD-Comprehensive Antibiotic Resistance Database-https://card.mcmaster.ca, ResFinder-https://cge.cbs.dtu.dk/ services/ResFinder (accessed on 27 February 2022)) and make in silico AMR detection possible. Recent studies showed high concordance between the resistance profiles obtained using WGS and those obtained using phenotypic susceptibility testing, demonstrating that data obtained from genome sequences can correlate well with phenotypic resistance in some cases [112, 113] . In addition to genome-based resistome analyses, RNA-mediated transcriptomic approaches have also been described [114, 115] . Despite all of the advantages, WGS is not routinely performed in clinical practice. Considering the turnaround times of WGS, the existence of unknown resistance mechanisms, and the elevated cost compared with traditional and emerging techniques, the use of WGS for AST is not yet part of routine practice in clinical microbiology [116, 117] . Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) was discovered in the 1980s and introduced into the microbiological routine as an effective tool for bacterial and yeast identification about 15 years ago. It has been applied to classify the specific bacterial protein contents and their matching protein biomarkers because of its rapid turnaround time, low sample volume requirements, and per-sample costs [118] . Several MALDI-TOF MS-based methods have been proposed for rapid detection of antimicrobial resistance, including monitoring antibiotic modification by bacterial culture (e.g., beta-lactam hydrolysis [65, 119] ), acetylation of fluoroquinolones [120] , direct detection of proteins involved in specific resistance mechanisms [121, 122] , and detection of stable isotope labelling that requires expensive, isotopically labelled media [123, 124] . The hydrolysis of the target beta-lactam antibiotic, as shown by peak disappearance, is used to detect beta-lactamase-producing bacteria using MALDI-TOF MS. As a result, the assay for detecting carbapenemase production [125] automatically determines sensitivity or resistance depending on the degree of antibiotic hydrolysis. The method had 98% sensitivity and 100% specificity after 30 min of incubation of bacteria with the antibiotic, with both reaching 100% after 60 min of incubation [126, 127] . However, beta-lactam resistance is only recognized when it is mediated by beta-lactamases; alternative resistance mechanisms have not been elucidated; therefore, other tests should confirm negative results. Another assay for the detection of carbapenemases is a rapid and novel method using detonation nanodiamonds (DNDs) as a platform for the concentration and extraction of A. baumannii carbapenemase-associated proteins before MALDI-TOF MS analysis [128] . The sensitivity and the specificity of the proposed platform could reach 96% and 73%, as compared with traditional imipenem susceptibility testing, and 100% compared with PCR results. This method may detect the carbapenemases produced by A. baumannii in 90 min and does not require the addition of a carbapenemase substrate, as other mass spectrometric methods do. It is efficient for detecting other carbapenemase-producing bacteria. MALDI Biotyper-Antibiotic Susceptibility Test Rapid Assay (MBT-ASTRA) is an alternative MS-based method for AST which utilises semi-quantitative MALDI-TOF MS to measure the relative growth rates of bacterial isolates exposed to antibiotics compared with untreated controls during a short incubation step. A software tool calculates and compares the area under the curves (AUCs) of spectra of bacteria either exposed or not to an antibiotic [129] . In this method, if the microbial strain is susceptible, the AUC of the bacterial suspension with the antibiotic will be reduced compared with that without antibiotics, whereas with a resistant strain the AUCs with or without antibiotics will be comparable. The main advantage of the MBT-ASTRA is that the assay does not depend on the resistance mechanism and is utilisable with any antibiotic [130] . Moreover, it does not require specialised media or instrumentation, beyond the MALDI-TOF mass spectrometer. However, a drawback of the MBT-ASTRA assay is that the concentration of antibiotics used and the incubation time must be optimised for each species and antibiotic combination [129] . MBT-Resist assay, based on the detection of peak shift after stable isotope labelling, is an approach that uses the following principle: bacteria are grown in parallel in two distinct culture mediums, one containing 12C as a carbon component and the other containing 13C [131] . The system compares the mass spectrum of bacteria grown on an isotope-labelled medium with antibiotics to the mass spectrum of the same strain grown on an unlabelled medium without antibiotics. Resistant strains can thrive in the presence of antibiotics, incorporating 13C into the polypeptide, causing a shift in the peak to a higher m/z in the mass spectrum [131] . Antibiotic resistance by direct-on-target microdroplet growth assay (DOT-MGA) is a novel approach for detecting antimicrobial susceptibility in bacteria treated with breakpoint concentrations of antibiotic on the target plate of MALDI-TOF MS [132] . The best performance was obtained by recovering bacteria from positive blood cultures and after a 4 h incubation of microdroplets with or without meropenem at the breakpoint concentration. Under these conditions, 96.3% validity, 91.7% sensitivity, and 100% specificity were achieved. Recently, a screening panel for the detection of ESBL and AmpC beta-lactamase activity was developed [133] . Compared with the PCR results, positive percentage agreement values for ESBL, AmpC, and ESBL + AmpC resistance were 94.4%, 94.4%, and 100%, and negative percentage agreement values were 100%, 93.7%, and 100%, respectively. The accuracy of the DOT-MGA achieved results incomparable with those of the BMD assay, with a time saving of about 14 h, and higher than combination disk tests. According to Yoon et al., due to the great speed and simple application, MALDI-TOF MS would be the most suitable for endemic AMR clinical strains in specific settings, i.e., MRSA, VRE, CRAB, CRPA, and ESBL-, AmpC-, and carbapenemase-producing Enterobacterales [134] . The advantages and disadvantages of the commonly used methods of antimicrobial susceptibility testing were summarised in Table 1 . Time-consuming Increased ability to detect slow-growing or non-cultivable organisms Challenging interpretation of results * Qualitative; results are expressed as susceptible (S), susceptible, increased exposure (I), or resistant (R) based on established criteria from EUCAST. ** Quantitative; results are expressed as minimal inhibitory concentration (MIC) for each drug. Susceptibility reports should include interpretation of MIC, such as S, I, or R. *** Semiquantitative; results are expressed as MIC using three to four antimicrobial dilutions for each drug. Precise MIC values cannot be established if the MIC falls below or above the three to four dilutions used in the test panel. Susceptibility reports include interpretation of breakpoint MIC as S, I, or R. MALDI-TOF MS-matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Since there are a large number of antibiotics in use, it would be irrational and pointless to test the susceptibility of isolates to all of them. Although each laboratory should establish its own panels of antibiotics depending on the bacterial isolate, for which susceptibility is tested, there are some common rules regarding the choice of antibiotics [36] . The number and selection of antimicrobials tested are primarily dependent on the organism isolated, infection site, type of infection (community or hospital-acquired), comorbidities, patient's age, and gender, but also the institution's formulary, physician requests, and the automated panel or other testing methodology used [135] . Interpretation of AST results and reporting of bacterial susceptibility categories to antibiotics is based on the breakpoints published by the two most commonly used systems worldwide: CLSI and EUCAST [42, 61] . CLSI provides recommendations for agents that are important to test routinely (group A) and those that may be tested or reported selectively based on the institution's formulary (groups B and C) [136] . Accurate identification of bacteria is crucial in the choice of antibiotics [137] . It is well known that resistance mecha-nisms have not been observed in some bacterial species so far, e.g., continued penicillin susceptibility of Streptococcus pyogenes [26] . It is also known that some bacterial species are intrinsically resistant to particular antibiotics or antibiotic classes. It is systematically presented in the EUCAST document "Intrinsic Resistance and Unusual Phenotypes" [138] . Therefore, it is unnecessary to test certain bacterial species for activity in vitro, but microbiologists should report intrinsic resistance to clinicians. This document also indicates when to avoid the use of antimicrobial agents that are likely to result in treatment failure. Exceptional or unusual resistance phenotypes are also described in the EUCAST document [138] , and knowledge of those phenotypes allows microbiologists to further investigate the obtained results. Examples of such phenotypes are penicillin-resistant group A streptococci or vancomycin-resistant staphylococci. These results should be confirmed and checked before reporting. Organisms with unusual phenotypes could be seen due to the emergence of antibiotic resistance, but it is more likely that an error occurred during the strain identification, antibacterial susceptibility testing, or when the mixed culture was tested [139] . Due to the similar in vitro activities of some antimicrobials belonging to the same class, a representative antibiotic should be tested to predict susceptibility to other class members [27] . Thus, for example, in beta-haemolytic streptococci, the susceptibility to penicillins can be inferred from benzylpenicillin susceptibility results [42] . Susceptibility of staphylococci to beta-lactams can be assessed by susceptibility to cefoxitin, while susceptibility of staphylococci and streptococci to macrolides and fluoroquinolones could be predicted using erythromycin and norfloxacin, respectively [135] . In enterococci, ampicillin is used as an indicator of resistance to ampicillin, amoxicillin, and piperacillin with and without beta-lactamase inhibitor. It is not commonly used in Gram-negative bacteria due to their more heterogeneity in resistance. Until recently, both EUCAST and CLSI use "susceptible" (S), "intermediate" (I), and "resistant" (R) categories and also shared their definitions. While categories S and R still do not differ in interpretation in both standards and are easily understood, category I was not clear since it represented few definitions into one: uncertain therapeutic effect, appropriate to use when antibiotic is physiologically concentrated at the site of infection, appropriate to use a high dosage of the drug, etc. [140] . The above-described interpretation of the category I led clinicians to either avoid or misuse certain antibiotics [140] . Consequently, in June 2018 the EUCAST Steering Committee changed the definitions of AST categories to susceptible, standard dosing regimen (S), susceptible, with increased exposure (I), and resistant (R) [141] . This is done to emphasise the close relationship between the susceptibility of the organism and the exposure of the organism to antibiotics at the site of infection [142] , encouraging physicians to use antibiotics from the I category at the appropriate dosage. EUCAST breakpoints are based on dose and mode of administration as indicated in rationale documents and files of the recommended AST breakpoints [143] . Tables of recommended doses for antibiotics, standard dose regimen for S, and high doses for the new I category were also published [42] . This relatively new rule still needs to be implemented in healthcare professional training and clinical use [31] . Antimicrobial susceptibility categories help clinicians to predict the outcome of antibiotic treatment and provide them information on the recommended dosage. For some infections that require high doses of antibiotics for reaching therapeutic concentrations (e.g., meningitis), EUCAST listed meningitis breakpoints that are different from breakpoints for non-meningeal isolates [144] . Furthermore, for the treatment of infections caused by Pseudomonas spp., an increased antibiotic concentration on the site of infection is required. Therefore, according to the EUCAST recommendations, the majority of isolates belong to the category "Susceptible, increased exposure" when evaluating susceptibility to fluoroquinolones, penicillins, and third and fourth generations of cephalosporins, aztreonam, and imipenem [42] . The principle of the automated systems for AST is to interpret the results of MICs to the susceptibility category that corresponds to built-in standards (e.g., EUCAST, CLSI). This allows concurrent reporting of categories of susceptibility, values of MICs, and even some resistance phenotypes, providing valuable information for patient treatment. It also allows clinicians and pharmacologists to determine the appropriate dosage for each patient, which represents an individual or personalised medicine. Such an approach enables the estimation of the degree of resistance or susceptibility to the tested agent [31] . However, this assumption should not be based on a direct comparison of the MIC values obtained during testing, nor under the assumption that lower MIC values indicate greater sensitivity of tested bacteria in all cases. Additionally, MICs values can indicate lower effectiveness of the antibiotic, despite the susceptibility category. For example, the MIC value of 2 mg/L for vancomycin predicts the risk of treatment failure in tested S. aureus strains, even though it is still in the susceptible category [145] . Although both EUCAST/CLSI standards are constantly being improved and published annually, there are still limitations in reporting. Thus, interpretations of susceptibility to some common pathogens (e.g., Acinetobacter spp. and Stenotrophomonas maltophilia) and key antibiotics are lacking, resulting in the absence of AST reporting. Overall, the general recommendation is to report susceptibility to narrow-spectrum antibiotics whenever possible and to avoid reporting of broad-spectrum antibiotics that pose higher AMR selective pressure (carbapenems, fluoroquinolones) [146, 147] . Selective reporting of AST results is recognized as a useful tool for antibiotic stewardship. It implies performing AST according to usual practices, but the results are reported to the physician only for a few antibiotics recognised as a first-line choice [14] . Therefore, microbiologists play an essential role in AMS [148, 149] . As Tebano G et al. [14] propose, selective reporting might achieve the following: firstly, reducing unnecessary antibiotic prescriptions in cases of the detection of contaminants or members of the normal microbiota. Secondly, reducing inappropriate antibiotic prescriptions by persuading clinicians to choose narrowover broad-spectrum antibiotics. Both goals could be obtained by reducing the number of reported antibiotics or selective exclusion of certain antibiotics from the reporting where appropriate. Some countries established their National Antimicrobial Committee to address the challenges of AMR [150, 151] . These committees adapt the recommendations of EUCAST/CLSI in accordance with the epidemiological specificities of their countries and provide laboratory guidelines for antibiotic selection, testing, and reporting for their national-level use. In conclusion, evidence-based data shows that selective reporting leads to improved antibiotic usage, reducing both unnecessary and inappropriate prescribing. It can be concluded that, although numerous interventions have been proposed to reduce antibiotic consumption and bacterial resistance, much effort is still needed, especially in middle-and low-income countries. Therefore, wider and more consistent support in the implementation of these strategies is highly needed on a global scale [152] . Soon after the introduction of AST in clinical laboratories, the need for standardization of the process was recognized. In the past, several of the European national antimicrobial breakpoint Committees developed their own AST standards. Consequently, the national Committees initiated the standardisation of AST performance and interpretation according to the EUCAST recommendations [153] . A much older and widely used standard is the American CLSI, formerly NCCLS [61] . Although the process of harmonization of these two organizations is ongoing, there are still significant differences. All AST needs to be evaluated in order to provide a certain quality of the results. For conventional methods, manufacturers of antibiotic disks or tablets, gradient tests, and commercial microdilutions plates with a predetermined concentration of antibiotics have to provide a certain quality of the products, as do suppliers of automated or semi-automated tests. In addition, the microbiology laboratories are responsible for the adequate performance of the antimicrobial susceptibility tests. The control of the quality of performance is ensured through internal and external controls. Internal quality assessment should be provided daily or less frequently if the performance is consistent. External quality assessment tests the overall performance of the laboratory by a collection of bacterial strains of undisclosed susceptibility, usually sent by central international laboratories such as NEQUAS in the UK. After the results are sent back, a detailed evaluation of the participating laboratory performance is provided by the central laboratory [154] . Quality control is responsible only for the analytical phase of susceptibility testing, monitoring only the performance of the used test(s). The troubleshooting algorithm for AST QC is shown in Table 2 , adapted from the previously published report [155] . Quality assurance (QA), on the other hand, provides the proper performance of all three phases of testing: pre-analytical, analytical, and post-analytical. Major components of the QA program for susceptibility testing are the following: clinically relevant testing strategies, testing of reference quality control (QC) strains, technical competency, organism antibiogram verification, supervisor review of results, procedure manual, proficiency surveys. Quality control includes the following: the precision (repeatability) and accuracy of AST procedures, the performance of reagents used in the tests, the performance of persons who carry out the tests and read the results [156] . The AST procedures include diffusion susceptibility methods, dilution susceptibility testing, and different automated systems. The results of commercial automated systems are obtained more rapidly and are highly reproducible; the use of the automated system software supports the interpretation according to different standards and expert rules. Unusual resistance phenotype or resistance types that are difficult to detect may represent a challenge [36] . As for automated antimicrobial susceptibility systems, which are used more and more frequently in microbiological laboratories, usually, every device has its QC procedure. The most commonly used automated systems for antimicrobial susceptibility are VITEK 2 System (bioMérieux, France), BD Phoenix System (BD Diagnostic Systems, Baltimore, MD, USA), MicroScan WalkAway SI (Siemens Healthcare Diagnostics, West Sacramento, CA, USA), and TREK Sensititre (ARIS 2X, Trek Diagnostic Systems, Cleveland, OH, USA). Quality controls of such devices are not included in standards such as EUCAST or CLSI [61, 156, 157] . Agencies that provide the clearance of commercial systems for general use, such as the FDA for the USA, should use reference methods before the clearance for use is given. To evaluate or compare different susceptibility tests, usually these types of errors are considered: very major error, major error, and minor error. Very major error refers to the characterisation of the resistant isolate as susceptible. A major error is made when a susceptible isolate is characterised as resistant. When an intermediate isolate is characterised as susceptible or resistant, as well as when susceptible or resistant isolate is reported as intermediate, a minor error is made. Errors in antimicrobial susceptibility testing should be monitored by laboratories and carefully analysed. Potential problems either with the identification or susceptibility testing may be revealed [158] . For evaluation of new susceptibility tests or devices, a certain number of susceptible as well as resistant strains should be tested against different antibiotics. Additionally, all types of the above-mentioned errors should be evaluated [36] . For new antimicrobial susceptibility devices, acceptable performance according to FDA considers that major errors should not exceed 3% based on the number of susceptible organisms tested [159] . The acceptable numbers of very major discrepancies rate should be 7.5% or less (for the upper 95% confidence limit) and 1.5% or less (for the lower 95% confidence limit), based on the number of resistant organisms tested. Additionally, growth failure rates in the system have to be less than 10%. The above-mentioned criteria may not be identical for acceptable accuracy compared with the international standards on susceptibility test device evaluation [160] . In the last decades, several innovative approaches for AST have been developed. Some of the most promising platforms for the forthcoming period-MALDI-TOF, flow cytometry, and isothermal microcalorimetry-will be briefly described. Application of MALDI-TOF MS as routine rapid AST requires additional validation, such as standardised protocols, test kits, and software [161] . Further research efforts are needed to refine and optimise MALDI-TOF MS-based assays to obtain accurate and reliable results in the shortest possible time. A major focus of future research in this field will be to achieve the standardisation of methods and simultaneous susceptibility testing of microbes to various classes of antimicrobials [162] . So far, only two commercially available kits with software for automated interpretation of spectra have been authorised in Europe to detect carbapenemase activity or resistance toward third-generation cephalosporins in clinical microbiology laboratories. By using flow cytometry, the changes in morphology, physiological and metabolic activity, and survival of microorganisms can be monitored after exposure to antimicrobials. The effects of treatment can be seen within a few hours, suggesting a strong potential for AST. Using nuclear dyes that do not penetrate the cell walls of healthy organisms, the amount of dying and dead cells can be rapidly determined. In order to determine the emission spectrum, the cells must go through a flow channel and be excited by a laser to release dye [163] . Nevertheless, the future challenge is to improve the ability of a system to distinguish cellular damage caused by -cidal vs. -static antibiotics, as well as to address the issue of autofluorescence of certain species of bacteria. In addition, an enormous amount of work is required for the verification of the clinical database and the method itself. Isothermal microcalorimetry is a dynamic technique allowing the measurement of heat generated by the metabolism of actively growing cells. This technique is not new; however, it has been successfully adapted for AST of bacteria [164] . Additional advantages of this technique are that it requires a small culture volume, the testing is performed in sealed ampules, and monitoring during testing does not require manual manipulation. The completed analysis can also provide information on the -static vs. -cidal activity of an agent. Up to now, various alternative approaches to combat AMR have been proposed. For instance, epigenetic agents in modulating antibiotic susceptibility [165] , agents that inhibit virulence-conferring factors such as multidrug efflux inhibitors [166] , or targets' essential processes for bacterial survival, such as cell division inhibitor [167] , have been recently developed. If such antibacterial agents become licensed for treatment, antimicrobial susceptibility techniques should be updated along with the new strategies developed to counter AMR. The timely administration of appropriate antimicrobial therapy based on accurate AST is widely recognised as a cornerstone of the management of infectious diseases. Despite the shortcomings of the traditional AST methods, such as BMD and DD, their use is required in clinical practice to obtain the correct results according to the standardised protocols of EUCAST and CLSI, or for comparison against the results of novel techniques. Molecular-based methods are rapid, efficient, and reliable, with high specificity and sensitivity. Nonetheless, additional efforts are needed to bring closer NGS technology to routine microbiology. MALDI-TOF MS appears to be a promising highly specific system with reduced time and low cost of consumables. However, this method is not yet validated for all species and antimicrobials, and the acquisition costs of the system and its maintenance are high. Hence, further improvement in the currently used and novel AST methods and instruments is mandatory for speeding up the determination of antimicrobial efficacy in clinical microbiology laboratories in the foreseeable future. Understanding the mechanisms and drivers of antimicrobial resistance Global burden of bacterial antimicrobial resistance in 2019: A systematic analysis Bacterial co-infections with SARS-CoV-2 Impact of the COVID-19 pandemic on the surveillance, prevention and control of antimicrobial resistance: A global survey SARS-CoV-2, bacterial co-infections, and AMR: The deadly trio in COVID-19? By 2050, Drug-Resistant Infections Could Cause Global Economic Damage on Par with Clinical perspective of antimicrobial resistance in bacteria Mechanisms of antibiotic resistance Antimicrobial use: Risk driver of multidrug resistant microorganisms in healthcare settings Executive Summary: Implementing an Antibiotic Stewardship Program: Guidelines by the Infectious Diseases Society of America and the Society for Healthcare Epidemiology of America Developmental roadmap for antimicrobial susceptibility testing systems Selective reporting of antibiotic susceptibility testing results: A promising antibiotic stewardship tool Monitoring and Evaluation of the Global Action Plan on Antimicrobial Resistance: Framework and Recommended Indicators Surveillance of antimicrobial resistance in low-and middle-income countries: A scattered picture Stewardship Interventions: A Practical Guide; World Health Organization Antimicrobial stewardship and antibiograms: Importance of moving beyond traditional antibiograms Demonstration of the weighted-incidence syndromic combination antibiogram: An empiric prescribing decision Aid Antibiotic regimens for treatment of infections due to multidrug-resistant Gram-negative pathogens: An evidence-based literature review Management of adults with hospital-acquired and ventilator-associated pneumonia: 2016 clinical practice guidelines by the Infectious Diseases Society of America and the M39-A4: Analysis and Presentation of Cumulative Antimicrobial Susceptibility Test Data: Approved Guideline A combination antibiogram evaluation for Pseudomonas aeruginosa in respiratory and blood sources from intensive care unit (ICU) and Non-ICU Settings Current and emerging methods of antibiotic susceptibility testing Current and emerging techniques for antibiotic susceptibility tests Current concepts in laboratory testing to guide antimicrobial therapy A fast impedance-based antimicrobial susceptibility test M07: Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically The agar microdilution method-A new method for antimicrobial susceptibility testing for essential oils and plant extracts Antibacterial effect of Juniperus communis and Satureja montana essential oils against Listeria monocytogenes in vitro and in wine marinated beef The minimum inhibitory concentration of antibiotics: Methods, interpretation, clinical relevance Methods for in vitro evaluating antimicrobial activity: A review Antibiotics 2022 Time-kill kinetics of antibiotics active against rapidly growing mycobacteria Issues in pharmacokinetics and pharmacodynamics of anti-infective agents: Kill curves versus MIC In vitro antibacterial and time-kill evaluation of the Erythrina caffra Thunb. extract against bacteria associated with diarrhoea Antimicrobial susceptibility testing: A review of general principles and contemporary practices Screening of the antibacterial effect of Juniperus sibirica and Juniperus sabina essential oils in a microtitre plate-based MIC assay A new rapid resazurin-based microdilution assay for antimicrobial susceptibility testing of Neisseria gonorrhoeae Evaluation of ComASP TM Colistin (formerly SensiTest TM Colistin), a commercial broth microdilution-based method to evaluate the colistin minimum inhibitory concentration for carbapenemresistant Klebsiella pneumoniae isolates Agar and broth dilution methods to determine the minimal inhibitory concentration (MIC) of antimicrobial substances Comparison of agar dilution and antibiotic gradient strip test with broth microdilution for susceptibility testing of swine Brachyspira species Breakpoint Tables for Interpretation of MICs and Zone Diameters, Version 12.0 Verification of Ceftazidime-Avibactam and Ceftolozane-Tazobactam Susceptibility Testing Methods against Carbapenem-Resistant Enterobacteriaceae and Pseudomonas aeruginosa Performance of ceftolozane-tazobactam etest, MIC test strips, and disk diffusion compared to reference broth microdilution for-Lactam-Resistant Pseudomonas aeruginosa isolates Tigecycline susceptibility of multidrug-resistant Acinetobacter baumannii from intensive care units in the western Balkans The search for a practical method for colistin susceptibility testing: Have we found it by going back to the future? Performance characteristics of newer MIC gradient strip tests compared with the Etest for antimicrobial susceptibility testing of Neisseria gonorrhoeae E-test versus agar dilution for antibiotic susceptibility testing of Helicobacter pylori: A comparison study Test: An Alternative Method for Susceptibility Testing of Mycobacterium tuberculosis Antifungal Susceptibly Testing by Concentration Gradient Strip Etest Method for Fungal Isolates: A Review Current antimicrobial susceptibility testing for beta-lactamase-producing Enterobacteriaceae in clinical settings Comparison of nine phenotypic methods for detection of extended-spectrum β-lactamase production by Enterobacteriaceae Evaluation of metallo-beta-lactamase production in multiple antibiotic-resistant Pseudomonas spp. and Acinetobacter baumannii strains EUCAST Guidelines for Detection of Resistance Mechanisms and Specific Resistances of Clinical and/or Epidemiological Importance Detection of AmpC beta-lactamase in Escherichia coli: Comparison of three phenotypic confirmation assays and genetic analysis Antibiotics 2022 Evaluation of a new etest vancomycin-teicoplanin strip for detection of glycopeptide-intermediate Staphylococcus aureus (GISA), in particular, heterogeneous GISA A method for the assay of penicillin M02: Performance Standards for Antimicrobial Disk Susceptibility Tests Routine and Extended Internal Quality Control for MIC Determination and Disk Diffusion as Recommended by EUCAST, Version 12.0 Evaluation of the Scan ® 1200 as a rapid tool for reading antibiotic susceptibility testing by the disc diffusion technique M100: Performance Standards for Antimicrobial Susceptibility Testing Standardisation of disk diffusion results for antibiotic susceptibility testing using the sirscan automated zone reader Rapid phenotypic detection of microbial resistance in gram-positive bacteria by a real-time laser scattering method Rapid automated antimicrobial susceptibility testing of Streptococcus pneumoniae by use of the bioMerieux VITEK 2 Rapid antimicrobial susceptibility testing of isolates from blood cultures by direct inoculation and early reading of disk diffusion tests Evaluation of a rapid Bauer-Kirby antibiotic susceptibility determination Shortening the incubation time for antimicrobial susceptibility testing by disk diffusion for Enterobacteriaceae: How short can it be and are the results accurate? Fully automated disc diffusion for rapid antibiotic susceptibility test results: A proof-of-principle study The EUCAST rapid disc diffusion method for antimicrobial susceptibility testing directly from positive blood culture bottles The European Committee on Antimicrobial Susceptibility Testing. Zone Diameter Breakpoints for Rapid Antimicrobial Susceptibility Testing (RAST) Directly from Blood Culture Bottles Screening for ESBL and Carbapenemases in E. coli and K. pneumoniae for Epidemiological Purposes as Part of the RAST Procedure. EUCAST Guidelines for Detection of Resistance Mechanisms and Specific Resistance of Clinical and/or Epidemiological Importance Rapid identification of microorganisms from positive blood cultures by MALDI-TOF mass spectrometry subsequent to very short-term incubation on solid medium Predictive value of direct disk diffusion testing from positive blood cultures in a children's hospital and its utility in antimicrobial stewardship The quality of antimicrobial discs from nine manufacturers-EUCAST evaluations in 2014 and 2017 Performance of penicillinase detection tests in Staphylococcus epidermidis: Comparison of different phenotypic methods New chromogenic identification and detection of Staphylococcus aureus and methicillin-resistant S. aureus Evaluation of a novel chromogenic agar medium for isolation and differentiation of vancomycin-resistant Enterococcus faecium and Enterococcus faecalis isolates Evaluation of a New Selective Chromogenic Agar Medium for Detection of Extended-Spectrum-Lactamase-Producing Enterobacteriaceae Evaluation of CHROMagar KPC for rapid detection of carbapenem-resistant Enterobacteriaceae Evaluation of CHROMagar Acinetobacter for detection of enteric carriage of multidrug-resistant Acinetobacter baumannii in samples from critically ill patients A decade of development of chromogenic culture media for clinical microbiology in an era of molecular diagnostics Trends in molecular characteristics and antimicrobial resistance of group B streptococci: A multicenter study in Serbia A new chromogenic medium for isolation of Bacteroides fragilis suitable for screening for strains with antimicrobial resistance CHROMagar COL-APSE: A selective bacterial culture medium for the isolation and differentiation of colistin-resistant Gram-negative pathogens Rapid Detection of Carbapenemase-producing Enterobacteriaceae Antimicrobial susceptibility testing: Currently used methods and devices and the near future in clinical practice Accuracy of three automated systems (MicroScan WalkAway, VITEK, and VITEK 2) for susceptibility testing of Pseudomonas aeruginosa against five broad-spectrum beta-lactam agents Real-time optical antimicrobial susceptibility testing Evaluation of the VITEK 2 system for identification and antimicrobial susceptibility testing of medically relevant gram-positive cocci Use of the VITEK 2 System for rapid identification of clinical isolates of Staphylococci from bloodstream infections Automated and Manual Systems for Antimicrobial Susceptibility Testing of Bacteria Comparison of five commonly used automated susceptibility testing methods for accuracy in the China Antimicrobial Resistance Surveillance System Comparison of meropenem MICs and susceptibilities for carbapenemase-producing Klebsiella pneumoniae isolates by various testing methods Comparison of BD Phoenix, Vitek 2, and MicroScan Automated Systems for detection and inference of mechanisms responsible for carbapenem resistance in Enterobacteriaceae Comparison of BD Phoenix and bioMérieux Vitek 2 automated systems for the detection of macrolide-lincosamide-streptogramin B resistance among clinical isolates of Staphylococcus Multicenter evaluation of the BD Phoenix Automated Microbiology System for antimicrobial susceptibility testing of Streptococcus species Evaluation of the Vitek 2, Phoenix, and MicroScan for antimicrobial susceptibility testing of Stenotrophomonas maltophilia In vitro antimicrobial susceptibility testing methods: Agar dilution to 3D tissue-engineered models Emerging mechanisms of antimicrobial resistance in bacteria and fungi: Advances in the era of genomics Clinical and molecular characteristics of OXA-72-producing Acinetobacter baumannii ST636 outbreak at a neonatal intensive care unit in Serbia Detection of the mecA gene and identification of Staphylococcus directly from blood culture bottles by multiplex polymerase chain reaction Multicenter Evaluation of the Acuitas AMR Gene Panel for Detection of an Extended Panel of Antimicrobial Resistance Genes among Bacterial Isolates Evaluation of three commercial assays for rapid detection of genes encoding clinically relevant carbapenemases in cultured bacteria Cepheid|Molecular Testing of Healthcare Associated Infections PCR-based detection methods for single-nucleotide polymorphism or mutation: Real-time PCR and its substantial contribution toward technological refinement Use of a commercially available microarray to characterize antibioticresistant clinical isolates of Klebsiella pneumoniae Evaluation of Verigene Blood Culture Test Systems for Rapid Identification of Positive Blood Cultures Evaluation of the Check-Points Check MDR CT103 and CT103 XL Microarray Kits by Use of Preparatory Rapid Cell Lysis Selection and validation of reference genes for gene expression studies in Klebsiella pneumoniae using Reverse Transcription Quantitative real-time PCR A rapid antimicrobial susceptibility test for determining Yersinia pestis susceptibility to doxycycline by RT-PCR quantification of RNA markers Using genomics to track global antimicrobial resistance. Front. Public Health Comparisons of whole-genome sequencing and phenotypic drug susceptibility testing for Mycobacterium tuberculosis causing MDR-TB and XDR-TB in Thailand RNA markers for ultra-rapid molecular antimicrobial susceptibility testing in fluoroquinolone-treated Klebsiella pneumoniae Simultaneous detection of genotype and phenotype enables rapid and accurate antibiotic susceptibility determination Practical issues in implementing whole-genome-sequencing in routine diagnostic microbiology The role of whole genome sequencing in antimicrobial susceptibility testing of bacteria: Report from the EUCAST Subcommittee Advances in mass spectrometry for the identification of pathogens Rapid direct detection of carbapenemase-producing Enterobacteriaceae in clinical urine samples by MALDI-TOF MS analysis Rapid detection of the plasmid-mediated quinolone resistance determinant AAC(6')-Ib-cr in Enterobacteriaceae by MALDI-TOF MS analysis Discrimination between wild-type and ampicillin-resistant Escherichia coli by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry Use of matrixassisted laser desorption ionization-time of flight mass spectrometry to identify vancomycin-resistant enterococci and investigate the epidemiology of an outbreak MALDI biotyper-based rapid resistance detection by stable-isotope labeling Rapid detection of antibiotic resistance based on mass spectrometry and stable isotopes Rapid detection of OXA-48-producing Enterobacteriaceae by matrix-assisted laser desorption ionization-time of flight mass spectrometry Performance evaluation of MALDI-TOF MS MBT STAR-BL versus in-house Carba NP testing for the rapid detection of carbapenemase activity in Escherichia coli and Klebsiella pneumoniae Strains Rapid detection of KPC-Producing Enterobacterales susceptible to imipenem/relebactam by using the MALDI-TOF MS MBT STAR-Carba IVD assay. Front Direct detection of carbapenemase-associated proteins of Acinetobacter baumannii using nanodiamonds coupled with matrix-assisted laser desorption/ionization time-of-flight mass spectrometry A suitable tool for fast antibiotic susceptibility testing? Methods Semi-quantitative MALDI-TOF for antimicrobial susceptibility testing in Staphylococcus aureus Establishing drug resistance in microorganisms by mass spectrometry New microbiological techniques in the diagnosis of bloodstream infections Rapid Detection of extended-spectrum β-lactamases (ESBL) and AmpC β-lactamases in Enterobacterales: Development of a screening panel using the MALDI-TOF MS-based direct-ontarget microdroplet growth assay MALDI-TOF mass spectrometry technology as a tool for the rapid diagnosis of antimicrobial resistance in bacteria Susceptibility testing and reporting of new antibiotics with a focus on tedizolid: An international working group report The antibiogram: Key considerations for its development and utilization. JAC-Antimicrob A guide to bacterial culture identification and results interpretation The European Committee on Antimicrobial Susceptibility Testing. Intrinsic Resistance and Unusual Phenotypes Interpretative reading: Recognizing the unusual and inferring resistance mechanisms from resistance phenotypes Point-Counterpoint: Differences between the European Committee on Antimicrobial Susceptibility Testing and Clinical and Laboratory Standards Institute Recommendations for reporting antimicrobial susceptibility results Breakpoint Tables for Interpretation of MICs and Zone Diameters, Version 9 Widespread implementation of EUCAST breakpoints for antibacterial susceptibility testing in Europe The clinical significance of vancomycin minimum inhibitory concentration in Staphylococcus aureus infections: A systematic review and meta-analysis RCPAQAP audit of antimicrobial reporting in Australian and New Zealand laboratories: Opportunities for laboratory contribution to antimicrobial stewardship Antimicrobial stewardship in the microbiology laboratory: Impact of selective susceptibility reporting on ciprofloxacin utilization and susceptibility of gram-negative isolates to ciprofloxacin in a hospital setting Microbiological surveillance and antimicrobial stewardship minimise the need for ultrabroad-spectrum combination therapy for treatment of nosocomial infections in a trauma intensive care unit: An audit of an evidence-based empiric antimicrobial policy Antimicrobial stewardship in South Africa: A scoping review of the published literature Antibiotics 2022 Recommendations of the Spanish Antibiogram Committee (COESANT) for selecting antimicrobial agents and concentrations for in vitro susceptibility studies using automated systems Antimicrobial stewardship in Australia: The role of qualitative research in programme development Impact of national interventions to promote responsible antibiotic use: A systematic review Development of the EUCAST disk diffusion antimicrobial susceptibility testing method and its implementation in routine microbiology laboratories Laboratory Guide: Methodologies for Antimicrobial Susceptibility Testing Quality Assurance in Antimicrobial Susceptibility Testing Disk Diffusion Test Manual, Version 10 Antimicrobial Susceptibility Testing. In Koneman's Color Atlas and Textbook of Diagnostic Microbiology FDA. Antimicrobial Susceptibility Test (AST) Systems-Class II Special Controls Guidance for Industry Clinical Laboratory Testing and In Vitro Diagnostic Test Systems-Susceptibility Testing of Infectious Agents and Evaluation of Performance of Antimicrobial Susceptibility Test Devices-Part 2: Evaluation of Performance of Antimicrobial Susceptibility Test Devices Against Reference Broth Micro-Dilution Advances in rapid identification and susceptibility testing of bacteria in the clinical microbiology laboratory: Implications for patient care and antimicrobial stewardship programs Detection of antibiotic-resistance by MALDI-TOF mass spectrometry: An expanding area The exclusive use of flow cytometry to evaluate the antibiotic-susceptibility Isothermal micro calorimetry-A new method for MIC determinations: Results for 12 antibiotics and reference strains of E. coli and S. aureus Antibiotic resistance and epigenetics: More to it than meets the eye New multidrug efflux inhibitors for gram-negative bacteria Bacterial cell division proteins as antibiotic targets Funding: This review received no external funding. Data Availability Statement: Not applicable. The authors declare no conflict of interest.