key: cord-0871044-by2stdj8 authors: Ohira, Shingo; Komiyama, Riho; Kanayama, Naoyuki; Sakai, Kayo; Hirata, Takero; Yoshikata, Kento; Ueda, Yoshihiro; Miyazaki, Masayoshi; Nakayama, Masashi; Koizumi, Masahiko; Konishi, Koji title: Improvement in bladder volume reproducibility using A‐mode portable ultrasound bladder scanner in moderate‐hypofractionated volumetric modulated arc therapy for prostate cancer patients date: 2022-02-02 journal: J Appl Clin Med Phys DOI: 10.1002/acm2.13546 sha: 1f48240da462d406372176e62d5655b39b036efc doc_id: 871044 cord_uid: by2stdj8 PURPOSE: This study introduced an A‐mode portable ultrasound bladder scanner, the Lilium® α‐200 (here after Lilium; Lilium Otsuka, Kanagawa, Japan), for the treatment of prostate cancer patients with hypofractionated volumetric modulated arc therapy to improve the reproducibility of bladder volume (BV). MATERIALS AND METHODS: Thirty patients were advised to maintain full BV prior to computed tomography (CT) simulation and daily treatment. Among these, the BV of 15 patients was measured using Lilium until a BV of 80% in the simulation was achieved (with the Lilium group). Daily cone‐beam CT (CBCT) was performed for treatment. The correlation between BV measured by CBCT and Lilium was assessed. The differences in the BV and dosimetric parameters of the bladder in the CBCT versus planning CT were compared between the groups with and without Lilium. RESULTS: There was a significantly strong relationship (r = 0.796, p < 0.05) between the BVs measured using CBCT and Lilium. The relative BV ratios to simulation CT < 0.5 and > 2 were observed in 10.3% and 12.7%, respectively, of treatment sessions without Lilium group, while these ratios were 1% and 2.8%, respectively, in the Lilium group. The mean absolute difference in the range of V(30Gy) to V(40Gy) without Lilium sessions was significantly larger (p < 0.05) than that in the Lilium group. CONCLUSION: The use of the A‐mode portable ultrasound bladder scanner significantly improved the reproducibility of the BV, resulting in few variations in the dosimetric parameters for the bladder. External beam radiotherapy is an effective treatment option for patients with prostate cancer. Recent irradiation techniques, such as volumetric modulated arc therapy (VMAT), in conjunction with image-guided radiotherapy, can generate a conformal dose to the target while minimizing exposure to the surrounding organs at risk (OARs) with high accuracy. [1] [2] [3] Using these techniques, moderate-(2.2-4 Gy per fraction) or ultra-hypofractionated (> 4 Gy) radiotherapy is increasingly applied in patients with prostate cancer owing to its superiority in terms of biological effect for improving therapeutic ratio compared with the standard fractionated radiotherapy (1.8-2 Gy), 4 shortened treatment time as an alternative to surgery, 5 and circumstance of COVID-19 pandemic. 6 Patients are usually advised to drink a fixed volume of water at a fixed time before the treatment simulation to acquire computed tomography (CT) images of patients with comfortably full bladders. The advantage of maintaining a full bladder is to decrease the radiation exposure to the bladder and small bowels. 7 However, Chen observed significant variation in the daily bladder volume (BV) on the daily cone-beam CT (CBCT) compared with the CT images used for treatment planning for prostate cancer radiotherapy, and the dosimetric parameters of the bladder changed proportionately depending on BV variations. 8 Because hypofractionated radiotherapy may induce side effects, 4 maintaining a constant BV during the treatment course is challenging. A new portable ultrasound bladder scanner (A-mode), named Lilium® α-200 (here after Lilium; Lilium Otsuka, Kanagawa, Japan), 9 was developed, and the device could monitor and record the BV continuously through a small probe. Majima et al. has shown the feasible accuracy of the BV measurement using the Lilium in the multicenter study. 10 Although the assessment of BV before daily radiotherapy using a CT scanner is timeconsuming, and patients need to be exposed to radiation, the A-mode device may act as a screening tool of the BV prior to performing CBCT scans. In addition, the bladder scanner using the A-mode is more cost-effective than that using the B-mode, which is introduced in clinical radiotherapy practice to assess the BV. 11, 12 The Lilium has the potential to reduce the daily variation of BV with the minimal workload in radiotherapy for patients with prostate cancer. This study assessed the relationship between the daily BV measured by CBCT and Lilium, as well as to investigate whether the use of Lilium improves the reproducibility of BV during the treatment course in prostate cancer patients. The Lilium was introduced in the clinical practice to reduce the daily BV variation in pelvic radiotherapy, and BV was measured for patients with prostate, bladder, and postoperative prostate cancer. This retrospective study was approved by the ethics committee of our institution and selected 30 patients with prostate cancer who underwent moderate-hypofractionated radiotherapy (before and after the introduction of Lilium). The CT images were reconstructed with a slice thickness of 2 mm and transferred to a treatment planning system (Eclipse; Varian Medical Systems, Palo Alto, CA). A clinical target volume (CTV) was defined as the prostate gland ± seminal vesicles, and an isotropic margin of 4 mm was added to the CTV to generate the planning target volume (PTV). A prescription dose of 45-60 Gy was delivered using VMAT technique to the PTV in 15-20 fractions (2.75-3 Gy/fraction). Varian C-arm treatment units with six degrees of freedom couches were used for dose delivery (Varian Medical Systems, Palo Alto, CA, USA). For treatment, 15 patients (without Lilium group) collected urine before approximately 1-2 h prior to treatment and drank a cup of water as in the simulation. Subsequently, they were immobilized just as in the simulation, and daily CBCT images were acquired before dose delivery for the target localization. The spatial integrity of the images acquired by the dual-energy CT scanner and CBCT was tested periodically using CATPHAN 600 phantom (Phantom Laboratory, Salem, NY), in which the air and Teflon rods are spaced in a given distance (50 mm). The phantom was chosen because it equips various modules that can test CT number uniformity, spatial resolution, and so on. 13 where P i and D i indicate amplitude and distance between the two graphical spikes in the amplitude of the ultrasound wave (anterior and posterior bladder wall). For treatment, BV was measured using the Lilium before CBCT acquisitions. In case the BV was less than 80% of that in the CT simulation,the patient was advised to drink more water and wait, and the measurement of BV using Lilium was repeated before treatment until the desired BV was achieved. However, in cases where the patients expressed intolerability of uresiesthesia, CBCT was per-formed, and doses were delivered accordingly. All BV measurements were performed by well-trained radiation therapists. The daily BVs were measured using CBCT images and Lilium. Pearson's correlation coefficient was used to measure the strength of the linear relationship between the BVs measured by CBCT and Lilium. The absolute value of Pearson's r was defined as very weak (0-0.19), weak (0.2-0.39), moderate (0.40-0.59), strong (0.6-0.79), and very strong (0.8-1). The linear regression analysis was also performed to obtain a regression equation. Subsequently, the relative daily BV ratio for each patient was defined as the ratio of BV in the treatment and simulation. The daily variation of dosimetric parameters for the bladder was assessed based on the dose distribution in the treatment plan, and the BVs F I G U R E 2 Relationship of bladder volume between that measured by cone-beam computed tomography (CBCT) and Lilium® α-200 contoured using CBCT. The daily volumes of bladder receiving a specific dose in the range of 30-40 Gy (V 30Gy -V 40Gy ), which were predictive indicators of genitourinary toxicity, 14, 15 were evaluated and compared with those in the treatment plan. The absolute difference in the dosimetric parameters between the groups with and without Lilium was compared using the Mann-Whitney U test. Statistical significance was set at p < 0.05. All statistical analyses were performed using the SPSS software (version 27; IBM, Armonk, NY). A patient discontinued treatment leaving one treatment session, and a BV could not be measured using Lilium due to malfunction.Consequently,289 pairs of BVs measured using CBCT and Lilium were analyzed for the Lilium group. Figure 2 shows the relationship between the BVs measured using CBCT and Lilium. The BVs measured using Lilium increased as measured using CBCT also increased. There was a strong positive relationship (r = 0.796, p < 0.05) between the BVs using the two devices. The daily variation of BVs with and without Lilium groups is shown in Figure 3 . A large variation in BV was observed when Lilium was not used, and the relative daily BV ratio to simulation CT ranged from 0.26 to 6.22. Without the Lilium group, relative BV ratios of < 0.5 and > 2 were observed in 10.3% and 12.7% of treatment sessions, respectively, and only 22.7% of treatment sessions were within 0.8-1.2. By contrast, the daily variation in BV was reduced with the Lilium group, and the minimum and maximum relative BV ratio was 0.39 and 3.19, respectively. The relative BV ratios of < 0.5 and > 2 were observed at 1% and 2.8% of treatment sessions, respectively. Moreover, 56.4% of treatment sessions were within 0.8-1.2. Table 2 summarizes the absolute difference in dosimetric parameters between the bladder in the treatment plan and daily CBCT. The mean absolute difference in dosimetric parameters in all evaluated dose regions was large when Lilium was not used. The respective mean ± standard deviation (SD) of absolute difference in V 30Gy (6.7% ± 6.3%), V 35Gy (5.1% ± 4.7%), and V 40Gy (4.0% ± 3.6%) without Lilium group were significantly larger (p < 0.05) than those with Lilium (3.7% ± 2.8%, 2.9% ± 2.2% and 2.4% ± 1.8% for V 30Gy , V 35Gy , and V 40Gy , respectively). In this study, we report the advantage of using a newly developed A-mode portable ultrasound bladder scanner to reduce daily variation in BV during a moderatehypofractionated VMAT treatment course for patients with prostate cancer. Yee et al. analyzed BV using the daily CBCT images for patients with bladder cancer and found significant variations in bladder and PTV volume and position occurred. 16 Chen et al. 8 reported the dosimetric impact of different bladder filling during prostate cancer radiotherapy and found that a 10% increase in BV causes a 5.6% decrease in the mean dose.Moreover, Mak et al. reported considerable interfractional displacement of seminal vesicles in prostate radiotherapy, and the displacement was correlated with bladder filling. 17 Therefore, maintaining BV consistency is imperative for reducing doses to the bladder and ensuring accurate dose delivery. The ultrasound scanner has been utilized as a simple and noninvasive BV measurement, and measurement in a supine position is considered as a gold standard. 18 Currently, several portable ultrasound bladder scanners that can measure BV with adequate accuracy are clinically available. Yoon et al. used a portable threedimensional ultrasound scanner (Biocon-700; Mcube Technology, Korea), in which the ultrasound transducer was rotated 120 • to obtain 12 images of B-mode. A strong correlation existed between the BV measured using the bladder scanner and planning CT. 12 Another report by Luo et al. also demonstrated the adequate accuracy of the BV measurements using a portable bladder scanner (Bladder-Scan BVI9400; Verathon Medical, The Netherlands) in which the internal transducer moved 360 • to scan 12 planes to produce a threedimensional B-mode image. 11 The use of biofeedback protocol with the B-mode portable ultrasound scanner is an easy to use and is an effective approach to reduce daily BV variations. 19 In our study, a more simplified portable bladder scanner utilizing the A-mode technique was used to measure the BV, and a significantly strong correlation was observed between the BV measured by the Lilium ultrasound scanner and CBCT images ( Figure 2 ). BV measured using CBCT was determined as the external line of the bladder wall to evaluate dosimetric parameters of bladder in the treatment plan,while the BV measured using Lilium was calculated based on the inside line of the anterior bladder wall and midline of the posterior bladder wall (Figure 1) . Thus, in this study, BV was underestimated in the A-mode device compared to CBCT. Compared with B-mode ultrasound devices, which map soft tissues in a two-dimensional image, A-mode devices are generally more costeffective and have less reliance on a technician's skill, training, and time. 20 However, we should be aware that the accuracy of the BV measurement using the Lilium was more affected by the patient's characteristics than that using the B-mode device. Yamaguchi et al. reported that flattened bladder shape and large prostates could be factors in measurement error in the BV measurement using the Lilium. 21 In fact, Luo et al. demonstrated the stronger linear relationships between the BV measured using B-mode device and CBCT (r = 0.91) compared with that obtained in our study (r = 0.796). 11 Because scanner malfunction can be a source of measurement error, a quality assurance (QA) program is essential for proper operation of the ultrasound scanner. Hykes et al. described the QA program for A-mode real-time ultrasound scanner to maintain long-term performance, and the transducers were recommended to be tested at least monthly using a phantom with various patterns of rods (AIUM 100 mm Standard Test Object). 22 Ideally, the QA program needs to be performed using a tissueequivalent phantom, which exhibits the same behavior (velocity, reflection scattering, absorption, and attenuation) with human anatomy. However, the Lilium ultrasound scanner cannot utilize the commercially available phantom, and the QA depends on maintenance by a service from the manufacturer. A phantom to test the spatial integrity of the Lilium ultrasound scanner should be developed, and the QA program needs to be performed periodically by users. The bladder is a highly distensible organ, and its position can move due to respiratory motion and bowel filling. Thus, the dosimetric parameter for the bladder in treatment planning is not accurate, and it is difficult to understand the actual delivered doses. 23 In fact, few studies have reported true three-dimensional bladder dosimetry in relation to toxicities in prostate radiotherapy. In our study, the absolute differences in dosimetric parameters between the BVs in simulation CT and daily CBCT were significantly reduced ( Table 2 ), suggesting that the A-mode portable ultrasound scanner can be an effective tool for improving the reliability of dosimetric parameters in the treatment plan. A single institutional DELINEATE trial showed that cumulative late radiation therapy oncology group grade >2 genitourinary toxicity at 1 year was observed in 0% and 10% of the standard (74 Gy in 37 fractions) and moderatehypofractionated (60 Gy in 20 fractions) prostate radiotherapy, respectively. 24 Thus, without rigorous management of BVs during hypofractionated prostate radiotherapy, unexpected radiation side effects may occur. 7 This study has some limitations as well. First, the number of patients was limited, and depth analysis could not be performed, such as the relationship between BV reproducibility and age. Second, BV was measured using Lilium only before treatment in this study; however, it was continuously changing. Nathoo et al. measured the BV using a portable diagnostic ultrasound scanner at 15-min intervals over 60 min and generated the individual bladder kinetic model for ensuring the reproducible BV. They concluded that the optimal urine correction is 60 min after voiding and drinking 500 ml of water. 25 Lotz et al. found a large variation in the urine inflow rate between individuals, ranging from 2.1 to 15 ml/min, and it was linearly correlated with age (negative slope). 26 Real-time monitoring of BV has the potential to improve the reproducibility of BV during treatment. The A-mode portable bladder scanner used in this study can measure BV continuously,while the B-mode devices require medical staff to provide BV measurements at regular intervals. The features of the A-mode device may improve the reproducibility of BV with the minimal workload in clinical radiotherapy. Third, the daily dosimetric parameter of the bladder was estimated based on the dose distribution on the planning CT image, while the patient's position and anatomical changes of OARs were not accounted for. Finally, Lilium was introduced only for patients with prostate cancer, although the other treatment sites in the pelvic region (such as cervical cancer and rectal cancer) also require a full bladder. 27, 28 Despite these limitations, our study demonstrates the potential of the A-mode portable bladder scanner for reducing the variation in BV during a treatment course and provides useful information for maintaining a full bladder. The BV measured by Lilium was significantly correlated with that measured by CBCT. The use of this portable ultrasound bladder scanner could improve the reproducibility of daily BV and reduce the dosimetric deviation between the simulation and treatment. The A-mode portable ultrasound scanner (Lilium) has potential to be a screening tool for the daily BV measurement prior to performing CBCT with feasible accuracy to maintain the BV during treatment course for prostate cancer patient. This study was supported by JSPS KAKENHI Grant (Grant-in-Aid for Scientific Research (C) 21K07742). Improved outcomes with dose escalation in localized prostate cancer treated with precision image-guided radiotherapy Reproducibility and accuracy of a target motion mitigation technique for dose-escalated prostate stereotactic body radiotherapy Dosimetric comparison of intensitymodulated radiotherapy and volumetric-modulated arc radiotherapy in patients with prostate cancer: a meta-analysis Hypofractionated radiotherapy for localized prostate cancer Ultrahypofractionation of localized prostate cancer : statement from the DEGRO working group prostate cancer The impact of the COVID-19 pandemic on radiotherapy services in England, UK: a population-based study Impact of bladder volume on acute genitourinary toxicity in intensity modulated radiotherapy for localized and locally advanced prostate cancer Dosimetric impact of different bladder and rectum filling during prostate cancer radiotherapy Feasibility of approximate measurement of bladder volume in male patients using the Lilium alpha-200 portable ultrasound bladder scanner Feasibility of the Lilium alpha-200 portable ultrasound bladder scanner for accurate bladder volume measurement Interfractional variation in bladder volume and its impact on cervical cancer radiotherapy:clinical significance of portable bladder scanner Evaluating variations of bladder volume using an ultrasound scanner in rectal cancer patients during chemoradiation:is protocol-based full bladder maintenance using a bladder scanner useful to maintain the bladder volume? An evaluation of the stability of image-quality parameters of Varian on-board imaging (OBI) and EPID imaging systems Different rectal toxicity tolerance with and without simultaneous conventionallyfractionated pelvic lymph node treatment in patients receiving hypofractionated prostate radiotherapy Late toxicity and quality of life with prostate only or whole pelvic radiation therapy in high risk prostate cancer (POP-RT): a randomised trial Cone beam CT imaging analysis of interfractional variations in bladder volume and position during radiotherapy for bladder cancer Seminal vesicle interfraction displacement and margins in image guided radiotherapy for prostate cancer Accuracy of total bladder volume and residual urine measurements: comparison between realtime ultrasonography and catheterization Bladder filling variation during radiation treatment of prostate cancer: can the use of a bladder ultrasound scanner and biofeedback optimize bladder filling? Comparison of A-mode and B-mode ultrasound for measurement of subcutaneous fat Comparative accuracy of the Lilium alpha-200 portable ultrasound bladder scanner and conventional transabdominal ultrasonography for postvoid residual urine volume measurement in association with the clinical factors involved in measurement errors Quality assurance for real-time ultrasound equipment Radiation dose-volume effects of the urinary bladder Standard and hypofractionated dose escalation to intraprostatic tumor nodules in localized prostate cancer: efficacy and toxicity in the DELINEATE trial A feasibility study on the role of ultrasound imaging of bladder volume as a method to improve concordance of bladder filling status on treatment with simulation Reproducibility of the bladder shape and bladder shape changes during filling The impact of daily bladder filling on small bowel dose for intensity modulated radiation therapy for cervical cancer Reduced pelvic field sparing anastomosis for postoperative radiotherapy in selected patients with mid-upper rectal cancer Improvement in bladder volume reproducibility using A-mode portable ultrasound bladder scanner in moderate-hypofractionated volumetric modulated arc therapy for prostate cancer patients The authors have no conflict of interest to declare in relation to this study.