key: cord-0958710-moenp6nb authors: Chen, Hsuan-Yu; Chen, Andrew; Chen, Chiachung title: Investigation of the Impact of Infrared Sensors on Core Body Temperature Monitoring by Comparing Measurement Sites date: 2020-05-19 journal: Sensors (Basel) DOI: 10.3390/s20102885 sha: d36c176175904edc6271bdcc461af2fc172c9f3a doc_id: 958710 cord_uid: moenp6nb Many types of thermometers have been developed to measure body temperature. Infrared thermometers (IRT) are fast, convenient and ease to use. Two types of infrared thermometers are uses to measure body temperature: tympanic and forehead. With the spread of COVID-19 coronavirus, forehead temperature measurement is used widely to screen people for the illness. The performance of this type of device and the criteria for screening are worth studying. This study evaluated the performance of two types of tympanic infrared thermometers and an industrial infrared thermometer. The results showed that these infrared thermometers provide good precision. A fixed offset between tympanic and forehead temperature were found. The measurement values for wrist temperature show significant offsets with the tympanic temperature and cannot be used to screen fevers. The standard operating procedure (SOP) for the measurement of body temperature using an infrared thermometer was proposed. The suggestion threshold for the forehead temperature is 36 °C for screening of fever. The body temperature of a person who is possibly ill is then measured using a tympanic infrared thermometer for the purpose of a double check. Body temperature is an indicator of human physiological activity and health, especially in pediatrics, surgery and general emergency departments [1] . Most of the early methods of measuring body temperature used contact mercury thermometers. Advances in electrical technology mean that contact-type electronic thermometers have become widely used. The measurement sites are the sublingual mouth, the rectum, the axillary and the base of the urethra. Some medical reports show that contact thermometers accurately measure body temperature [2, 3] . To screen for an illness, the body temperature of many individuals must be measured, so noncontact type infrared thermometers (IRT) are used to measure the tympanic (ear) and forehead temperature. These IRTs are fast, convenient and safe to use. Medical IRTs for ear temperature and forehead temperature are used in medicine. When the temperature of a natural object is higher than the absolute temperature, the surface emits thermal radiation. The infrared thermometer (IRT) detects this radiant energy that is released by the object via sensing elements and converts it into an electrical signal. After signal processing, 0.74 • C and for the tympanic temperature and forehead temperature, the difference was −1.48 • C to 2.52 • C. The correlation coefficients for the forehead temperature, the esophageal temperature, and ear temperature are small. Measuring the temperature of the forehand using an IRT is convenient, but some studies show that the measurements are not accurate. There are three types of forehead thermometer: a thermistor probe, a liquid crystal strip and an IR thermometer. The thermistor probe and the liquid crystal strip are contact-type thermometers. The thermistor probe is a deep body thermometer because it is inserted into the tissue of the forehead. Studies show that a contact-type deep skin forehead thermometer is suited to clinical use [29] [30] [31] [32] [33] . This device was very accurate. Duran et al. [8] compared the performance of a contact-type forehead thermometer with that of an axillary glass-mercury thermometer and found that both thermometers give similar readings. Zeiner et al. [34] compared the body temperature that was measured using a deep forehead thermometer and a bladder thermometer and found no significant difference between the two readings. Liquid crystal thermometers have been used to measure forehead temperature [35] [36] [37] [38] [39] . These studies show that liquid crystal devices are not as accurate as forehead thermistor thermometers and do not give as accurate a reading for core temperature as a rectal or tympanic measurement. Patel et al. [27] compared esophageal, tympanic and forehead temperature for adults. The mean difference between the esophageal and the forehead temperature was 0.3 • C and the mean difference between the tympanic and the forehead temperature was 0.5 • C. The forehead temperature was measured using a contact-type, liquid crystalline thermometry strip. The authors noted that the forehead temperature was not an accurate representation of the standard core temperature. Asadian et al. [40] measured the oral, axillary, tympanic and forehead temperature. The central nasopharyngeal measurement was used as a standard value. The results show that the forehead and tympanic readings were the least and most accurate, respectively. The forehead temperature was measured using a contact-type strip thermometer. Berksoy et al. [41] used a noncontact IRT to measure forehand temperature and compared the reading with those for tympanic and rectal temperature. The results show that measurements of the tympanic temperature are more practical than axillary thermometry. Duncan et al. [42] showed that the difference between the oral temperature and the temperature that was recorded using an IRT was 0.89 • C ± 0.58 • C. The authors commented that the thermometers were not interchangeable. Kocoglu et al. [10] compared the accuracy of a tympanic IRT reading with the readings for axillary and rectal temperature that were measured using a glass thermometer. The highest temperature was the rectal temperature and the lowest temperature was the axillary temperature. The difference between the mean values for tympanic and rectal temperature was <0.2 • C. Yeoh et al. [43] studied the tympanic membrane vicinity as the measurement site of core body temperature and found the mean temperature of rectum, esophagus, left ear and right ears were 37.2 • C, 36.8 • C, 36.2 • C and 36.1 • C, respectively. Oguz et al. [44] measured the tympanic and axillary body temperature of 1364 children and found the mean axillary and tympanic temperatures were 36.04 • C and 36.91 • C, respectively. The axillary and tympanic temperature that are considered as fever are >37.0 • C and >37.8 • C, respectively. Cutuli et al. [45] reported the inconsistent opinions of the unit protocol for measuring body temperature and suggested the requirement of detailed studies of the noninvasive temperature measurement methods. Mogensen et al. [46] used the rectal temperature as a standard value. The study measured the tympanic temperature using an IRT and the forehead temperature using a contact-type temporal thermometer. The results show that the tympanic measurement was adequate for screening purposes but the forehead temperature was not. Kistemaker et al. [47] tested two types of contact forehead thermometers and compared the readings with those for a rectal thermistor. The results showed that a contact-type forehead thermometer registered a higher temperature than a rectal thermistor. Dante et al. [48] compared the performance for contact-type forehead, tympanic and axillary thermometers. The mean value for tympanic temperature was higher than that for the forehead and axillary thermometers, and these devices were not interchangeable. Ng et al. [13] used a noncontact, handheld IRT to measure the forehead skin temperature. The reference temperature was measured using an electronic thermometer. The surface temperature was underestimated by 2 • C, so a temperature of 35.6 • C was the index temperature for fever if the forehead temperature was measured using a noncontact, handheld IRT. Ng et al. [14] studied the relationship between the facial skin temperature that was detected by an IRT, and the actual temperature was detected using a direct thermometer. They recommend that the measured temperature of 35.5 • C signified a fever. Measuring body temperature is an effective technique for screening SARS and other flus. The coronavirus (COVID-19) has spread globally and temperature measurement is used to rapidly screen people. Handheld forehead IRTs are easy to use, rapid, noncontact and inexpensive, so they are widely used. This IRTs was reported to measure wrist temperature and forehead temperature. However, these devices have unproven reliability [15] . Many studies compare temperature measurements for different parts of the human body using different thermometers. Most of the forehead thermometers are contact-type. The measurements of body temperature depend on the type of thermometer, the sensing elements and the manufacturers' claimed performance for each thermometer. This study uses two commercial tympanic IRTs and an industrial IRT to measure the tympanic, forehead and wrist temperatures, and the measurement data is analyzed using statistical techniques. The results for the measurement of body temperature are used to define the threshold for fever screening using a forehead IRT. The received energy from the measuring target is as follows: σε is the Stefan-Boltzmann constant, ε is the emissivity of this target and T k is the absolute temperature of this target in K. If the reflection of the ambient radiation and the self-radiation of the infrared thermometer are considered, Equation (1) is revised as follows: where C 1 is the constant of infrared device, T amb is the ambient temperature in K and T dev is the device temperature in K. If the infrared thermometers do not cover all ware length range, an exponent constant, n is used to replace the power constant of 4. The target temperature is then determined as follows: To ensure the accuracy of the T k measurement, the T dev value must be detected with a thermometer of highly accuracy and the T amb is kept stable. Two types of tympanic IRTs and one industrial IRT were used for this study: a BRAUN IRT-3020 Thermoscm model (Braun Co., Melsungen, Germany), an OMRON MC-510 Gentle Temp model (OMRON Co., Kyoto, Japan) and a THI 301 IRT (Tasco Co., OSAKA, Japan). The specifications of these IRTs are listed in Table 1 . Nonlinearity and repeatability 37 . The temperature calibrator that is used for this study is a TC-2000 Scan Sense (Instrutek Co., Larvik, Norway). The temperature of an oil bath was detected using a PT-100 thermometer and the manufacturer's specification signifies an uncertainty of 0.03 • C. The IRTs were calibrated using theTC-2000 temperature calibrator at temperatures of 34 • C, 36 • C and 38 • C. The calibration methods are described in a previous study [49, 50] . Two experiments were performed for this study. The BRAUN IRT-3020 thermometer was used to measure the tympanic temperature in the right and left ears, the forehead temperature and the wrist temperature of the same subject. The test used 614 males and females, aged from 16 to 60 years. The ambient temperature was measured with a PT-100 hand-held thermometer (Electronic Temperature Instrument Ltd., West Sussex, UK). The accuracy of this thermometer is ±0.1 • C. The temperature of the measurement room was different in relation to its air-conditioning devices and the testing time of day. All subjects waited in the room for five minutes after entrance and before measurement. For the same subject, The BRAUN IRT-3020 thermometer and the OMERN MC-510 thermometer were used to measure the tympanic temperature in both ears. The THI-301 thermometer was used to measure the temperature at the left and right wrists and the forehead temperature. Due to the different periods of experiments, the subjects of the first and second experiments were different. 168 individuals aged 18-42 years were used for this test. All temperature readings were analyzed using Excel software. Data are expressed as a mean and a standard deviation. The coefficient of variance (CV) value is calculated to determine the accuracy of measurement. CV = (standard deviation)/mean (5) Significant differences between the two sets of data were tested using a paired t-test. A p value equal to or less than 0.05 denotes statistical significance. The relationship between measurement values is determined using a correlation analysis and a regression analysis. Six hundred fifty-nine randomly selected citizens were recruited for this study. The results of the tests are listed in Table 2 . The CV value is less than 1.0% for tympanic temperature. The CV value for forehead is 1.129%, and for wrist temperature is 1.332%. A CV of 5% indicates good precision [51] . The distribution for forehead and wrist temperature is higher than for tympanic temperature. The correlation between these measurement values is shown in Table 3 . The correlation coefficients are small, except for the measurement values for two ears. A paired t-test for the tympanic temperature in both ears showed that there is no significant difference (t = 0.474, p < 0.05) in the readings for tympanic temperature for left and right ears. The relationship between the air temperature and the body temperature as measured at the ear, the forehead and the wrist are shown in Figure 1 . There is no significant correlation between air temperature and body temperature. That is, the environmental temperature does not have a significant effect on the body temperature measurement. The sensing element of BRAUN IRTs is pyroelectrical detectors, and the uncertainty of this device is low [49] , so the average of the left and right tympanic temperature (Ear ave ) is denoted as the standard body temperature. This experiment used a tympanic IRT to directly measure the forehead and wrist temperatures. The different between Ear ave and the forehead is termed Ear ave -forehead and that between Ear ave and the wrist temperature is termed Ear ave -wrist. The relationship between Ear ave -forehead and Ear ave is shown in Figure 2 . The relationship between Ear ave -wrist and Ear ave is shown in Figure 3 . The lowest possible reading for the BRAUN thermometer is 34 • C, so some of the data in the two figures show a fixed pattern. Earave-forehead is the difference between average ear temperature (Earave) and forehead temperature. Earave-forehead is the difference between average ear temperature (Earave) and forehead temperature. Earave-wrist is the difference between the average ear temperature (Earave) and wrist temperature. This device cannot measure a temperature of less than 34 °C. A statistical correlation analysis and regression analysis shows that there is no significant correlation between tympanic temperature and forehead and wrist temperature. The mean forehead temperature is 34.71 °C ± 0.392 °C and the mean wrist temperature is 34.16 °C ± 0.355 °C. The mean difference between the average tympanic temperature and the forehead temperature is 2.2 °C ± 0.411 °C. These results are similar to those of a previous study [18, 52, 53] . The second experiment measured the tympanic temperature in the left and right ears using a BRAUN and an OMRON IRT, and the temperature of the forehead and two wrists was measured using an industrial THI-301 IRT for the same subject at the same time. The results of the measurements are listed in Table 4 . A paired t-test for the temperature in the left and right ears gives a value of t = 2.38 (p < 0.02) for the BRAUN IRT and of t = 2.48 (p < 0.02) for the OMRON IRT. The paired t-test shows that there is no significant difference at the level of p < 0.02. The difference in the readings of the two IRTs is 0.22 °C ± 0.201 °C, which is acceptable for practical applications. This device cannot measure a temperature of less than 34 • C. A statistical correlation analysis and regression analysis shows that there is no significant correlation between tympanic temperature and forehead and wrist temperature. The mean forehead temperature is 34.71 • C ± 0.392 • C and the mean wrist temperature is 34.16 • C ± 0.355 • C. The mean difference between the average tympanic temperature and the forehead temperature is 2.2 • C ± 0.411 • C. These results are similar to those of a previous study [18, 52, 53] . The second experiment measured the tympanic temperature in the left and right ears using a BRAUN and an OMRON IRT, and the temperature of the forehead and two wrists was measured using an industrial THI-301 IRT for the same subject at the same time. The results of the measurements are listed in Table 4 . A paired t-test for the temperature in the left and right ears gives a value of t = 2.38 (p < 0.02) for the BRAUN IRT and of t = 2.48 (p < 0.02) for the OMRON IRT. The paired t-test shows that there is no significant difference at the level of p < 0.02. The difference in the readings of the two IRTs is 0.22 • C ± 0.201 • C, which is acceptable for practical applications. The measurement results for the THI-301 IRT are listed in Table 5 . The CV values for the three measurements are lower than 2%. The criterion for dispersion is 5% [51] . These CV values demonstrate that the device has good precision. The minimum respective values for the forehead, left wrist and right wrist temperature are 33.8 • C, 32.4 • C and 32.2 • C. The tympanic IRT cannot read a temperature of less than 34 • C so the tympanic IRT is not sufficiently accurate at this temperature to measure the wrist temperature. The BRAUN IRT has the lowest values of combined uncertainty [52] , so the average value for the right and left tympanic temperatures as measured using the BRAUN IRT (noted as BRAUN ave ) are served as the standard values for this study. The results are listed in Table 6 . B ave -Forehead: the difference between the average tympanic temperatures using a BRAUN tympanic infrared thermometer and the forehead temperatures using a THI-301 infrared thermometer. B ave -Wrist right : the difference between the average tympanic temperatures using a BRAUN tympanic infrared thermometer and the right wrist temperatures using a THI-301 infrared thermometer. B ave -Wrist left : the difference between the average tympanic temperatures using a BRAUN tympanic infrared thermometer and left wrist temperatures using a THI-301 infrared thermometer. Wrist right -Wrist left : the difference between the wrist temperatures of two hands using a THI-301 infrared thermometer. The difference between the average temperature that is measured by the BRAUN and the temperature at the forehead is denoted as BRAUN ave -Forehead and is 2.107 • C ± 0.301 • C. The difference between the average temperature that is measured by the BRAUN and the temperature at the right wrist is denoted as BRAUN ave -Wrist right and is 3.2781 • C ± 0.527 • C. The difference between the average temperature that is measured by the BRAUN and the temperature at the left wrist, which is denoted as BRAUN ave -Wrist left , is 3.308 • C ± 0.534 • C. The mean forehead temperature is 2.017 • C lower than the mean tympanic temperature, and the mean wrist temperature is 3.3 • C lower than the mean tympanic temperature. The paired t-test for the wrist temperatures of the left and right hands, measured using a THI-301 industrial IRT, is insignificant (t = 0.71, at p < 0.05 criterion). A study by Werner et al. [53] showed that the difference between the core temperature and the forehead temperature is 2.1 • C-2.3 • C and the difference between the core temperature and the wrist temperature is 3.0 • C-3.3 • C. The result of this study is similar to that of the previous study [53] . The correlation analysis for the measurement results for the second experiment is shown in Table 7 . All correlation coefficients for the observed values are low, except for the temperatures in left and right ears which were detected using two tympanic IRTs. The relationship between BRAUN ave -Forehead and BRAUN ave is shown in Figure 4 . The relationships between BRAUN ave -Wrist right , BRAUN ave -Wrist left and BRAUN ave are shown in Figure 5 . Correlation analysis and regression analysis show that there is no significant correlation. The relationship between BRAUNave-Forehead and BRAUNave is shown in Figure 4 . The relationships between BRAUNave-Wristright, BRAUNave-Wristleft and BRAUNave are shown in Figure 5 . Correlation analysis and regression analysis show that there is no significant correlation. Temp. diff., ℃ The average ear temperature using a BRAUN infrared thermometer, ℃ Figure 4 . The relationship between B ave -Forehead and B ave : B ave -Forehead is the difference between the average era temperature (Ear ave ) using a BRAUN infrared thermometer and the forehead temperature using a THI-301 infrared thermometer. The relationship between the Bave-Wristright, the Bave-Wristleft and the Bave. Bave-Wristright is the difference between the average ear temperature (Earave) using a BRAUN infrared thermometer and the right wrist temperature using a THI-301. These results show the diversity of the body temperature. There is a high correlation between the tympanic temperature in the left and right ears. However, a paired t-test shows different results for the two experiments. In the first experiment, there is no significant difference between the temperature in the right and left ears, as measured using a BRAUN IRT (p < 0.05). In the second experiment, there is a significant difference in the temperature in the right and left ears for the two types of IRT (p < 0.002). The subjects for the two experiments were different. The first experiment involved 659 subjects and the second experiment involved 114. The size of the sample and a different age distribution affects the standard deviation, which affects the calculation of t values. In the first experiment, the difference between the mean tympanic temperature and the forehead temperature is 2.20 °C ± 0.711 °C. This experiment used a tympanic IRT to detect the forehead temperature. The emissivity of this device cannot be adjusted. In the second experiment, the difference between the mean tympanic temperature and the forehead temperature is 2.1 °C ± 0.301 °C. This experiment used an industrial IRT to measure the forehead temperature. The emissivity of this industrial device was adjusted to 0.98. Both experiments produce similar results. The forehead temperature is 2.1 °C or 2.2 °C lower than the tympanic temperature. In the second experiment, the wrist temperature that is measured using a THI-301 IRT is 3.3 °C lower than the tympanic temperature. These measurements for wrist temperature do not represent the actual body temperature, so measurements of the wrist temperature cannot be used to establish a threshold for screening for fever. Measurements of forehead temperature are convenient, fast and involve a low risk of infection, so they are used to screen for fever to identify possible victims of the novel coronavirus (COVID-19) . However, there is no international standard for the threshold of forehead temperature that indicates fever. The Department of Health of Hong Kong [54] recommended a threshold level for fever of >38 °C for ear temperature and >36 °C for forehead temperature. The difference between the two critical temperatures is 2 °C. The results of this study are close to these figures. Liu et al. [12] determined the accuracy of alternative IR techniques for fever screening during the 2003 SARS epidemic. Two types of IRTs were used to measure tympanic and forehead temperature. The results for 276 subjects showed a tympanic temperature of 36 Temp. diff., ℃ The average ear temperature using a BRAUN infrared thermometer, ℃ Tbraun-left Tbraun -right Figure 5 . The relationship between the B ave -Wrist right , the B ave -Wrist left and the B ave . B ave -Wrist right is the difference between the average ear temperature (Ear ave ) using a BRAUN infrared thermometer and the right wrist temperature using a THI-301. These results show the diversity of the body temperature. There is a high correlation between the tympanic temperature in the left and right ears. However, a paired t-test shows different results for the two experiments. In the first experiment, there is no significant difference between the temperature in the right and left ears, as measured using a BRAUN IRT (p < 0.05). In the second experiment, there is a significant difference in the temperature in the right and left ears for the two types of IRT (p < 0.002). The subjects for the two experiments were different. The first experiment involved 659 subjects and the second experiment involved 114. The size of the sample and a different age distribution affects the standard deviation, which affects the calculation of t values. In the first experiment, the difference between the mean tympanic temperature and the forehead temperature is 2.20 • C ± 0.711 • C. This experiment used a tympanic IRT to detect the forehead temperature. The emissivity of this device cannot be adjusted. In the second experiment, the difference between the mean tympanic temperature and the forehead temperature is 2.1 • C ± 0.301 • C. This experiment used an industrial IRT to measure the forehead temperature. The emissivity of this industrial device was adjusted to 0.98. Both experiments produce similar results. The forehead temperature is 2.1 • C or 2.2 • C lower than the tympanic temperature. In the second experiment, the wrist temperature that is measured using a THI-301 IRT is 3.3 • C lower than the tympanic temperature. These measurements for wrist temperature do not represent the actual body temperature, so measurements of the wrist temperature cannot be used to establish a threshold for screening for fever. Measurements of forehead temperature are convenient, fast and involve a low risk of infection, so they are used to screen for fever to identify possible victims of the novel coronavirus (COVID-19). However, there is no international standard for the threshold of forehead temperature that indicates fever. The Department of Health of Hong Kong [54] recommended a threshold level for fever of >38 • C for ear temperature and >36 • C for forehead temperature. The difference between the two critical temperatures is 2 • C. The results of this study are close to these figures. Liu et al. [12] determined the accuracy of alternative IR techniques for fever screening during the 2003 SARS epidemic. Two types of IRTs were used to measure tympanic and forehead temperature. The results for 276 subjects showed a tympanic temperature of 36.44 • C ± 0.37 • C and a forehead temperature of 35.63 • C ± 0.36 • C. Ng et al. [13] measured forehead temperature for 1000 healthy subjects using three handheld IRTs. The temperature was measured as 33.3 • C ± 1.18 • C using IR thermometers, but the results of the three IR thermometers were inconsistent. Their study noted that a forehead temperature of more than 35.5 • C suggests fever. Ng et al. [14] determined that the forehead and tympanic temperature for 567 patients, measured using an IRT, differs by 2.34 • C. However, the study does not report the specifications of the sensors or the calibration technique. Williams et al. [55] measured tympanic and forehead temperature using different IRTs. The tympanic was measured as 36.7 • C ± 0.5 • C and the forehead temperature as 34.0 • C ± 0.7 • C, and they concluded that measurements of forehead temperature are not as representative of the actual body temperature as tympanic temperature. In our study, the tympanic and forehead temperatures measured by BRAUN IRT were 36.9 • C ± 0.286 • C and 34.714 • C ± 0.392 • C, respectively. The results of our study were similar to those of Ng et al. [14] . There is inconsistency in the results of studies. Ganio et al. [56] and Edling et al. [57] reported that a forehead IRT has good repeatability but cannot be applied as a clinical device because there is a fixed offset. Langham et al. [39] reported a 2 • C difference in forehead temperature, as measured using an IRT, and the reference value was the bladder temperature. Chiappini et al. [58] and Teran et al. [59] determined that the use of an IRT to measure forehead temperature gives accurate results, so these are suited to clinical use. However, the fixed offset could be explained as the physiological phenomenon explained by the different temperature distribution between the inner and surface forehead skin. The precision of forehead IRT could be evaluated with the CV values in our study. The mean difference between a tympanic IRT and a contact-type deep skin forehead thermometer was less than 1 • C [27, 28, 34] . If an IRT is used to measure the forehead temperature, the surface temperature of the skin is measured, not the muscle temperature near the superficial temporal artery [47] . Core body temperature is the standard temperature of the body. It is defined as the temperature of the blood perfusing the thermoregulatory receptor in the hypothalamus [4] . The tympanic membrane has the same vascular supply that perfuse the hypothalamus, so it is the best site to measure the core body temperature [60] . The superficial temporal is a side drench of the carotid artery and feeds beneath the vascular bed of the head's skin [47] . If a contact type thermometer is used to measure the forehead temperature, the sensing element, which is a thermistor, must touch the skin of the forehead. The sensors must be inserted into the muscle. The effect of the environment, such as wind speed, is neglected, so the measurement of forehead temperature using a contact-type thermometer is close to the tympanic temperature [33, 34] . However, the fixed difference between the measurement values of tympanic and forehead IRTs needs to be considered. A tympanic IRT is painless in use, does not involve contact with the membrane and gives a rapid result. However, the cover must be frequently replaced and the probe position must be adjusted to reduce distress to subjects. The measurement of forehead temperature using an IRT is more convenient than that of tympanic IRT. The performance of an IRT is affected by the calibration method and environmental factors, such as solar radiation and wind speed over the skin [61, 62] . Standard calibration practices [6, 7, 62, 63] and proper operational procedures ensure the accuracy of the measurement of the forehead temperature. The critical tympanic temperature that is used to screen for fever is 38 • C [1, 54] . The Hong Kong government's threshold level for fever is 38 • C for tympanic temperature and 36 • C for forehead temperature. Ng et al. [13, 14] recommended a temperature of 35.5 • C to screen for fever. Previous studies show that the mean difference between forehead and tympanic temperature is 1.9 • C [64]; 2.7 • C [55] ; and 2.0 • C [39] . This study determines that the mean difference between tympanic temperature and forehead temperature is 2.2 • C. The uncertainty of three IRTs was calculated by Chen et al. [55] . At 36 • C, the combined uncertainty for the BRAUN is 0.15 • C, for the OMERO is 0.23 • C and for the THI-301 is 0.29 • C. This study determines that the tympanic temperature that is measured by the BRAUN IRT for the two experiments is 36.9 • C ± 0.286 • C and 36.937 • C ± 0.301 • C. The forehead temperature that is measured by the BRAUN is 34.714 • C ± 0.392 • C and by the THI-301 IRT is 34.80 • C ± 0.343 • C. The difference temperature between tympanic and forehead temperature for the two experiments is 2.20 • C ± 0.411 • C and 2.107 • C ± 0.311 • C. The difference between tympanic temperature and forehead temperature ranges from 2.1 • C to 2.2 • C for the two experiments. The threshold for fever level for tympanic temperature is 38 • C. The measurement uncertainty for an IRT means that the threshold for fever for forehead temperature that is measured using an IRT could be 36 • C. Therefore, forehead temperature does not give as accurate a representation of actual body temperature as the tympanic temperature, and as such, forehead IRTs are suitable for quick screening but are not as accurate as a traditional body temperature thermometer. The combined uncertainty of two IRTs at 36 • C for the BRAUN and the THI-301 was 0.42 • C and 0.45 • C, respectively. For the BRAUN IRT, the mean forehead temperature was 34.714 • C. The 95th and 99th percentiles were 35.40 • C and 35.69 • C, respectively. For the THI-301 IRT, the mean forehead temperature was 34.802 • C. The 95th and 99th percentiles were 35.54 • C and 35.85 • C, respectively. If the 35.5 • C is selected as the threshold value for fever screening, the percentiles are 96.94% and 93.96% for BRAUN IRT and THI IRT, respectively. That is, there could be a false-positive region of 3.06% or 6.04%, respectively, and this will cause trouble in the fever screening of the masses. If the 36.0 • C is selected as the threshold value for fever screening, the percentiles are 96.89% and 99.61% for BRAUN IRT and THI IRT, respectively and the region of false are small. Considering the measurement uncertainty of IRTs and the requirement of practical operation (ease of use, speed and convenience), the standard operating procedure (SOP) is proposed to screen for fever using measurements of forehead temperature. Tympanic and forehead IRTs should be used. b. Both IRTs must be calibrated using a black box method to ensure accuracy [6, 7, 47, 63 ]. c. The threshold level for fever is 36.0 • C for a forehead temperature measurement using an IRT. d. If the forehead temperature of a subject is measured to be higher than 36.0 • C, the tympanic temperature must be measured. e. If the tympanic temperature is >38 • C. The subject is considered to have a fever condition and clinical treatment is advised. Forehead temperature measurement using an infrared thermometer is used to rapidly screen for fever to identify victims of the coronavirus (COVID-19). The performance of this type of thermometer and the threshold temperature for screening for fever is studied. This study uses two medical tympanic IRTs and an industrial thermometer IRT to measure the tympanic temperature in both ears and the temperature at the wrist and forehead. The results show that the tympanic temperature that is measured using the BRAUN IRT is 36.9 • C ± 0.286 • C and 36.937 • C ± 0.301 • C for the two experiments. The forehead temperature that is measured by the BRAUN is 34.714 • C ± 0.392 • C and by the THI-301 IRT is 34.80 • C ± 0.343 • C. The temperature differences between the tympanic temperature and the forehead temperature for the two experiments are 2.20 • C ± 0.411 • C and 2.107 • C ± 0.311 • C, respectively. The difference between the tympanic temperature and the forehead temperature ranges from 2.1 • C to 2.2 • C for the two experiments. The measurement uncertainty for an IRT means that the threshold for fever when measuring forehead temperature is 36 • C. This study proposes a standard operating procedure (SOP) to screen for fever by measuring forehead temperature using an IRT. Forehead IRTs are suited to quick screening but are not be used to represent the actual body temperature as tympanic temperature measurements. The Significance of Core Temperature-Pathophysiology and Measurement Methods A comparative study of infrared tympanic thermometry and rectal mercury thermometry. Scand Infrared ear thermometry compared with rectal thermometry in child: A systematic review An assessment of infrared tympanic thermometers for body temperature measurement Study of facial skin and aural temperature IR ear thermometers: What do they measure and how do they comply with the EU technical regulation? Infrared ear thermometers-parameters influencing their reading and accuracy Comparison of temporal artery, mid-forehead skin and axillary temperature recordings in preterm infants <1500 g of birthweight Infrared ear thermometry in the critically ill patient: An alternative to axillary thermometry Infrared tympanic thermometer can accurately measure the body temperature in children in an emergency room setting US increases Ebola screening at five airports Limitations of forehead infrared body temperature detection for fever screening for severe acute respiratory syndrome A brief report on the normal range of forehead temperature as determined by noncontact, handheld, infrared thermometer Non-contact infrared thermometry temperature measurement for screening fever in children Thermometer Guns' on Coronavirus Front Lines are "Notoriously not Accurate". The New York Times Body temperature measurement to prevent pandemic COVID-19 in hospitals in Taiwan: Repeated measurement is necessary Infrared camera assessment of skin surface temperature-Effect of emissivity Comparative study of circadian variation in oral, tympanic, forehead, axillary and elbow pit temperatures measured in a cohort of young university students living their normal routines An indirect skin emissivity measurement in the infrared thermal range through reflection of a CO 2 laser beam Novel approach to assess the emissivity of the human skin Image processing based body temperature estimation using thermal video sequence Non-invasive forehead segmentation in thermographic imaging Measurement of core body temperature using graphene-inked infrared thermopile sensor Comparison of infrared thermal detection systems for mass fever screening in a tropical healthcare setting Errors in body temperature assessment related to individual variation, measuring technique and equipment Which thermometer? Factors influencing best choice for intermittent clinical temperature assessment Comparison of ear-based, bladder, oral, and axillary methods for core temperature measurement Comparison of esophageal, tympanic, and forehead skin temperatures in adult patients Thermistor recording of forehead skin temperature as an index of carotid artery disease Improved probe for a deep body thermometer Comparison of distal esophageal temperature with "deep" and tracheal temperatures Deep-forehead" temperature correlates well with blood temperature Insufficiency in a new temporal-artery thermometer for adult and pediatric patients Non-invasive continuous cerebral temperature monitoring in patients treated with mild therapeutic hypothermia: An observational pilot study Continuous monitoring of skin temperature using a liquid-crystal thermometer during anesthesia An evaluation of liquid-crystal thermometry as a screening device for intraoperative hyperthermia An evaluation of a plastic strip thermometer Does forehead liquid crystal temperature accurately reflect "core" temperature? Noninvasive temperature monitoring in post anesthesia care units Accuracy and precision of four common peripheral temperature measurement methods in intensive care patients Comparison of infrared tympanic, non-contact infrared skin, and axillary thermometer to rectal temperature measurements in a pediatric emergency observation unit Can a non-contact infrared thermometer be used interchangeably with other thermometers in an adult emergency department? Re-visiting the tympanic membrane vicinity as core body temperature measurement site Axillary and tympanic temperature measurement in children and normal values for ages Body temperature measurement methods and targets in Australian and New Zealand intensive care units Forehead or ear temperature measurement cannot replace rectal measurements, except for screening purposes Reliability of an infrared forehead skin thermometer for core temperature measurements Evaluating the interchangeability of forehead, tympanic, and axillary thermometers in Italian pediatric clinical settings: Results of a multicentre observational study Evaluation of performance and uncertainty of infrared tympanic thermometers Performance evaluation of the infrared thermocouple Classical and Modern Regression with Applications Experimental study on skin temperature and thermal comfort of the human body in a recumbent posture under uniform thermal environments Dynamics and topography of human temperature regulation in response to thermal and work load Proper Use of Thermometers to Measure Body Temperature Thermal screening of facial skin arterial hot spots using non-contact infrared radiometry Validity and reliability of devices that assess body temperature during indoor exercise in the heat Forehead or axillary temperature measurement is not reliable. Methods and thermometers compared with rectal temperature measurement as a reference Performance of non-contact infrared thermometer for detecting febrile children in hospital and ambulatory settings Clinical accuracy of a non-contact infrared skin thermometer in pediatric practice. Child Care Health Dev Clinical temperature: New physiological basis An evaluation of the performance of an infrared tympanic thermometer Accuracy and precision of noninvasive temperature measurement in adult intensive care patients Traceability and calibration in temperature measurement: A clinical necessity Clinical evaluation of liquid crystal skin thermometers This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license The authors declare no conflict of interest. Author Contributions: H.-Y.C. drafted the proposal, executed the statistical analysis, interpreted the results and revised the manuscript. A.C. performed experiments and interpreted some results. C.C. reviewed the proposal, performed some experiments, interpreted some results and criticized the manuscript and participated in its revision. All authors have read and approved the final manuscript.Funding: This research received no external funding.