key: cord-0855817-f3kso3jw authors: Han, Ping; Go, Maybelle K.; Chow, Jeng Yeong; Xue, Bo; Lim, Yan Ping; Crone, Michael A.; Storch, Marko; Freemont, Paul S.; Yew, Wen Shan title: A high-throughput pipeline for scalable kit-free RNA extraction date: 2021-12-01 journal: Sci Rep DOI: 10.1038/s41598-021-02742-w sha: db54ef078e5b97b2de99a94eefb61b110a4d2c62 doc_id: 855817 cord_uid: f3kso3jw An overreliance on commercial, kit-based RNA extraction in the molecular diagnoses of infectious disease presents a challenge in the event of supply chain disruptions and can potentially hinder testing capacity in times of need. In this study, we adapted a well-established, robust TRIzol-based RNA extraction protocol into a high-throughput format through miniaturization and automation. The workflow was validated by RT-qPCR assay for SARS-CoV-2 detection to illustrate its scalability without interference to downstream diagnostic sensitivity and accuracy. This semi-automated, kit-free approach offers a versatile alternative to prevailing integrated solid-phase RNA extraction proprietary systems, with the added advantage of improved cost-effectiveness for high volume acquisition of quality RNA whether for use in clinical diagnoses or for diverse molecular applications. www.nature.com/scientificreports/ tube format is incompatible with high-throughput applications. Here we demonstrate that a well-established, TRIzol-based, non-kit RNA extraction protocol can be adapted into a semi-automated, high-throughput format using equipment readily available in research and hospital laboratories. Moreover, recent establishment of Biofoundries around the world, with their versatile and scalable automation pipelines, is poised to lend firm infrastructural support to our approach 4 . Nonetheless, even where automated liquid handling capabilities are lacking, the adapted protocol for a 96-well format can still be carried out manually using a multi-channel pipette, albeit at a slower pace yet remaining cost-effective for acquisition of RNA from large numbers of samples. The TRIzol method, while widely proven effective in RNA extraction, poses challenges to adaptation for highthroughput processing performed on automated liquid handling instruments due to the multiple steps involved. The need for phase separation, precipitation by centrifugation and potential phenol carryover further complicates the workflow. In this study, the conventional protocol was adapted to a 96-well format with emphasis on miniaturization and reduced human intervention (Fig. 1 ). To validate that the semi-automated, non-kit extraction method was able to isolate RNA at sufficient yield and quality for sensitive downstream applications, biological samples were spiked with non-infectious, synthetic SARS-CoV-2 virus-like-particles (VLPs) and heat-inactivated 2019 novel coronavirus and the extracted RNA evaluated for SARS-CoV-2 detection via real-time reverse transcription qPCR (RT-qPCR). While prevailing clinical diagnoses of SARS-CoV-2 are mostly conducted on nasopharyngeal or oropharyngeal specimens, several studies have shown that saliva presents a low-cost, noninvasive and reliable alternative for SARS-CoV-2 detection [5] [6] [7] . As such, saliva and pharyngeal swabs were used as diagnostic samples in this study due to the ease of self-collection from volunteers, negating the need for skilled personnel for sample collection. Importantly, saliva can be highly viscous and difficult to pipet and was presented as an assessment on the automated liquid handlers' capability to exact extraction from specimens with varying viscosity. For comparison, analyses were also carried out on oropharyngeal nasal mid-turbinate (OP-NMT) swabs performed by a trained medical personnel. Two laboratory automation platforms were employed, namely, the Opentrons OT-2 lab robot and the Eppendorf epMotion® 5075, both which were equipped with an 8-channel volumetric dispensing arm and a thermal block for 96-well plates. In order to perform RNA isolation in 96-well format, the reaction volume was scaled down proportionally in a 3:1 TRIzol reagent-to-sample ratio, reducing the volume of reagent required by sevenfold to 105-μL, thereby directly lowering the consumable cost per sample extraction. With a concomitant decrease in input sample to a volume of just 45-μL, glycogen was added as a co-precipitant to aid in RNA recovery. In a kit-free extraction method, precipitation and purification of RNA by centrifugation remain ineludible steps, making manual intervention obligatory in the workflow. The addition of glycogen as an inert carrier for nucleic acids was crucial in overcoming the challenge of low centrifugal force when using swing-out rotors for plates. The adapted procedure successfully extracted 1 to 4 ng/μl of RNA from both saliva and throat swabs with the yield slightly higher from saliva ( Fig. 2A) . During transfer of the aqueous phase following phase separation, cross-contamination with proteins, lipids, DNA and phenol can occur. Indeed, ensuring high quality of isolated RNA, particularly in a miniaturized format, proved challenging. Phenol carryover was identified in sample extractions using both automated liquid handlers as indicated by a characteristic absorbance maximum at 270 nm. Despite the relatively low purity and quantity, the extracted RNA was tested for its suitability for use in sensitive downstream applications such as real time RT-qPCR. The phenol contamination did not appear to interfere with the efficacy of the RT-qPCR targeted at www.nature.com/scientificreports/ the ubiquitous human Ribonuclease P (RNase P) from both the saliva and throat (Fig. 2B) . To determine the reproducibility in analytical efficiency, the validation was repeated using three commercially available one-step RT-qPCR master mixes (Fig. 2C ). The Ct values showed slight variability for the KAPA qPCR mix compared to the Luna and iTaq mix but the results, nonetheless, suggested that the extracted RNA was of sufficient yield and quality for downstream diagnostic application irrespective of the molecular reagent kits. (B) Threshold cycles (C T ) of one-step RT-qPCR analysis on total RNA extracted from saliva (blue cirlce) and throat swabs (red triangle) using Opentron-2 (OT2) and Eppendorf epMotion (epM). Analysis targeting human Ribonuclease P (RNase P) was performed on five technical replicates. (C) Detection of RNase P using multiple qPCR master mixes. Analysis was performed with Luna (NEB; blue diamond), iTaq (Bio-Rad; red square) and KAPA (Kapa Biosystems; green inverted triangle) one-step SYBR RT-qPCR kits. Total RNA was extracted from throat swabs using Eppendorf epMotion. Statistical differences were analysed using unpaired, two-tailed t-test where ns indicates non-significance (p > 0.05). Scientific Reports | (2021) 11:23260 | https://doi.org/10.1038/s41598-021-02742-w www.nature.com/scientificreports/ To assess viral RNA extraction efficiency of the miniaturised, semi-automated TRIzol-based extraction protocol against commercial RNA purification kits, biological samples were spiked with heat-inactivated 2019 novel coronavirus (nCoV) and RNA extracted using the QIAamp Viral RNA Mini Kit specific for viral RNA isolation from body fluids, the RNeasy Mini Kit for generic purification of total RNA from cells as well as the TRIzol-based method performed manually at standard volume. To ensure consistent sample input, 45 ul of biological samples were each spiked with a fixed quantity of 4,500 copies of nCoV, and topped up with DMEM to the varying starting sample volumes recommended for each kit. Kit-extracted RNA was eluted according to manufacturer's instructions, followed by volume reduction by evaporation to 10 ul to standardize with that used in the miniaturised TRIzol-based method before performing RT-qPCR assay targeted at RNaseP and SARS-CoV-2 N gene. The CDC 2019-nCoV N1 primer set was used as it has been shown to be more sensitive than the N2 or N3 sets, typically generating lower Ct values from positive samples 8, 9 . The standard TRIzol-based method gave RNA recovery comparable to the QIAamp Viral RNA kit and RNeasy Mini kit with Ct values of 24.3, 24.1, 24.7 for RNaseP and 27.7, 28.1, 28.1 for nCoV-N1 detection, respectively (Fig. 3A, B) . The results were similar to that reported by Won et. al., 2020, indicating that TRIzol was equally effective in extracting RNA compared to commercial available kits 10 . While the miniaturised, 96-well-adapted TRIzol-based extraction yielded adequate RNA to amplify the target genes, the method showed a lower sensitivity as observed by the higher Ct values for RNaseP (Ct = 26.3) and nCoV-N1 (Ct = 30.6). The limit of detection of the semi-automated workflow was also examined with tenfold serial dilutions (10 1 -10 5 copies per reaction) of SARS-CoV-2 VLPs and nCoV. The highest quantified concentration in the standard curve for VLPs was limited to 10 4 copies per reaction as the available stock concentration was 1700 copies per μL. The extracted RNA showed a detection limit of 10 copies of VLP per reaction, while that of nCoV was 100 copies per reaction, suggesting good efficiency of extraction using the high-throughput platform (Fig. 4) . Upon demonstrating the efficiency of the miniaturised, TRIzol-based RNA extraction using the semi-automated workflow, the platform was further validated on mock patient samples loaded with SARS-CoV-2 RNA. www.nature.com/scientificreports/ Saliva and throat swab samples were collected from 24 volunteers and SARS-CoV-2 VLPs and nCoV were separately added into 8 samples each at 1 × 10 5 copies/ml to simulate clinically relevant concentrations 11 . In the SYBR RT-qPCR assay, up to 3 false positive signals with significant Ct values (Ct < 37) was detected out of the expected 16 positive spiked samples (Fig. 5) . Examination of the melt profiles revealed nonconcordant melt peaks, indicating the occurrence of amplification artefacts in both saliva and throat negative control samples. Amplification of nonspecific products in quantitative PCR using SYBR Green-based detection is not uncommon and presents compromises in specificity and reproducibility 12 . Negative detection was observed for the internal control primer set targeted at RNaseP, especially in throat swab samples, in which up to 50% of the 24 samples did not yield signals. This could largely be due to the variability in skill of swab sampling since this was self-administered by untrained volunteers. Non-detection of internal control was noticeably reduced in saliva samples, occurring in 1 to 3 out of 24 samples. These results lend support to the use of saliva as a reliable alternative for detection of SARS-CoV-2 as it alleviates the need for trained personnel for specimen collection and reduces the reliance on swabs and viral transport media. To circumvent the challenge of false positives, which puts the diagnostic reliability into question, the analysis was repeated using a multiplex, probe-based RT-qPCR kit deploying three primer-probe sets targeting the ORF1ab region 1 and 2 of the SARS-CoV-2 genome as well as an internal control (Fig. 6) . Expectedly, the probebased assay displayed higher specificity compared to the SYBR Green-based method, producing no false positives in any of the SARS-CoV-2-negative samples. Indeed, all biological samples spiked with VLPs did not generate positive signals as these recombinant VLPs were engineered to carry the SARS-CoV-2 N gene and lacked the ORF1ab region targeted by the RT-qPCR assay used. In a bid to verify if the partial detection of internal control was indeed the result of inconsistencies in selfsampling by untrained volunteers, sampling was repeated by a medically trained personnel familiar with oropharyngeal nasal mid-turbinate swabbing performed in routine Covid-19 testing. This time, detection for the RNaseP internal control in the SYBR RT-qPCR assay was positive for all 24 volunteers (Fig. 7A) . The result suggests that the RNA extraction platform is reliable in a realistic, diagnostic setting. Instead of VLPs, 8 biological samples were spiked with 1 × 10 5 copies/ml nCoV and another 8 with a tenfold lower concentration of 1 × 10 4 copies/ml nCoV. Occurrence of amplification artefacts recurred as observed in 4 random, false positive signals www.nature.com/scientificreports/ (Ct ≤ 37) in the negative, water-spiked controls, underlining the shortcoming in specificity of SYBR Green-based detection, particularly for diagnostic analyses (Fig. 7A) . Analyses by both SYBR Green-and probe-based RT-qPCR gave positive detection in all 16 SARS-CoV-2-spiked samples, with Ct values higher for those spiked with 1 × 10 4 copies/ml nCoV (Fig. 7A, B) . As exemplified by the recent, unanticipated surge in Covid-19 testing across the world, supply chain issues in commercial RNA extraction kits can present a rate-limiting obstacle to swift and extensive clinical diagnosis. In a bid to ease the reliance on proprietary, high-throughput RNA extraction kits and their dedicated instruments, a semi-automated, high-throughput kit-free workflow was designed with miniaturization and minimal manual operation in mind (Fig. 1) . This approach is notably more cost-effective as kit-based RNA isolation is more expensive than TRIzol-based RNA extraction (Table 1) . By further lowering the sample volume from hundreds down to less than 50 µl and by processing in the 96-well format, the semi-automated workflow significantly reduces processing time, but, more importantly, lowers the operational cost which can translate to substantial savings for healthcare test facilities. Based on actual prices for all consumables used, including general laboratory plastic wares such as collection tubes, pipette tips and 96-well plates, as well as chemicals and reagents purchased from local vendors, the cost per sample was as low as less than USD 5 to up to USD 15 without taking into consideration the capital cost for instruments. The price disparity is primarily attributable to the broad price range between the affordable universal SYBR-Green dye-based kits and the more specific but costlier probe-based assays. Even in the absence of automated liquid handlers, the 96-well adapted protocol can still be performed manually using an 8-channel pipette at comparable speed, albeit more labour-intensive. It can be argued that emerging diagnostic strategies involving extraction-free direct RT-qPCR detection offers an even more attractive option that will not only drive the cost lower but also greatly reduce time required for sample processing [13] [14] [15] . While holding great promise for SARS-CoV-2 detection, its prospective application for medical surveillance of future novel infectious diseases will require re-validation, thereby necessitating the reliance on the assured reliability of conventional, prior RNA isolation, at least in the early stage of routine clinical testing. Sample collection. Pharyngeal, saliva and oropharyngeal nasal mid-turbinate samples were collected from 24 volunteers. Approval was sought from the ethics committee/institutional review board, the National University of Singapore Office of Safety, Health and Environment (NUS OSHE). All experiments were performed in accordance with NUS OSHE guidelines and regulations. Informed consent was obtained from all volunteers. The pharyngeal, or throat, swab was performed using a sterile polyester swab with a polystyrene shaft (Deltalab, Spain) according to instructions detailed in Won et al. 2020 10 . After collection, the swab was immediately suspended in 400 μl Dulbecco's Modified Eagle Medium (DMEM, Gibco, USA) and mixed vigorously to transfer the specimen. Saliva collection was carried out by briefly swirling the saliva in the mouth, pooling and spitting into a sterile 5 ml tube. Oropharyngeal nasal mid-turbinate swabs were conducted by a medically trained healthcare worker using a sterile polyester swab with a polystyrene shaft (Deltalab, Spain) which was suspended directly in 400 μl Dulbecco's Modified Eagle Medium (DMEM, Gibco, USA) and mixed vigorously to transfer the specimen. Consumption of food, water and brushing of teeth was avoided at least an hour before sampling. All samples were kept on ice until RNA extraction on the same day. Positive controls for SARS-CoV-2 were prepared using MS2 bacteriophage SARS-CoV-2 virus-like-particles (VLPs, 1700 copies/μl) provided by Crone et al. 2020 8 and heat-inactivated 2019 novel coronavirus (nCoV, 3.75 × 10 5 genome copies/μl, VR-1986HK, ATCC) with serial dilutions made using diethylpyrocarbonate (DEPC)treated water as diluent. For mock SARS-CoV-2 positive samples in the 24 healthy volunteers, both saliva and throat samples from 8 volunteers were spiked with VLPs and another 8 volunteers with nCoV at 1 × 10 5 copies/ ml concentration before RNA extraction. Separately, the oropharyngeal nasal mid-turbinate samples were spiked with 1 × 10 4 and 1 × 10 5 copies/ml nCoV for 8 volunteers each. www.nature.com/scientificreports/ by pipetting 20 times and allowed to incubate for 10 min at room temperature before centrifugation at 3200 × g at 4 °C for 20 min. The supernatant was aspirated and the remaining pellet washed with 150 μl of 75% ethanol by slowly pipetting 5 times before centrifugation again at 3200 × g at 4 °C for 10 min. The washing step was carried out twice and the pellet dried at 65 °C for 5 min on a heating block. To solubilise the RNA pellet, 10 μl of DEPCtreated water pre-warmed to 65 °C was added and mixed slowly by pipetting 20 times. The RNA suspension was allowed to incubate for 10 min at room temperature to further improve solubility before use in PCR analysis. The RNA concentration was determined with a microplate reader (BioTek, USA) using the Quant-iT RNA assay kit (Thermo Fisher Scientific, USA). RNA purification by commercial kits, Qiagen QIAamp Viral RNA kit and RNeasy Mini kit, was carried out according to manufacturer's instructions using the recommended starting sample volume and final elution volume. The volume of the RNA elute was brought down to a standardised volume of 10 ul by vacuum centrifugation before RT-qPCR analysis. Quantitative reverse transcription PCR (RT-qPCR). For SYBR Green-based detection, RT-qPCR was carried out using Luna Universal One-Step RT-qPCR Kit (New England Biolabs, USA), iTaq Universal SYBR Green One-Step Kit (Bio Rad, USA) and KAPA SYBR FAST One-Step qRT-PCR Master Mix Kit (Kapa Biosystems, USA). In the reaction setup for all kits, 1 μl of extracted RNA was added to half the standard reaction, i.e. 10 μl, volume. The following primer sets were used: human RNase P (5′-CGG TGT TTG CAG ATT TGG AC, 5′-CGA TAT GAT TGA TAG CAA CAA CTG AAT AGC ), US CDC SARS-CoV-2 N1 (5′-GAC CCC AAA ATC AGC GAA AT, 5′-TCT GGT TAC TGC CAG TTG AAT CTG ). RT-qPCR was performed following manufacturer's thermal cycling protocol using the Bio Rad CFX96 Real-Time PCR Detection System (Bio Rad, USA) with the threshold setting determined by the CFX Maestro software. In addition, qualitative detection specific for SARS-CoV-2 was conducted using the MiRXES Fortitude Kit 2.0 RUO (MiRXES, Singapore), which is a one-step, multiplex real-time RT-PCR kit with three primer-probe sets targeting the ORF1ab region 1 and 2 of the SARS-CoV-2 genome as well as an internal control. The PCR reaction setup and run profile was carried out according to manufacturer's instruction with the recommended threshold setting at 50. All data generated or analysed during this study are included in this published article (and its Supplementary Information file). Received: 16 May 2021; Accepted: 16 November 2021 PCR-based diagnostics for infectious diseases: uses, limitations, and future applications in acute-care settings Leveraging open hardware to alleviate the burden of COVID-19 on global health systems Material shortages hamper testing for novel coronavirus Building a global alliance of biofoundries Saliva is a reliable, non-invasive specimen for SARS-CoV-2 detection Evaluating the use of posterior oropharyngeal saliva in a point-of-care assay for the detection of SARS-CoV-2 Saliva or nasopharyngeal swab specimens for detection of SARS-CoV-2 A role for Biofoundries in rapid development and validation of automated SARS-CoV-2 clinical diagnostics Analytical sensitivity and efficiency comparisons of SARS-CoV-2 RT-qPCR primer-probe sets Development of a Laboratory-safe and Low-cost Detection Protocol for SARS-CoV-2 of the Coronavirus Disease 2019 (COVID-19) Viral load of SARS-CoV-2 in clinical samples Amplification of nonspecific products in quantitative polymerase chain reactions (qPCR) Extraction-free SARS-CoV-2 detection by rapid RT-qPCR universal for all primary respiratory materials Saliva sampling and its direct lysis, an excellent option to increase the number of SARS-CoV-2 diagnostic tests in settings with supply shortages Evaluating the efficacy of RT-qPCR SARS-CoV-2 direct approaches in comparison to RNA extraction generated the VLPs used in the study. The manuscript was written with contributions from all authors We would like to thank the Global BioFoundries Alliance (https:// biofo undri es. org/) for facilitating this collaborative work between the London BioFoundry at the Imperial College Translation and Innovation Hub and the Singapore BioFoundry at the NUS Synthetic Biology for Clinical and Technological Innovation. We would also like to express our deep appreciation to Yvonne Low from the National University Health System for her skills and time in conducting the oropharyngeal nasal mid-turbinate swabbing of the volunteers. This work was supported by a grant from the National Research Foundation, Singapore. M.A.C., M.S. and P.S.F. acknowledge funding from UKRI-EPSRC (EP/R014000/1, EP/S001859/1), UKRI-BBSRC (BB/M025632/1) and the UK Dementia Research Institute. The authors declare no competing interests. Correspondence and requests for materials should be addressed to W.S.Y.Reprints and permissions information is available at www.nature.com/reprints.Publisher's note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http:// creat iveco mmons. org/ licen ses/ by/4. 0/.