key: cord-0862427-kp23em25 authors: Dutta, Nirmita; Lillehoj, Peter B.; Estrela, Pedro; Dutta, Gorachand title: Electrochemical Biosensors for Cytokine Profiling: Recent Advancements and Possibilities in the Near Future date: 2021-03-23 journal: Biosensors (Basel) DOI: 10.3390/bios11030094 sha: baae774f7370e468a715b1b28500e89c7c41f50c doc_id: 862427 cord_uid: kp23em25 Cytokines are soluble proteins secreted by immune cells that act as molecular messengers relaying instructions and mediating various functions performed by the cellular counterparts of the immune system, by means of a synchronized cascade of signaling pathways. Aberrant expression of cytokines can be indicative of anomalous behavior of the immunoregulatory system, as seen in various illnesses and conditions, such as cancer, autoimmunity, neurodegeneration and other physiological disorders. Cancer and autoimmune diseases are particularly adept at developing mechanisms to escape and modulate the immune system checkpoints, reflected by an altered cytokine profile. Cytokine profiling can provide valuable information for diagnosing such diseases and monitoring their progression, as well as assessing the efficacy of immunotherapeutic regiments. Toward this goal, there has been immense interest in the development of ultrasensitive quantitative detection techniques for cytokines, which involves technologies from various scientific disciplines, such as immunology, electrochemistry, photometry, nanotechnology and electronics. This review focusses on one aspect of this collective effort: electrochemical biosensors. Among the various types of biosensors available, electrochemical biosensors are one of the most reliable, user-friendly, easy to manufacture, cost-effective and versatile technologies that can yield results within a short period of time, making it extremely promising for routine clinical testing. Cytokines are a class of soluble signaling proteins, ranging in size from 8 to 40 kDa, which act as mediators of innate and adaptive immunity. Cytokines are secreted in response to an inflammatory stimulus by nearly all nucleated cells, particularly immune cells or leucocytes. While structurally versatile, cytokines are grouped together based on their biological functions, which are similar in principle and often orchestrated in an interdependent manner [1, 2] . Various types of inflammatory stimulus, such as antigenic stimulation, UV light, heat-shock or other forms of stress inducer, can result in the production of cytokines [1] . Antigenic pathogen associated molecular patterns (PAMPS) and damage associated molecular patterns (DAMPS) trigger an immune response when encountered by immune cells at the site of an injury or pathogen invasion. PAMPS are recognized by pathogen recognition receptors (PRRs), the most important of which are Toll-like receptors (TLRs) on macrophages and dendritic cells. The result is a cascade of molecular signals transduced through common pathways, such as nuclear factor kappa-light-chain-enhancer of activated B cells (NF-κB) and mitogen-activated protein kinase (MAPK) pathways, which activate transcription factors responsible for switching on cytokine encoding genes. Major cytokineproducing sources are T cells, B cells, macrophages, dendritic cells and natural killer cells. cytokine-producing sources are T cells, B cells, macrophages, dendritic cells and natural killer cells. Gene expression shows transient behavior, and the expressed cytokines, upon secretion, bind to their respective receptors on the surface of target cells, leading to further signal transduction via common pathways, such as Janus kinase-signal transducer and activator of transcription (JAK-STAT), NF-κB, MAPK and apoptotic Caspase pathways [2] . The cytokine family includes interleukins (ILs), chemokines (CXCLs), interferons (IFNs), tumor necrosis factors (TNFs), mesenchymal growth factors, transforming growth factors (TGFs), adipocytokines and the hematopoietic growth factors [2, 3] . Cytokines do not function independently, but act in sync with each other in a tight-knit immunoregulatory system (shown in Figure 1 [4] ). Cytokine function is accomplished by multiple mechanisms, including pleiotropy (cytokines conveying different signals in different target cells), redundancy (multiple cytokines conveying similar signals), synergism (multiple cytokines conveying a signal combinatorially), antagonism (cytokines conveying opposing or inhibitory signals) or cascade induction (a network of cytokines successively inducing the production of more types of cytokines) [2, 5] . Cytokines facilitate crosstalk between immune cells at all stages of an inflammatory response, which can either promote (proinflammatory) and or reduce (anti-inflammatory) this response [1, 6] . Cytokines also work in close association with antigen recognition molecules, such as T cell receptors, B cell receptors, major histocompatibility complex classes I and II (MHC I and II) and PRRs, on the surface of antigen presenting cells. Some cytokines can induce their own production and amplify the target immune response, while others can regulate the class switching of antibody encoding genes in differentiating B cells [5] . In addition to playing an indispensable role in inflammation, cytokines play an important role in hematopoietic differentiation, embryonic development and degenerative processes [7] . [4] with permission from PubMed Central. Cytokines show aberrant upregulation, or downregulation, as part of an immune response due to a disease or physiological condition, including a bacterial/viral infection, cancer, autoimmunity or neurodegeneration. In pathogenic infections, the differential expression of cytokines can provide some information about the type and extent of an acute immune response. The cytokine expression profile can also help distinguish between the [4] with permission from PubMed Central. Cytokines show aberrant upregulation, or downregulation, as part of an immune response due to a disease or physiological condition, including a bacterial/viral infection, cancer, autoimmunity or neurodegeneration. In pathogenic infections, the differential expression of cytokines can provide some information about the type and extent of an acute immune response. The cytokine expression profile can also help distinguish between the immune response patterns in bacterial and viral infections [8, 9] . In this respect, it is Biosensors 2021, 11, 94 3 of 42 important to emphasize the potential role of cytokines in the assessment of the pathogenesis of emerging infectious diseases, such as the novel coronavirus disease , which is one of the deadliest pandemics in the 21st century. Recently, Dhar et al. conducted a meta-analysis to understand the correlation between the cytokines overexpressed in COVID-19 patients and disease pathogenesis. The results suggested that two cytokines, IL-6 and IL-10, showed significantly raised levels and could be quantified to estimate the risk and severity of the disease [10] . While all physiological conditions generate an immune response, cancers and autoimmune diseases are two illnesses that systematically manipulate the immune system to support their own survival [11, 12] . In cancer, an inadequate or altered tumor-suppressive cytokine expression renders the immune system debilitated of its killing competence. In a study by Ashizawa et al., high serum levels of IL-6 was associated with gastric cancer and lymph node metastasis [13, 14] . Yun et al. showed the similarity between the STAT phosphorylation pattern induced by the regulatory T cell-subset (T reg ) specific cytokine milieu between colorectal cancer tumor-microenvironment and peripheral blood [15] . A study performed on human lung cancer cell lines and biopsies showed an overexpression of IL-6 and IL-10 mRNAs typical of the helper T cell-subset 2 (T H 2) type response, overexpression of immunosuppressive TGFβ mRNA, and high secretion levels of IFNγ, granulocyte-macrophage colony-stimulating factor (GM-CSF) with reduced levels of IL-2, typical of the helper T cell-subset 1 (T H 1) type response [16] . Another study on human bladder cancer cells established a link between increased secretion of macrophage migration inhibitory factor and cell proliferation and survival [17] . Pancreatic carcinoma cells were shown to overexpress mRNAs of various cytokines (IL-1β, IL-6, IL-10, IFNγ, TGF-β1, etc.) and IL-1β, TGF-β2, TGF-β3 proteins [18] . It has also been shown that cytokines can significantly change the tumor microenvironment by recruiting immune cells and increasing the infiltration of cytotoxic and phagocytic cells; thus, it has been proposed to use cytokines as therapeutic targets for their tumor-suppressive activity, including IFNγ, IL-2, IL-10, TGFβ and GM-CSF [19] . In autoimmunity, the immune system is triggered by self-antigens, which results in a systematic reaction that attacks healthy tissue. It is often mediated by autoreactive T cells which escape common checkpoint mechanisms (e.g., thymic deletion, development into natural T reg or peripheral anergy), leading to an imbalance between immunostimulatory and immunosuppressive cytokines. A study by Ioannou and Isenberg demonstrated an association between high levels of IL-2 secreted by T H cells with thyroiditis, rheumatoid arthritis and other arthropathies [20] . IL-12 subunit p40 encoding gene polymorphism has been associated with type I diabetes in humans [21] . Autoimmune diseases are found to be inhibited by the increased secretion of anti-inflammatory IL-10 and TGFβ by T reg cells [22] . Neurodegeneration is the result of rapid damage and eventual death of neurons in a delayed response to injury. Inflammatory pathways might have a role to play here, though infiltration by immune cells in the central nervous system is barred, but inducible. There has been evidence of the involvement of IL-1, TNFα (promoting) and TGFβ (protective) in neurodegenerative processes [23] . The manipulation of cytokine expression and serum concentration can alter the direction and severity of an immune response, and serve as a potential target for immunotherapeutic procedures [24] . (i) Interleukin-6 (IL-6) is a 21-26 kDa glycoprotein, 212 amino acid residues long, with a pleiotropic effect, secreted in higher levels in response to an inflammatory stimulus [25, 26] . It can mediate both pro-and anti-inflammatory responses by binding to its specific receptor IL-6R on target cells [27] . Secreted by leucocytes, osteoblasts, fibroblasts, mesenchymal, endothelial and many other cells [26] , IL-6 is found in the serum of a healthy human in concentrations of~2.91 ± 6.45 pg/mL [28] . IL-6 induces naïve B cells to proliferate and differentiate into plasma cells, induces the proliferation of T cells and pluripotent hematopoietic cells, mediates acute phase response in the liver and stimulates the produc-tion of C-reactive protein and fibrinogen [26, 27] . Altered levels of IL-6 have been shown to be associated with various diseases and conditions, such as Alzheimer's disease [29] , major depressive disorders [30] , cardiovascular diseases [31] , cancers such as breast, prostate and gastric cancers [32] , and autoimmune diseases, such as rheumatoid arthritis, chronic inflammatory proliferative disease and systemic lupus erythematosus [33] . (ii) Tumor necrosis factor alpha (TNFα) is a 157 amino acid residues long [34] ,~17 kDa polypeptide that exists as a serum soluble homotrimer [35] . Primarily produced by activated macrophages, TNFα binds to its receptor TNFαR on the surface of target cells, promoting the activation of proinflammatory genes, differentiation of immune cells into effectors and even the induction of apoptotic pathways in some cells [35] . The serum level of TNFα in healthy individuals is~3.21 ± 4.04 pg/mL [28] , and altered levels of TNFα have been linked with several diseases and conditions, including autoimmune diseases such as chronic inflammatory bowel diseases (Crohn's disease) [36] , rheumatoid arthritis [20, 37] , ankylosing spondylitis [38] and psoriasis [39] , and cancers [40] . (iii) Interferon gamma (IFNγ) is a~20 kDa, 146 amino acid residues long glycosylated protein of the type II interferon family, which exists as a homodimer in blood [41, 42] . Primarily secreted by NK cells and activated T cells, IFNγ binds to the IFNγ R 1 and R 2 receptor complex, and can inhibit viral replication, activate macrophages as part of the innate immunity, and is a major modulator of the T cell mediated immune response [42, 43] . The serum level of IFNγ in healthy individuals is~13.1 ± 22.7 pg/mL [28] . Elevated levels of IFNγ is associated with several illnesses, such as rheumatoid arthritis and multiple sclerosis [44, 45] , and is widely considered a clinical biomarker for early-stage tuberculosis diagnosis [46] . (iv) Transforming growth factor-beta 1 (TGF-β1) is a~25 kDa protein [47] involved in regulatory or homeostatic function upon effector T cells and T reg cells [48] . The physiological level of TGF-β1 in healthy individuals is 4.1 ±2.0 ng/mL [49] . Secretion of TGF-β1 in blood serum at elevated levels is linked to several conditions, including liver fibrosis [50] , bladder carcinoma [51] and renal disease [52] . TGF-β1 is also thought to alleviate inflammation in chronic neurodegeneration [53] . Levels of cytokines in healthy and diseased individuals generally range from pg/mL to ng/mL in blood/serum, and can be at much lower concentrations (fg/mL) in other bodily fluids, such as saliva, sweat and urine [28, 54] , and therefore requires highly sensitive detection techniques. Conventional methods for cytokine detection and quantification include enzyme-linked immunosorbent assay (ELISA) [55] [56] [57] [58] [59] [60] [61] , Western Blot [56, 62] , flow cytometer assisted bead-based immunoassays [63, 64] , enzyme-linked immuno spot assay (ELISPOT) and fluorophore-linked immuno spot assay (FluoroSpot) [65] . ELISA is currently the gold standard for quantitative detection of cytokines used in clinical laboratories [66] . Other detection techniques, including those still in the process of development, include polymerase chain reaction (PCR)-based amplification of cytokine mRNAs [67] , capillary electrophoresis [68] , surface plasmon resonance-based assays [69] , electrochemiluminescent assays [70] [71] [72] , fluoroimmunoassays [73, 74] , radioimmunoassays [58, 75] , immune-PCR assay [76] and electrochemical assays. Several other review articles have recently been published focusing on various aspects of cytokine detection. Stenken et al. reviewed common analytical techniques for cytokine detection [65] , and Liu et al. and Mobed et al. discussed strategies for ultrasensitive detection of cytokines using immunosensors [77, 78] . Loo et al. discussed the prospect of using paper-based electrochemical biosensors for cytokine and cancer biomarkers detection [13] . Recently, Campuzano et al. reviewed electrochemical biosensors, with the exception of FET-biosensors, for the detection of inflammatory cytokines published in the past five years [79] . Here, we present a comprehensive review on electrochemical biosensors for quantitative detection of cytokines for the diagnosis and assessment of human diseases and conditions, including cancers, autoimmune diseases and neurodegenerative diseases, published within the past two decades. Electrochemical detection harnesses chemical/biochemical reactions taking place at or near an electrode surface, and the changes in the electrochemical double layer or in electron transfer characteristics at the electrode interface. The biochemical reactions that are most widely exploited are enzymatic reactions, highly specific antigen-antibody binding, and antigen-specific aptamer folding interactions. The information regarding the quantifiable biochemical reaction is relayed in the form of an electrical signal, and read in measurable electrical parameters, such as current (I), voltage (V), resistance (R), conductance (G), capacitance (C) or phase shift (θ). A simplified schematic of a typical electrochemical biosensor is shown in Figure 2 . Biosensors 2021, 11, x FOR PEER REVIEW 5 of 46 diseases and conditions, including cancers, autoimmune diseases and neurodegenerative diseases, published within the past two decades. Electrochemical detection harnesses chemical/biochemical reactions taking place at or near an electrode surface, and the changes in the electrochemical double layer or in electron transfer characteristics at the electrode interface. The biochemical reactions that are most widely exploited are enzymatic reactions, highly specific antigen-antibody binding, and antigen-specific aptamer folding interactions. The information regarding the quantifiable biochemical reaction is relayed in the form of an electrical signal, and read in measurable electrical parameters, such as current (I), voltage (V), resistance (R), conductance (G), capacitance (C) or phase shift (θ). A simplified schematic of a typical electrochemical biosensor is shown in Figure 2 . The most common setup used to construct an electrochemical biosensor is the threeelectrode system comprising a working electrode (WE, where the biochemical reactions take place to generate a proportionate current); counter electrode (CE, which completes the current loop); and a reference electrode (RE, electrically isolated, with a constant potential, against which the potential of WE is calibrated). All three electrodes are immersed in and connected by an electrolyte solution. For some electrochemical techniques, the electrolyte solution contains one or more electroactive redox species to enhance the electron transfer rate between the electrochemical reaction and sensor surface for signal amplification. Several detection techniques are applied to electrochemical biosensors that differ in the mode of application of potential and signal generation, which can be broadly classified as potential step or potential sweep methods. Potential step methods, such as chronoamperometry (CA), involve stepping the potential applied to the WE from a base potential. The resulting current signal is correlated with the analyte concentration. Alternatively, potential sweep methods involve the gradual scanning of the WE potential from a base value, driving the redox reaction either unidirectionally or bidirectionally. Some common methods are cyclic voltammetry (CV), differential pulse voltammetry (DPV), square wave voltammetry (SWV) and anodic stripping voltammetry (ASV). CV makes use of linear scanning of the WE potential in a triangular waveform (cyclic scanning) at a constant rate. DPV uses voltage pulses of fixed amplitude superimposed on a linear potential ramp; while in SWV, the square waveform voltage of a fixed, but large, amplitude is superimposed on the potential ramp, such as to induce reversible redox reaction in each pulse. In ASV, a metal serves as the redox species, which is first deposited at the cathode at a negative potential, followed by scanning the potential in a pulse waveformpositive sweep. In the latter step, reduced metals are reoxidized and stripped out of the WE. In each of these techniques, the peak current across the potential sweep is correlated to the analyte concentration. Electrochemical impedance spectroscopy (EIS) is another technique that involves the application of an alternating, sinusoidal voltage over a range of frequencies to the WE, and the resulting electrical impedance or phase shift is correlated to the analyte concentration [80] . Apart from these, other electrochemical techniques include potentiometric approaches based on field-effect transistors (FETs). FET biosensors utilize a semiconductor-based transducer as a current path across source and drain electrodes, to which an external voltage is applied capacitatively through a gate electrode [81, 82] . The authors of this review have utilized the above-mentioned electrochemical techniques to achieve ultrasensitive detection of analytes in many biosensors [81, [83] [84] [85] [86] [87] [88] [89] [90] [91] [92] . With the increasing discovery of new clinically relevant cytokines, and the multifunctional nature of their action, there is a pressing need for new technologies that can rapidly detect and quantify cytokines with high sensitivity and accuracy. Recent advances in nanotechnology, such as the discovery of metallic and magnetic nanoparticles, carbon nanotubes, and quantum dots [93] , has enabled ultrasensitive electrochemical detection of cytokines. Various electrochemical biosensors have been demonstrated for cytokine detection which can be categorized by their sensing strategy or transduction mechanism, such as aptasensors [94] , enzyme-labeled immunosensors [95] , self-assembled monolayer (SAM)-enabled impedimetric immunosensors [96] , heavy metal nanoparticle-labeled immunosensors [97] , redox-labeled immunosensors [98] , magnetoimmunosensors [99] and FET-based biosensors [100] . Additionally, biosensors exhibiting an overlap of sensing strategies or combined with other technologies (e.g., microfluidics) have been demonstrated for cytokine detection [101] . For each category of electrochemical biosensors, the design and sensor preparation process as well as sensor performance parameters are presented in Tables 1-18 . Among all sensor performance parameters, the most common parameters used to evaluate the sensor performance are the limit of detection (LOD), detection range, specificity, reproducibility and stability. The LOD is the lowest concentration that can be detected and is typically calculated as three times the standard deviation (SD) above the mean detection signal at zero concentration of the target biomarker [102] . A different approach to estimating the LOD is by calculating three times the SD of the blank signal, divided by the slope of the calibration plot. As different methods of calculating LOD exist, care should be taken when comparison LODs from different references. The detection range is the span of concentrations that can be detected and is often expressed as a linear regression equation. Specificity provides an indication of how accurate the biosensor can detect the target biomarker in the presence of other biomarkers and interfering species in the sample. Reproducibility provides an indication of the biosensor accuracy over multiple measurements (commonly using different sensing electrodes) and is often expressed in terms of statistical parameters, such as the SD, relative standard deviation (RSD) or coefficient of variation (CV). Most biosensors are tested using buffer samples or biofluids (blood, serum, saliva, etc.) obtained from healthy volunteers spiked with cytokines, which can provide a suitable assessment of device functionality. Further evaluation of biosensor performance is performed using clinical samples, and the results are typically compared with a gold standard technique. Aptasensors employ aptamers (Apts), which are short, single-stranded oligonucleotides, as biorecognition elements. Compared with antibodies, Apts are an attractive candidate for biorecognition, owing to the ease of in vitro synthesis, modification and stability over a range of temperature and pH conditions [103] . Tertis et al. developed an impedimetric aptasensor for IL-6 detection consisting of anti-IL-6 Apts immobilized on polypyrrole and gold nanoparticle (AuNP)-coated screen printed carbon electrodes (SPCEs). Polypyrrole is an electroactive polymer with a low polymerization potential and, in combination with AuNPs, exhibits excellent biocompatibility. AuNPs provide a large surface area for Apt immobilization, thereby enhancing electron transfer [104] . This study showed how the immobilization of negatively charged oligonucleotides (i.e., Apt) on AuNPs can be accomplished much faster when coupled with a pulsed, externally applied oscillating potential compared with a constant applied potential [94] . Kumar et al. demonstrated an impedimetric aptasensor, which was used for measurements of IL-6 in sweat and serum. This sensor consisted of AuNPs and thiolated Apts immobilized on an Au electrode [105] . Liu et al. developed an electrochemical aptasensor using a methylene blue (MB)-conjugated Apt specific for TNFα onto an Au electrode. At room temperature, these Apts formed a hairpin structure, bringing the 5' conjugated MB very close to the electrode, enabling passage of faradaic currents ( Figure 3 ). Upon TNFα binding, the Apt straightens causing the MB to be situated far away from the electrode surface, reducing the passage of current in a proportionate manner [106] . that can be detected and is typically calculated as three times the standard deviation (SD) above the mean detection signal at zero concentration of the target biomarker [102] . A different approach to estimating the LOD is by calculating three times the SD of the blank signal, divided by the slope of the calibration plot. As different methods of calculating LOD exist, care should be taken when comparison LODs from different references. The detection range is the span of concentrations that can be detected and is often expressed as a linear regression equation. Specificity provides an indication of how accurate the biosensor can detect the target biomarker in the presence of other biomarkers and interfering species in the sample. Reproducibility provides an indication of the biosensor accuracy over multiple measurements (commonly using different sensing electrodes) and is often expressed in terms of statistical parameters, such as the SD, relative standard deviation (RSD) or coefficient of variation (C̅ V). Most biosensors are tested using buffer samples or biofluids (blood, serum, saliva, etc.) obtained from healthy volunteers spiked with cytokines, which can provide a suitable assessment of device functionality. Further evaluation of biosensor performance is performed using clinical samples, and the results are typically compared with a gold standard technique. Aptasensors employ aptamers (Apts), which are short, single-stranded oligonucleotides, as biorecognition elements. Compared with antibodies, Apts are an attractive candidate for biorecognition, owing to the ease of in vitro synthesis, modification and stability over a range of temperature and pH conditions [103] . Tertis et al. developed an impedimetric aptasensor for IL-6 detection consisting of anti-IL-6 Apts immobilized on polypyrrole and gold nanoparticle (AuNP)-coated screen printed carbon electrodes (SPCEs). Polypyrrole is an electroactive polymer with a low polymerization potential and, in combination with AuNPs, exhibits excellent biocompatibility. AuNPs provide a large surface area for Apt immobilization, thereby enhancing electron transfer [104] . This study showed how the immobilization of negatively charged oligonucleotides (i.e., Apt) on AuNPs can be accomplished much faster when coupled with a pulsed, externally applied oscillating potential compared with a constant applied potential [94] . Kumar et al. demonstrated an impedimetric aptasensor, which was used for measurements of IL-6 in sweat and serum. This sensor consisted of AuNPs and thiolated Apts immobilized on an Au electrode [105] . Liu et al. developed an electrochemical aptasensor using a methylene blue (MB)-conjugated Apt specific for TNFα onto an Au electrode. At room temperature, these Apts formed a hairpin structure, bringing the 5' conjugated MB very close to the electrode, enabling passage of faradaic currents ( Figure 3 ). Upon TNFα binding, the Apt straightens causing the MB to be situated far away from the electrode surface, reducing the passage of current in a proportionate manner [106] . . Schematic representation of an TNF-α aptasensor with methylene blue (MB) as a redox label. In the absence of the target, Apt hairpins remain folded with MB reporters in proximity to the electrode, ensuring efficient electron transfer and a measurable faradaic current. Upon target binding, the redox tag moves far away from the electrode and the redox current decreases. Reprinted from [106] with permission from The Royal Society of Chemistry. . Schematic representation of an TNF-α aptasensor with methylene blue (MB) as a redox label. In the absence of the target, Apt hairpins remain folded with MB reporters in proximity to the electrode, ensuring efficient electron transfer and a measurable faradaic current. Upon target binding, the redox tag moves far away from the electrode and the redox current decreases. Reprinted from [106] with permission from The Royal Society of Chemistry. Using a similar approach, Liu et al. developed an aptasensor comprised of 5' MBconjugated anti-IFNγ Apt immobilized on an Au electrode for IFNγ detection. The Apt-IFNγ bonds were disrupted using a buffer containing urea, and the sensor could be reused multiple times [107] . An aptasensor utilizing the unwinding mechanism of hairpins combined with four DNA sequences in successive partial hybridization was demonstrated by Zhao et al. DNA capture probes were immobilized on an Au electrode, and Apts specific to IFNγ were used as recognition probes. In the absence of IFNγ, Apts hybridized with the capture probes. However, IFNγ in the sample competitively bound and hindered this hybridization step. Subsequently, the surface immobilized, hybridized recognition probes induced the unwinding of the otherwise stable DNA hairpin H1 (biotinylated), followed by the hybridization of H1 to the sticky end of the recognition probes. This event further led to the unwinding and hybridization of DNA hairpin H2 (biotinylated) to the free end of H1. A cascade of successive hybridization events followed until H1 and H2 were quenched in the solution. Upon addition of streptavidin-conjugated alkaline phosphatase (ALP) and its substrate, 1-naphthyl phosphate (1-NPP), the latter was dephosphorylated to 1-naphthol (1-NP), which was irreversibly oxidized at the anode ( Figure 4 ). The large assembly of ALP, generated by the induced cascade of hybridizations, resulted in high signal amplification [108] . Min et al. demonstrated an aptasensor for IFNγ detection by immobilizing DNA or RNA Apts on Au electrodes and compared their performance. Pentanethiol was used as a filler to ensure sufficient interspacing between Apts, so that the individual IFNγ proteins could bind to the Apt without hindrance. It was shown that RNA aptasensors could detect IFNγ at fM levels, while DNA aptasensors could only detect IFNγ in the pM range [109] , presumably due to the different affinities of the Apts. Using a similar approach, Liu et al. developed an aptasensor comprised of 5' MBconjugated anti-IFNγ Apt immobilized on an Au electrode for IFNγ detection. The Apt-IFNγ bonds were disrupted using a buffer containing urea, and the sensor could be reused multiple times [107] . An aptasensor utilizing the unwinding mechanism of hairpins combined with four DNA sequences in successive partial hybridization was demonstrated by Zhao et al. DNA capture probes were immobilized on an Au electrode, and Apts specific to IFNγ were used as recognition probes. In the absence of IFNγ, Apts hybridized with the capture probes. However, IFNγ in the sample competitively bound and hindered this hybridization step. Subsequently, the surface immobilized, hybridized recognition probes induced the unwinding of the otherwise stable DNA hairpin H1 (biotinylated), followed by the hybridization of H1 to the sticky end of the recognition probes. This event further led to the unwinding and hybridization of DNA hairpin H2 (biotinylated) to the free end of H1. A cascade of successive hybridization events followed until H1 and H2 were quenched in the solution. Upon addition of streptavidin-conjugated alkaline phosphatase (ALP) and its substrate, 1-naphthyl phosphate (1-NPP), the latter was dephosphorylated to 1-naphthol (1-NP), which was irreversibly oxidized at the anode ( Figure 4 ). The large assembly of ALP, generated by the induced cascade of hybridizations, resulted in high signal amplification [108] . Min et al. demonstrated an aptasensor for IFNγ detection by immobilizing DNA or RNA Apts on Au electrodes and compared their performance. Pentanethiol was used as a filler to ensure sufficient interspacing between Apts, so that the individual IFNγ proteins could bind to the Apt without hindrance. It was shown that RNA aptasensors could detect IFNγ at fM levels, while DNA aptasensors could only detect IFNγ in the pM range [109] , presumably due to the different affinities of the Apts. . Schematic illustration of an electrochemical DNA biosensor for IFNγ based on hybridization chain reaction with enzyme-amplification. IFNγ partly quenches the recognition probes, while unbound recognition probes hybridize with the capture probes, followed by successive hybridization with biotinylated DNA hairpins. Tagged ALP dephosphorylates 1-NPP, which is further oxidized to 1-NP. Reprinted from [108] with permission from Elsevier. An aptasensor consisting of MB-conjugated DNA hairpin Apt immobilized on Au electrodes for TGF-β1 detection was developed by Matharu et al. This biosensor was placed in a hepatic stellate cell culture, which was stimulated with platelet derived growth factor (PDGF) to produce TGF-β1 in the culture media. In order to prevent electrode fouling from the stellate cells, polydimethylsiloxane (PDMS) microcups were used as covering, which were only lifted during electrochemical measurements [110] . A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 1-2. . Schematic illustration of an electrochemical DNA biosensor for IFNγ based on hybridization chain reaction with enzyme-amplification. IFNγ partly quenches the recognition probes, while unbound recognition probes hybridize with the capture probes, followed by successive hybridization with biotinylated DNA hairpins. Tagged ALP dephosphorylates 1-NPP, which is further oxidized to 1-NP. Reprinted from [108] with permission from Elsevier. An aptasensor consisting of MB-conjugated DNA hairpin Apt immobilized on Au electrodes for TGF-β1 detection was developed by Matharu et al. This biosensor was placed in a hepatic stellate cell culture, which was stimulated with platelet derived growth factor (PDGF) to produce TGF-β1 in the culture media. In order to prevent electrode fouling from the stellate cells, polydimethylsiloxane (PDMS) microcups were used as covering, which were only lifted during electrochemical measurements [110] . A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 1 and 2. Enzymatic immunosensors are one of the most widely explored types of electrochemical biosensors for cytokine detection, where an enzyme is used as a reporter molecule, which catalyzes the conversion of a substrate to product by reducing its activation energy. Horseradish peroxidase (HRP) is one of the most commonly used enzyme reporters due to its stability, small size and high turnover rate. In a study published by Wang et al., an amperometric sandwich-type immunosensor was developed for the detection of IL-6 using an anti-IL-6 capture antibody (Ab 1 ) and anti-IL-6 signal antibody (Ab 2 ). Ab 1 was immobilized on a surface modified indium tin oxide (ITO) electrode, decorated with polydopamine (PDOP) and AuNPs. Ab 2 was attached to carbon nanotubes (CNTs) functionalized with PDOP and AuNPs. HRP adsorbed on the CNTs catalyzed the oxidation of 1,2-phenylenediamine in the presence of H 2 O 2 , followed by cathodic reduction which generated the detection signal ( Figure 5 ) [111] . CNTs exhibit favorable properties for electrochemical detection such as enhanced electron transfer kinetics, excellent conductivity and a large surface area to volume ratio, available for functionalization [112] . Munge et al. prepared and compared two biosensor platforms where CNTs and AuNPs were posed against each other. One sensor was functionalized with glutathione capped AuNPs, and the other sensor was coated with single-walled carbon nanotubes (SWCNTs) forests. The LOD of the former (10 pg/mL) was 3 times lower than the latter (30 pg/mL). For the AuNP sensor, AuNPs were immobilized on a poly(diallyldimethylammonium chloride) (PDDA) functionalized pyrolytic graphite electrode, with covalently attached anti-IL-6 Ab 1 . The antigen binding specificity was enhanced using biotinylated Ab 2 (Bt-Ab 2 ). In the presence of IL-6, streptavidin conjugated HRP (SAv-HRP) was attached to the sensor surface, generating an amperometric current with the redox cycling of hydroquinone in the presence of H 2 O 2 [95] . chemical detection such as enhanced electron transfer kinetics, excellent conductivity and a large surface area to volume ratio, available for functionalization [112] . Munge et al. prepared and compared two biosensor platforms where CNTs and AuNPs were posed against each other. One sensor was functionalized with glutathione capped AuNPs, and the other sensor was coated with single-walled carbon nanotubes (SWCNTs) forests. The LOD of the former (10 pg/mL) was 3 times lower than the latter (30 pg/mL). For the AuNP sensor, AuNPs were immobilized on a poly(diallyldimethylammonium chloride) (PDDA) functionalized pyrolytic graphite electrode, with covalently attached anti-IL-6 Ab1. The antigen binding specificity was enhanced using biotinylated Ab2 (Bt-Ab2). In the presence of IL-6, streptavidin conjugated HRP (SAv-HRP) was attached to the sensor surface, generating an amperometric current with the redox cycling of hydroquinone in the presence of H2O2 [95] . Zhang et al. developed an immunosensor for IFNγ detection by utilizing AuNPs as transducers, coated with PDDA and Ab1, attached to ITO electrodes. AuNPs also formed the basis of bionanolabels, coated with Bt-Ab2 and SAv-HRP enzyme label, and hydroquinone and H2O2 were used as substrates [113] . Bt-Ab2 and SAv-HRP labels were also used for the construction of a sandwich type immunosensor by Sanchez-Tirado et al., wherein Zhang et al. developed an immunosensor for IFNγ detection by utilizing AuNPs as transducers, coated with PDDA and Ab 1 , attached to ITO electrodes. AuNPs also formed the basis of bionanolabels, coated with Bt-Ab 2 and SAv-HRP enzyme label, and hydroquinone and H 2 O 2 were used as substrates [113] . Bt-Ab 2 and SAv-HRP labels were also used for the construction of a sandwich type immunosensor by Sanchez-Tirado et al., wherein SPCE grafted with electrochemically reduced p-aminobenzoic acid (p-ABA) diazonium salt served as a transducer, to which anti-IFNγ Ab 1 was attached [114] . The first amperometric biosensor for measuring TGF-β1 was developed by Sanchez-Tirado et al., which employed molecular beads functionalized with Ab 1 using a Mix & Go polymer, Bt-Ab 2 and a SAv-HRP (catalyst)-hydroquinone, H 2 O 2 (substrates) system. This sensing platform could detect TGF-β1 in plasma and urine samples at concentrations of 0.1-25 ng/mL and 10-50 pg/mL, respectively [115] . Two other amperometric immunosensors for TGF-β1 detection were reported using alternative approaches for antibody immobilization. The first approach, demonstrated by Sanchez-Tirado et al., employed a free radical grafting method with carboxyl aryl diazonium on screen printed carbon electrodes (SPCEs) containing SWCNTs, similar to the work described in [114] . Ab 2 was immobilized onto the SPCE via Bt-SAv, and Ab 1 and HRP were immobilized onto SWCNTs, which also contained 1-(3-aminoethyl)-4,4'-bipyridinium bromide, a viologen, as a redox mediator to amplify the redox signal generated by hydroquinone and H 2 O 2 ( Figure 6 ). Viologens are known to have three oxidation states: +2, +1, 0 at negative potentials; the first two oxidation states are utilized here to shuttle electrons between the electrode and proteins [116] . The second approach for antibody immobilization involved azide-alkyne cycloaddition between azide-functionalized multi-walled carbon nanotubes (MWCNTs) and ethynylated fragment crystallizable (F c ) region of anti-Ab 1 IgG, in the presence of Cu(I) catalyst. Developed by Sanchez-Tirado et al., this approach also employed Bt-Ab 2 and a SAv-HRP (catalyst)-hydroquinone, H 2 O 2 (substrate) system. 11-Azide-3,6,9-trixaundecan-1-amine was used for azide (-N 3 gr.) functionalization, and sodium periodate (NaIO 4 ) was used to oxidize 1,2-diols of F c region of carbohydrates into alkyne groups [117] . between azide-functionalized multi-walled carbon nanotubes (MWCNTs) and ethyn ated fragment crystallizable (Fc) region of anti-Ab1 IgG, in the presence of Cu(I) catal Developed by Sanchez-Tirado et al., this approach also employed Bt-Ab2 and a SAv-H (catalyst)-hydroquinone, H2O2 (substrate) system. 11-Azide-3,6,9-trixaundecan-1-am was used for azide (-N3 gr.) functionalization, and sodium periodate (NaIO4) was used oxidize 1,2-diols of Fc region of carbohydrates into alkyne groups [117] . Figure 6 . Schematic illustration of the different steps involved in the construction of an amperometric immunosensor for TGF-β1 using V-Phe-SWCNT hybrids. Ab1, HRP and a viologen were immobilized on SWCNT. Ab2 was immobilized via Bt-SAv on a screen printed carbon elec trode (SPCE) grafted with carboxyl aryl diazonium. The signal generated by HRP catalyzed H2O reduction was amplified by the redox mediator viologen. Reprinted from [116] with permission from Elsevier. A relatively unique attempt was made by Moschou et al. to adapt an ELISA platfo onto a printed circuit board (PCB) platform for amperometric measurements of IFNγ. cept for the replacement of the capture antibody (Ab1) by an antigen binding fragm Figure 6 . Schematic illustration of the different steps involved in the construction of an amperometric immunosensor for TGF-β1 using V-Phe-SWCNT hybrids. Ab 1 , HRP and a viologen were immobilized on SWCNT. Ab 2 was immobilized via Bt-SAv on a screen printed carbon electrode (SPCE) grafted with carboxyl aryl diazonium. The signal generated by HRP catalyzed H 2 O 2 reduction was amplified by the redox mediator viologen. Reprinted from [116] with permission from Elsevier. A relatively unique attempt was made by Moschou et al. to adapt an ELISA platform onto a printed circuit board (PCB) platform for amperometric measurements of IFNγ. Except for the replacement of the capture antibody (Ab 1 ) by an antigen binding fragment (F ab1 ) of the capture antibody, the rest of the setup was identical, mounted onto an Au electrode for electrochemical transduction (Figure 7) . In ELISA, a chromogenic 3,3',5,5'tetramethylbenzidine (TMB red ) solution (bluish-green hue) is oxidized by HRP in the presence of H 2 O 2 , resulting in a 3,3',5,5'-tetramethylbenzidine diimine (TMB ox ) solution (blue hue). Adding an acid/stop reagent turns the solution yellow. These color changes are read by a spectrophotometer and correlated with the IFNγ concentration. In the amperometric biosensor, the reduction current of TMB ox was used for IFNγ quantification. The calibration plot obtained was comparable to that of ELISA [118] . tetramethylbenzidine (TMBred) solution (bluish-green hue) is oxidized by HRP in the presence of H2O2, resulting in a 3,3',5,5'-tetramethylbenzidine diimine (TMBox) solution (blue hue). Adding an acid/stop reagent turns the solution yellow. These color changes are read by a spectrophotometer and correlated with the IFNγ concentration. In the amperometric biosensor, the reduction current of TMBox was used for IFNγ quantification. The calibration plot obtained was comparable to that of ELISA [118] . Figure 7 . Schematic of (A) IFNγ ELISA assay followed in 96-or 384-well plates and (B) IFNγ amperometric assay followed in gold Au/PMMA/PCB plates. Oxidization of TMBred by HRP induces photometric change measured in ELISA, and a reduction current of TMBox is measured in the amperometric biosensor. Reprinted from [118] with permission from Elsevier. Alkaline phosphatase (ALP) is another enzyme commonly used as a reporter for electrochemical biosensors. Yin et al. proposed an ALP-containing bioconjugate label whereby poly(styrene-acrylic acid) (PSA) spheres were functionalized with polyallylamine hydrochloride (PAH). These spheres were also coated with AuNPs and anti-TNFα Ab2. Anti-TNFα Ab1 was covalently immobilized on a glassy carbon electrode (GCE) modified with poly(acrylic acid)-doped polyaniline, and 1-naphthyl phosphate was used as a substrate [119] . In addition to the use of an enzymatic label, polymers have been generously used in sensor fabrication, which furnish with an ample amount of a variety of functional groups, as well as high loading capacity. Arya et al. proposed a unique approach to biosensor fabrication by immobilizing anti-TNFα Ab1 away from the electrode surface to avoid degradation or alteration of antibodies upon the application of an external bias potential, referred to as an off-matrix modified sensor. This matrix was prepared by punching laser-engraved wells on polymethyl methacrylate (PMMA) sheets, which were then coated with Ab1 via a 4-fluoro-3-nitroazidobenzene linker and attached to Au comb electrodes. TNFα samples, Bt-Ab2-AuNP and SAv-ALP were added to the sensor through two fluidic channels, and 4-aminophenylphosphate was used as a substrate [120] . Arya and Estrela further optimized this approach by using carboxyl polypyrrole films on a combshaped Au electrode microarray to detect TNFα, which could be measured at concentrations as low as 78 pg/mL in undiluted serum samples [121] . Guerrero et al. developed an immunosensor for the detection of IL-1β, a 17.5 kDa protein, in saliva samples [122] . They used azide-alkyne cycloaddition (electro-click chemistry) in the presence of a Cu(I) catalyst for the immobilization of ethynylated anti-Ab1 IgG Figure 7 . Schematic of (A) IFNγ ELISA assay followed in 96-or 384-well plates and (B) IFNγ amperometric assay followed in gold Au/PMMA/PCB plates. Oxidization of TMB red by HRP induces photometric change measured in ELISA, and a reduction current of TMB ox is measured in the amperometric biosensor. Reprinted from [118] with permission from Elsevier. Alkaline phosphatase (ALP) is another enzyme commonly used as a reporter for electrochemical biosensors. Yin et al. proposed an ALP-containing bioconjugate label whereby poly(styrene-acrylic acid) (PSA) spheres were functionalized with polyallylamine hydrochloride (PAH). These spheres were also coated with AuNPs and anti-TNFα Ab 2 . Anti-TNFα Ab 1 was covalently immobilized on a glassy carbon electrode (GCE) modified with poly(acrylic acid)-doped polyaniline, and 1-naphthyl phosphate was used as a substrate [119] . In addition to the use of an enzymatic label, polymers have been generously used in sensor fabrication, which furnish with an ample amount of a variety of functional groups, as well as high loading capacity. Arya et al. proposed a unique approach to biosensor fabrication by immobilizing anti-TNFα Ab 1 away from the electrode surface to avoid degradation or alteration of antibodies upon the application of an external bias potential, referred to as an off-matrix modified sensor. This matrix was prepared by punching laser-engraved wells on polymethyl methacrylate (PMMA) sheets, which were then coated with Ab 1 via a 4-fluoro-3-nitroazidobenzene linker and attached to Au comb electrodes. TNFα samples, Bt-Ab 2 -AuNP and SAv-ALP were added to the sensor through two fluidic channels, and 4-aminophenylphosphate was used as a substrate [120] . Arya and Estrela further optimized this approach by using carboxyl polypyrrole films on a comb-shaped Au electrode microarray to detect TNFα, which could be measured at concentrations as low as 78 pg/mL in undiluted serum samples [121] . Guerrero et al. developed an immunosensor for the detection of IL-1β, a 17.5 kDa protein, in saliva samples [122] . They used azide-alkyne cycloaddition (electro-click chemistry) in the presence of a Cu(I) catalyst for the immobilization of ethynylated anti-Ab 1 IgG on azide-functionalized MWCNTs, similar in principle to the TGF-β1 sensor described in [117] . Bt-Ab 1 and a SAv-ALP (catalyst)-1-naphthyl phosphate (substrate) system were utilized for voltammetric detection on an SPCE (Figure 8 ). This sensor claimed to offer an improved detection range (10-200 pg/mL) compared to ELISA (15.6-500 pg/mL), as well as shorter assay duration (2 h 30 min) compared with ELISA (3 h 45 min) [123] . on azide-functionalized MWCNTs, similar in principle to the TGF-β1 sensor described in [117] . Bt-Ab1 and a SAv-ALP (catalyst)-1-naphthyl phosphate (substrate) system were utilized for voltammetric detection on an SPCE (Figure 8 ). This sensor claimed to offer an improved detection range (10-200 pg/mL) compared to ELISA (15.6-500 pg/mL), as well as shorter assay duration (2 h 30 min) compared with ELISA (3 h 45 min) [123] . Upon sandwich immunoreaction, ALP dephosphorylates 1naphthyl phosphate to 1-naphthol, which is oxidized at the electrode. Reprinted from [123] with permission from Elsevier. Sun et al. developed a sandwich immunosensor for TNFα detection using glucose oxidase (GOx) as an enzyme label, which converts glucose to gluconolactone. This biosensor utilized a PDDA-coated Au electrode containing ferrocene (Fc) carboxylic acid-conjugated peptide nanowires (diphenylalanine), AuNPs and Ab1. Ab2 was functionalized with gold nanorods (GNRs) and glucose oxidase (GOx) via a cysteine-glutaraldehyde bridge ( Figure 9 ). Peptide nanowires (PNWs) exhibit the properties of self-assembly, hydrophobicity and can load Abs in high proportions. The signal generated by glucose oxidation was greatly amplified by the immobilized ferrocene mediators [124] . Table 1 . Ab2 functionalized with gold nanorods (GNRs) and glucose oxidase (GOx) are brought close to the electrode upon sandwich immunoreaction. GOx converts glucose to gluconolactone. Reprinted from [124] with permission from Elsevier. Upon sandwich immunoreaction, ALP dephosphorylates 1-naphthyl phosphate to 1-naphthol, which is oxidized at the electrode. Reprinted from [123] with permission from Elsevier. Sun et al. developed a sandwich immunosensor for TNFα detection using glucose oxidase (GOx) as an enzyme label, which converts glucose to gluconolactone. This biosensor utilized a PDDA-coated Au electrode containing ferrocene (Fc) carboxylic acid-conjugated peptide nanowires (diphenylalanine), AuNPs and Ab 1 . Ab 2 was functionalized with gold nanorods (GNRs) and glucose oxidase (GOx) via a cysteine-glutaraldehyde bridge ( Figure 9 ). Peptide nanowires (PNWs) exhibit the properties of self-assembly, hydrophobicity and can load Abs in high proportions. The signal generated by glucose oxidation was greatly amplified by the immobilized ferrocene mediators [124] . on azide-functionalized MWCNTs, similar in principle to the TGF-β1 sensor described in [117] . Bt-Ab1 and a SAv-ALP (catalyst)-1-naphthyl phosphate (substrate) system were utilized for voltammetric detection on an SPCE (Figure 8 ). This sensor claimed to offer an improved detection range (10-200 pg/mL) compared to ELISA (15.6-500 pg/mL), as well as shorter assay duration (2 h 30 min) compared with ELISA (3 h 45 min) [123] . Upon sandwich immunoreaction, ALP dephosphorylates 1naphthyl phosphate to 1-naphthol, which is oxidized at the electrode. Reprinted from [123] with permission from Elsevier. Sun et al. developed a sandwich immunosensor for TNFα detection using glucose oxidase (GOx) as an enzyme label, which converts glucose to gluconolactone. This biosensor utilized a PDDA-coated Au electrode containing ferrocene (Fc) carboxylic acid-conjugated peptide nanowires (diphenylalanine), AuNPs and Ab1. Ab2 was functionalized with gold nanorods (GNRs) and glucose oxidase (GOx) via a cysteine-glutaraldehyde bridge ( Figure 9 ). Peptide nanowires (PNWs) exhibit the properties of self-assembly, hydrophobicity and can load Abs in high proportions. The signal generated by glucose oxidation was greatly amplified by the immobilized ferrocene mediators [124] . Table 1 . Ab2 functionalized with gold nanorods (GNRs) and glucose oxidase (GOx) are brought close to the electrode upon sandwich immunoreaction. GOx converts glucose to gluconolactone. Reprinted from [124] with permission from Elsevier. Table 1 . Ab 2 functionalized with gold nanorods (GNRs) and glucose oxidase (GOx) are brought close to the electrode upon sandwich immunoreaction. GOx converts glucose to gluconolactone. Reprinted from [124] with permission from Elsevier. A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 3 and 4 . Impedimetric immunosensors consist of a transducing electrode coated with a biorecognition element (Ab or Apt) where binding of the target protein results in a change in the electrical impedance. One of the main advantages of impedimetric immunosensors is that they are label free and do not require an enzyme label or reporter molecule, thereby simplifying the detection process. Yang et al. developed an impedimetric immunosensor for IL-6 detection consisting of a Si/SiO 2 electrode modified with SWCNTs and AuNPs immobilized with anti-IL-6 Ab. A combination of SWCNTs and AuNPs exhibited efficient and enhanced electron transfer rate. Using this immunosensor, IL-6 could be detected at concentrations as low as 0.01 fg/mL [125] . Bellagambi et al. developed an impedimetric immunosensor for the detection of TNFα in human saliva. Anti-TNFα Ab was immobilized on an Au electrode using a 4-carboxymethyl aniline linker. Measurements were performed using spiked samples with known TNFα concentrations to generate a calibration curve, which was used for measurements of human saliva samples. Using this sensor, TNFα could be detected in human saliva within the range of 1-15 pg/mL [126] . Aydin et al. constructed an impedimetric immunosensor for TNFα detection consisting of a hydroxylated ITO electrode functionalized with a semiconducting poly(3-thiophene acetic acid) linker. The poly(3thiophene acetic acid) modification process resulted in carboxylic acid groups on the surface, facilitating the immobilization of anti-TNFα Ab ( Figure 10 ). Measurements in clinical saliva samples revealed that this sensor could detect TNFα at concentrations as low as 3.7 fg/mL [96] . where anti-IFNγ Ab was immobilized on an Au electrode using a SAM of N-acetylcysteine (NAC). Both impedimetric and amperometric measurements were performed, and impedimetric quantification was found to yield a lower LOD. The authors utilized cumulative out-of-phase impedance (Σ∆Z") shift as the calibration parameter, instead of charge transfer resistance (R ct ) (utilized more frequently), and the measurements were taken with multiple successive flow-injections at a constant flow-rate [127] . Yao et al. developed an immunosensor using polyethylene glycol (PEG)-linked Ab onto interdigitated Au electrodes for the impedimetric detection of TGF-β1 [128] . Baraket et al. developed an impedimetric immunosensor for IL-10 detection, which is a potential biomarker for estimating the extent of inflammation post left ventricular assisted device (LVAD) implantation in end-stage heart failure patients [129] . This immunosensor utilizes carboxyl diazonium-free radical grafting on an Au-polyimide electrode for anti-IL-10 Ab immobilization [130] . Another IL-10 biosensor developed by Lee et al. employed a hafnium oxide (HfO 2 ) functionalized Si substrate, where Ab immobilization was facilitated using a 11-(triethoxysilyl)undecanal (TESUD) SAM. Using this immunosensor, IL-10 could be detected at concentrations as low as 0.1 pg/mL [131] . Pui et al. devised an integrated system combining an impedimetric sensor with cytokines-producing cell cultures for TNFα quantification via EIS. Anti-TNFα Ab was linked to Au electrodes (in arrays) using a reduced dithiobis-succinimidyl propionate SAM layer, and ethanolamine was used for surface-blocking. The charge transfer resistance (R ct ) was observed to decrease with increasing TNFα concentration, which could be detected up to a concentration of 57 fM [132] . Sharma et al. developed an impedimetric immunosensor for IL-8 detection using a recombinant antibody-mimetic protein as the biorecognition element, which was selected by Phage display, and expressed using a pET11 vector. IL-8 (or CXCL8) is a chemokine assigned with the task of recruiting neutrophils chemotactically at the site of injury [133] . This sensor consisted of Au electrodes deposited on a silicon substrate functionalized with a monothiol-alkane-PEG-acid capacitative SAM for Ab attachment. The impedance phase shift was monitored in response to samples containing varying concentrations of IL-8, which could be detected at concentrations as low as 90 fg/mL [134] . Russel et al. demonstrated a unique microelectrode-based immunosensor for the real-time detection of IL-6, comprising an array of eight Au discs, with needle-shaped microelectrodes on a silicon substrate. Anti-IL-6 Ab was covalently immobilized on Au discs. Both DPV and EIS were employed for real-time detection of IL-6, using potassium ferricyanide/ferrocyanide (K 3 [Fe(CN) 6 ]/K 4 [Fe(CN) 6 ]) as the redox couple. Interestingly, the authors observed a decrease in impedance and increase in peak current upon antigen binding, which contradicts the typical sensor response for macroelectrodes. This could likely be due to the exposure of channels through the densely packed SAM layer on the microelectrode surface upon Agn-Ab binding, rendering better accessibility to the redox couple. Furthermore, microelectrodes offer advantages of reduced IR s drop and enhanced signal-to-noise ratio. Grafting of additional conductive carbon-based or nanoparticle layers is also not required [135] . Deng et al. developed a unique sandwich type immunosensor based on post-processing of nanolabels for enhanced signal generation. The working principle of this immunosensor relied on the growth and positively charged-capping of AuNPs, which were conjugated to anti-IL-6 Ab 2 for signal generation. The detection signal is the redox current generated at the electrode by K 3 [Fe(CN) 6 ]/K 4 [Fe(CN) 6 ] pair. In the absence of the IL-6-AuNP-Ab 2 immunocomplex, the densely packed SAM layer of 11-mercaptoundecanoic acid prevents [Fe(CN) 6 ] 4−/3− from reaching the Au electrode, thus resulting in no current generation. IL-6 mediated immobilization of AuNP-Ab 2 , followed by growth and positively chargedcapping with cetyltrimethylammonium bromide (CTAB) led to the formation of tunnels in the SAM layer. This gives the redox couple a better chance of undergoing redox reaction at the electrode surface [136] . A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 5 and 6. Non-enzymatic labels have also been employed as reporters for electrochemical immunosensors. Heavy metal nanoparticles are commonly used as non-enzymatic reporters due to their enhanced stability and reproducibility compared with enzymes. Peng et al. reported an IL-6 immunosensor using a unique nanocomposite label consisting of a hollow titanium phosphate (TiP) shell coated with polystyrene, silver nanoparticles (AgNP) and anti-IL-6 Ab 2 . TiP was an ideal choice for dense AgNP deposition, owing to the ion exchange of Ag + with H + present in a large number of -OH groups in TiP. The other component of this sensor was Fe 3 O 4 nanoparticles functionalized with anti-IL-6 Ab 1 . In the presence of IL-6 Agn, Ab 2 -AgNP-TiP spheres were assembled via an external magnet, through an Ab 2 -Agn-Ab 1 /Fe 3 O 4 bridge. The captured AgNPs were subsequently dissociated and quantified using stripping voltammetry, and the signal was proportional to the IL-6 concentration [97] . Another unique label was reported by Zhang et al., which comprised of CNTs coated with CdTe quantum dots and a single capture Ab, for IL-6 detection. This sensor also employed a GCE modified with a PDDA-graphene oxide (GO) SAM. Polydopamine (PDA) coated AuNPs and IL-6 (in saturating concentrations) were adsorbed on this layer. Sample IL-6 competed with surface-bound IL-6 for the binding sites of captured Ab. This led to the depletion of CNT@CdTe QDs-Ab on the sensor surface with increasing sample IL-6 concentration ( Figure 11 ). Voltammetric measurements were performed on a separate heated carbon paste electrode, and Cd + stripping gave off the oxidation current used for the corresponding IL-6 quantification. A heated electrode was found to enhance the stripping kinetics as opposed to a non-heated one [137] . PDA was also utilized by Shi et al. for coating a polystyrene label, which was functionalized with AgNPs and Ab 2 . PDA offers the advantages of self-polymerization, good substrate adhesion and good biocompatibility [138] . Ab 1 was attached to graphene nanoribbons (GNR) stacked on a heated SPCE. Ag + stripping was performed for voltammetric measurements of IL-6, similar to the previous study [139] . Silica nanoparticles (SiNPs) were used as a unique label in the study reported by Yuan et al. Two constructs were used for the detection of TNFα. The first consisted of SiNPs coated with fibrous poly(glycidyl methacrylate) (PGMA) chains, grown with surface-initiated atom transfer radical polymerization (SI-ATRP). CdTe quantum dots were covalently embedded onto this layer alongside Ab 2 . The second construct consisted of a poly(abscisic acid) coated Au substrate, loaded with Ab 1 . The SiNPs exhibited a smooth surface, enabling uniform and consistent molecular adsorption. Cd 2+ ions were electrodeposited, followed by stripping on a bismuth film layered GCE for generating the detection signal [140] . Anodic stripping of Cd + was also used by Wang et al. for IFNγ detection. Magnetic nanoparticles (MNPs) coated with polyaniline (PANI) and Ab 1 were used to magnetically capture IFNγ from the sample. This step was followed by the attachment of CdS NPs via Ab 2 binding, which were conjugated to AuNPs. Upon dissolution of Cd 2+ on a SPCE with HNO 3 solution, Cd 2+ was quantified with anodic stripping voltammetry [141] . A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 7-8. A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 7 and 8. Redox labels are molecular species that can undergo redox reaction close to an electrode surface to generate an electrochemical signal. By uniquely designing redox labels, the reaction kinetics can be optimized to enhance the sensor performance. Due to the broad range of nanoparticles available, the performance characteristics of the redox labels can be fine-tuned based on the nanoparticle properties, such as electrical conductivity, reaction kinetics, band-gap, etc. A unique nanocomposite label was reported by Li et al. for IL-6 detection, comprising a CaCO 3 core coated with a porous polyelectrolyte layer. The porous layer consisted of PDDA and poly(styrene sulfonate) (PSS), which was loaded with Fc and Ab 2 . In the presence of IL-6, the redox current of Fc generated an electrochemical signal on an electrode functionalized with GO and Ab 1 [98] . Redox mediators are often used to enhance the electrochemical signal and detection sensitivity by undergoing coupled redox cycling. Li et al. reported an immunosensor for TNFα detection utilizing two redox mediators, Fe 2+/3+ of Prussian blue nanoparticles (PB) and Ce 3+/4+ of a CeO 2 coating, which served as an artificial peroxidase for generating current from H 2 O 2 reduction. Anti-TNFα Ab 1 was immobilized on a GCE modified with AuNP-CNT complexes, and anti-TNFα Ab 2 was immobilized onto chitosan-coated CeO 2 /PB. PB has shown to be an excellent choice as an electron transfer mediator, when coupled with H 2 O 2 reduction [142] . Weng et al. reported an immunosensor for TNFα detection that utilized Fe 2+/3+ as a redox label for signal generation. This sensor consisted of a GCE with K 3 [Fe(CN) 6 ] entrapped within a chitosan (CS) layer. CS was cross-linked with glutaraldehyde, and the entire assembly was secured in place with a Nafion coating. The Nafion coating also facilitated the adsorption of the captured Ab. When subject to cyclic voltammetry, [Fe(CN) 6 ] 4−/3− undergoes a redox reaction, and its anodic current decreased with an increasing TNFα concentration because of the barrier in charge transfer posed by the non-conducting TNFα-anti-TNFα Ab pairs [143] . In another immunosensor proposed by Wang et al., ZnO NPs, loaded with Ab and sealed with a Nafion coating, were used for impedimetric detection of IFNγ on a GCE. [Fe(CN) 6 ] 4−/3− served as the redox species [144] . Wang et al. prepared a redox label with SiNPs coated with poly(guanine) and avidin (Av), which attracted Bt-Ab 2 . Bt-Av affinity was used to attract Ab 1 to the Av-functionalized SPCE ( Figure 12) . When subjected to a potential sweep in the positive range, electroactive guanine was irreversibly oxidized, and quickly exhausted in the vicinity of the electrode surface. To prolong this effect, a mediator such as Ru(bpy) 3 2+ was used to act as an electron messenger. The transient nature of the mediator ensured its effectivity even in very low concentrations. This immunosensor was used for measurements of TNFα [145] . Redox labels are molecular species that can undergo redox reaction close to an electrode surface to generate an electrochemical signal. By uniquely designing redox labels, the reaction kinetics can be optimized to enhance the sensor performance. Due to the broad range of nanoparticles available, the performance characteristics of the redox labels can be fine-tuned based on the nanoparticle properties, such as electrical conductivity, reaction kinetics, band-gap, etc. A unique nanocomposite label was reported by Li et al. for IL-6 detection, comprising a CaCO3 core coated with a porous polyelectrolyte layer. The porous layer consisted of PDDA and poly(styrene sulfonate) (PSS), which was loaded with Fc and Ab2. In the presence of IL-6, the redox current of Fc generated an electrochemical signal on an electrode functionalized with GO and Ab1 [98] . Redox mediators are often used to enhance the electrochemical signal and detection sensitivity by undergoing coupled redox cycling. Li et al. reported an immunosensor for TNFα detection utilizing two redox mediators, Fe 2+/3+ of Prussian blue nanoparticles (PB) and Ce 3+/4+ of a CeO2 coating, which served as an artificial peroxidase for generating current from H2O2 reduction. Anti-TNFα Ab1 was immobilized on a GCE modified with AuNP-CNT complexes, and anti-TNFα Ab2 was immobilized onto chitosan-coated CeO2/PB. PB has shown to be an excellent choice as an electron transfer mediator, when coupled with H2O2 reduction [142] . Weng et al. reported an immunosensor for TNFα detection that utilized Fe 2+/3+ as a redox label for signal generation. This sensor consisted of a GCE with K3[Fe(CN)6] entrapped within a chitosan (CS) layer. CS was cross-linked with glutaraldehyde, and the entire assembly was secured in place with a Nafion coating. The Nafion coating also facilitated the adsorption of the captured Ab. When subject to cyclic voltammetry, [Fe(CN)6] 4−/3− undergoes a redox reaction, and its anodic current decreased with an increasing TNFα concentration because of the barrier in charge transfer posed by the non-conducting TNFα-anti-TNFα Ab pairs [143] . In another immunosensor proposed by Wang et al., ZnO NPs, loaded with Ab and sealed with a Nafion coating, were used for impedimetric detection of IFNγ on a GCE. [Fe(CN)6] 4−/3− served as the redox species [144] . Wang et al. prepared a redox label with SiNPs coated with poly(guanine) and avidin (Av), which attracted Bt-Ab2. Bt-Av affinity was used to attract Ab1 to the Av-functionalized SPCE ( Figure 12) . When subjected to a potential sweep in the positive range, electroactive guanine was irreversibly oxidized, and quickly exhausted in the vicinity of the electrode surface. To prolong this effect, a mediator such as Ru(bpy)3 2+ was used to act as an electron messenger. The transient nature of the mediator ensured its effectivity even in very low concentrations. This immunosensor was used for measurements of TNFα [145] . A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 9 and 10. Magnetoimmunosensors utilize magnetic beads (Mbs) to facilitate protein purification, Agn-Ab binding and the transport of Agn-Ab immunocomplexes to the electrode surface using a magnet. Kongsuphol et al. developed a magnetoimmunosensor for the impedimetric detection of TNFα in undiluted serum using magnetic beads coated with anti-TNFα Ab. This sensing strategy involved three main steps. First, the undiluted serum sample was incubated with Mbs coated with anti-albumin and anti-IgG Abs, and the Mbimmunocomplexes were separated using a magnetic rack in order to remove albumin and IgG, which account for a major percentage of background interference. Second, the sample was incubated with Mbs coated with anti-TNFα Ab, and the Mb-immunocomplexes were separated with a magnetic rack to isolate TNFα. Third, TNFα was eluted from Mbs via SDS denaturation, followed by adsorption and impedimetric quantification on a comb structured gold microelectrode (CSGM) (Figure 13 ). One advantage of this approach is that the specific removal of albumin and IgG in the serum sample prior to TNFα detection greatly minimizes the likelihood of a false negative result [99] . Mbs were used by Eletxigerra et al. for the preparation of a sandwich type amperometric immunosensor for TNFα detection. For this sensor, a HRP (enzyme)-hydroquinone, H 2 O 2 (substrates) system was used for signal generation on a SPCE. Anti-TNFα Ab 1 was immobilized on Mbs, and Ab 2 -Bt-SAv-HRP was used as an enzymatic label [146] . Another magnetoimmunosensor developed by Bettazzi et al. employed a unique approach for the immobilization of anti-TNFα Ab 1 on Mb coated with Protein G, which is known to bear affinity for the F c region of IgG of many species. Ab 1 /Protein G/Mbs were transported to a screen printed graphite electrode (SPGE) using a surface magnet, and upon TNFα binding, formed a sandwich immunocomplex with Ab 2 -Bt-SAv-ALP. An ALP (enzyme)-1-naphthyl phosphate/1-naphthol (substrate) system was used for signal generation. This immunosensor could detect TNFα down to 0.044 ng/mL [147] . immunocomplexes were separated using a magnetic rack in order to remove albumin and IgG, which account for a major percentage of background interference. Second, the sample was incubated with Mbs coated with anti-TNFα Ab, and the Mb-immunocomplexes were separated with a magnetic rack to isolate TNFα. Third, TNFα was eluted from Mbs via SDS denaturation, followed by adsorption and impedimetric quantification on a comb structured gold microelectrode (CSGM) (Figure 13 ). One advantage of this approach is that the specific removal of albumin and IgG in the serum sample prior to TNFα detection greatly minimizes the likelihood of a false negative result [99] . Mbs were used by Eletxigerra et al. for the preparation of a sandwich type amperometric immunosensor for TNFα detection. For this sensor, a HRP (enzyme)-hydroquinone, H2O2 (substrates) system was used for signal generation on a SPCE. Anti-TNFα Ab1 was immobilized on Mbs, and Ab2-Bt-SAv-HRP was used as an enzymatic label [146] . Another magnetoimmunosensor developed by Bettazzi et al. employed a unique approach for the immobilization of anti-TNFα Ab1 on Mb coated with Protein G, which is known to bear affinity for the Fc region of IgG of many species. Ab1/Protein G/Mbs were transported to a screen printed graphite electrode (SPGE) using a surface magnet, and upon TNFα binding, formed a sandwich immunocomplex with Ab2-Bt-SAv-ALP. An ALP (enzyme)-1-naphthyl phosphate/1naphthol (substrate) system was used for signal generation. This immunosensor could detect TNFα down to 0.044 ng/mL [147] . Figure 13 . Schematic illustration of a magnetoimmunosensor for TNFα detection (a) Serum background is removed using magnetic beads coated with albumin and IgG antibodies. (b) The background removed sample is transferred to a new tube, and TNFα is then captured using magnetic bead coated with TNFα antibody. (c) The captured TNFα is eluted from the magnetic beads using 2% SDS at 63℃, and (d) the magnetic beads are separated out of the eluted sample. (e) Impedance spectra of the eluted sample are then analyzed using an EIS technique in which the CSGM electrode is employed as a platform for TNFα adsorption and for EIS analysis. R, W and C labeled on CSGM electrodes denote reference, working and counter electrodes, respectively. Nyquist plots represent impedance spectra recorded from (i) blank chip and (ii) eluted sample of 1pg/ml TNFα spiked in human serum. Reprinted from [99] with permission from Elsevier. A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 11-12. Figure 13 . Schematic illustration of a magnetoimmunosensor for TNFα detection (a) Serum background is removed using magnetic beads coated with albumin and IgG antibodies. (b) The background removed sample is transferred to a new tube, and TNFα is then captured using magnetic bead coated with TNFα antibody. (c) The captured TNFα is eluted from the magnetic beads using 2% SDS at 63°C, and (d) the magnetic beads are separated out of the eluted sample. (e) Impedance spectra of the eluted sample are then analyzed using an EIS technique in which the CSGM electrode is employed as a platform for TNFα adsorption and for EIS analysis. R, W and C labeled on CSGM electrodes denote reference, working and counter electrodes, respectively. Nyquist plots represent impedance spectra recorded from (i) blank chip and (ii) eluted sample of 1pg/ml TNFα spiked in human serum. Reprinted from [99] with permission from Elsevier. A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 11 and 12. Common abbreviations: PBS = Phosphate-buffered saline, BSA = Bovine Serum Albumin, hIg = Human immunoglobulin protein. While all of the electrochemical biosensors discussed above are capable of quantitative cytokine detection, most require sample processing and/or sample dilution, which increases the assay complexity and detection time. One approach to simplify this process is to combine a flow-injection or microfluidic system with the sensor to facilitate sample loading and processing. One such flow-injection disposable electrochemical immunosensor was developed by Liang and Mu for the detection of IL-6, by utilizing AuNPs and SPGE for enhanced conductivity, Agn-Ab binding specificity and a HRP (enzyme)-thionine, H 2 O 2 (substrates) catalytic system ( Figure 14) . The generated signal reduced proportionately with the increase in sample IL-6 concentration. The binding of IL-6 to HRP-Ab hindered the accessibility of the active center of HRP for oxidization [101] . A four-electrode sensor with a flow-injection system was designed by Berggren et al. for potentiostatic capacitance measurements of IL-6. This sensor consisted of an Au WE coated with anti-IL-6 Ab via a 1,4-butanediol diglycidyl ether SAM and two REs: a platinum wire reference, with a drifting potential, which was compared and calibrated against an external Ag/AgCl electrode with a stable potential. The SAM comprised of cysteamine and 1-dodecanethiol and helped reduce electron transfer through SAM tunnels and increased the capacitive component (C SAM ) of the electrode surface [148] . Further enhancement in the sensor performance can be achieved through the use of a microfluidic system, as demonstrated by Messina et al. This platform consisted of a microfluidic channel containing monoclonal Ab 1 immobilized on 3-aminopropyl-modified controlled-pore glass (APCPG). The sample solution was flow-injected into the microchannel, followed by the injection and incubation of solutions containing polyclonal Bt-Ab 2 and SAv-ALP. The end of the microchannel contained an Au electrode, and an anodic current signal was generated with an ALP (enzyme)-p-aminophenyl phosphate (substrate) system [149] . Ruecha et al. developed a microfluidic immunosensor for IFNγ detection utilizing a screen-printed paper electrode (SPPE). The electrode consisted of graphene coated with polyaniline (PANI), enabling the covalent immobilization of anti-IFNγ Ab. One unique advantage of this paper-based sensor is its ability to be mass produced at low costs [150] . SPGE for enhanced conductivity, Agn-Ab binding specificity and a HRP (enzyme)-thionine, H2O2 (substrates) catalytic system ( Figure 14) . The generated signal reduced proportionately with the increase in sample IL-6 concentration. The binding of IL-6 to HRP-Ab hindered the accessibility of the active center of HRP for oxidization [101] . A four-electrode sensor with a flow-injection system was designed by Berggren et al. for potentiostatic capacitance measurements of IL-6. This sensor consisted of an Au WE coated with anti-IL-6 Ab via a 1,4-butanediol diglycidyl ether SAM and two REs: a platinum wire reference, with a drifting potential, which was compared and calibrated against an external Ag/AgCl electrode with a stable potential. The SAM comprised of cysteamine and 1-dodecanethiol and helped reduce electron transfer through SAM tunnels and increased the capacitive component (CSAM) of the electrode surface [148] . Further enhancement in the sensor performance can be achieved through the use of a microfluidic system, as demonstrated by Messina et al. This platform consisted of a microfluidic channel containing monoclonal Ab1 immobilized on 3-aminopropyl-modified controlled-pore glass (APCPG). The sample solution was flow-injected into the microchannel, followed by the injection and incubation of solutions containing polyclonal Bt-Ab2 and SAv-ALP. The end of the microchannel contained an Au electrode, and an anodic current signal was generated with an ALP (enzyme)-p-aminophenyl phosphate (substrate) system [149] . Ruecha et al. developed a microfluidic immunosensor for IFNγ detection utilizing a screenprinted paper electrode (SPPE). The electrode consisted of graphene coated with polyaniline (PANI), enabling the covalent immobilization of anti-IFNγ Ab. One unique advantage of this paper-based sensor is its ability to be mass produced at low costs [150] . HRP-labeled anti-IL-6 Ab immobilized on a SPCE coated with AuNPs generate redox signal with thionine and H2O2 substrates. This coupled reaction is hindered with IL-6 binding. Reprinted from [101] with permission from Elsevier. A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 13-14. Figure 14 . Schematic illustration of the flow-injection immunoassay system for IL-6 determination. HRP-labeled anti-IL-6 Ab immobilized on a SPCE coated with AuNPs generate redox signal with thionine and H 2 O 2 substrates. This coupled reaction is hindered with IL-6 binding. Reprinted from [101] with permission from Elsevier. A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 13 and 14. Common abbreviations: PBS = Phosphate-buffered saline, BSA = Bovine Serum Albumin. Recently, FET-based biosensors have gained much attention due to their fast response time, miniaturized size and large amplification of potentiometric signal. A liquid gated FET biosensor for IL-6 detection was reported by Huang et al., which employed GO on a silicon substrate for transduction, and anti-IL-6 Ab for biorecognition. The edges of GO were grown using ethanol-chemical vapor deposition for enhanced surface uniformity and higher electrical conductivity [151] . Khosravi et al. designed a FET microarray biosensor for IL-6 detection consisting of a SWCNT/SiO 2 /Si gate coated with RNA Apts. The interaction between IL-6 and RNA Apts caused a change in the conductance proportional to the IL-6 concentration ( Figure 15 ). Using this approach, IL-6 could be detected in real time, with an increase in IL-6 concentration inducing a characteristic spike, followed by a decrease in the sensor conductance [100] . An FET-like aptasensor was devised by Farid et al. for IFNγ detection, which utilized pyrene-terminated Apt stacked on a single layer of graphene (G) on a PDMS substrate. The IFNγ sample on PDMS/G/Apt formed the liquid gate of the FET. A unique feature about this configuration is the small size of the Apt, which ensured that the change in surface charge induced by bound IFNγ does not exceed the Debye length, beyond which charges tend to be masked and ineffective in influencing the gate voltage (V G ). The parameter used here is current (I Dirac ) at the Dirac point (voltage of minimum conductance). The right side of Dirac point constitutes the n-conduction region (major charge carriers: electrons), and the left side constitutes the p-conduction region (major charge carriers: holes). Upon IFNγ binding, there was not only an increase in I Dirac , but also a shift in the Dirac point, which was successfully correlated with IFNγ concentration in a decaying exponential plot [152] . formity and higher electrical conductivity [151] . Khosravi et al. designed a FET microarray biosensor for IL-6 detection consisting of a SWCNT/SiO2/Si gate coated with RNA Apts. The interaction between IL-6 and RNA Apts caused a change in the conductance proportional to the IL-6 concentration ( Figure 15 ). Using this approach, IL-6 could be detected in real time, with an increase in IL-6 concentration inducing a characteristic spike, followed by a decrease in the sensor conductance [100] . An FET-like aptasensor was devised by Farid et al. for IFNγ detection, which utilized pyrene-terminated Apt stacked on a single layer of graphene (G) on a PDMS substrate. The IFNγ sample on PDMS/G/Apt formed the liquid gate of the FET. A unique feature about this configuration is the small size of the Apt, which ensured that the change in surface charge induced by bound IFNγ does not exceed the Debye length, beyond which charges tend to be masked and ineffective in influencing the gate voltage (VG). The parameter used here is current (IDirac) at the Dirac point (voltage of minimum conductance). The right side of Dirac point constitutes the n-conduction region (major charge carriers: electrons), and the left side constitutes the p-conduction region (major charge carriers: holes). Upon IFNγ binding, there was not only an increase in IDirac, but also a shift in the Dirac point, which was successfully correlated with IFNγ concentration in a decaying exponential plot [152] . A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 15-16. While most of the biosensors described above were designed to detect a single cytokine, researchers have also developed sensors capable of detecting multiple cytokines simultaneously. Typically, this is achieved by using two or more WEs which are coated with different biorecognition elements and a shared RE and CE. Alternatively, a single WE could be used containing different electroactive labels coupled to different biorecognition elements. The obtained electrochemical signal can be resolved for the quantification of different analytes. Li et al. developed an electrochemical immunosensor for multiplexed detection of IL-6 and IL-17 using PDDA-coated polystyrene beads labeled with different metal nanoparticles. Beads conjugated to anti-IL-6 Ab 2 were labeled with Cd 2+ , and beads conjugated to anti-IL-17 Ab 2 were labeled with Fc ( Figure 16 ). SWV measurements were performed using an AuNP-functionalized GCE coated with anti-IL-6 and anti-IL-17 Ab 1 . This sensor successfully resolved two current peaks for IL-6 and IL-17 at two different voltage points. The current peak was found to sharpen upon the dissolution of polystyrene with tetrahydrofuran, in comparison to simple polystyrene incubation prior to detection [153] . Pui et al. developed an FET biosensor for the continuous detection of IL-6 and TNFα secreted by a macrophage cell culture, when stimulated with lipopolysaccharide. APTESglutaraldehyde pair was used to immobilize anti-TNFα Ab and anti-IL-6 Ab on two different silicon nanowire (SiNW) gates. The gates were connected to individual drains, but shared a common source. Changes in conductance were used to determine the antigen concentration. IL-6 and TNFα were also determined from rat serum samples stimulated with lipopolysaccharide, and the results were compared with that of ELISA [154] . Baraket et al. reported a biosensor for multiplexed IL-10 and IL-1β detection using two 4-carboxymethyl aryl diazonium-grafted Au WEs, one functionalized with anti-IL-10 Ab and the other with anti-IL-1β Ab. Cytokine concentration was detected using EIS [155] . Torrente-Rodriguez et al. designed a magnetoimmunosensor for the quantification of IL-8 mRNA and IL-8 protein in saliva samples. Test solutions were spiked with IL-8 Figure 16 . Schematic illustration of a PS-Cd 2+ -anti-IL-6 Ab, PS-Fc-anti-IL-17 Ab immunosensor for multiplexed detection of IL-6 and IL-17. Upon sandwich immunoreaction, two current peaks are generated with SWV, corresponding to the reduction in Cd 2+ and Fc. Reprinted from [153] with permission from Elsevier. Pui et al. developed an FET biosensor for the continuous detection of IL-6 and TNFα secreted by a macrophage cell culture, when stimulated with lipopolysaccharide. APTES-glutaraldehyde pair was used to immobilize anti-TNFα Ab and anti-IL-6 Ab on two different silicon nanowire (SiNW) gates. The gates were connected to individual drains, but shared a common source. Changes in conductance were used to determine the antigen concentration. IL-6 and TNFα were also determined from rat serum samples stimulated with lipopolysaccharide, and the results were compared with that of ELISA [154] . Baraket et al. reported a biosensor for multiplexed IL-10 and IL-1β detection using two 4-carboxymethyl aryl diazonium-grafted Au WEs, one functionalized with anti-IL-10 Ab and the other with anti-IL-1β Ab. Cytokine concentration was detected using EIS [155] . Torrente-Rodriguez et al. designed a magnetoimmunosensor for the quantification of IL-8 mRNA and IL-8 protein in saliva samples. Test solutions were spiked with IL-8 protein and an associated synthetic DNA oligonucleotide representing IL-8 mRNA. Anti-IL-8 Ab and cDNA (complementary to IL-8 mRNA) were immobilized on magnetic beads. A HRP (catalyst)-hydroquinone, H 2 O 2 (substrates) system was employed for amperometric detection on a screen printed dual carbon electrode (SPdCE), coupled with a neodymium magnet for Mb separation (Figure 17 ). The detection limit of IL-8 protein obtained here (72.4 pg/mL) was about 23 times lower than the cutoff (600 pg/mL) set for oral cancer patients. The ability to measure the transient expression of cytokines, both mRNA and protein simultaneously, can provide meaningful insight into the progression of an immunological response [156] . ferent silicon nanowire (SiNW) gates. The gates were connected to individual d shared a common source. Changes in conductance were used to determine th concentration. IL-6 and TNFα were also determined from rat serum samples st with lipopolysaccharide, and the results were compared with that of ELISA [154 et al. reported a biosensor for multiplexed IL-10 and IL-1β detection using two ymethyl aryl diazonium-grafted Au WEs, one functionalized with anti-IL-10 Ab other with anti-IL-1β Ab. Cytokine concentration was detected using EIS [155] . Torrente-Rodriguez et al. designed a magnetoimmunosensor for the quan of IL-8 mRNA and IL-8 protein in saliva samples. Test solutions were spiked protein and an associated synthetic DNA oligonucleotide representing IL-8 mRN IL-8 Ab and cDNA (complementary to IL-8 mRNA) were immobilized on magne A HRP (catalyst)-hydroquinone, H2O2 (substrates) system was employed for am ric detection on a screen printed dual carbon electrode (SPdCE), coupled with a ium magnet for Mb separation (Figure 17 ). The detection limit of IL-8 protein here (72.4 pg/mL) was about 23 times lower than the cutoff (600 pg/mL) set for or patients. The ability to measure the transient expression of cytokines, both mR protein simultaneously, can provide meaningful insight into the progression of a nological response [156] . Reprinted from [156] with permission from Elsevier. A summary of design and sensor preparation process as well as sensor performance parameters are presented on Tables 17 and 18 . Table 17 . Biosensors for multiplexed cytokine detection: Design and sensor preparation. Ref. Blocking Agent Preparation and Immobilization (Brief) 1. Blocking Agent Preparation and Immobilization (Brief) 2. [154] (1) SiNW/APTES-Glu-anti-IL-6 Ab gate Cytokines are an important class of molecules that hold tremendous potential for the prognosis and diagnosis of a broad range of conditions and illnesses, including cancer, autoimmune diseases and neurodegenerative diseases. Within the past decade, a broad range of electrochemical sensors have been reported for the quantification of various cytokines, including IL-6, TNFα and IFNγ, which hold promise as disease biomarkers. A wide variety of nanomaterials, biomolecules and sensing schemes have been employed to enhance the sensor performance, such as the detection sensitivity, specificity and reproducibly. For instance, enzyme-labeled immunosensors are some of the earliest electrochemical biosensors that were designed for detecting cytokines. Replacing enzymes with non-enzymatic labels, such as redox labels and heavy metal nanoparticles, improved the stability of the sensor to a favorable extent. Magnetoimmunosensors, on the other hand, provided a separate site for the immunoreaction to take place, away from the electrode. This led to further enhancement of stability and also ensured hassle-free regeneration and reusability of the magnetoimmunosensors. Label-free detection offered by impedimetric and FETbiosensors are by far the most simple and cost-effective sensing technologies available. It is also noteworthy that aptasensors offer better shelf life and flexibility of design than immunosensors, due to the stability and versatility of synthetically manufactured and in vitro selected aptamers. Furthermore, FET-biosensors and screen-printed biosensors are compact and miniaturized platforms, and present the benefits of cost-effective production, portability and ease of handling. Based on these approaches, many of which have been utilized in works discussed in this review, we see that much progress has been made toward the development of simple, low-cost electrochemical biosensors for sensitive cytokine measurements. However, there are still several challenges that need to be overcome in order for this technology to be acceptable for routine clinical testing. To the best of our knowledge, no electrochemical biosensors have been made commercially available, so far, for the detection of cytokines. The stability (shelf life) of most electrochemical sensors ranges from only a few days to weeks. In comparison, the shelf life of many commercial ELISA tests ranges from several months up to one year when stored at 4 • C/−20 • C. The reason that electrochemical sensors exhibit such lower stability is because the transducing electrodes are typically preassembled with various SAMs, nanomaterials and biomolecules, and the integrity of these assemblies can degrade over time. For example, most of the biosensors discussed in this review were tested immediately after being fabricated, and some reported a significant loss in the detection signal intensity when stored for a few weeks. In contrast, most ELISA tests require biomolecular assembly prior to each measurement. Therefore, future efforts should focus on extending the shelf life of electrochemical biosensors through the use of shelf-stable sensor components and biomolecular stabilizing agents. Another challenge is obtaining high-sensitivity cytokine measurements in raw biofluid samples, such as blood, saliva and urine. Many of the cytokine biosensors that have been reported were tested using buffer samples or diluted matrices due to the high content of interfering species present in raw biofluids. Additional research is necessary to enhance the detection sensitivity and specificity in order to detect cytokines at clinically relevant levels in raw biofluids. As discussed earlier, cytokines are expressed transiently, and their pathways intersect at several points. Therefore, cytokine profiling can offer valuable information about the type, severity and route of progression of various pathophysiological conditions, which can also be used to assess the efficacy of immunotherapeutic treatments. Cytokine profiling can be especially helpful in the assessment of disease conditions accompanied by the overactivation of the immune system, as a primary source of physiological damage. In particular, the cases of COVID-19 infections exhibit such phenomenon, marked by cytokine storms, acute lung inflammation and, in extreme cases, multi-organ dysfunction and failure. With this in mind, the development of electrochemical biosensors for multiplexed detection of cytokines is a promising research direction. Furthermore, the utility of multiplexed biosensors can be further extended to studying the effect of cytokines as therapeutic drugs in cell lines. This brings us to another promising application of electrochemical sensors: monitoring cell lines. Immune cell lines are routinely cultured and maintained in cancer research, immunology and infectious disease laboratories. These fields could greatly benefit from the development of electrochemical biosensors that can be integrated with cell cultures through microfluidic platforms. A couple of works discussed in this review have already attempted this. An integrated microfluidic sensing platform will not only offer ease of handling, but also reduce the chances of contamination, and enable continuous cytokine screening. However, there are some practical aspects, such as limited stability/viability of cell lines in such integrated platforms, electrode biofouling and so on, that have yet to be addressed. Lastly, accurate cytokine detection in noninvasive bodily fluids, such as saliva, sweat or urine, can be useful, particularly for infants and individuals with poor blood clotting (hemophilia). Nomenclature and Biologic Significance of Cytokines Involved in Inflammation and the Host Immune Response Study on Certain Biomarkers of Inflammation in Psoriasis Through Anti-inflammatory cytokines A Historical Review of Cytokines Evaluation of potential biomarkers for the discrimination of bacterial and viral infections Cytokines and chemokines as biomarkers of community-acquired bacterial infection IL-6 and IL-10 as predictors of disease severity in COVID-19 patients: Results from meta-analysis and regression Immune escape as a fundamental trait of cancer: Focus on IDO Immune modulation of some autoimmune diseases: The critical role of macrophages and neutrophils in the innate and adaptive immunity Cytokine and cancer biomarkers detection: The dawn of electrochemical paper-based biosensor Clinical significance of interleukin-6 (IL-6) in the spread of gastric cancer: Role of IL-6 as a prognostic factor A novel type of blood biomarker: Distinct changes of cytokine-induced stat phosphorylation in blood t cells between colorectal cancer patients and healthy individuals Quantitative analysis of Th1, Th2 and TGF-β1 cytokine expression in tumor, TIL and PBL of non-small cell lung cancer patients Inhibition of macrophage migration inhibitory factor decreases proliferation and cytokine expression in bladder cancer cells Cytokine expression profile in human pancreatic carcinoma cells and in surgical specimens: Implications for survival Current evidence for the induction of autoimmune rheumatic manifestations by cytokine therapy Linkage disequilibrium of a type 1 diabetes susceptibility locus with a regulatory IL12B allele Cytokines and autoimmunity Cytokines and acute neurodegeneration Immunotherapy of autoimmunity and cancer: The penalty for success 1.9 A crystal structure of interleukin 6: Implications for a novel mode of receptor dimerization and signaling IL-6 in Inflammation, Immunity, and Disease A panoramic review of IL-6: Structure, pathophysiological roles and inhibitors Serum cytokine profiles in healthy young and elderly population assessed using multiplexed bead-based immunoassays A meta-analysis of cytokines in Alzheimer's disease Elevated Neuroimmune Biomarkers in Sweat Patches and Plasma of Premenopausal Women with Major Depressive Disorder in Remission: The POWER Study Inflammation, obesity, stress and coronary heart disease: Is interleukin-6 the link Systematic review of the association between circulating interleukin-6 (IL-6) and cancer IL-6 in autoimmune disease and chronic inflammatory proliferative disease Mutational analysis of structure-activity relationships in human tumor necrosis factor-alpha Human pro-tumor necrosis factor is a homotrimer Tumour necrosis factor and Crohn's disease Cytokines in chronic inflammatory arthritis. V. Mutual antagonism between interferon-gamma and tumor necrosis factor-alpha on HLA-DR expression, proliferation, collagenase production, and granulocyte macrophage colony-stimulating factor production by rheu Comparison of serum IL-1β, sIL-2R, IL-6, and TNF-α levels with disease activity parameters in ankylosing spondylitis Role of increased production of monocytes TNF-α, IL-1β and IL-6 in psoriasis: Relation to focal infection, disease activity and responses to treatments TNF-α in promotion and progression of cancer Interferon-gamma: A Review Crystal structure of a complex between interferon-γ and its soluble high-affinity receptor Interferons, interferon-like cytokines, and their receptors Interferon-γ-induced calcium influx in T lymphocytes of multiple sclerosis and rheumatoid arthritis patients: A complementary mechanism for T cell activation? Randomized study of antibodies to IFN-γ and TNF-α in secondary progressive multiple sclerosis Interferon gamma and interferon gamma inducible protein-10 in detecting tuberculosis infection Recombinant TGF-β1 is synthesized as a two-component latent complex that shares some structural features with the native platelet latent TGF-β1 complex TGFβ1 and Treg cells: Alliance for tolerance Transforming growth factor-beta1 circulates in normal human plasma and is unchanged in advanced metastatic breast cancer TGFβ1 in liver fibrosis: Time to change paradigms? TGF-β1 promotes the migration and invasion of bladder carcinoma cells by increasing fascin1 expression Suppression of microRNA-29 expression by TGF-β1 promotes collagen expression and renal fibrosis TGFβ1 regulates the inflammatory response during chronic neurodegeneration Cytokines as Biomarkers and Their Respective Clinical Cutoff Levels Gingival fluid IL-1 and IL-6 levels in refractory periodontitis Regulatory role of C5a in LPS-induced IL-6 production by neutrophils during sepsis Tumor Necrosis Factor (TNF) α quantification by ELISA and bioassay: Effects of TNFα-soluble TNF receptor (p55) complex dissociation during assay incubations Assaying tumor necrosis factor concentrations in human serum a WHO International Collaborative Study Development and validation of an enzyme-linked immunosorbent assay (ELISA) for recombinant human gamma interferon Differential clinical symptoms among acute phase Indian patients revealed significant association with dengue viral load and serum IFN-gamma level Elevated levels of circulating interleukin-6 and transforming growth factor-β1 in patients with metastatic prostatic carcinoma Elevated IL-1b and IL-6 levels in lumbar herniated discs in patients with sciatic pain Multiplexed fluorescent bead-based immunoassays for quantitation of human cytokines in serum and culture supernatants Simultaneous Quantification of Six Human Cytokines in a Single Sample Using Microparticle-based Flow Cytometric Technology Bioanalytical chemistry of cytokines-A review Elisa and Multiplex Technologies for Cytokine A polymerase chain reaction assay for the detection and quantitation of cytokine gene expression in small numbers of cells Analysis of recombinant cytokines in human body fluids by immunoaffinity capillary electrophoresis Microfluidic platform for multiplexed cell sampling and time-resolved spr-based cytokine sensing New quantitative electrochemiluminescence method (ECLIA) for interleukin-6 (IL-6) measurement Carbon nanotube microwell array for sensitive electrochemiluminescent detection of cancer biomarker proteins Development of electrogenerated chemiluminescence-based enzyme linked immunosorbent assay for sub-pM detection Functionalized fluorescent core-shell nanoparticles used as a fluorescent labels in fluoroimmunoassay for IL-6 High-sensitivity miniaturized immunoassays for tumor necrosis factor α using microfluidic systems Radioimmunoassay for the measurement of serum IL-6 and its correlation with tumour cell mass parameters in multiple myeloma Detection of human serum tumor necrosis factor-α in healthy donors, using a highly sensitive immuno-PCR assay Recent advances in cytokine detection by immunosensing Biosensors: A novel approach to and recent discovery in detection of cytokines Revisiting electrochemical biosensing in the 21st century society for inflammatory cytokines involved in autoimmune, neurodegenerative, cardiac, viral and cancer diseases Field effect detection of biomolecular interactions Detection principles of biological and chemical FET sensors An ultrasensitive enzyme-free electrochemical immunosensor based on redox cycling amplification using methylene blue Label-free impedimetric aptasensor with antifouling surface chemistry: A prostate specific antigen case study Enzyme-assisted glucose quantification for a painless Lab-on-PCB patch implementation Low-Interference Washing-Free Electrochemical Immunosensor Using Glycerol-3-phosphate Dehydrogenase as an Enzyme Label Enzyme-free electrochemical immunosensor based on methylene blue and the electro-oxidation of hydrazine on Pt nanoparticles Washing-free heterogeneous immunosensor using proximity-dependent electron mediation between an enzyme label and an electrode Label-free electrochemical detection of S. mutans exploiting commercially fabricated printed circuit board sensing electrodes Electrical Impedance Detection of Bacterial Pathogens Using Immobilized Antimicrobial Peptides A self-pumping lab-on-a-chip for rapid detection of botulinum toxin Label-free sub-picomolar protein detection with field-effect transistors Recent advances in nanotechnology applied to biosensors Label-free electrochemical aptasensor based on gold and polypyrrole nanoparticles for interleukin 6 detection Electrochemical immunosensors for interleukin-6. Comparison of carbon nanotube forest and gold nanoparticle platforms A highly sensitive immunosensor based on ITO thin films covered by a new semiconductive conjugated polymer for the determination of TNFα in human saliva and serum samples Synthesis of silver nanoparticle-hollow titanium phosphate sphere hybrid as a label for ultrasensitive electrochemical detection of human interleukin-6 Electrochemical sensor utilizing ferrocene loaded porous polyelectrolyte nanoparticles as label for the detection of protein biomarker IL-6 EIS-based biosensor for ultra-sensitive detection of TNF-α from non-diluted human serum Ultrasensitive label-free sensing of IL-6 based on PASE functionalized carbon nanotube micro-arrays with RNA-aptamers as molecular recognition elements Flow-injection immuno-bioassay for interleukin-6 in humans based on gold nanoparticles modified screen-printed graphite electrodes Limit of blank, limit of detection and limit of quantitation Perspective on the Future Role of Aptamers in Analytical Chemistry Biosensors based on gold nanostructures Label free nano-aptasensor for interleukin-6 in protein-dilute bio fluids such as sweat An aptasensor for electrochemical detection of tumor necrosis factor in human blood Aptamer-based electrochemical biosensor for interferon gamma detection An electrochemical aptasensor based on hybridization chain reaction with enzyme-signal amplification for interferon-gamma detection A simple and direct electrochemical detection of interferon-γ using its RNA and DNA aptamers Detecting transforming growth factor-β release from liver cells using an aptasensor integrated with microfluidics Dual amplification strategy for the fabrication of highly sensitive interleukin-6 amperometric immunosensor based on poly-dopamine Carbon nanotube based electrochemical sensors for biomolecules Electrochemical immunosensor for interferon-γ based on disposable ITO detector and HRP-antibody-conjugated nano gold as signal tag Electrochemical immunosensor for the determination of the cytokine interferon gamma (IFN-γ) in saliva Electrochemical immunosensor for sensitive determination of transforming growth factor (TGF)-β1 in urine Viologen-functionalized single-walled carbon nanotubes as carrier nanotags for electrochemical immunosensing. Application to TGF-β1 cytokine Carbon nanotubes functionalized by click chemistry as scaffolds for the preparation of electrochemical immunosensors. Application to the determination of TGF-beta 1 cytokine Amperometric IFN-γ immunosensors with commercially fabricated PCB sensing electrodes Electrochemical immunosensor of tumor necrosis factor α based on alkaline phosphatase functionalized nanospheres Off surface matrix based on-chip electrochemical biosensor platform for protein biomarker detection in undiluted serum Electrochemical ELISA Protein Biosensing in Undiluted Serum Using a Polypyrrole-Based Platform Interleukin 1β: A proinflammatory target for preventing atherosclerotic heart disease Design of electrochemical immunosensors using electro-click chemistry. Application to the detection of IL-1β cytokine in saliva Sensitive immunosensor for tumor necrosis factor α based on dual signal amplification of ferrocene modified self-assembled peptide nanowire and glucose oxidase functionalized gold nanorod An electrochemical impedance sensor for the label-free ultrasensitive detection of interleukin-6 antigen Electrochemical biosensor platform for TNF-α cytokines detection in both artificial and human saliva: Heart failure Development of an electrochemical immunosensor for direct detection of interferon-γ at the attomolar level Biomarkers of liver fibrosis detecting with electrochemical immunosensor on clinical serum Early expression of pro-and anti-inflammatory cytokines in left ventricular assist device recipients with multiple organ failure syndrome Cytokine detection using diazonium modified gold microelectrodes onto polyimide substrates with integrated AG/AGCL reference electrode A novel biosensor based on hafnium oxide: Application for early stage detection of human interleukin-10 Detection of tumor necrosis factor (TNF-α) in cell culture medium with label free electrochemical impedance spectroscopy Pathophysiological roles of interleukin-8/CXCL8 in pulmonary diseases Label-free electrochemical impedance biosensor to detect human interleukin-8 in serum with sub-pg/ml sensitivity Development of a needle shaped microelectrode for electrochemical detection of the sepsis biomarker interleukin-6 (IL-6) in real time Sensitive electrochemical immunosensor based on enlarged and surface charged gold nanoparticles mediated electron transfer Proof-of-principle" concept for ultrasensitive detection of cytokines based on the electrically heated carbon paste electrode Mussel-Inspired Surface Chemistry for Multifunctional Coatings Ultrasensitive multi-analyte electrochemical immunoassay based on GNR-modified heated screen-printed carbon electrodes and PS@PDA-metal labels for rapid detection of MMP-9 and IL-6 Polymer-functionalized silica nanosphere labels for ultrasensitive detection of tumor necrosis factor-alpha Measurement of Interferon Gamma Concentration Using an Electrochemical Immunosensor Prussian Blue-functionalized ceria nanoparticles as label for ultrasensitive detection of tumor necrosis factor-α Label-free electrochemical immunosensor based on K3[Fe(CN)6] as signal for facile and sensitive determination of tumor necrosis factor-alpha Fabrication of electrochemical immunosensor for interferon-γ determination and its application of tuberculosis diagnosis Sensitive Immunoassay of a Biomarker Tumor Necrosis Factor-r Based on Poly(guanine)-Functionalized Silica Nanoparticle Label Amperometric magnetoimmunoassay for the direct detection of tumor necrosis factor alpha biomarker in human serum Electrochemical bioassay for the detection of TNF-α using magnetic beads and disposable screen-printed array of electrodes An immunological interleukine-6 capacitive biosensor using perturbation with a potentiostatic step Microfluidic immunosensor design for the quantification of interleukin-6 in human serum samples Label-free paper-based electrochemical impedance immunosensor for human interferon gamma detection Novel biosensor for interleukin-6 detection Detection of Interferon gamma using graphene and aptamer based FET-like electrochemical biosensor Ultrasensitive multiplexed protein biomarker detection based on electrochemical tag incorporated polystyrene spheres as label Nanoelectronic detection of triggered secretion of pro-inflammatory cytokines using CMOS compatible silicon nanowires A fully integrated electrochemical biosensor platform fabrication process for cytokines detection Electrochemical bioplatforms for the simultaneous determination of interleukin (IL)-8 mRNA and IL-8 protein oral cancer biomarkers in raw saliva The authors declare no conflict of interest.