key: cord-0898152-rccyzavj authors: Kadja, Tchamie; Liu, Chengkun; Sun, Yvonne; Chodavarapu, Vamsy P. title: Low-Cost, Real-Time Polymerase Chain Reaction System for Point-of-Care Medical Diagnosis date: 2022-03-17 journal: Sensors (Basel) DOI: 10.3390/s22062320 sha: 4c00d6ae8c61610859a33f4e3cb60cc8ee2a15c8 doc_id: 898152 cord_uid: rccyzavj Global health crises due to the prevailing Coronavirus Disease 2019 (COVID-19) pandemic have placed significant strain on health care facilities such as hospitals and clinics around the world. Further, foodborne and waterborne diseases are not only spreading faster, but also appear to be emerging more rapidly than ever before and are able to circumvent conventional control measures. The Polymerase Chain Reaction (PCR) system is a well-known diagnostic tool for many applications in medical diagnostics, environmental monitoring, and food and water quality assessment. Here, we describe the design, development, and testing of a portable, low-cost, and real-time PCR system that can be used in emergency health crises and resource-poor situations. The described PCR system incorporates real-time reaction monitoring using fluorescence as an alternative to gel electrophoresis for reaction analysis, further decreasing the need of multiple reagents, reducing sample testing cost, and reducing sample analysis time. The bill of materials cost of the described system is approximately $340. The described PCR system utilizes a novel progressive selective proportional–integral–derivative controller that helps in reducing sample analysis time. In addition, the system employs a novel primer-based approach to quantify the initial target amplicon concentration, making it well-suited for food and water quality assessment. The developed PCR system performed DNA amplification at a level and speed comparable to larger and more expensive commercial table-top systems. The fluorescence detection sensitivity was also tested to be at the same level as commercially available multi-mode optical readers, thus making the PCR system an attractive solution for medical point-of-care and food and water quality assessment. Since 2020, the world has been experiencing a severe Coronavirus Disease 2019 pandemic caused by the Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2). Across the world, local, state/provincial and central governments have struggled to address the extraordinary human, economic, and social tragedy caused by the pandemic. The current pandemic revealed the vulnerabilities in healthcare systems where, in addition to critical shortages of healthcare personnel, ventilators, and personal protective equipment, there is a widespread need for diagnostic tools to quickly assess patient health. Medical equipment such as Polymerase Chain Reaction (PCR) systems have played a crucial role during the COVID-19 pandemic. The PCR test for COVID-19 has been frequently used to analyze the upper respiratory specimens from people, looking for genetic material (ribonucleic acid or RNA) of SARS-CoV-2. The pandemic has caused a dire impact on all countries, but emerging and resource-poor countries which lack adequate healthcare infrastructure have faced more difficulties. Here, low-cost and simple-to-use Sensors 2022, 22 , 2320 3 of 14 OR, USA) of size 40 mm × 40 mm × 3.3 mm, in conjugation with a heat sink 394-2AB (Wakefield-Vette, Nashua, NH, USA) of size 139.7 mm × 38.1 mm × 127 mm. Here, the TECs would work in junction with the thermal block to heat the specimen during the high temperature phase. TECs are powered with a 12 V, 20 A DC power supply, which also serves as power source for the rest of the system. High-performance thermal paste (Corsair XTM50) was used for maximum heat conduction between all thermal cycler objects. Two fans (120 mm × 120 mm × 25 mm, 12 V) were added to the system to keep the heat sink from overheating and to accelerate the cooling process. A structural diagram of the path of heat flow between various objects is shown in Figure 1 . supply. Meanwhile, the cooling process during thermal cycling is achieved using two thermoelectric cooler (TEC) Peltier modules CP105433H (CUI Devices, Lake Oswego, OR, USA) of size 40 mm × 40 mm × 3.3 mm, in conjugation with a heat sink 394-2AB (Wakefield-Vette, Nashua, NH, USA) of size 139.7 mm × 38.1 mm × 127 mm. Here, the TECs would work in junction with the thermal block to heat the specimen during the high temperature phase. TECs are powered with a 12 V, 20 A DC power supply, which also serves as power source for the rest of the system. High-performance thermal paste (Corsair XTM50) was used for maximum heat conduction between all thermal cycler objects. Two fans (120 mm × 120 mm × 25 mm, 12 V) were added to the system to keep the heat sink from overheating and to accelerate the cooling process. A structural diagram of the path of heat flow between various objects is shown in Figure 1 . Real-time PCR, also known as Quantitative PCR (qPCR) is achieved by the detection of emitted light from an excited fluorophore which is embedded in the DNA amplicons [27] during the PCR annealing phase, using an LED as excitation source. This allows for real-time DNA assay without relying solely on gel electrophoresis. The proposed low-cost optical system is composed of an LED as light source, with a 3D-printed receptacle that can hold a PCR tube, an optical filter, and a chip-scale spectrometer. In operation, the incoming excitation light travels from underneath the PCR tube, while the emitted light is detected at 90° from the excitation spectrum to minimize background noise. The emission wavelength is detected using a low-cost chip-scale 10-channel spectrometer (AMS AG, model: AS7341) with an orange optical filter (546 nm center wavelength, Full Width at Half Maximum (FWHM) of 8 nm) in the foreground. The spectrometer utilizes a light sensor that can detect center wavelengths at 415 nm, 445 nm, 480 nm, 515 nm, 555 nm, 590 Real-time PCR, also known as Quantitative PCR (qPCR) is achieved by the detection of emitted light from an excited fluorophore which is embedded in the DNA amplicons [27] during the PCR annealing phase, using an LED as excitation source. This allows for realtime DNA assay without relying solely on gel electrophoresis. The proposed low-cost optical system is composed of an LED as light source, with a 3D-printed receptacle that can hold a PCR tube, an optical filter, and a chip-scale spectrometer. In operation, the incoming excitation light travels from underneath the PCR tube, while the emitted light is detected at 90 • from the excitation spectrum to minimize background noise. The emission wavelength is detected using a low-cost chip-scale 10-channel spectrometer (AMS AG, model: AS7341) with an orange optical filter (546 nm center wavelength, Full Width at Half Maximum (FWHM) of 8 nm) in the foreground. The spectrometer utilizes a light sensor that can detect center wavelengths at 415 nm, 445 nm, 480 nm, 515 nm, 555 nm, 590 nm, 630 nm, and 680 nm (all with FWHM of 8 nm), which is useful for different types of fluorescence applications. Figure 2 shows the diagram of the optical system. Figure 2 shows the diagram of the optical system. Real-time fluorescence detection data from the photodetector was transmitted to the user via the network module. A low-cost ESP8266 WiFi module was used in the system design. This setting makes it compatible with most of the current WiFi compatible devices such as phones, computers, and tablets. In the proposed system, the temperature tracking, heating, cooling, and data transmission is performed under the control of a low-cost 8-bit RISC-based ATmega2560 microcontroller. The microcontroller incorporates a multichannel 10-bit Analog to Digital Converter (ADC) as well as an 8-bit Pulse Width Modulation (PWM) capability. This microcontroller was used as part of an Arduino Mega 2560 board for programming purposes and to take advantage of the 16 MHz embedded clock oscillator offered by the board. In that regard, 490 Hz PWM signals can be generated and used for precise temperature control of the thermal block and lid. At such high frequency, and because the pulse rate is higher than the rate of temperature variation, only the average PWM voltage becomes relevant. In addition, power MOSFETs were used to control the electrical current flow through each conductor. Temperatures of the thermal block and lid are measured by acquiring the resistance of the embedded thermistors in both parts. For a heater with known conductor material, the resistance as a function of temperature is given as, where, Conductor resistance at temperature , Conductor resistance at reference temperature Temperature coefficient of resistance for conductor material Conductor temperature in degrees Celsius Reference temperature at which is specified ∆ Temperature difference between Conductor temperature and reference temperature Real-time fluorescence detection data from the photodetector was transmitted to the user via the network module. A low-cost ESP8266 WiFi module was used in the system design. This setting makes it compatible with most of the current WiFi compatible devices such as phones, computers, and tablets. In the proposed system, the temperature tracking, heating, cooling, and data transmission is performed under the control of a low-cost 8-bit RISC-based ATmega2560 microcontroller. The microcontroller incorporates a multichannel 10-bit Analog to Digital Converter (ADC) as well as an 8-bit Pulse Width Modulation (PWM) capability. This microcontroller was used as part of an Arduino Mega 2560 board for programming purposes and to take advantage of the 16 MHz embedded clock oscillator offered by the board. In that regard, 490 Hz PWM signals can be generated and used for precise temperature control of the thermal block and lid. At such high frequency, and because the pulse rate is higher than the rate of temperature variation, only the average PWM voltage becomes relevant. In addition, power MOSFETs were used to control the electrical current flow through each conductor. Temperatures of the thermal block and lid are measured by acquiring the resistance of the embedded thermistors in both parts. For a heater with known conductor material, the resistance as a function of temperature is given as, where, R(T) = Conductor resistance at temperature T, R(T 0 ) = Conductor resistance at reference temperature T 0 α = Temperature coefficient of resistance for conductor material T = Conductor temperature in degrees Celsius T 0 = Reference temperature at which α is specified ∆T = Temperature difference between Conductor temperature and reference temperature Temperature can be derived from the previous expression for a measured resistance value. A custom Printed Circuit Board (PCB) holds the main electronic components, and the whole system fits in a 3D-printed enclosure. The overall dimensions are 21 cm × 16 cm × 20 cm and the unit weighs 500 g. Figure 3 shows the exterior design of the system, and Figure 4 illustrates the schematic diagram of the full system architecture. Temperature can be derived from the previous expression for a measured resistance value. A custom Printed Circuit Board (PCB) holds the main electronic components, and the whole system fits in a 3D-printed enclosure. The overall dimensions are 21 cm × 16 cm × 20 cm and the unit weighs 500 g. Figure 3 shows the exterior design of the system, and Figure 4 illustrates the schematic diagram of the full system architecture. Temperature can be derived from the previous expression for a measured resistance value. A custom Printed Circuit Board (PCB) holds the main electronic components, and the whole system fits in a 3D-printed enclosure. The overall dimensions are 21 cm × 16 cm × 20 cm and the unit weighs 500 g. Figure 3 shows the exterior design of the system, and Figure 4 illustrates the schematic diagram of the full system architecture. First, a resistance-temperature curve was generated for the metallic thermal block that was harvested from a GeneAmp PCR System 2400 and the data was used to build a look-up table. As mentioned previously, the thermal block could either be harvested from an old PCR system or custom manufactured. The resistance was recorded using a benchtop ohmmeter and its corresponding temperature value was acquired with an infrared thermometer. The curve of the temperature as a function of resistance has a First, a resistance-temperature curve was generated for the metallic thermal block that was harvested from a GeneAmp PCR System 2400 and the data was used to build a look-up table. As mentioned previously, the thermal block could either be harvested from an old PCR system or custom manufactured. The resistance was recorded using a benchtop ohmmeter and its corresponding temperature value was acquired with an infrared thermometer. The curve of the temperature as a function of resistance has a logarithmic trend with a coefficient of correlation of 0.98. The resistance-temperature characteristics of the thermal block are shown in Figure 5 . PCR thermal cycling requires precise temperature control to achieve optimal results. In that regard, a novel Proportional-Integral-Derivative (PID) scheme was used as the thermal control system. Because the system could be characterized as not "well-behaved" due to the extended delay in reaction time, regular tuning techniques such as Ziegler-Nichols and Cohen-Coon are not suitable [28, 29] . Using traditional PID control for a PCR reaction would imply switching between temperature setpoints and making the system react accordingly. Using such controllers on a highly delayed system would fail to meet the typically 30 s (or less) PCR step requirement because of the elongated settling time. For that purpose, we propose a new method called the Progressive Selective PID Controller (PSPC), to overcome the time delay limitation. In our method, three PID controllers are executed based on a selector command. Each PID controls the denaturation, the annealing, and the extension phase, respectively. The control variable of each controller is saved in memory and progressively updated as the process continues. This prevents loop recalculation of PID values and allows continuous fine tuning. Therefore, the system becomes more precise over time. The initial controller setting was performed using a heuristic approach, depending on the type of material used. The system can be characterized as a second-degree PID control system. In regular control, the PID controller can be modeled as, . ( The transfer controller transfer function can be approximated as [30, 31] , The equation of the resistance-temperature characteristics curve was obtained as, PCR thermal cycling requires precise temperature control to achieve optimal results. In that regard, a novel Proportional-Integral-Derivative (PID) scheme was used as the thermal control system. Because the system could be characterized as not "well-behaved" due to the extended delay in reaction time, regular tuning techniques such as Ziegler-Nichols and Cohen-Coon are not suitable [28, 29] . Using traditional PID control for a PCR reaction would imply switching between temperature setpoints and making the system react accordingly. Using such controllers on a highly delayed system would fail to meet the typically 30 s (or less) PCR step requirement because of the elongated settling time. For that purpose, we propose a new method called the Progressive Selective PID Controller (PSPC), to overcome the time delay limitation. In our method, three PID controllers are executed based on a selector command. Each PID controls the denaturation, the annealing, and the extension phase, respectively. The control variable u(t) of each controller is saved in memory and progressively updated as the process continues. This prevents loop recalculation of PID values and allows continuous fine tuning. Therefore, the system becomes more precise over time. The initial controller setting was performed using a heuristic approach, depending on the type of material used. The system can be characterized as a second-degree PID control system. In regular control, the PID controller can be modeled as, The transfer controller transfer function can be approximated as [30, 31] , By defining, θ T = K p K I K D , we can rewrite (2) as, With our improved PID controller, the controller transfer function can be redefined as, With, SLR = selector value, D = Denaturation step, A = Annealing step, E = Extension step, θ T n = PID gains at specified selector value Figure 6 shows the temperature controller schematic diagram. we can rewrite (2) as, ∅ With our improved PID controller, the controller transfer function can be redefined as, With, selector value, Denaturation step, Annealing step, Extension step, PID gains at specified selector value Figure 6 shows the temperature controller schematic diagram. Fluorescence measurements can be performed during thermal cycling after pausing the process at one of its stages, removing the control PCR tube from the heating block, and placing it into the optical system. Performing such measurement during the extension stage will be optimal, as the temperature is low enough to prevent evaporation inside the tube after removing the heated lid, as well as suitable to further allow new DNA strands formations. Prolonged exposure to an excitation source has adverse effects on the fluorescence potency of the fluorophore used for real-time PCR monitoring [21, 32] . This effect, known as photobleaching, is intensified by sub-optimal irradiance [33] . The optical system used in our design has a controlled light emission of 3 s and uses optimal light intensity to reduce photobleaching effects. The various fluorophores used in PCR have different excitation wavelengths, which accordingly determine the required excitation LED. Fluorescence measurements can be performed during thermal cycling after pausing the process at one of its stages, removing the control PCR tube from the heating block, and placing it into the optical system. Performing such measurement during the extension stage will be optimal, as the temperature is low enough to prevent evaporation inside the tube after removing the heated lid, as well as suitable to further allow new DNA strands formations. Prolonged exposure to an excitation source has adverse effects on the fluorescence potency of the fluorophore used for real-time PCR monitoring [21, 32] . This effect, known as photobleaching, is intensified by sub-optimal irradiance [33] . The optical system used in our design has a controlled light emission of 3 s and uses optimal light intensity to reduce photobleaching effects. The various fluorophores used in PCR have different excitation wavelengths, which accordingly determine the required excitation LED. Primers with longer amplicon size tend to yield less amplification efficiency because there is need for more time to complete DNA elongation, therefore they reach the reaction plateau at later cycles (high C T ). Using that property, it was determined experimentally that with primers of product size between 300 and 600 base pairs (bp), and a PCR protocol of 3 steps and 40 cycles, starting point amplicons with concentrations at different orders of magnitude yield final DNA concentrations at a gradient. This protocol was used to quantify listeria concentration in water by applying fluorescence on the PCR final products. As mentioned previously, PCR reaction settings largely depend on the type of DNA and expected outcome. Using fixed temperatures in thermal cycling would not work well for all assays and must be tailored accordingly. The PSPC was designed to allow user input and custom-made experiments. Users can define the number of cycles, number of steps for each cycle, the step temperatures, step hold times, and the system will automatically adapt without affecting its performance. In addition, that PSPC was designed to operate autonomously with minimal human interaction during the process. The autonomy and versatility of the system were achieved by developing a C++ program for the ATmega2560 microcontroller that would allow processing the thermal cycling and data communication. An on-board user-friendly interface with keypad, buttons and display is provided for input/output interaction without the need for an external device or computer, even though the user could also remotely access real-time fluorescence data on an external WiFi connected device. The remote data is provided and presented via an automatically generated HTML interface on the device connected to the PSPC. The current system was designed to be a cost-effective solution with use in resourcepoor countries. The bill-of-materials cost (see Table 1 ) for the system was about $340. Following the PCR reactions, amplicon quantification was performed using a 1% or 1.2% gel (100 mL Tris-Borate-EDTA (TBE) buffer, 1 g or 1.2 g agarose) locally prepared for the experiment, run at 100 volts for 30 min, loading 2 µL 6X loading dye and 10 µL PCR sample. The ladder used was a 1 kb DNA ladder (New England Biolabs, cat. no. N3232L). The images were captured using UVP Multispectral Imaging System (Biospectrum 500, LM-26, BioChemi 500 Camera f/1.2). The experiments described in this work compare the efficiency of our PSPC system versus a commercially available PCR system (Applied Biosystems, model: StepOnePlus). The developed PCR thermal cycling was conducted concurrently on PSPC and the StepOnePlus systems. Fluorescence measurement was compared with the commercial device Synergy 0.2 mL tubes in the PSPC as they allow better adhesion with the heated lid. Tubes used on both PSPC and StepOnePlus were low-profile polypropylene tubes, although capillary glass tubes could be used as well. Fresh Listeria monocytogenes colonies were collected from a streaked plate and placed in sterile water. The initial bacterial solution (Tube 1) was diluted further in a serial 1:10 dilution down to 1:10,000,000 (Tube 8) diluted solution. Tubes 8 and 7 were plated on LB agar plates for colony count, and count results could be extrapolated for higher concentrations. Amplicon concentrations were mixed with Select MasterMix (Applied Biosystems SYBR, Mfr. No. 4472908) and combined with different primer sets (each at 20 µM working concentrations). Table 2 summarizes the target, amplicon size, primer sequence, and reaction reagents for gel electrophoresis analysis and fluorescence analysis, respectively. The PCR protocols were identical on both PSPC and StepOnePlus systems, which were as follows: 10 min hot-start at 95 • C, followed by 40 cycles of 95 • C (30 s denaturation), 55 • C (30 s annealing), and 72 • C (60 s extension), followed by 10 min final elongation at 72 • C. Table 2 . PCR target, analysis type and reaction components. Listeria (115 bp The heating and cooling speed recorded for the PSPC was comparable to the StepOne-Plus system. With the thermal block fully loaded (24 tubes), we recorded an average of 40 s to cool down from 95 • C to 55 • C, 9 s to heat up from 55 • C to 72 • C, and 13 s from 72 • C to 95 • C. Table 3 summarizes the thermal transition rate between PCR steps at the preset temperatures. For a 40-cycle reaction, it took an average of 2 h 23 min on PSPC, versus 2 h 10 min on the StepOnePlus system. Table 3 summarizes the heating and cooling rates of the system. Table 3 . Temperature rates of change during cycles. Transition Rate With the selective progressive controller, the temperature curve shows low slope and reduced fluctuations at each temperature step, which is appropriate for temperature selective applications. Figure 7 shows the temperature curve recorded for the PSPC at 95 • C-55 • C-72 • C cycle. Figure 8 shows the setpoint error at different cycles for the extension step. Listeria (629 bp) Fluorescence 1 µL F. primer 5′-GTA GCG GTG AAA TGC GTA GA -3′ 1 µL R. primer 5′-GCC TAC AAT CCG AAC TGA GAA TA -3′ 8 µL template (8 µL deionized H20 for negative control) The heating and cooling speed recorded for the PSPC was comparable to the StepOnePlus system. With the thermal block fully loaded (24 tubes), we recorded an average of 40 s to cool down from 95 °C to 55 °C, 9 s to heat up from 55 °C to 72 °C, and 13 s from 72 °C to 95 °C. Table 3 summarizes the thermal transition rate between PCR steps at the preset temperatures. For a 40-cycle reaction, it took an average of 2 h 23 min on PSPC, versus 2 h 10 min on the StepOnePlus system. Table 3 summarizes the heating and cooling rates of the system. With the selective progressive controller, the temperature curve shows low slope and reduced fluctuations at each temperature step, which is appropriate for temperature selective applications. Figure 7 shows the temperature curve recorded for the PSPC at 95 °C-55 °C-72 °C cycle. Figure 8 shows the setpoint error at different cycles for the extension step. The error at different cycles was compared using the area under the curve at each of the cycles. The definite integral computation resulted in values of 14,675.24, 8552.74, and 545.32 for cycle 1, cycle 2, and cycle 3, respectively. This shows how the system improves at tracking the reference point over time. This is the result of the-progressive selective control. We used PSPC for listeria detection by amplifying DNA from a listeria colony as described in the PCR conditions. The total runtime was 2 h and 23 min for PSPC for 40 cycles The error at different cycles was compared using the area under the curve at each of the cycles. The definite integral computation resulted in values of 14,675.24, 8552.74, and 545.32 for cycle 1, cycle 2, and cycle 3, respectively. This shows how the system improves at tracking the reference point over time. This is the result of the-progressive selective control. We used PSPC for listeria detection by amplifying DNA from a listeria colony as described in the PCR conditions. The total runtime was 2 h and 23 min for PSPC for 40 cycles (30 s, 30 s, 60 s),10 min hot-start, and 10 min final elongation. Because the heating block is not fully enclosed, a little condensation and evaporation could be observed in the exposed area of tubes, but it does not adversely affect the reaction outcome. Figure 9 shows the gel electrophoresis capture with both PSPC and commercial reaction amplicons for tubes 5 to 8 (1/10,000 to 1/10,000,000 dilution). Colony count from plated bacteria gave an average bacterial presence of 280 CFU/mL in tube 8, and 2160 CFU/mL in tube 7. This result can be extrapolated at higher concentrations to 20,000 CFU/mL in tube 6 and 200,000 CFU/mL in tube 5. The PCR result shows that the right amplicon sizes were amplified, and results are comparable on both systems. Even in instances such as 629 bp product size, the PSPC produced a brighter gel signal compared to the StepOnePlus system. The error at different cycles was compared using the area under the curve at each of the cycles. The definite integral computation resulted in values of 14,675.24, 8552 .74, and 545.32 for cycle 1, cycle 2, and cycle 3, respectively. This shows how the system improves at tracking the reference point over time. This is the result of the-progressive selective control. We used PSPC for listeria detection by amplifying DNA from a listeria colony as described in the PCR conditions. The total runtime was 2 h and 23 min for PSPC for 40 cycles (30 s, 30 s, 60 s),10 min hot-start, and 10 min final elongation. Because the heating block is not fully enclosed, a little condensation and evaporation could be observed in the exposed area of tubes, but it does not adversely affect the reaction outcome. Figure 9 shows the gel electrophoresis capture with both PSPC and commercial reaction amplicons for tubes 5 to 8 (1/10,000 to 1/10,000,000 dilution). Colony count from plated bacteria gave an average bacterial presence of 280 CFU/mL in tube 8, and 2160 CFU/mL in tube 7. This result can be extrapolated at higher concentrations to 20,000 CFU/mL in tube 6 and 200,000 CFU/mL in tube 5. The PCR result shows that the right amplicon sizes were amplified, and results are comparable on both systems. Even in instances such as 629 bp product size, the PSPC produced a brighter gel signal compared to the StepOnePlus system. Longer product size amplicon amplification was not successful on both systems as the process would require a longer elongation time than the one used in the protocol. Nevertheless, the other results were similar on both systems. PSPC was used to correlate Listeria amplicon population, or colony counts, in water, using the aforementioned PCR protocol. A 2 s integration time was used on the sensor for fluorescence measurement. Colony count shows an estimate of less than 10 CFU per PCR tube for 1/10,000,000 dilution. Figures 10 and 11 show the summary of the fluorescence level as a function of amplicon population on the StepOnePlus and PSPC systems. PSPC was used to correlate Listeria amplicon population, or colony counts, in water, using the aforementioned PCR protocol. A 2 s integration time was used on the sensor for fluorescence measurement. Colony count shows an estimate of less than 10 CFU per PCR tube for 1/10,000,000 dilution. Figures 10 and 11 show the summary of the fluorescence level as a function of amplicon population on the StepOnePlus and PSPC systems. The parallel operation on the StepOnePlus and PSPC systems showed similar level of sensitivity. Fluorescence signal recordings show a gradient from tubes 1 to tube 4 (1/1 to 1/1000 dilution) in both settings. Below that threshold, the detection no longer follows a gradient pattern. The experiment did not perform any type of bacteria preservation method, which could explain a random decay of amplicon population at lower concentrations. Also, the negative control showed a higher fluorescence level as compared to samples with lower bacteria concentrations. These suggest that the reagents used in the reactions contain background fluorescence. Nevertheless, the other results were similar on both systems. PSPC was used to correlate Listeria amplicon population, or colony counts, in water, using the aforementioned PCR protocol. A 2 s integration time was used on the sensor for fluorescence measurement. Colony count shows an estimate of less than 10 CFU per PCR tube for 1/10,000,000 dilution. Figures 10 and 11 show the summary of the fluorescence level as a function of amplicon population on the StepOnePlus and PSPC systems. The parallel operation on the StepOnePlus and PSPC systems showed similar level of sensitivity. Fluorescence signal recordings show a gradient from tubes 1 to tube 4 (1/1 to 1/1000 dilution) in both settings. Below that threshold, the detection no longer follows a gradient pattern. The experiment did not perform any type of bacteria preservation method, which could explain a random decay of amplicon population at lower concentrations. Also, the negative control showed a higher fluorescence level as compared to samples with lower bacteria concentrations. These suggest that the reagents used in the reactions contain background fluorescence. The parallel operation on the StepOnePlus and PSPC systems showed similar level of sensitivity. Fluorescence signal recordings show a gradient from tubes 1 to tube 4 (1/1 to 1/1000 dilution) in both settings. Below that threshold, the detection no longer follows a gradient pattern. The experiment did not perform any type of bacteria preservation method, which could explain a random decay of amplicon population at lower concentrations. Also, the negative control showed a higher fluorescence level as compared to samples with lower bacteria concentrations. These suggest that the reagents used in the reactions contain background fluorescence. We described the design, development and testing of a portable, low-cost, and realtime PCR system that can be utilized for food and water quality assessment. In addition, the system can be used in emergency health crises and point-of-care diagnostics. The described PCR system incorporated real-time reaction monitoring using fluorescence detection and eliminated the need of gel electrophoresis for reaction analysis, further decreasing the need of multiple reagents, reducing sample testing cost, and reducing sample analysis time. The bill of materials cost of the described system was approximately $340. The described PCR system utilized a novel progressive selective proportional-integral-derivative controller that helps in improving system accuracy and reducing sample analysis time. In addition, the system employs a novel primer-based approach to quantify the initial amplicon target concentration. The system is versatile and provided a clinically relevant performance, and can be used with a wide range of DNA samples. The present work focused on DNA, but the instrument could easily be adapted to RNA-type surveillance by incorporating an additional cDNA synthesis step prior to PCR. Author Contributions: Conceptualization, T.K. and V.P.C.; methodology, T.K. and Y.S.; software, T.K. and C.L.; validation, T.K. and Y.S.; formal analysis, T.K. and Y.S.; investigation, T.K.; resources, V.P.C. and Y.S.; data curation, T.K.; writing-original draft preparation, T.K.; writing-review and editing, T.K., V.P.C. and Y.S.; visualization, T.K.; supervision, V.P.C.; project administration, V.P.C.; funding acquisition, T.K. All authors have read and agreed to the published version of the manuscript. We would like to acknowledge the financial support given by University of Dayton School of Engineering. Molecular epidemiology of foodborne diseases Prevalence and antimicrobial resistance profiles of foodborne pathogens isolated from dairy cattle and poultry manure amended farms in northeastern Ohio, the United States Linking between contamination of environmental water and salmonella foodborne illness: A Review Advances in understanding of the innate immune response to human norovirus infection using organoid models Global prevalence and genotype distribution of norovirus infection in children with gastroenteritis: A meta-analysis on 6 years of research from 2015 to 2020 PCR past, present and future Polymerase chain reaction Progress report: Next-generation sequencing (NGS), multiplex polymerase chain reaction (PCR), and broad-range molecular assays as diagnostic tools for fever of unknown origin (FUO) investigations in adults Underwater Application of quantitative PCR on an ocean mooring Integrated portable polymerase chain reaction-capillary electrophoresis microsystem for rapid forensic short tandem repeat typing Predictive growth model of listeria monocytogenes under fluctuating temperature conditions in pasteurized milk by using real-time polymerase chain reaction Thermal Cycler Model TC 9639 with Multi-Format Block *Upgraded* Now with Gradient Capabilites Customize Your Chai Open qPCR Techne Prime Pro 48 Real-Time qPCR System-PCR Equipment and Supplies, Thermal Cyclers :text=A%20simple%20PCR%20machine%20like,over%2090%2C000%24%20for%20QuantStudio%2012k A rapid and low-cost PCR thermal cycler for infectious disease diagnostics Low-cost battery-powered and user-friendly real-time quantitative PCR system for the detection of multigene A Micromachined Low-Power-Consumption Portable PCR System Miniaturized flow-through PCR with different template types in a silicon chip thermocycler Battery-powered portable rotary real-time fluorescent qPCR with low energy consumption, low cost, and high throughput Convective polymerase chain reaction around Micro Immersion Heater Infrared-mediated thermocycling for ultrafast polymerase chain reaction amplification of DNA Parallel Picoliter RT-PCR assays using Microfluidics Polymerase chain reaction in microfluidic devices Listeria (listeriosis). U.S. Food and Drug Administration Real-time PCR: Overview and applications Revisiting the ziegler-nichols step response method for PID control A Better Option to Ziegler Nichols-PID Tuning Method Online controller tuning via frit and recursive least-squares Adaptive PID controller using RLS for Siso stable and unstable systems Regulation of intracellular pH in the rabbit cortical collecting tubule Photobleaching of fluorescent dyes under conditions used for single-molecule detection: evidence of two-step photolysis Institutional Review Board Statement: Not applicable. Data Availability Statement: Not applicable. The authors declare no conflict of interest.