key: cord-0922150-qt04ureq authors: Ibrahim, Essam H.; Ghramh, Hamed A.; Kilany, Mona title: Development of rapid and cost-effective top-loading device for the detection of anti-SARS-CoV-2 IgG/IgM antibodies date: 2021-07-21 journal: Sci Rep DOI: 10.1038/s41598-021-94444-6 sha: e720c994cd421c84ad6fc8bfda2a8efcbf9c1dd4 doc_id: 922150 cord_uid: qt04ureq Infection with SARS-CoV-2, the Betacoronavirus, caused a pandemic that affected the globe negatively. The gold method, RT-PCR, can detect SARS-CoV-2 but it is time-consuming and needs sophisticated equipment and professional personnel. On the other hand, rapid tests offer fast results and can detect anti-SARS-CoV-2 antibodies (Abs). The aim of this study is to develop a new rapid and cost-effective method for the detection of anti-SARS-CoV-2 IgG/IgM Abs. A new top-loading detection device was developed and composed of a small piece of plastic (25 × 25 × 0.5 mm) with an opening in the center, a piece of nitrocellulose (NC) membrane enough to block the opening from one side and adhesive tape to affix the NC to the plastic piece. The NC is blotted with anti-human IgG/IgM and rabbit serum. The device was evaluated against a commercially available IgG/IgM ELISA detection kit using normal, Covid-19-positive, HCV, HBV, and Cytomegalovirus-positive sera. Outcomes demonstrated simplicity, reproducibility, and accuracy of the new device and results can be obtained in less than 5 min. We anticipate our developed assay method to be used widely in point of care before deciding on the use of expensive nucleic acid assays. IgM and IgG are not recognized at all. Cellular reactions against the S protein have been portrayed and found to associate with IgG and IgA antibody titers 13, 14 . Total lymphocytes (CD4 + /CD8 + cells, B-lymphocytes and NK cells) showed a significant correlation with inflammatory status in SARS-CoV-2 infection, with special regards to CD8 + T cells and CD4 + /CD8 + ratio 15 . Decreased absolute numbers of T cells (CD4 + /CD8 + ) were clear in mild cases and severe cases but exaggerated in the severe cases. The CD4 + T cells IFN-γ production was shown to be decreased in severe cases than in moderate cases 16 . The rapid genomic sequence analysis of SARS-CoV-2 facilitated the design of reverse-transcription polymerase chain reaction (RT-PCR) methods and antibody-based diagnostics 17 . RT-PCR methods can detect the presence SARS-CoV-2 in upper respiratory tract specimens. The majority of the antibody-based methods depend on recombinant S protein. Some use the nucleoprotein (N), but taking into account that N is more conserved among coronaviruses and the possibility of cross-reactivity. Several laboratory assays are validated, and several antibody data were accrued [18] [19] [20] [21] [22] . There are several formats of antibody tests, including rapid lateral flow devices, enzyme-immunoassay techniques, and virus neutralization assays. Antibody diagnostic tests are useful in discovering the asymptomatic and/or recovered persons 23 . In respect of its low-cost, fast, sensitive, low limit of detection and specific diagnostic power, the lateral flow immunoassay (LFA) technique became the popular diagnostic method in recent years 24 . This technique can be helpful in the detection of the SARS-CoV-2 infection, offering a rapid diagnosis of suspected patients. Thousands of lateral flow assay kits have been developed since the emergence of SARS-CoV-2 25 . Different formats are adopted in LFA but all have a basic structure. LFA contains four main components; sample application pad, conjugate pad, nitrocellulose membrane and adsorbent pad 26 . Sample application pad is made of cellulose or glass fiber. To that pad, the sample is applied to start the assay. Conjugate pad is the place where labeled biomolecules are dispensed. Liquid sample diffusing from the sample pad reacts with materials found in the conjugate pad, releases labeled biomolecules and moves them toward the nitrocellulose membrane. Nitrocellulose membrane has test and control lines. It is ideal for providing good support and binding to capture probes (antibodies or antigen). Adsorbent pad is found at the end of the strip and works as a sink. It works to keep a continuous flow of the liquid over the membrane and prevents back flow of the sample. Taking into consideration the studies completed on SARS and MERS, the immune responses of exposed individuals would be focused on antibody responses to the S protein [27] [28] [29] . So, in this study, we aimed to design a new format of colloidal gold-based immune-diagnosis using the S protein and a top-loading method. Characterization of synthesized AuNPs. Gold nanoparticles (AuNPs) were shown to be synthesized at a peak of 520-570 nm (Fig. 1A ,B) with spherical shape (Fig. 1C) . The powder X-ray diffraction (PXRD) pattern ( Fig. 2 ) displayed some diffraction peaks at ca. 38.29°, 44.52°, 64.69° and 77.45°, which are well-indexed to the (111), (200), (220) and (311) facets of the typical fcc phase of gold (JCPDS no. 04-0784), respectively 30 . Other diffraction peaks may be attributed to the organic materials (citrate) during preparation. The average grain size was calculated to be 17.11 nm (Table 1) 31 . www.nature.com/scientificreports/ Optimization of protein concentration and pH value. The concentration of the S protein that is capable of preventing the clumping of AuNPs in the presence of 10% NaCl was determined. Protein concentration of 10 µg/mL started to protect AuNPs from precipitation by 10% NaCl. Protein protection continued to increase until the concentration of 50 µg/mL (Fig. 3 ). Optimal pH for S:AuNPs conjugation was tested in the pH range of 4.2-10.6 by using 40 µg S protein/mL AuNPs. Optimal pH for S:AuNPs conjugation was shown to be stable at all tested pH values (4.2-10.6) (Supplementary Fig. S1 online). We continued using the pH 8.2 for further work. S-colloidal gold conjugate probe. The S protein was mixed with colloidal AuNPs to get S-colloidal gold conjugate. The final S-gold conjugate was tested using SEM analysis and the SEM image revealed aggregates of spherical coated AuNPs ( Supplementary Fig. S2 online) . www.nature.com/scientificreports/ Sensitivity of TLTD. Sera were serially diluted in wash buffer up to 1:128 and its reactivity against S protein was tested. Results showed that antibodies were detectable up to the dilution 1:16 in some sera, but it was good for all sera at the dilution 1:4 ( Fig. 4) . were tested for anti-SARS-CoV-2 IgG and IgM antibodies using the commercially available ELISA kit. The results of ELISA greatly matched those obtained by the top-loading test device (TLTD), but with the discrepancy in the intensity of the reaction in some sera. All control samples showed no IgG and IgM reactive antibodies against SARS-CoV-2. Patient sera showed different reactivity regarding IgG and IgM depending on the case, but most of them were IgM reactive, either alone or showing faint dot at IgG place. Many samples were negative for both IgM and IgG and only 2 samples were positive IgG alone ( Fig. 5 and Supplementary Fig. S3 online). To determine specificity of the TLTD for COVID-19 detection, normal and HCV, HBV, HIV, and cytomegalovirus-infected sera were tested using TLTD. No positive reaction was seen using these www.nature.com/scientificreports/ sera ( Supplementary Fig. S4 online) indicating the specificity of TLTD for Covid-19. Also, the TLTD was tested against normal (syphilis, malaria, HCV, HBV, HIV, HTLV and cytomegalovirus-free) sera and was found to be non-reactive to any of the tested sera ( Supplementary Fig. S5 online). Reproducibility and stability of VID. Results indicated reproducibility and stability of the TLTD formats 2-4. Format 1 showed non-reproducible results as dots either appeared with different intensities or did not appeared at all. SARS-CoV-2, the positive RNA virus, caused global health and economic disturbances. A new genetic mutation appeared in the Great Britain to begin a new episode of the pandemic. Accurate and rapid diagnosis is required in any disease including Covid-19. At the onset of COVID-19 symptoms, sera from infected persons contained detectable anti-SARSA-CoV-2 Abs 32 . qPCR, the method which depends on the detection of the viral RNA, is used as the gold-standard method for the diagnosis of SARS-CoV-2 infection 33 . Notwithstanding, numerous labs and medical clinics do not have the facilities and expertise needed to run the qPCR. Point-of-care (POC) needs rapid and simple diagnostic tests that are designed to shorten the sample-to-result timeline, enabling fast diagnosis and right treatment 34 . Hence, there is a growing utilization of rapid tests for the diagnosis of Covid-19, giving the advantage to quickly affirm the infection. Lateral flow assays are the most widely used form in POC because of its low-cost and ease of use. In addition to rapid assay tests, ELISA-based assays are useful in detecting the antibodies directed against the SARS-CoV-2. Here, we tried to create a new method for the detection of coronavirus infection at antibody level. We developed a new device that uses fewer materials to build up and does not need highly professional personnel to perform the test at any POC location. A lateral flow device is composed of sample pad, conjugate pad, NC membrane and absorbing filter paper affixed together on a plastic backing and packed into a plastic device 35 . TLTD is composed only of NC membrane affixed to adhesive tape and a piece of square plastic with an opening. This demonstrates how TLTD is simple, economic and easy to assemble. We performed the test using four formats. Format one had been previously used to detect other diseases such as Leptospirosis 36 , food allergen 37 , HIV-1 38 , etc. Before evaluating the formats described in this study, all samples were tested for COVID-19 IgG and IgM using a commercially available ELISA kit. The plates of the kit are coated with the S protein. This matches our created TLTD as both will measure only IgG and IgM directed against the S protein. Format 1 showed non-reproducible results as dots either appeared with different intensities or did not appear at all. This may be due to the manual assembly of the device, no enough time for antibodies interactions or this method is not good enough. So we decided to use another formats. Format 3 and 4 uses fewer amounts of liquids (serum and conjugate) as compared to format 2 as the thickness of the plastic used to build the body of formats 3 and 4 is less than that of format 2. The antibody detection results of formats 2-4 were comparable to that obtained by ELSA. Most of the qPCR-positive samples were IgM-positive, some were IgG/IgM-negative, little IgG-positive and many IgG/IgM-positive. The positive sera for anti-S IgM antibodies may be due to variables such as most of the patients are in early phase of immune response where the IgM still dominant and the IgG antibodies started to appear or have still not appeared. All HCV, HBV, HIV, and cytomegalovirus-infected sera gave negative results when tested using TLTD indicating the specificity of TLTD for Covid-19. Also, when testing the TLTD against syphilis, malaria, HCV, HBV, HIV, HTLV and cytomegalovirus-free sera, the TLTD was found to be non-reactive to any of the tested sera indicating that the device in not reactive with any component of SARS-CoV-2-free human sera. So, we can conclude that TLTD does not cross react with other viruses and is negative for the uninfected sera. The purpose of making the TLTD in these formats is to use it in more remote areas or POC before deciding to choose qPCR assay. The transport and storage without the use of a cold chain is a great advantage of any product. Here, we tested the stability of TLTD after incubating it at different temperatures. The results reflected the ability of TLTD to tolerate high temperature (37 • C) for a long time (30 days), giving an advantage to the newly designed device. Results indicated reproducibility and stability of the TLTD formats 2-4 but format 1 was non-reproducible. A new design of rapid antibody detection device was developed. The new device is simple, accurate, reproducible, and cost-effective. It is very suitable for the detection of IgG and IgM arisen in SARS-CoV-2-infected persons in point of care and remote areas. The device is easy to use and does not need highly professional personnel to perform. It overcomes the need for sophisticated equipment like the PCR system for the diagnosis in POC and in remote areas and can be completed within minutes. Preparation of colloidal gold nanoparticles. Nano www.nature.com/scientificreports/ a scanning electron microscope (SEM, JSM-7500 F; JOEL-Japan) at 80 kV. The powder X-ray diffraction (PXRD) measurement was carried out by X-ray diffractometer (Shimadzu LabX-XRD-6000 diffractometer) run at 40 kV and 35 mA and a spectrum was recorded by CuKα radiation with a wavelength of 1.5406 Å in the 2θ range of 10°-80° with Shimadzu software with the pdf-2 library for the analysis of XRD data. The average crystallite size (D), was estimated by Debye-Scherer's Eq. 40 : where the shape factor k is constant, equals 0.9 (k is dimensionless), X-ray's beam wavelength λ in nm, β is known as the full width at half maximum (FWHM) of the peak in radian and the Braggs' diffraction angle is θ in degree. Optimization of pH and protein concentration for the AuNPs conjugation. The concentration of the S protein (Cusabio Technology LLC., China) that is capable of preventing the clumping of AuNPs in the presence of 10% NaCl (Sigma-Aldrich) was determined and this concentration was selected for further use. First the pH of the colloidal gold was adjusted to 8.2 with 0.1 M K 2 CO 3 (Sigma-Aldrich). Various amounts of S protein (0, 5, 10, 15, 20, 25, 30, 35, 40, 45 and 50 µg) were individually mixed with 1 mL AuNPs and incubated for 10 min with shaking at room temperature. After that, 10% NaCl (100 µL/tube) was added, continued shaking for 10 min and then left standing for 2 h and each tube was scanned spectrophotometrically (460-700 nm). The amount of protein showed maximum absorbance at about 520 nm was considered as the minimum quantity of S protein required to prevent the precipitation of colloidal gold. Optimal pH for S:AuNPs conjugation was tested in the pH range of 4.2-10.6 by using 40 µg S protein/mL AuNPs and testing the proper conjugation as described above. Preparations of S-AuNPs conjugate. The S protein (at a final concentration of 40 µg/mL) was mixed with 10 mL of AuNPs (pH 8.2) and gently shaken for 30 min. The conjugation process was stabilized via the addition of 2.5 mL of 10% BSA (Sigma-Aldrich) while gently shaking for 30 min. After that, 250 µL of 5% PEG20,000 (Sigma-Aldrich) solution (with a final concentration of 0.1%) was added to the mixture and left at room temperature while gently stirring for 16 min. The S-AuNPs conjugate was centrifuged at 14,000 rpm (10 min/4 °C), the supernatant was discarded and the pellet was resuspended in 1 mL storage buffer (phosphate-buffered saline (PBS, 137 mM NaCl, 2.7 mM KCl, 10 mM Na 2 HPO 4 , and 1.8 mM KH 2 PO 4 , pH 7.4) containing 5% sucrose, 2% bovine serum albumin (BSA), 0.5% Tween-20, and 0.03% NaN 3 ). The final S-gold conjugate was stored at 4 • C for further study. The conjugate was tested using SEM analysis. We prepared four formats of top-loading test device (TLTD) for the detection of human IgG and IgM antibodies (Abs) against S protein of SARS-CoV-2 as described. Format 1: A container with a screw cap was used to assemble the TLTD (Fig. 6) . The cap has a ring opening of 12 mm diameter at the center. The container was filled with several layers of square absorbing filter papers (Thomas Scientific). A nitrocellulose (NC) piece (0.45 µm, GE Healthcare-Life Sciences) is laid on the top of the absorbing filter papers and the cap is closed to hold the NC paper firmly in contact with the absorbing filter papers. Format 2: only the cap of format 1 was used but the NC paper was sealed with an adhesive tape to block the hole from the inside of the cap. Format 3: A square piece of flat plastic with a hole in its center was used. One side of the hole is blocked using a piece of NC membrane and adhesive tape. Format 4: The same design as that of format 3 but with rectangular opening. Principle and mechanism of TLTD workings. Nitrocellulose paper is manually blotted with 1 µL of goat anti-human IgG Fc Abs (1 mg/mL, R&D Systems), 1 µL of goat anti-human IgM Abs (1 mg/mL, Vector Laboratories, Inc.) and 1 µL of rabbit serum (Sigma-Aldrich) as a positive control and left to dry at 37 • C for 1 h. The NC was blocked with blocking buffer (PBS containing 2% BSA, 0.5% Teween-20) for 3 h. The NC was washed with washing buffer (PBS containing 0.5% Teween-20) and left to dry. The NC was then assembled into the device as described above. The principle of the detection of anti-SARS-CoV-2 IgG and IgM Abs using all formats is shown in Fig. 7 . Format 1: Human serum is being added to the hole of the TLTD directly onto NC membrane and left to pass through by the action of gravity and the absorbing filter papers below the NC. IgG and IgM which are present in the human serum will be captured by the corresponding blotted antibodies. The prepared probe (S-AuNPs) mixed with the commercially available goat anti-rabbit IgG − gold antibody (Sigma-Aldrich) are added to the hole of the TLTD directly to NC membrane and left to pass through by the action of gravity and absorbing filter papers below the NC. Approximately 2-3 drops of wash buffer are added to the hole of the TLTD if needed. If any of the captured human IgG or IgM are reactive toward the S protein of SARS-CoV-2, the S-Gold probe will be captured by these antibodies and indicated by the characteristic reddish-brown color of the gold NPs. Anti-rabbit IgG-gold will react with IgG present in the blotted rabbit serum giving a reddish-brown dot (Supplementary video 1) . Regarding formats 2-4, the serum is applied to the hole of the TLTD, left for 1-2 min, liquid is absorbed using absorbing filter paper, mixed gold conjugates are applied to the hole of the TLTD, left for 1-2 min, liquid is absorbed using absorbing filter paper and washed twice by adding wash buffer to the hole and absorbing the buffer with absorbing filter paper (Supplementary video 2). Where NC: Nitrocellulose membrane covering the plastic opening; FP: absorbing filter papers. Format 1 is formed of a cap with a circular opening, a nitrocellulose membrane put into the inside of this cap to cover its opening (1A), FP put over the NC (1B), and the cap containing the NC and FP are put on the container body to get the final shape of the TLTD format 1 (1C). Format 2 is formed of only a cap (2A) and a NC (2B) affixed with a tape into the inside of the cap to cover its opening (2C). Format 3 is formed of only a square piece of plastic with circular hole in its center (3A) and a NC affixed with a tape to into one side to cover its opening (3B). Format 4 is formed of only a rectangular piece of plastic (4A) with a centered rectangular opening and a NC affixed with a tape at one side to cover its opening (4B). www.nature.com/scientificreports/ Samples. The Research Ethics Committee at King Khalid University (HAPO-06-B-001) reviewed and agreed on this study. All experiments were performed in accordance with the above mentioned committee guidelines and regulations. Patient informed consent was not a need as we got coded serum samples without patient information. We used patient sera that were diagnosed positive for COVID-19 via RT-PCR (n = 89). We have stored coded sera (n = 7267 random healthy blood donor volunteers) for several years before the onset of the pandemic and were used in a previous study (https:// journ als. ekb. eg/ artic le_ 13699. html). From these samples, negative control sera (n = 100, HIV, HCV, HBV, HTLV and cytomegalovirus-free), HCV positive (antibody and PCR confirmed, n = 5), HBV positive (surface antigen and PCR confirmed, n = 50) and cytomegalovirus positive (antibody confirmed, n = 20) were selected. All samples used in this work were tested for anti-SARS-CoV-2 IgG and IgM Abs using COVID-19 Human IgM IgG Assay Kit (Abnova). The ELISA plates of the Abnova kit are coated with the S protein, similar to the S-Gold probe used in this work. All samples were also tested for the presence of anti-SARS-CoV-2 IgG and IgM Abs using the four TLTD formats as described above. For titration purpose, sera were serially diluted in wash buffer (1:2, 1:4, 1:8, 1:16, 1:32, 1:64 and 1:128). The outcomes were decided as positive or negative by visual review and the test detection limit was identified as the minimum concentration of the serum at which the test dot was unambiguous. To determine the specificity of the TLTD for COVID-19 detection, normal and HCV, HBV, HIV, and cytomegalovirus-infected sera were tested using TLTD. TLTD stability and reproducibility. COVID-19 (positive and negative) confirmed samples were examined 3 times to explore the reproducibility of TLTD. For stability, TLTDs were kept at 4 °C, ambient temperature and 37 °C (accelerated heat stability) up to 30 days and examined using positive/negative samples at the end of incubations. Innate immunity in vertebrates: An overview Inhibition of toll like receptor immune responses by microbial pathogens Toll-like receptor control of the adaptive immune responses The immortality of humoral immunity Human T Cell Development, Localization, and Function throughout Life A novel coronavirus outbreak of global health concern Structural and functional properties of SARS-CoV-2 spike protein: potential antivirus drug development for COVID-19 A novel coronavirus genome identified in a cluster of Pneumonia Cases -Wuhan Structure, function, and antigenicity of the SARS-CoV-2 spike glycoprotein Early enhanced expression of interferon-inducible protein-10 (CXCL-10) and other chemokines predicts adverse outcome in severe acute respiratory syndrome Serum IgA, IgM, and IgG responses in COVID-19 Antibody responses to SARS-CoV-2 in patients with novel coronavirus disease 2019 Targets of T cell responses to SARS-CoV-2 coronavirus in humans with COVID-19 disease and unexposed individuals Phenotype and kinetics of SARS-CoV-2-specific T cells in COVID-19 patients with acute respiratory distress syndrome Characteristics of peripheral lymphocyte subset alteration in COVID-19 pneumonia Clinical and immunological features of severe and moderate coronavirus disease 2019 A new coronavirus associated with human respiratory disease in China Detection of SARS-CoV-2-specific humoral and cellular immunity in COVID-19 convalescent individuals Neutralizing antibody responses to severe acute respiratory syndrome coronavirus 2 in coronavirus disease 2019 inpatients and convalescent patients Neutralizing antibody responses to SARS-CoV-2 in a COVID-19 recovered patient cohort and their implications. SSRN Electron A serological assay to detect SARS-CoV-2 seroconversion in humans Preexisting and de novo humoral immunity to SARS-CoV-2 in humans Interpreting Diagnostic Tests for SARS-CoV-2 Nanoparticle-based lateral flow biosensors Covid-19 pandemic: Review of contemporary and forthcoming detection tools Designs, formats and applications of lateral flow assay: A literature review Longitudinally profiling neutralizing antibody response to SARS coronavirus with pseudotypes Antibody responses against SARS coronavirus are correlated with disease outcome of infected individuals Potent and persistent antibody responses against the receptor-binding domain of SARS-CoV spike protein in recovered patients Gold nanoparticle-containing membranes from in situ reduction of a gold(III)-aminoethylimidazolium aurate salt Biodistribution of colloidal gold nanoparticles after intravenous administration: Effect of particle size Antibody responses to SARS-CoV-2 in patients with COVID-19 SARS-CoV-2 and the COVID-19 disease: A mini review on diagnostic methods Point-of-care testing (POCT): Current techniques and future perspectives Development of colloidal gold-based immunochromatographic assay for rapid detection of goose parvovirus Sensitivity and Specificity of a New Vertical Flow Rapid Diagnostic Test for the Serodiagnosis of Human Leptospirosis A critical comparison between flow-through and lateral flow immunoassay formats for visual and smartphone-based multiplex allergen detection Development of a new limiting-antigen avidity dot immuno-gold filtration assay for HIV-1 incidence Controlled Nucleation for the Regulation of the Particle Size in Monodisperse Gold Suspensions Zinc-substituted hydroxyapatite: A biomaterial with enhanced bioactivity and antibacterial properties The authors extend their appreciation to King Abdulaziz City for Science and Technology (KACST) for funding this work through the Research Project (5-20-01-010-0009). All data are available in the manuscript and supplementary files. The authors declare no competing interests. The online version contains supplementary material available at https:// doi. org/ 10. 1038/ s41598-021-94444-6.Correspondence and requests for materials should be addressed to E.H.I.Reprints and permissions information is available at www.nature.com/reprints.Publisher's note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations. License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http:// creat iveco mmons. org/ licen ses/ by/4. 0/.